Index: firmware/App/Services/SystemCommMessages.c =================================================================== diff -u -r62b202c33e90b032326da24aa4067e2f5032f448 -r1b4447ba6fdc3d90a9013571b5710a52295c43d3 --- firmware/App/Services/SystemCommMessages.c (.../SystemCommMessages.c) (revision 62b202c33e90b032326da24aa4067e2f5032f448) +++ firmware/App/Services/SystemCommMessages.c (.../SystemCommMessages.c) (revision 1b4447ba6fdc3d90a9013571b5710a52295c43d3) @@ -1144,11 +1144,8 @@ cmdResponse.rejectCode = DG_CMD_REQUEST_REJECT_REASON_INVALID_PARAMETER; } } - else if ( ( getCurrentOperationMode() != DG_MODE_HEAT ) && ( getCurrentOperationMode() != DG_MODE_CHEM ) ) + else { - // The trimmer heater is controlled by HD except when DG is running heat disinfect or chemical disinfect. - // If the trimmer heater is run by DG in the cleaning modes but HD is expecting the heater to be off, it will - // keep sending the stop command all the time. stopHeater( DG_TRIMMER_HEATER ); }