Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r66f77cd1004d7bba851bd980edf41d63da8118ff -r65a72f68f34ed506bf99c8cf52b339cb5bad11db --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 66f77cd1004d7bba851bd980edf41d63da8118ff) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 65a72f68f34ed506bf99c8cf52b339cb5bad11db) @@ -354,7 +354,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -1296,8 +1296,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {