Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r791b25bd20e80fce8c5d7e8a3ac37d8395f2315c -r1406dcedbc65edd2c134cf2e0861bb5aa2e569da --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 791b25bd20e80fce8c5d7e8a3ac37d8395f2315c) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 1406dcedbc65edd2c134cf2e0861bb5aa2e569da) @@ -954,8 +954,6 @@ { setOcclusionInstallLevel(); // Record occlusion pressure level after a new cartridge is installed doorClosedRequired( TRUE, TRUE ); - signalAllowDGFillRes1(); - signalAllowDGFillRes2(); state = DRY_SELF_TESTS_WAIT_FOR_DOOR_CLOSE_STATE; } } @@ -1467,19 +1465,6 @@ setupForSelfTestsStop(); } -#ifndef _RELEASE_ - if ( ( SW_CONFIG_ENABLE_VALUE == getSoftwareConfigStatus( SW_CONFIG_DISABLE_AIR_PUMP ) ) || // Allow res 1&2 fills now if air pump not disabled - ( SW_CONFIG_ENABLE_VALUE == getSoftwareConfigStatus( SW_CONFIG_DISABLE_DRY_SELF_TESTS ) ) ) - { - if ( state != DRY_SELF_TESTS_SYRINGE_PUMP_PRIME_STATE ) - { - signalAllowDGFlushFills(); - signalAllowDGFillRes1(); - signalAllowDGFillRes2(); - } - } -#endif - return state; }