Index: AlarmDefs.h =================================================================== diff -u -r0921cc68c4dbc05d1db12b71b9d15587e38efbd9 -rcd79ae102a12d446abf0a4e8518f9abf5a9740b8 --- AlarmDefs.h (.../AlarmDefs.h) (revision 0921cc68c4dbc05d1db12b71b9d15587e38efbd9) +++ AlarmDefs.h (.../AlarmDefs.h) (revision cd79ae102a12d446abf0a4e8518f9abf5a9740b8) @@ -300,8 +300,8 @@ { 24 , ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Primary Alarm Current | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, { 25 , ALARM_ID_DD_VALVE_CONTROL_FAILURE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Valve Control | DD FPGA not accepting commanded valve states | If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition - { 26 , ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Primary Alarm Current | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, - { 27 , ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Primary Alarm Current | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, + { 26 , ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Arterial Pressure Sensor Error | TD arterial pressure sensor reporting error or not being read successfully | Arterial pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, + { 27 , ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Venous Pressure Sensor Error | TD venous pressure sensor reporting error or not being read successfully | Venous pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, }; #endif // manual alarms