Index: firmware/App/Modes/BloodPrime.c =================================================================== diff -u -r0a4dcd288d4347b85baaa0b07da568b6add5eac7 -r33c3a81272ce8c9bae64ae20537e9be45b11edd0 --- firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 0a4dcd288d4347b85baaa0b07da568b6add5eac7) +++ firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 33c3a81272ce8c9bae64ae20537e9be45b11edd0) @@ -37,7 +37,6 @@ // ********** private definitiions *********** #define TUBING_BLOOD_PRIME_VOLUME_ML 80.0F ///< Target blood prime volume to prime the blood side circuit (in mL). -#define DEFAULT_BLOOD_PRIME_VOLUME_ML 300.0F ///< Default blood prime volume (in mL). #define MAX_BLOOD_PRIME_VOLUME_ERROR_ML 60.0F ///< Maximum error in total additional blood prime volume (20% of total). #define MIN_RAMP_TIME_SEC 60 ///< Minimum ramp time for blood prime (in seconds). @@ -94,7 +93,6 @@ bloodPrimePublishTimerCtr = BLOOD_PRIME_DATA_PUBLISH_INTERVAL - 2; // setup so publish will occur time after next cumulativeBloodPrimeVolume_mL.data = 0.0; expectedBloodPrimeVolume_mL = 0.0; - bloodPrimeTargetVolume_mL = DEFAULT_BLOOD_PRIME_VOLUME_ML; resetBloodPrimeFlags(); bloodPrimeTargetVolume_mL = (F32)TUBING_BLOOD_PRIME_VOLUME_ML + (F32)getDialyzerBloodVolume();