Index: firmware/App/Modes/ModePreTreat.c =================================================================== diff -u -rf1253893206dece006b6bdb47ccdc22574cdbda9 -r985af7efc50a79e95511a50770897e321f881a41 --- firmware/App/Modes/ModePreTreat.c (.../ModePreTreat.c) (revision f1253893206dece006b6bdb47ccdc22574cdbda9) +++ firmware/App/Modes/ModePreTreat.c (.../ModePreTreat.c) (revision 985af7efc50a79e95511a50770897e321f881a41) @@ -629,8 +629,7 @@ signalBloodPumpHardStop(); signalDialOutPumpHardStop(); - //setDialInPumpTargetFlowRate( DIP_PATIENT_CONNECTION_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); - setDialInPumpTargetFlowRate( 250, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); // TODO remove this line once the new flow control is implemented + setDialInPumpTargetFlowRate( DIP_PATIENT_CONNECTION_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); cmdStartDGTrimmerHeater(); } @@ -1178,6 +1177,7 @@ rsrvrCmd.reservoirID = (U32)DG_RESERVOIR_1; rsrvrCmd.useLastTrimmerHeaterDC = FALSE; reservoirFlushedStatus[ DG_RESERVOIR_1 ] = TRUE; + cmdSetDGActiveReservoir( &rsrvrCmd ); } } @@ -1188,6 +1188,7 @@ rsrvrCmd.reservoirID = (U32)DG_RESERVOIR_2; rsrvrCmd.useLastTrimmerHeaterDC = FALSE; reservoirFlushedStatus[ DG_RESERVOIR_2 ] = TRUE; + cmdSetDGActiveReservoir( &rsrvrCmd ); } }