Index: firmware/App/HDCommon.h =================================================================== diff -u -r03b93873cb60bba706f38a387565af2f5e4c4c09 -r0cc09b5d47c4018e6bfa8fb5f5c1a5d341512106 --- firmware/App/HDCommon.h (.../HDCommon.h) (revision 03b93873cb60bba706f38a387565af2f5e4c4c09) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision 0cc09b5d47c4018e6bfa8fb5f5c1a5d341512106) @@ -51,12 +51,11 @@ // #define RUN_DPI_OPEN_LOOP 1 // Run dialysate inlet pump in open loop mode #define WORN_OUT_CARTRIDGE 1 // Running with an old worn out cartridge (max wear) // #define PBA_ESTIMATION 1 // Estimate arterial pressure rather than look at PBA sensor - #define DISABLE_MOTOR_CURRENT_CHECKS 1 // Do not error on HD pump current checks - #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks - #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks - #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks +// #define DISABLE_MOTOR_CURRENT_CHECKS 1 // Do not error on HD pump current checks +// #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks +// #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks +// #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks // #define DISABLE_SYRINGE_PUMP 1 // Disable syringe pump functionality - #define ALWAYS_ALLOW_SYRINGE_PUMP_CMDS 1 // Allow syringe pump commands at any time except when pump is busy #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks #define DISABLE_ARTERIAL_PRESSURE_CHECK 1 // Do not error on arterial pressure out of range