Index: AlarmDefs.h =================================================================== diff -u -r0ff5c360c04887aaf2634ba31bd6402cfe1b3297 -rdbf8d7f8686e298e90ee3842319977a4805a71d6 --- AlarmDefs.h (.../AlarmDefs.h) (revision 0ff5c360c04887aaf2634ba31bd6402cfe1b3297) +++ AlarmDefs.h (.../AlarmDefs.h) (revision dbf8d7f8686e298e90ee3842319977a4805a71d6) @@ -102,10 +102,10 @@ ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON = 71, ///< DD trimmer heater is On when no fluid is present ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT = 72, ///< DD water inlet input pressure sensor freshness/error timeout fault ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT = 73, ///< DD water inlet output pressure sensor freshness/error timeout fault - ALARM_ID_AVAILABLE_74 = 74, ///< Available alarms - ALARM_ID_AVAILABLE_75 = 75, ///< Available alarms - ALARM_ID_AVAILABLE_76 = 76, ///< Available alarms - ALARM_ID_AVAILABLE_77 = 77, ///< Available alarms + ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR = 74, ///< DD UF pump speed control error + ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE = 75, ///< DD spent chamber state 1 fill pressure out of range + ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE = 76, ///< DD spent chamber state 2 fill pressure out of range + ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT = 77, ///< DD spent chamber fill timeout fault ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT = 78, ///< TD air trap fill timeout alarm during treatment ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS = 79, ///< TD air trap illegal levels ALARM_ID_TD_PINCH_VALVE_FAULT = 80, ///< TD pinch valve fault @@ -120,7 +120,7 @@ ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE = 89, ///< DD balance chamber state2 fill start pressure not in range ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE = 90, ///< DD balance chamber state2 fill end pressure not in range ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE = 91, ///< DD generate dialysate hydraulics chamber ngative pressure out of range - ALARM_ID_AVAILABLE_92 = 92, ///< Available alarms + ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT = 92, ///< DD spent chamber fill timeout fault ALARM_ID_AVAILABLE_93 = 93, ///< Available alarms ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE = 94, ///< DD balance chamber state1 fill initiated and pressure drop is not in range ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE = 95, ///< DD balance chamber state2 fill initiated and pressure drop is not in range @@ -403,11 +403,11 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_74 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_75 }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_76 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_77 }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 13, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_PINCH_VALVE_FAULT }, @@ -422,7 +422,7 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_92 }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, @@ -470,30 +470,30 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_PRESSURE_RELIEF_OUT_LOW_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_PRESSURE_RELIEF_OUT_HIGH_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_POWER_BUDGET_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_FLOW_RATE_LOW_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_FLOW_RATE_LOW_RANGE }, { ALARM_PRIORITY_MEDIUM, 432, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_TEMPERATURE_OUT_LOW_RANGE }, { ALARM_PRIORITY_MEDIUM, 433, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_TEMPERATURE_OUT_HIGH_RANGE }, { ALARM_PRIORITY_MEDIUM, 434, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_PRESSURE_OUT_LOW_RANGE }, { ALARM_PRIORITY_MEDIUM, 435, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_PRESSURE_OUT_HIGH_RANGE }, { ALARM_PRIORITY_MEDIUM, 436, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_HIGH_RANGE }, { ALARM_PRIORITY_MEDIUM, 437, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_LOW_RANGE }, - { ALARM_PRIORITY_HIGH, 110, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_OUTLET_CONDUCTIVITY_HIGH_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_CONDUCTIVITY_HIGH_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_REJECTION_RATIO_LOW_RANGE }, - { ALARM_PRIORITY_HIGH, 110, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_REGULATOR_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_FILTER_HEALTH_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_REGULATOR_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_FILTER_HEALTH_PRESSURE_OUT_OF_RANGE }, // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID - { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 150, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , ALARM_ID_DD_BLOOD_LEAK_DETECTED }, { ALARM_PRIORITY_HIGH, 2, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, { ALARM_PRIORITY_HIGH, 2, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, { ALARM_PRIORITY_HIGH, 2, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { ALARM_PRIORITY_MEDIUM, 698, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, - { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, }; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID /// Table of alarms and their rank. @@ -608,8 +608,13 @@ { 110, ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT }, { 110, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, { 110, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, - { 110, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, { 110, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, + { 110, ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR }, + { 110, ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, + { 110, ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, + { 110, ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT }, + { 110, ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT }, + { 110, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, { 111, ALARM_ID_TD_DD_COMM_TIMEOUT }, { 111, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_FP }, @@ -638,11 +643,6 @@ { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, - { 999, ALARM_ID_AVAILABLE_74 }, - { 999, ALARM_ID_AVAILABLE_75 }, - { 999, ALARM_ID_AVAILABLE_76 }, - { 999, ALARM_ID_AVAILABLE_77 }, - { 999, ALARM_ID_AVAILABLE_92 }, { 999, ALARM_ID_AVAILABLE_93 }, { 999, ALARM_ID_UI_RESERVED_116 }, { 999, ALARM_ID_UI_RESERVED_117 }, @@ -758,11 +758,11 @@ { 71 , ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Heater On With No Fluid | DD D45 heater is On with low fluid level | If the Balance air separation chamber fluid level is low for a certain period of time. | "}, { 72 , ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Water Inlet Input Press Sensor Error | DD water inlet input pressure sensor reporting error or not being read successfully | Water Inlet Input pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 73 , ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Water Inlet Output Press Sensor Error | DD water inlet output pressure sensor reporting error or not being read successfully | Water Inlet Output pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, - { 74 , ALARM_ID_AVAILABLE_74 , " | | | | | | "}, - { 75 , ALARM_ID_AVAILABLE_75 , " | | | | | | "}, + { 74 , ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Pump Speed Fault | DD UF pump D76_PUMP speed control error | If the Ultrafiltration pump D76_PUMP speed > 0.02 tolerance for a certain period of time. | "}, + { 75 , ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD Spent chamber fill state 1 start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 24 psig or > 28 psig) not in range for spent chamber fill state1 start. | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition - { 76 , ALARM_ID_AVAILABLE_76 , " | | | | | | "}, - { 77 , ALARM_ID_AVAILABLE_77 , " | | | | | | "}, + { 76 , ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD Spent chamber fill state 2 start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 24 psig or > 28 psig) not in range for spent chamber fill state2 start. | "}, + { 77 , ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Spent Chamber fill timeout Error | DD Spent chamber fill timeout error | If the number of spent chamber fill > = 7 counts | "}, { 78 , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, { 79 , ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Air Trap Sensor | Air trap level sensors reporting illegal combination of air/fluid | If the air trap level readings are not valid for more than 2 seconds. | "}, { 80 , ALARM_ID_TD_PINCH_VALVE_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Pinch Valve | Pinch valve reporting a fault | If the valve detects a driver, direction or encoder fault condition. | "}, @@ -777,7 +777,7 @@ { 89 , ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill start. | "}, { 90 , ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill end. | "}, { 91 , ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Hyd Cham Neg Pres not in range | DD hydraulics chamber degassing pressure is not in range | If the hydraulics chamber pressure is not less than or equal to calculated target degas pressure psig. | "} - { 92 , ALARM_ID_AVAILABLE_92 , " | | | | | | "}, + { 92 , ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Bicarb Chamber fill timeout Error | DD Bicarb chamber fill timeout error | If Bicarb chamber fill time > = 1 seconds | "}, { 93 , ALARM_ID_AVAILABLE_93 , " | | | | | | "}, { 94 , ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state1 fill in progress. | "}, { 95 , ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state2 fill in progress. | "},