Index: firmware/App/Modes/BloodPrime.c =================================================================== diff -u -r41eaf78372c0ee2a79b38d0a06bd3f02d54baed4 -r38355442b06187fe5d57deca647b3adf2fa26b89 --- firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 41eaf78372c0ee2a79b38d0a06bd3f02d54baed4) +++ firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 38355442b06187fe5d57deca647b3adf2fa26b89) @@ -8,7 +8,7 @@ * @file BloodPrime.c * * @author (last) Michael Garthwaite -* @date (last) 01-Mar-2023 +* @date (last) 15-May-2023 * * @author (original) Sean Nash * @date (original) 06-Feb-2021 @@ -36,8 +36,7 @@ // ********** private definitiions *********** -#define TUBING_BLOOD_PRIME_VOLUME_ML 80.0F ///< Target blood prime volume to prime the blood side circuit (in mL). -#define MIN_RAMP_TIME_SEC 60 ///< Minimum ramp time for blood prime (in seconds). +#define MIN_RAMP_TIME_SEC 60 ///< Minimum ramp time for blood prime (in seconds). /// Initial flow rate for blood pump when starting blood prime operation. #define BLOOD_PRIME_INIT_BP_FLOW_RATE_ML_MIN 100 @@ -140,6 +139,7 @@ setBloodPumpTargetFlowRate( (U32)bloodPrimeRampFlowRate_mL_min, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); setDialInPumpTargetFlowRate( DIALYSATE_FLOW_RATE_FOR_RECIRC, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); cmdStartDGTrimmerHeater(); + setCurrentSubState( NO_SUB_STATE ); // Start air trap control startAirTrapControl();