Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -rec55a70045556e618cb27526754a010147b21106 -r9851e65ac25088cce280989d0b95a78101834f46 --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision ec55a70045556e618cb27526754a010147b21106) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 9851e65ac25088cce280989d0b95a78101834f46) @@ -374,19 +374,19 @@ Message = DD Fault: Pump Speed Fault [75] -Title = Incoming Water Pressure Input High -ListTitle = DD inlet water pressure input is too high -Message = DD: Inlet Water Pres Input High +Title = Service Required: Dialysate Device +ListTitle = DD Spent chamber fill state 1 start pressure is not in range +Message = DD Fault: BC Pressure not in range [76] -Title = Incoming Water Pressure Output Low -ListTitle = DD Inlet water pressure output too low -Message = DD: Inlet Water Pres Output Low +Title = Service Required: Dialysate Device +ListTitle = DD Spent chamber fill state 2 start pressure is not in range +Message = DD Fault: BC Pressure not in range [77] -Title = Incoming Water Pressure Output High -ListTitle = DD inlet water pressure output is too high -Message = DD: Inlet Water Pres Output High +Title = Service Required: Dialysate Device +ListTitle = DD Spent chamber fill timeout error +Message = DD Spent Chamber fill timeout Error [78] Title = Blood Pump Stopped: Venous Air Detector @@ -459,9 +459,9 @@ Message = DD Fault: Hyd Cham Neg Pres not in range [92] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DD Bicarb chamber fill timeout error +Message = DD Bicarb Chamber fill timeout Error [93] Title = @@ -773,3 +773,33 @@ ListTitle = DD dialysate temperature sensors drift timeout. Message = DD Fault: D28 and D30 Delta +[155] +Title = Blood Leak +ListTitle = DD blood leak detected alarm +Message = Blood Stop: Dialyzer Blood Leak + +[156] +Title = Service Required: Hemodialysis Device +ListTitle = DD blood leak sensor invalid calibration record +Message = DD POST: BLD Cal + +[157] +Title = Service Required: Hemodialysis Device +ListTitle = DD blood leak sensor embedded mode communication failed +Message = DD POST: BLD Config Emb Mode + +[158] +Title = Service Required: Hemodialysis Device +ListTitle = DD blood leak sensor set point set failure +Message = DD POST: BLD Config Set Point + +[159] +Title = Blood Leak: Bypass +ListTitle = DD blood leak detected and it is in the process of attempting to remove the bubbles/foam +Message = Blood Stop: Dialyzer Blood Leak Bypass + +[160] +Title = FP: Permeate Underfill +ListTitle = FP tank level is too low +Message = FP Tank: Tank is not filling +