Index: shared/scripts/configuration/config.py =================================================================== diff -u -r9e8d0a1f125fa5b8fa103adc30489bd55893ad47 -r11b2970e98eb80e8bd299d25b607b8ff00a2043b --- shared/scripts/configuration/config.py (.../config.py) (revision 9e8d0a1f125fa5b8fa103adc30489bd55893ad47) +++ shared/scripts/configuration/config.py (.../config.py) (revision 11b2970e98eb80e8bd299d25b607b8ff00a2043b) @@ -6,9 +6,9 @@ # THIS CODE MAY NOT BE COPIED OR REPRODUCED IN ANY FORM, IN PART OR IN # WHOLE, WITHOUT THE EXPLICIT PERMISSION OF THE COPYRIGHT OWNER. # -# @file utils.py +# @file config.py # -# @author (last) Joseph varghese +# @author (last) LTTS # @date (last) 18-Jan-2022 # ############################################################################ @@ -19,29 +19,292 @@ AUT_NAME = "denaliSquish" -COMMON_PATH = os.environ['HOME']+"/Projects" - COLOR_CODES = {"Aqua": "#81ffff", "Blue": "#017dea", "Blue 2": "#1090d5", "Green": "#29fd2d", "Grey": "#d3d3d3", "Lavender": "#db98f5", "Light Blue": "#acd7f1", "Light Teal": "#29f1da","Lime": "#b7fc36", "Magenta":"#fd28fd", "Orange": "#f2721c", "Peach":"#f1979a", "Red": "#c53b33", "Rose":"#fc178d", "Slate blue":"#7f7ffa", "Violet": "#6435c9", "White": "#ffffff", "Yellow": "#fcfc4d"} + CURRENT_COLOR = '#000000' COMPLETE_COLOR= '#4290ec' ENABLED_COLOR = '#fcfcfc' INCOMPLETE_COLOR = '#607a91' +TREATMENT_UNHIGHLIGHTED_COLOR = "#000000" +TREATMENT_HIGHLIGHTED_COLOR = "#135088" + +COMMON_PATH = os.environ['HOME']+"/Projects" + +#StandBy Mode +BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" + +#instructions_imagelocation +INSTRUCTION_CONF_LOCATION = "/home/denali/Projects/application/resources/settings/Instructions/Instructions.conf" +IMAGE_LOCATION = "file:///home/denali/Projects/application/resources/settings/Instructions/" +EXPECTED_IMAGE_LOCATION = "/home/denali/Projects/application/resources/settings/Instructions/" +SERVICE_CONF_LOCATION = os.environ['HOME']+"/Projects/application/resources/settings/Service.conf" +TREATMENT_LOG_LOCATION = '/home/denali/Desktop/sd-card/treatment/*.log' + +ACCEPTED = True +REJECTED = False +ENABLED = True +DISABLED = False +CURRENT_COLOR = '#000000' +COMPLETE_COLOR = '#4290ec' +ENABLED_COLOR = '#fcfcfc' +INCOMPLETE_COLOR = '#607a91' OUT_OF_RANGE_COLOR = "#c53b33" +#main treatment screen +IN_RANGE_COLOR = "#fcfcfc" +SYSTOLIC_LOWER_LIMIT = 60 +SYSTOLIC_UPPER_LIMIT = 250 +DIASTOLIC_LOWER_LIMIT = 40 +DIASTOLIC_UPPER_LIMIT = 200 +HEART_RATE_LOWER_LIMIT = 40 +HEART_RATE_UPPER_LIMIT = 180 +KEYBOARD_SWITCH = ["ABC", "&123"] + +HEPARIN_TYPE = "UFH 1,000 IU/mL" +ACID_CONCENTRATE = ["08-1251-1", "08-2251-0", "08-3251-9"] +BICARBONATE_CONCENTRATE = "Fres. Centrisol" +DIALYZER_TYPE = ["BB Diacap Pro 13H", "BB Diacap Pro 16H", "BB Diacap Pro 19H", "F Optiflux F160NRe", "F Optiflux F180NRe"] + +COLOR_CODES = {"Aqua": "#81ffff", "Blue": "#017dea", "Blue 2": "#1090d5", "Green": "#29fd2d", "Grey": "#d3d3d3", + "Lavender": "#db98f5", "Light Blue": "#acd7f1", "Light