Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r6b74d8fdb09c6d5468f8aaff9cb376b058fedab5 -rb1c0e4c327daaf809cbcb7e5667616ab9e6f2f92 --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 6b74d8fdb09c6d5468f8aaff9cb376b058fedab5) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision b1c0e4c327daaf809cbcb7e5667616ab9e6f2f92) @@ -354,7 +354,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -1312,8 +1312,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {