Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r764c4865feb788998136c6129a7ec656b563047d -r45fb982b55fa1895d400d5ae3d31b0237e8a32db --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 764c4865feb788998136c6129a7ec656b563047d) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 45fb982b55fa1895d400d5ae3d31b0237e8a32db) @@ -43,7 +43,7 @@ #define PUMP_SELF_TEST_FLOW_RATE_ML_MIN 100 ///< Self-test pump flow rate in mL/min. #define BLOOD_PUMP_PRESSURE_SELF_TEST_FLOW 100 #define DIAL_IN_PUMP_PRESSURE_SELF_TEST_FLOW 400 -#define SYRINGE_PUMP_OCCLUSION_CHECK_DELAY ( 1 * MS_PER_SECOND ) ///< Delay 3 seconds then check for syringe pump prime occlusion. +#define SYRINGE_PUMP_OCCLUSION_CHECK_DELAY ( 1 * MS_PER_SECOND ) ///< Delay 1 second then check for syringe pump prime occlusion. #define BLOOD_PUMP_RUN_TIME_PRESSURE_SELF_TEST ( 20 * MS_PER_SECOND ) ///< Pressure self-test time to run blood pump in ms. #define DIAL_IN_PUMP_RUN_TIME_PRESSURE_SELF_TEST ( 20 * MS_PER_SECOND ) @@ -354,7 +354,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -1306,8 +1306,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {