Index: firmware/App/Modes/ModeStandby.c =================================================================== diff -u -r3e5c16b70fb34da7d58f5cf6efff896fdb3d00c3 -rfd89ed3ac616c0ee40e3ffc357d379cb2f4aa12a --- firmware/App/Modes/ModeStandby.c (.../ModeStandby.c) (revision 3e5c16b70fb34da7d58f5cf6efff896fdb3d00c3) +++ firmware/App/Modes/ModeStandby.c (.../ModeStandby.c) (revision fd89ed3ac616c0ee40e3ffc357d379cb2f4aa12a) @@ -130,6 +130,11 @@ setCPLDCleanLEDColor( CPLD_CLEAN_LED_OFF ); + // Upon transition to mode standby set CD1 and CD2 calibration records to be picked to the normal + // table. If the chemical disinfect fails, the mode transitions back to mode standby. + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD1_SENSOR, CAL_DATA_CD1_COND_SENSOR ); + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR ); + #ifndef _RELEASE_ setHeatNelsonSupportMode( NELSON_NONE ); setChemNelsonSupportMode( NELSON_NONE );