Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -rc48ea51c06c9d002f1b9b6a1d7b72d69756f183d -rbaa9fb9bb10eed6cd211abaa80e9f9acffdeb20d --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision c48ea51c06c9d002f1b9b6a1d7b72d69756f183d) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision baa9fb9bb10eed6cd211abaa80e9f9acffdeb20d) @@ -86,9 +86,9 @@ #ifndef V2_0_SYSTEM #define DIP_SPEED_ADC_TO_RPM_FACTOR 1.751752 ///< Conversion factor from ADC counts to RPM for dialIn pump motor. - #define DIP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.00025 ///< ~40 BP motor RPM = 1% PWM duty cycle + #define DIP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.000198 ///< ~40 BP motor RPM = 1% PWM duty cycle #else - #define DIP_SPEED_ADC_TO_RPM_FACTOR 1.280938 ///< Conversion factor from ADC counts to RPM for blood pump motor + #define DIP_SPEED_ADC_TO_RPM_FACTOR 1.280938 ///< Conversion factor from ADC counts to RPM for dialIn pump motor #define DIP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.0003125 ///< ~32 BP motor RPM = 1% PWM duty cycle #endif #define DIP_CURRENT_ADC_TO_MA_FACTOR 3.002 ///< Conversion factor from ADC counts to mA for dialIn pump motor.