Index: firmware/App/HDCommon.h =================================================================== diff -u -r708c23e585468515a8467d0056d712b2846b56da -r85b9b8c296b53a034da675bf76948a5075f5c1b0 --- firmware/App/HDCommon.h (.../HDCommon.h) (revision 708c23e585468515a8467d0056d712b2846b56da) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision 85b9b8c296b53a034da675bf76948a5075f5c1b0) @@ -39,7 +39,7 @@ // #define TASK_TIMING_OUTPUT_ENABLED 1 // Re-purposes alarm lamp pins for task timing #define DISABLE_ALARM_AUDIO 1 // Disable alarm audio #define SKIP_POST 1 // Skip POST tests - all pass - #define DONT_SKIP_NV_POST 1 // Do not skip NV Data POST +// #define DONT_SKIP_NV_POST 1 // Do not skip NV Data POST // #define USE_LIBRARY_TIME_FUNCTIONS 1 // Use the C library functions mktime() and gmtime() for epoch<=>date conversions #define DISABLE_AIR_TRAP_LEVELING_ALARM 1 // Disable air trap level control alarms // #define DISABLE_3WAY_VALVES 1 // Disable 3-way valves @@ -64,20 +64,20 @@ #define ALARMS_DEBUG 1 // Triggered alarms sent to debug UART #define ALARM_VOLUME_DEFAULT_LOW 1 // Set default alarm volume to lowest // #define TEMP_UI_ALARM_SILENCE_FIX 1 // Temporary UI fix for handling alarm silence request msg -// #define SKIP_PRIMING 1 // Skip Pre-treatment Prime + #define SKIP_PRIMING 1 // Skip Pre-treatment Prime #define SKIP_WET_SELF_TESTS 1 // Skip Pre-treatment prime wet self-tests - #define V2_0_SYSTEM 1 // Build for v2.0 system +// #define V2_0_SYSTEM 1 // Build for v2.0 system #define SKIP_SAMPLE_WATER 1 // Skip pre-treatment sample water #define SKIP_CONSUMABLE_TESTS 1 // Skip pre-treatment consumable Self-tests #define SKIP_DRY_SELF_TESTS 1 // Skip pre-treatment dry self-tests -// #define SKIP_UI_INTERACTION 1 // Skip UI interaction. + #define SKIP_UI_INTERACTION 1 // Skip UI interaction. #define DISABLE_BATT_COMM 1 // Disable battery communication. #define SKIP_AIR_BUBBLE_CHECK 1 // Skip air bubble detector self-test. #define DISABLE_OCCLUSION_SELF_TEST 1 // Skip occlusion sensor self-test. -// #define SKIP_CARTRIDGE_REMOVAL 1 // Skip cartridge removal check -// #define SKIP_EMPTY_RES_CHECK 1 // Skip reservoir empty check -// #define DISABLE_FPGA_COUNTER_CHECKS 1 // Disable alarms associated with FPGA read/error counters -// #define DISABLE_VOLTAGE_MONITOR 1 // Disable voltage monitoring/alarms + #define SKIP_CARTRIDGE_REMOVAL 1 // Skip cartridge removal check + #define SKIP_EMPTY_RES_CHECK 1 // Skip reservoir empty check + #define DISABLE_FPGA_COUNTER_CHECKS 1 // Disable alarms associated with FPGA read/error counters + #define DISABLE_VOLTAGE_MONITOR 1 // Disable voltage monitoring/alarms #define ALLOW_1_MIN_TREATMENT_DURATION 1 // Allow user to change treatment duration to as low as 1 minute #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently #define NO_PUMP_FLOW_LIMITS 1 // Allow any commanded flow rate for peristaltic pumps