Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -r8c6a692af6ef8136e84e8a4fb189b014bd1711e7 -rb0014aea833fc805312567f2f59c7a877bc8c500 --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 8c6a692af6ef8136e84e8a4fb189b014bd1711e7) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision b0014aea833fc805312567f2f59c7a877bc8c500) @@ -40,8 +40,8 @@ // ********** private definitions ********** -#define MAX_UF_VOLUME_ACCURACY_ERROR_ML 50.0F ///< Maximum ultrafiltration volume accuracy error in mL over the entire treatment. -#define MAX_UF_RATE_ACCURACY_ERROR_ML_MIN 5.0F ///< Maximum ultrafiltration rate accuracy error in mL/min over use of single reservoir during treatment. +#define MAX_UF_VOLUME_ACCURACY_ERROR_ML 100.0F ///< Maximum ultrafiltration volume accuracy error in mL over the entire treatment. +#define MAX_UF_RATE_ACCURACY_ERROR_ML_MIN 10.0F ///< Maximum ultrafiltration rate accuracy error in mL/min over use of single reservoir during treatment. /// Saline bolus data broadcast interval (ms/task time) count. static const U32 SALINE_BOLUS_DATA_PUB_INTERVAL = ( MS_PER_SECOND / TASK_GENERAL_INTERVAL );