Teal": "#29f1da","Lime": "#b7fc36", + "Magenta":"#fd28fd", "Orange": "#f2721c", "Peach":"#f1979a", "Red": "#c53b33", "Rose":"#fc178d", + "Slate blue":"#7f7ffa", "Violet": "#6435c9", "White": "#ffffff", "Yellow": "#fcfc4d"} + +TREATMENT_UNHIGHLIGHTED_COLOR = "#000000" +TREATMENT_HIGHLIGHTED_COLOR = "#135088" + +#create custom treatment +HIGHER_VALUE_OF_BLOOD_FLOW_RATE = 500 +LOWER_VALUE_OF_BLOOD_FLOW_RATE = 100 +STEP_VALUE_OF_BLOOD_FLOW_RATE = 25 +BUFFER_LOW_AND_HIGH_LIMITS = 30 +ARTERIAL_VALUE_0 = 0 +ARTERIAL_VALUE_NEGATIVE_300 = -300 +ARTERIAL_VALUE_NEGATIVE_20 = -20 +ARTERIAL_VALUE_NEGATIVE_270 = -270 +VENOUS_VALUE_600 = 600 +VENOUS_VALUE_20 = 20 +VENOUS_VALUE_50 = 50 +VENOUS_VALUE_570 = 570 +POSITIVE_BUFFER = 10 +NEGATIVE_BUFFER = -10 + + +NUM_OF_ALARM_PRIORITIES = 4 # Total number of alarm priorities + + + + +NUM_OF_MSG_ID = 289 + + +ALARM_ID_NO_ALARM = 0 +ALARM_ID_OUT_OF_RANGE = 290 + +# alarm colors +ALARMS_COLORS_HEADER = { + "ALARM_PRIORITY_NONE" : "#438feb", + "ALARM_PRIORITY_LOW" : "#db8f00", + "ALARM_PRIORITY_MEDIUM" : "#db8f00", + "ALARM_PRIORITY_HIGH" : "#831913" +} + +ALARMS_COLORS_BODY = { + "ALARM_PRIORITY_NONE" : "#254670", + "ALARM_PRIORITY_LOW" : "#f5a623", + "ALARM_PRIORITY_MEDIUM" : "#f5a623", + "ALARM_PRIORITY_HIGH" : "#c53b33" +} + +ALARM_MUTE_FLAG = 0B0000001000000000 +ALARM_UNMUTE_FLAG = 0B0000000000000000 + +ALARM_BUTTONS_OK = 0B0000000010111000 + +ALARM_BUTTONS_RESUME = 0B0000000000110000 +ALARM_BUTTONS_RESUME_RINSEBACK = 0B0000000000100000 +ALARM_BUTTONS_RESUME_END = 0B0000000000010000 + +ALARM_BUTTONS_RINSEBACK = 0B0000000000101000 +ALARM_BUTTONS_RINSEBACK_RESUME = ALARM_BUTTONS_RESUME_RINSEBACK +ALARM_BUTTONS_RINSEBACK_END = 0B0000000000001000 + +ALARM_BUTTONS_END = 0B0000000000011000 +ALARM_BUTTONS_END_RESUME = ALARM_BUTTONS_RESUME_END +ALARM_BUTTONS_END_RINSEBACK = ALARM_BUTTONS_RINSEBACK_END + +ALARM_BUTTONS_RESUME_RINSEBACK_END = 0B0000000000000000 +ALARM_BUTTONS_RINSEBACK_RESUME_END = ALARM_BUTTONS_RESUME_RINSEBACK_END +ALARM_BUTTONS_END_RESUME_RINSEBACK = ALARM_BUTTONS_RESUME_RINSEBACK_END + +#standby mode & ultrafiltration +SALINE_BOLUS_CUMULATIVE = 300 +UF_MINIMUM_VALUE = 0 +UF_MAXIMUM_VALUE = 600 + + #standby mode GOODMORNING_START_TIME_SEC = 0 GOODEVENING_START_TIME_SEC = 43200 -BLOOD_PRIMING_TEXT = "Blood Priming" -SALINE_UNIT = "mL" -BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" +#pressure +PRESSURE_STEPS = 10 +ARTERIAL_PRESSURE_MINIMUM = -400 +ARTERIAL_PRESSURE_MAXIMUM = +600 +ARTERIAL_PRESSURE_LOW_MIN = -300 +ARTERIAL_PRESSURE_LOW_DEF = -300 +ARTERIAL_PRESSURE_LOW_MAX = +200 +ARTERIAL_PRESSURE_HIGH_MIN = -300 +ARTERIAL_PRESSURE_HIGH_DEF = +100 +ARTERIAL_PRESSURE_HIGH_MAX = +200 + +VENOUS_PRESSURE_MINIMUM = -100 +VENOUS_PRESSURE_MAXIMUM = +700 +VENOUS_PRESSURE_LOW_MIN = -100 +VENOUS_PRESSURE_LOW_DEF = -100 +VENOUS_PRESSURE_LOW_MAX = +600 +VENOUS_PRESSURE_HIGH_MIN = +100 +VENOUS_PRESSURE_HIGH_DEF = +400 +VENOUS_PRESSURE_HIGH_MAX = +600 + +#main_treatment_flows +BLOOD_FLOW_MINIMUM_VALUE = 100 +BLOOD_FLOW_BUFFER = 25 +FLOW_SLIDER_WIDTH_BUFFER = 37 +BLOOD_FLOW_MAXIMUM_VALUE = 500 +DIALYSATE_FLOW_MINIMUM_VALUE = 0 +DIALYSATE_FLOW_MAXIMUM_VALUE = 0 # Seems to be bug. need change +DIALYSATE_FLOW_BUFFER = 50 +DIALYSATE_FLOW_SLIDER_WIDTH_BUFFER = 60 + +#main treatment screen +OUT_OF_RANGE_COLOR = "#c53b33" +IN_RANGE_COLOR = "#fcfcfc" +SYSTOLIC_LOWER_LIMIT = 60 +SYSTOLIC_UPPER_LIMIT = 250 +DIASTOLIC_LOWER_LIMIT = 40 +DIASTOLIC_UPPER_LIMIT = 200 +HEART_RATE_LOWER_LIMIT = 40 +HEART_RATE_UPPER_LIMIT = 180 +BLOOD_PRIMING_RANGE = 300 +BLOOD_PRIMING_TARGET_MAXIMUM = 300 + +#tst_pretreatment_screens +NUM_OF_PRETREATMENT_BULLETS = len(PRE_TREATMENT_SCREENS) +SALINE_BOLUS_CUMULATIVE = 300 + +#patient id +KEYBOARD_SWITCH = ["ABC", "&123"] +PATIENT_ID_COMBINATION = { + "qwerty" : "qwerty", + "sampleid" : "sampleid", + "abcdefghijklmnop" : "abcdefghijklmnop", + "patientid" : "patientid", + "username" : "username", + "sA@MpLE#123" : "sAMpLE123", + "A@BH&321" : "ABH321", + "5432@qwerty" : "5432qwerty" +} + +#Random selection for operational parameter for custom dta +OPERATIONAL_PARAMETERS = {"Heparin Type": HEPARIN_TYPE, "Acid Concentrate": "08-1251-1", + "Bicarbonate Concentrate": BICARBONATE_CONCENTRATE, "Dialyzer Type": "BB Diacap Pro 16H"} + +#dictionary consist of custom treatment values +CREATE_CUSTOM_TREATMENT = { + "Blood Flow Rate": 250, + "Dialysate Flow Rate": 400, + "Duration": 270, + "Heparin Dispensing Rate": 0.5, + "Heparin Bolus Volume": 1, + "Heparin Stop Time": 100, + "Saline Bolus Volume": 200, + "Dialysate Temperature": 37, + "Blood Pressure Measurement Interval": 40, + "Rinseback Flow Rate": 100, + "Arterial Pressure Limit Low" : -30, + "Arterial Pressure Limit High" : 0, + "Blood Pressure Measure Interval" : 40, + "Rinseback Rate" : 100, + "Venous Pressure Limit High" : 590, + "Venous Pressure Limit Low": 560 +} + +#dictionary consist of slider values of treatment parameters +CREATE_TREATMENT_PARAMETER_RANGE = { + "Blood Flow Rate": [100, 125, 150, 175, 200, 225, 250, 275, 300, 325, 350, 375, 400, 425, 450, 475, 500], + "Dialysate Flow Rate": [100, 150, 200, 250, 300, 350, 400, 450, 500, 550, 600], + "Duration": [60, 75, 90, 105, 120, 135, 150, 165, 180, 195, 240, 255, 270, 285, 300, 315, 330, 345, 360, 375, 390, 405, 420, 435, 450, 465, 480], + "Heparin Dispensing Rate": [0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1, 0], + "Heparin Bolus Volume": [0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.1, 1.2, 1.3, 1.4, 1.5, 1.6, 1.7, 1.8, 1.9, 2, 0], + "Heparin Stop Time": [0, 10, 20, 30, 40, 50, 60, 70, 80, 90, 100, 110, 120, 130, 140, 150, 160, 170, 180, 190, 200, 210, 220, 230, 240, 250, 260, 270, 280, 290, 300, 310, 320, 330, 340, 350, 360, 370, 380, 390, 400, 410, 420, 430, 440, 450, 460, 470, 0], + "Saline Bolus Volume": [100, 200, 300], + "Dialysate Temperature": [35.0, 35.5, 36.0, 36.5, 37.0], + "Blood Pressure Measurement Interval": [0, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60, 0], + "Rinseback Flow Rate": [50, 75, 100, 125, 150], +} + +#dictionary consist of slider minimum value, slider buffer value, width buffer of slider +CREATE_TREATMENT_SLIDER_VALUES = { + + "blood_flow_rate" : [100, 25, 46], + "dialysate_flow_rate" : [100, 