Index: firmware/App/Modes/ModeRecirculate.c =================================================================== diff -u -r5a36a768d11cc597a36b894c1fb3a5e5590130f1 -r2fff37fa585181917705645494549b5fd4a4d522 --- firmware/App/Modes/ModeRecirculate.c (.../ModeRecirculate.c) (revision 5a36a768d11cc597a36b894c1fb3a5e5590130f1) +++ firmware/App/Modes/ModeRecirculate.c (.../ModeRecirculate.c) (revision 2fff37fa585181917705645494549b5fd4a4d522) @@ -44,7 +44,7 @@ #define TARGET_RO_FLOW_RATE_L 0.3 ///< Target flow rate for RO pump. #define TARGET_FLUSH_LINES_RO_FLOW_RATE_L 0.6 ///< Target flow rate for RO pump. -#define FLUSH_LINES_VOLUME_L 0.1 ///< Water volume (in Liters) to flush when starting re-circulate mode. +#define FLUSH_LINES_VOLUME_L 0.01 ///< Water volume (in Liters) to flush when starting re-circulate mode. // ********** private data ********** @@ -91,16 +91,28 @@ setValveState( VRC, VALVE_STATE_DRAIN_C_TO_NO ); setValveState( VDR, VALVE_STATE_DRAIN_C_TO_NO ); setValveState( VPO, VALVE_STATE_NOFILL_C_TO_NO ); - setROPumpTargetFlowRate( TARGET_FLUSH_LINES_RO_FLOW_RATE_L, TARGET_RO_PRESSURE_PSI ); signalDrainPumpHardStop(); - stopPrimaryHeater(); requestConcentratePumpsOff( CONCENTRATEPUMPS_CP1_ACID ); requestConcentratePumpsOff( CONCENTRATEPUMPS_CP2_BICARB ); - // UV on + // UV reactors on turnOnUVReactor( INLET_UV_REACTOR ); turnOnUVReactor( OUTLET_UV_REACTOR ); + + // NOTE: The target flow rate should be set prior to setting the start primary heater + // because the initial guess in the heaters driver needs the target flow to calculate + // the new PWMs for the main and small primary heaters + setROPumpTargetFlowRate( TARGET_FLUSH_LINES_RO_FLOW_RATE_L, TARGET_RO_PRESSURE_PSI ); + startPrimaryHeater(); + +#ifndef _VECTORCAST_ + { // TODO - test code to start the fan since we're turning the heater on + F32 fanPWM = 0.25; + etpwmSetCmpA( etpwmREG6, (U32)( (S32)( ( fanPWM * (F32)(etpwmREG6->TBPRD) ) + FLOAT_TO_INT_ROUNDUP_OFFSET ) ) ); + etpwmSetCmpB( etpwmREG6, (U32)( (S32)( ( fanPWM * (F32)(etpwmREG6->TBPRD) ) + FLOAT_TO_INT_ROUNDUP_OFFSET ) ) ); + } +#endif } /*********************************************************************//** @@ -166,7 +178,9 @@ // when enough water volume has flowed to flush the lines, transition to re-circ state if ( flushLinesVolumeL >= FLUSH_LINES_VOLUME_L ) { +#ifndef SKIP_RECIRC setValveState( VDR, VALVE_STATE_RECIRC_C_TO_NC ); +#endif setROPumpTargetFlowRate( TARGET_RO_FLOW_RATE_L, TARGET_RO_PRESSURE_PSI ); result = DG_RECIRCULATE_MODE_STATE_RECIRC_WATER; }