Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -ref6283257df7c1f993d58fb934da57ea3e0a7067 -r726c340ed2bdedd8532a2ee26422e1b70643397f --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision ef6283257df7c1f993d58fb934da57ea3e0a7067) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 726c340ed2bdedd8532a2ee26422e1b70643397f) @@ -902,17 +902,9 @@ // Switch to saline bag setValvePosition( VBA, VALVE_POSITION_C_CLOSE ); // Start blood pump at saline bolus rate -#ifndef RUN_PUMPS_OPEN_LOOP setBloodPumpTargetFlowRate( SALINE_BOLUS_FLOW_RATE, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); -#else - setBloodPumpTargetFlowRate( SALINE_BOLUS_FLOW_RATE, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); -#endif // Start dialysate inlet pump at re-circ rate -#ifndef RUN_PUMPS_OPEN_LOOP setDialInPumpTargetFlowRate( DIALYSATE_FLOW_RATE_FOR_RECIRC, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); -#else - setDialInPumpTargetFlowRate( DIALYSATE_FLOW_RATE_FOR_RECIRC, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); -#endif // Begin saline bolus result = SALINE_BOLUS_STATE_IN_PROGRESS; }