Index: AlarmDefs.h =================================================================== diff -u -r2f9aacdc29711409fba96c2b5eaccbcbeaaeb71a -r16067a34bbe2f63c78340dc92c6416bc849fcc81 --- AlarmDefs.h (.../AlarmDefs.h) (revision 2f9aacdc29711409fba96c2b5eaccbcbeaaeb71a) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 16067a34bbe2f63c78340dc92c6416bc849fcc81) @@ -52,6 +52,7 @@ ALARM_ID_DD_FPGA_POST_TEST_FAILED = 21, ///< DD FPGA POST failure ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE = 22, ///< DD processor clock speed checks against FPGA clock failure ALARM_ID_DD_FPGA_COMM_TIMEOUT = 23, ///< DG FPGA communication down for too long + ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE = 24, ///< TD alarm audio failed POST NUM_OF_ALARM_IDS ///< Total number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -208,6 +209,7 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FPGA_POST_TEST_FAILED }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FPGA_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, }; // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID /// Table of alarms and their rank. @@ -216,6 +218,7 @@ { 1, ALARM_ID_TD_FPGA_POST_TEST_FAILED }, { 1, ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED }, { 1, ALARM_ID_TD_UI_COMM_POST_FAILED }, + { 1, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, { 10, ALARM_ID_TD_SOFTWARE_FAULT }, { 10, ALARM_ID_TD_UI_COMM_TIMEOUT }, { 10, ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS }, @@ -260,29 +263,30 @@ { // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition { 0 , ALARM_ID_NO_ALARM , "N/A | | | | Indication for when no alarms is a possible situation | N/A | "}, - { 1 , ALARM_ID_TD_SOFTWARE_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: SW Invalid State | HD Software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, + { 1 , ALARM_ID_TD_SOFTWARE_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: SW Invalid State | HD Software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, { 2 , ALARM_ID_DD_SOFTWARE_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Software Invalid State | DD software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, { 3 , ALARM_ID_RO_SOFTWARE_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Software Invalid State | DD software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, - { 4 , ALARM_ID_TD_FPGA_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: FPGA Compatibility | HD FPGA POST failure | Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. | "}, - { 5 , ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: WD Safestate | HD Watchdog POST failure | If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA | "}, - { 6 , ALARM_ID_TD_UI_COMM_POST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc- UI Comm | UI communication POST failure. UI failed to communicate within a reasonable time after power up | If the UI failed to checkin via the CAN bus within 40 seconds | "}, - { 7 , ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Proc- FPGA Clock | HD processor clock speed checks against FPGA clock failure | If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. | "}, - { 8 , ALARM_ID_TD_FPGA_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: FPGA Com | HD FPGA communication down for too long | If there are more than 3 comm errors in a 1 min period. | "}, - { 9 , ALARM_ID_TD_UI_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI Comm | User interface communication timeout | If the sub-system communication timeout has occurred. UI is 5 seconds. DD is 2 seconds. | "}, - { 10 , ALARM_ID_TD_DD_COMM_TIMEOUT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DD Comm (Timeout) | Dialysate generator communication timeout | If DD communication timeout is detected for more than 2 seconds. | "}, - { 11 , ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN CRC Com | Too many bad CRCs detected on received system messages | If too many bad messages CRCs are received within a set period of time | "}, - { 12 , ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN ACK UI Com | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, - { 13 , ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DD Comm (DD ACK) | HD to DD comm error. | HD message to DD that requires acknowledge was not acknowledged within a certain time. | "}, - { 14 , ALARM_ID_TD_CRITICAL_DATA_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Critical Data Err | HD critical data integrity check failed | If the critical data failed range check. | "}, + { 4 , ALARM_ID_TD_FPGA_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: FPGA Compatibility | HD FPGA POST failure | Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. | "}, + { 5 , ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: WD Safestate | HD Watchdog POST failure | If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA | "}, + { 6 , ALARM_ID_TD_UI_COMM_POST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc- UI Comm | UI communication POST failure. UI failed to communicate within a reasonable time after power up | If the UI failed to checkin via the CAN bus within 40 seconds | "}, + { 7 , ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Proc- FPGA Clock | HD processor clock speed checks against FPGA clock failure | If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. | "}, + { 8 , ALARM_ID_TD_FPGA_COMM_TIMEOUT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: FPGA Com | HD FPGA communication down for too long | If there are more than 3 comm errors in a 1 min period. | "}, + { 9 , ALARM_ID_TD_UI_COMM_TIMEOUT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI Comm | User interface communication timeout | If the sub-system communication timeout has occurred. UI is 5 seconds. DD is 2 seconds. | "}, + { 10 , ALARM_ID_TD_DD_COMM_TIMEOUT , "TD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DD Comm (Timeout) | Dialysate generator communication timeout | If DD communication timeout is detected for more than 2 seconds. | "}, + { 11 , ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN CRC Com | Too many bad CRCs detected on received system messages | If too many bad messages CRCs are received within a set period of time | "}, + { 12 , ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN ACK UI Com | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, + { 13 , ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD , "TD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DD Comm (DD ACK) | HD to DD comm error. | HD message to DD that requires acknowledge was not acknowledged within a certain time. | "}, + { 14 , ALARM_ID_TD_CRITICAL_DATA_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Critical Data Err | HD critical data integrity check failed | If the critical data failed range check. | "}, { 15 , ALARM_ID_DD_CRITICAL_DATA_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Critical Data Err | DD critical data integrity check failed | If the critical data failed range check. | "}, - { 16 , ALARM_ID_TD_WATCHDOG_EXPIRED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: WD Error | HD watchdog expired | If the watchdog signal is low while not in POST. | "}, + { 16 , ALARM_ID_TD_WATCHDOG_EXPIRED , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: WD Error | HD watchdog expired | If the watchdog signal is low while not in POST. | "}, { 17 , ALARM_ID_DD_WATCHDOG_EXPIRED , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: WD Error | DD watchdog expired | If the watchdog signal is low while not in POST. | "}, { 18 , ALARM_ID_DD_COMM_TOO_MANY_BAD_CRCS , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: CAN CRC Com | Too many bad CRCs detected on received system messages | If too many bad messages CRCs are received within a set period of time | "}, { 19 , ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_RO , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of RO Comm (RO ACK) | DD to RO comm error | DD message to RO that requires acknowledge was not acknowledged within a certain time. | "}, { 20 , ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of TD Comm (TD ACK) | DD to TD comm error | DD message to TD that requires acknowledge was not acknowledged within a certain time. | "}, { 21 , ALARM_ID_DD_FPGA_POST_TEST_FAILED , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD POST: FPGA Compatibility | DD FPGA POST failure | Minimum DD FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header | "}, { 22 , ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Proc- FPGA Clock | FPGA clock speed failure | If the FPGA clock speed is relatively close to the processor clock for a certain period of time. | "}, { 23 , ALARM_ID_DD_FPGA_COMM_TIMEOUT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: FPGA Com | DD FPGA communication down for too long | If retries for commands exceeds limit or FPGA reports comm error. | "}, + { 24 , ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Primary Alarm Current | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition };