Index: build/AlarmMapping.py =================================================================== diff -u -r508c7d7a5dd45f297931788a9279bb7c430a01ff -r3a236c634d5f6787f7471e7137a00b08038de520 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 508c7d7a5dd45f297931788a9279bb7c430a01ff) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 3a236c634d5f6787f7471e7137a00b08038de520) @@ -46,7 +46,7 @@ 44 : "HD UI communication timeout." 45 : "HD too many bad communications CRC." 46 : "HD didn't get ACK on message to UI that required acknowledgment." - 47 : "Available for use." + 47 : "DG maximum RO Pump PWM exceeded." 48 : "HD ultrafiltration volume accuracy error during treatment." 49 : "HD FPGA communication down for too long." 50 : "DG FPGA not accepting commanded valve states." @@ -62,7 +62,7 @@ 60 : "HD valve not functional." 61 : "HD valve current out of range." 62 : "HD valve position out of target." - 63 : "Alarm ID available." + 63 : "HD syringe pump prime timeout." 64 : "DG barometric pressure sensor out of range." 65 : "HD DG requests DG command with invalid parameter fault." 66 : "HD blood leak sensor set point set failure." @@ -257,7 +257,7 @@ 255 : "HD and UI software builds are not compatible." 256 : "HD processor is in Disinfect Chemical Flush mode." 257 : "Alarm ID available for use." - 258 : "Alarm Id available." + 258 : "HD UI POST OS version compatibility failure." 259 : "HD temperatures out of range." 260 : "HD UI POST Application Integrity (Sha256Sum) failure." 261 : "HD UI POST CANBus failure." Index: build/Alarms.xlsx =================================================================== diff -u -r23fb7111bb54e6b0bcb52a0427300cba57a6ceac -r3a236c634d5f6787f7471e7137a00b08038de520 Binary files differ