50, 75], + "duration" : [60, 15, 26.6], + "heparin_dispensing_rate" : [0, 0.1, 75], + "heparin_bolus_volume" : [0, 0.1, 37], + "heparin_stop_time" : [0, 10, 15.9], + "saline_bolus" : [100, 100, 375], + "dialysate_temperature" : [35, 0.5, 187.5], + "arterial_pressure_limit_high" : [20, 10, 11.931], + "blood_pressure_measurement_interval" : [0, 5, 62], + "rinseback_flow_rate" : [50, 25, 187], +} + + +#ultrafilteration volume based on flow parameters. Python list contains Rotational speed, Motor speed, +# MC speed, MC current and PWM. +TREATMENT_ULTRAFILTRATION_FLOW_PARAMETERS = { + 0 : [100, 0, 0, 0, 0, 0, 0], + 1 : [220, 1.45, 1.45, 1.45, 1.45, 1.45, 1.45], + } + +#dictionary contains uf maximum and minimum value's to set. +TREATMENT_ULTRAFILTRATION_TESTING_OPTION = { + 0 : {"uf_maximum" : 600, "uf_minimum" : 0}, + 1 : {"uf_maximum" : 480, "uf_minimum" : 250}, + } + +#tst_pretreatment_screens +NUM_OF_PRETREATMENT_BULLETS = len(PRE_TREATMENT_SCREENS) + +#dictionary contains saline cumulative and saline delivered value +#key : saline cumulative value. +#value: saline delivered value +SALINE_BOLUS_VALUES = { + 0 : 0, + 1.0 : 1.0, + 125.56 : 100.56, + 165.87 : 145.87, + 220.65 : 220.65, + 325.25 : 300.25, + 400.86 : 380.86, + 500.00 : 500.25, + 634.76 : 634.76, + 823.65 : 800.65, + 999.99 : 999.99, + 1000 : 1000 +} + +#testing options for main treatment saline +SALINE_BOLUS_TARGET = [0, 1, 150, 155, 299, 300] #Boundary Value Analysis + # pretreatment_patient_connection SYSTOLIC_PRESSSURE_120 = 120 SYSTOLIC_PRESSSURE_113 = 113 @@ -70,17 +333,11 @@ HEART_RATE_VAL_110 = 110 HEART_RATE_VAL_181 = 181 HEART_RATE_VAL_200 = 200 -SYSTOLIC_LOWER_LIMIT = 60 -SYSTOLIC_UPPER_LIMIT = 250 -DIASTOLIC_LOWER_LIMIT = 40 -DIASTOLIC_UPPER_LIMIT = 200 -HEART_RATE_LOWER_LIMIT = 40 -HEART_RATE_UPPER_LIMIT = 180 UF_MINIMUM_SLIDER_WIDTH = -2 UF_MAXIMUM_SLIDER_WIDTH = 638.00 UF_VALID_RANGE = 700 TRAINING_INDICATOR_SCREEN = 4 -ULTRAFILTRATION_VOLUME_COMBINATION_FROM_ULTRAFILTERATION_SETUP_SCREEN = { +ULTRAFILTRATION_VOLUME_COMBINATION_FROM_ULTRAFILTRATION_SETUP_SCREEN = { "OPTION_1" : {"uf_maximum" : 800, "uf_minimum" : 0}, "OPTION_2" : {"uf_maximum" : 500, "uf_minimum" : 0}, "OPTION_3" : {"uf_maximum" : 550, "uf_minimum" : 0}, @@ -95,6 +352,99 @@ "OPTION_12" : {"uf_maximum" : 1250, "uf_minimum" : 0}, } +#tst_pretreatment_screens color palettes +CURRENT_COLOR = '#000000' +COMPLETE_COLOR= '#4290ec' +ENABLED_COLOR = '#fcfcfc' +INCOMPLETE_COLOR = '#607a91' + + +#pre-treatment_priming +NOT_VISIBLE = False +VISIBLE = True +FOCUS = True +NOT_IN_FOCUS = False +TIME_OUT_DURATION_300 = 300 +TIME_OUT_DURATION_200 = 200 +TIME_OUT_DURATION_100 = 100 +COUNT_DOWN_TIME_100 = 100 +MINIMUM_COUNTDOWN_TIME = 0 +ENABLED = True +DISABLED = False +CONSUMABLE_SELF_TEST_BICARB_PUMP_CHECK_STATE = 3 + + +#pre-treatment_priming +NOT_VISIBLE = False +VISIBLE = True +FOCUS = True +NOT_IN_FOCUS = False + +#Filter Flush + +MAXIMUM_COUNTDOWN_TIME = 300 +MINIMUM_COUNTDOWN_TIME = 0 +TIME_OUT_DURATION_300 = 300 +TIME_OUT_DURATION_200 = 200 +TIME_OUT_DURATION_100 = 