Index: firmware/App/Modes/ModeStandby.c =================================================================== diff -u -r437ac340cfe0bd90bc016372113bae3d594b7526 -rf249310122b6543b4145bb4824966535c93515d0 --- firmware/App/Modes/ModeStandby.c (.../ModeStandby.c) (revision 437ac340cfe0bd90bc016372113bae3d594b7526) +++ firmware/App/Modes/ModeStandby.c (.../ModeStandby.c) (revision f249310122b6543b4145bb4824966535c93515d0) @@ -75,8 +75,6 @@ /// Interval (in task intervals) at which to publish standby mode data to CAN bus. static OVERRIDE_U32_T standbyModePublishInterval = { DISINFECTS_DATA_PUB_INTERVAL, DISINFECTS_DATA_PUB_INTERVAL, DISINFECTS_DATA_PUB_INTERVAL, 0 }; -static const U32 SERVICE_TIME_INTERVAL_S = (U32)( 365 * 0.5 * SECONDS_IN_A_DAY ); ///< HD/DG 6-month service interval in seconds. - // ********** private function prototypes ********** static HD_STANDBY_STATE_T handleStandbyModeStartState( void ); @@ -1008,7 +1006,7 @@ currentDGCleaningMode.alarmID = ALARM_ID_NO_ALARM; currentDGCleaningMode.startRequestCmdID = DG_CMD_NONE; currentDGCleaningMode.dgOpMode = DG_MODE_FAUL; - currentDGCleaningMode.sampleRqstTrgrd = FALSE; + currentDGCleaningMode.sampleRqstTrgrd = FALSE; currentDGCleaningMode.sampleRequestID = GENERIC_CONFIRM_ID_NONE; currentDGCleaningMode.stopRequestCmdID = GENERIC_CONFIRM_ID_NONE; } @@ -1160,10 +1158,10 @@ BOOL isChemFlushComplete = ( lastChemFlushCompleteDate > lastChemCompleteDate ? TRUE : FALSE ); // If either of the basic flush, heat disinfect, or chemical disinfect flush have been done within the interval, it means the filters have been flushed - BOOL isBasicFlushValid = ( lastFlushCompleteDate < ( getRTCTimestamp() - FLUSH_TIME_INTERVAL_S ) ? TRUE : TRUE ); - BOOL isChemFlushValid = ( lastChemFlushCompleteDate < ( getRTCTimestamp() - FLUSH_TIME_INTERVAL_S ) ? TRUE : TRUE ); - BOOL isHeatDisCoolFlushValid = ( lastHeatCoolCompleteDate < ( getRTCTimestamp() - FLUSH_TIME_INTERVAL_S ) ? TRUE : TRUE ); - BOOL isFlushValid = ( isBasicFlushValid || isChemFlushValid || isHeatDisCoolFlushValid ); + BOOL isBasicFlushValid = ( lastFlushCompleteDate < ( getRTCTimestamp() - FLUSH_TIME_INTERVAL_S ) ? TRUE : FALSE ); + BOOL isChemFlushValid = ( lastChemFlushCompleteDate < ( getRTCTimestamp() - FLUSH_TIME_INTERVAL_S ) ? TRUE : FALSE ); + BOOL isHeatDisCoolFlushValid = ( lastHeatCoolCompleteDate < ( getRTCTimestamp() - FLUSH_TIME_INTERVAL_S ) ? TRUE : FALSE ); + BOOL isFlushValid = ( ( TRUE == isBasicFlushValid ) || ( TRUE == isChemFlushValid ) || ( TRUE == isHeatDisCoolFlushValid ) ); // If all of the above conditions are true, it means we can start a treatment if ( ( TRUE == hasDisBeenDone ) && ( TRUE == isHeatDisValid ) && ( TRUE == isChemFlushComplete ) && ( TRUE == isFlushValid ) ) @@ -1197,8 +1195,8 @@ { U32 dgSrvcElapsedTimeS = getRTCTimestamp() - dgData.dgServiceRecord.lastServiceEpochDate; U32 hdSrvcElapsedTimeS = getRTCTimestamp() - hdServiceRecord.lastServiceEpochDate; - BOOL hasDGSrvcBeenExpired = ( dgSrvcElapsedTimeS > SERVICE_TIME_INTERVAL_S ? TRUE : FALSE ); - BOOL hasHDSrvcBeenExpied = ( hdSrvcElapsedTimeS > SERVICE_TIME_INTERVAL_S ? TRUE : FALSE ); + BOOL hasDGSrvcBeenExpired = ( dgSrvcElapsedTimeS > dgData.dgServiceRecord.serviceIntervalSeconds ? TRUE : FALSE ); + BOOL hasHDSrvcBeenExpied = ( hdSrvcElapsedTimeS > hdServiceRecord.serviceIntervalSeconds ? TRUE : FALSE ); if ( TRUE == hasDGSrvcBeenExpired ) {