Index: firmware/App/HDCommon.h =================================================================== diff -u -ra654052095e4c8e801bad6e8097decc7ee415914 -r5c2822d5c0c20c0bb3c668ee2dba6eaeafdce986 --- firmware/App/HDCommon.h (.../HDCommon.h) (revision a654052095e4c8e801bad6e8097decc7ee415914) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision 5c2822d5c0c20c0bb3c668ee2dba6eaeafdce986) @@ -53,7 +53,7 @@ #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks #define DISABLE_SYRINGE_PUMP 1 // Disable syringe pump functionality #define ALWAYS_ALLOW_SYRINGE_PUMP_CMDS 1 // Allow syringe pump commands at any time except when pump is busy - #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks +// #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks #define DISABLE_UF_ALARMS 1 // Do not error on HD ultrafiltration checks #define DISABLE_VALVE_ALARMS 1 // Do not error on HD valve position #define DISABLE_CAL_CHECK 1 // Disable calibration checks @@ -64,13 +64,13 @@ #define ALARMS_DEBUG 1 // Triggered alarms sent to debug UART #define ALARM_VOLUME_DEFAULT_LOW 1 // Set default alarm volume to lowest // #define TEMP_UI_ALARM_SILENCE_FIX 1 // Temporary UI fix for handling alarm silence request msg -// #define SKIP_PRIMING 1 // Skip Pre-treatment Prime + #define SKIP_PRIMING 1 // Skip Pre-treatment Prime #define SKIP_WET_SELF_TESTS 1 // Skip Pre-treatment prime wet self-tests // #define V2_0_SYSTEM 1 // Build for v2.0 system -// #define SKIP_SAMPLE_WATER 1 // Skip pre-treatment sample water + #define SKIP_SAMPLE_WATER 1 // Skip pre-treatment sample water #define SKIP_CONSUMABLE_TESTS 1 // Skip pre-treatment consumable Self-tests #define SKIP_DRY_SELF_TESTS 1 // Skip pre-treatment dry self-tests -// #define SKIP_UI_INTERACTION 1 // Skip UI interaction. + #define SKIP_UI_INTERACTION 1 // Skip UI interaction. #define DISABLE_BATT_COMM 1 // Disable battery communication. #define SKIP_AIR_BUBBLE_CHECK 1 // Skip air bubble detector self-test. #define DISABLE_OCCLUSION_SELF_TEST 1 // Skip occlusion sensor self-test. @@ -81,6 +81,7 @@ #define ALLOW_1_MIN_TREATMENT_DURATION 1 // Allow user to change treatment duration to as low as 1 minute #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently #define NO_PUMP_FLOW_LIMITS 1 // Allow any commanded flow rate for peristaltic pumps + #define DISABLE_BUBBLE_ALARMS 1 // Disable bubble alarms #include #include