Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -rd4bc221d9817b488c2dcd00868386d7036a5c9a1 -rb20421d84b9049977c04f4752f4fc5697425f25c --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision d4bc221d9817b488c2dcd00868386d7036a5c9a1) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision b20421d84b9049977c04f4752f4fc5697425f25c) @@ -238,7 +238,7 @@ } #endif - setDialInPumpTargetFlowRate( setDialysateFlowRate, MOTOR_DIR_FORWARD, mode, 0.0F ); + setDialInPumpTargetFlowRate( setDialysateFlowRate, MOTOR_DIR_FORWARD, mode ); setDialOutPumpTargetRate( setDialysateFlowRate + (S32)setUFRate, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); // Start Heparin pump as appropriate @@ -327,7 +327,7 @@ mode = PUMP_CONTROL_MODE_OPEN_LOOP; } #endif - setDialInPumpTargetFlowRate( setDialysateFlowRate, MOTOR_DIR_FORWARD, mode, 0.0F ); + setDialInPumpTargetFlowRate( setDialysateFlowRate, MOTOR_DIR_FORWARD, mode ); setDialOutPumpTargetRate( setDialysateFlowRate + (S32)setUFRate, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); } @@ -867,7 +867,7 @@ salineBolusStartRequested = FALSE; // Cmd all pumps to stop setBloodPumpTargetFlowRate( 0, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); - setDialInPumpTargetFlowRate( 0, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP, 0.0F ); + setDialInPumpTargetFlowRate( 0, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); setDialOutPumpTargetRate( 0, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); stopSyringePump(); // Begin saline bolus @@ -909,7 +909,7 @@ #endif // Start dialysate inlet pump at re-circ rate #ifndef RUN_PUMPS_OPEN_LOOP - setDialInPumpTargetFlowRate( DIALYSATE_FLOW_RATE_FOR_RECIRC, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP, 0.0F ); + setDialInPumpTargetFlowRate( DIALYSATE_FLOW_RATE_FOR_RECIRC, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); #else setDialInPumpTargetFlowRate( DIALYSATE_FLOW_RATE_FOR_RECIRC, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP, 0.0F ); #endif