Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -rd4c4881844398626062215e8974f9f9c02890dbf -rce8cdf93f8b3fe230bdc16e4e112d4af01024bab --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision d4c4881844398626062215e8974f9f9c02890dbf) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision ce8cdf93f8b3fe230bdc16e4e112d4af01024bab) @@ -445,7 +445,7 @@ { accept = TRUE; salineBolusStartRequested = TRUE; - signalUserRateChange(); // so pressure limits re-stabilize + signalInitiatePressureStabilization(); } // Send response @@ -486,7 +486,7 @@ { accept = TRUE; salineBolusAbortRequested = TRUE; - signalUserRateChange(); // so pressure limits re-stabilize + signalInitiatePressureStabilization(); } // Send response @@ -1129,7 +1129,7 @@ sendOperationStatusEvent(); } - signalUserRateChange(); // so pressure limits re-stabilize + signalInitiatePressureStabilization(); // Resume dialysis transitionToDialysis(); }