Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -rab14e70c7bac24b0c3ebd02c6763285cdd3951ce -rb0a5452ec4b9be9be937095b1249079a2fce5ce5 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision ab14e70c7bac24b0c3ebd02c6763285cdd3951ce) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision b0a5452ec4b9be9be937095b1249079a2fce5ce5) @@ -89,13 +89,13 @@ #define HEAT_DISINFECT_START_TEMPERATURE_C 81.0F ///< Heat disinfect minimum acceptable temperature in C. // R1 to R2 & R2 to R1 heat disinfect circulation -#define HEAT_DISINFECT_TARGET_RO_FLOW_LPM 1.5F // TODO this was 1.3 ///< Heat disinfect target RO flow rate in L/min. -#define HEAT_DISINFECT_MAX_RO_PRESSURE_PSI 45 // TODO this was 30 ///< Heat disinfect maximum RO pressure in psi. -#define HEAT_DISINFECT_TARGET_DRAIN_PRES_PSI 15.0F // TOD was 12 ///< Heat disinfect target drain outlet pressure in psi. +#define HEAT_DISINFECT_TARGET_RO_FLOW_LPM 1.3F // TODO this was 1.3 ///< Heat disinfect target RO flow rate in L/min. +#define HEAT_DISINFECT_MAX_RO_PRESSURE_PSI 30 // TODO this was 30 ///< Heat disinfect maximum RO pressure in psi. +#define HEAT_DISINFECT_TARGET_DRAIN_PRES_PSI 12.0F // TODO was 12 ///< Heat disinfect target drain outlet pressure in psi. #define HEAT_DISINFECT_TIME_MS ( 10 * SEC_PER_MIN * MS_PER_SECOND ) ///< Heat disinfect time for each section in milliseconds. #define HEAT_DISINFECT_START_TEMP_TIMOUT_MS ( 4 * MIN_PER_HOUR * SEC_PER_MIN * MS_PER_SECOND ) ///< Heat disinfect reaching to minimum temperature timeout in milliseconds. #define RSRVRS_TARGET_VOL_OUT_TIMEOUT_MS ( 0.5F * SEC_PER_MIN * MS_PER_SECOND ) ///< Reservoirs 1 & 2 maximum volume out of range timeout during heat disinfect. TODO change this to 5 seconds -#define RSRVRS_MAX_TARGET_VOL_CHANGE_ML 600.0F ///< Reservoirs 1 & 2 maximum allowed volume change when full during heat disinfect. TODO original value is 100 mL +#define RSRVRS_MAX_TARGET_VOL_CHANGE_ML 150.0F ///< Reservoirs 1 & 2 maximum allowed volume change when full during heat disinfect. #define POST_HEAT_DISINFECT_WAIT_TIME_MS ( 3 * SEC_PER_MIN * MS_PER_SECOND ) ///< Heat disinfect final wait time before flushing the system in milliseconds. // Cool down RO filter