Index: AlarmDefs.h =================================================================== diff -u -r1fe68a51b58a177dce67d19c4b4924ae2aee6081 -r7bdebb83418eb57b2ce574eadeecb8677b7afc0a --- AlarmDefs.h (.../AlarmDefs.h) (revision 1fe68a51b58a177dce67d19c4b4924ae2aee6081) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 7bdebb83418eb57b2ce574eadeecb8677b7afc0a) @@ -64,7 +64,11 @@ ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE = 33, ///< TD voltage out of range fault ALARM_ID_TD_AC_POWER_LOST = 34, ///< TD AC power lost ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT = 35, ///< TD AC power lost during treatment - ALARM_ID_TD_VENOUS_BUBBLE_DETECTED = 36, ///< TD venous bubble detected + ALARM_ID_DD_CD1_SENSOR_FPGA_FAULT = 36, ///< DD CD1 conductivity sensor freshness/error timeout fault + ALARM_ID_DD_CD2_SENSOR_FPGA_FAULT = 37, ///< DD CD2 conductivity sensor freshness/error timeout fault + ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT = 38, ///< DD CD3 conductivity sensor freshness/error timeout fault + ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT = 39, ///< DD CD4 conductivity sensor freshness/error timeout fault + ALARM_ID_TD_VENOUS_BUBBLE_DETECTED = 40, ///< TD venous bubble detected NUM_OF_ALARM_IDS ///< Total number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -234,6 +238,10 @@ { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_LOW , 852, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_AC_POWER_LOST }, { ALARM_PRIORITY_HIGH, 100, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CD1_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CD2_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT }, { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED, }, }; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID @@ -271,6 +279,10 @@ { 110, ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT }, { 110, ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT }, { 110, ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT }, + { 110, ALARM_ID_DD_CD1_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_DD_CD2_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT }, { 111, ALARM_ID_TD_DD_COMM_TIMEOUT }, { 111, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_RO }, @@ -336,7 +348,11 @@ { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Voltage Monitor | HD monitored voltage is out of range | If the HD monitored voltages are out of range for a certain period of time. | "}, { 34 , ALARM_ID_TD_AC_POWER_LOST , "TD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, { 35 , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT , "TD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, - { 36 , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED , "TD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, + { 36 , ALARM_ID_DD_CD1_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD1) Error | DD conductivity sensor located before balancing chamber reporting error or not being read successfully | Conductvity Sensor (CD1) is reporting an error or FPGA is failing to read sensor. | "}, + { 37 , ALARM_ID_DD_CD2_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD2) Error | DD conductivity sensor located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (CD2) is reporting an error or FPGA is failing to read sensor. | "}, + { 38 , ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD3) Error | DD conductivity sensor located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (CD3) is reporting an error or FPGA is failing to read sensor. | "}, + { 39 , ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD4) Error | DD conductivity sensor located in spent dialysate path reporting error or not being read successfully | Conductvity Sensor (CD4) is reporting an error or FPGA is failing to read sensor. | "}, + { 40 , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED , "TD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, }; #endif // manual alarms