Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r25a6eff69c9371f7a320751adf5359aaf070e5b0 -rd384eb875bc3cb4b41f12249a024072f0f0f29e1 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 25a6eff69c9371f7a320751adf5359aaf070e5b0) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision d384eb875bc3cb4b41f12249a024072f0f0f29e1) @@ -568,11 +568,11 @@ if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) { signalBloodPumpHardStop(); // Stop pump immediately - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + //SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) } else if ( bpOccl < ( OCCLUSION_CLEAR_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) { - clearAlarmCondition( ALARM_ID_OCCLUSION_BLOOD_PUMP ); + //clearAlarmCondition( ALARM_ID_OCCLUSION_BLOOD_PUMP ); } #endif }