Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -rf2ea5ccbee9e1e56b2b8bf089d384e43364ae01d -rd3dfc11f7021ca991e5fd8ba552d0ddfe4a22936 --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision f2ea5ccbee9e1e56b2b8bf089d384e43364ae01d) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision d3dfc11f7021ca991e5fd8ba552d0ddfe4a22936) @@ -354,7 +354,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -1312,8 +1312,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {