Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -rc7ea3969643ead419cbfcd34c3cb203c45bcad71 -r6ebbc7eb52b9b6fa0ef128e86e7a4be29971ad4c --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision c7ea3969643ead419cbfcd34c3cb203c45bcad71) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 6ebbc7eb52b9b6fa0ef128e86e7a4be29971ad4c) @@ -78,7 +78,7 @@ // 2m long tubing to cap = 19.5 mL (acid line) + 7.92 mL/m * 2 m (tubing to cap) + 20.82 mL (straw) = 56.15 mL // Prime time in seconds = ( 56.15 mL / 48 mL/min ) x 60 second/min + 25 seconds margin time = 95 seconds. -#define PRIME_CONCENTRATE_LINES_TIME_OUT_MS ( 95 * MS_PER_SECOND ) ///< Time required to prime the concentrate lines. +#define PRIME_CONCENTRATE_LINES_TIME_OUT_MS ( 60 * MS_PER_SECOND ) ///< Time required to prime the concentrate lines. #define FLUSH_BUBBLES_PUMP_TIME_OUT_MS ( 2 * MS_PER_SECOND ) ///< RO pump on during flush bubble interval in ms. #define DIALYSATE_TEMPERATURE_TOLERANCE_C 2.0F ///< Dialysate temperature tolerance in degree C. #define DATA_PUBLISH_COUNTER_START_COUNT 63 ///< Data publish counter start count. @@ -614,6 +614,9 @@ setConcentratePumpTargetSpeed( CONCENTRATEPUMPS_CP2_BICARB, BICARB_PUMP_40_ML_PER_MIN ); requestConcentratePumpOn( CONCENTRATEPUMPS_CP2_BICARB ); + // Prior to collecting data for bicarb test, change the CD2 calibration table + // setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR_FILL_BICARB_TEST ); // TODO uncomment + // State transition concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_BICARB_CONDUCTIVITY; @@ -680,6 +683,10 @@ setROPumpTargetFlowRateLPM( RO_PUMP_800_ML_PER_MIN / MILLILITERS_PER_LITER, TARGET_RO_PRESSURE_PSI ); setConcentratePumpTargetSpeed( CONCENTRATEPUMPS_CP1_ACID, ACID_PUMP_20_ML_PER_MIN ); requestConcentratePumpOn( CONCENTRATEPUMPS_CP1_ACID ); + + // Set back the conductivity of CD2 calibration table to the normal calibration table + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR ); + // State transition concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_ACID_CONDUCTIVITY;