100 +COUNT_DOWN_TIME_100 = 100 +MINIMUM_COUNTDOWN_TIME = 0 +CONSUMABLE_SELF_TEST_BICARB_PUMP_CHECK_STATE = 3 + +VALID = True +INVALID = False + +#post treatment review +ACID_CONCENTRATE_POST = ["Fres. Naturalyte", "08-1251-1", "08-2251-0", "08-3251-9", "08-3251-9"] +BRIGHTNESS_MIN_VAL = 20 +BRIGHTNESS_MAX_VAL = 100 +VOLUME_MIN_VAL = 20 +VOLUME_MAX_VAL = 100 + +#dictionary consist of review parameters values of post treatment +POST_TREATMENT_REVIEW_PARAMETER_RANGE = { + "Patient ID": ["abcd", "xyz"], + "Heparin Stop": [0, 101], + "Device ID": [1234, 1234], + "Heparin Delivered Volume": [10.1, 10.5], + "Blood Flow Rate": [100, 125], + "Treatment Start DateTime": [1654519251, 1623382321], + "Dialysate Flow Rate": [100, 150], + "Treatment End DateTime": [1654519251, 1623382321], + "Treatment Duration": [70, 75], + "Water Sample Test Result": [0, 1], + "Actual Treatment Duration": [62, 126], + "Dialysate Volume Used": [4.0, 6.0], + "Acid ConcentrateType": [0, 1], + "Prescribed UF Volume": [0.3, 0.5], + "Bicarbonate Concentrate Type": [0, 0], + "Target UF Volume": [10.1, 10.5], + "Potassium Concentration": [60, 75], + "Actual UF Volume": [0.3, 0.5], + "Calcium Concentration": [3, 56], + "Prescribed UF Rate": [0.1, 0.55], + "Bicarbonate Concentration": [14, 50], + "Target UF Rate": [0.2, 0.56], + "Sodium Concentration": [4, 6], + "Actual UF Rate": [0.5, 0.6], + "Dialysate Temperature": [35.0, 35.5], + "Saline Bolus Volume": [300, 250], + "Dialyzer Type": [0, 1], + "Average Blood Flow": [54.3, 67.3], + "Heparin Type": [0, 0], + "Average Dialysate Flow": [0.6, 0.5], + "Average Dialysate Temp": [35.0, 35.5], + "Heparin Bolus Volume": [0.0, 0.1], + "Average Arterial Pressure": [0.0, 11.56], + "Heparin Dispense Rate": [0.0, 0.1], + "Average Venous Pressure": [20.0, 101.45], +} + + + +#standby mode +GOODMORNING_START_TIME_SEC = 0 +GOODEVENING_START_TIME_SEC = 43200 NUM_OF_REQUEST_REJECT_REASONS = 46 #Number of settings change reject codes Index: shared/scripts/configuration/strings.py =================================================================== diff -u -r9e8d0a1f125fa5b8fa103adc30489bd55893ad47 -r11b2970e98eb80e8bd299d25b607b8ff00a2043b --- shared/scripts/configuration/strings.py (.../strings.py) (revision 9e8d0a1f125fa5b8fa103adc30489bd55893ad47) +++ shared/scripts/configuration/strings.py (.../strings.py) (revision 11b2970e98eb80e8bd299d25b607b8ff00a2043b) @@ -1,4 +1,6 @@ +# -*- coding: utf-8 -*- + ########################################################################### # # Copyright (c) 2019-2021 Diality Inc. - All Rights Reserved. @@ -8,20 +10,32 @@ # # @file strings.py # -# @author (last) LTTS -# @date (last) 18-Jan-2022 -# +# @author (last) LTTS +# @date (last) 14-June-2022 ############################################################################ -#standby mode -BLOOD_PRIMING_TEXT = "Blood Priming" +#pretreatment_screen_bullets +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] +POST_TREATMENT_SCREENS = ["Disconnection", "Review", "Disposables", "Disinfection"] + + SALINE_UNIT = "mL" BLOOD_PRIMING_UNIT = "mL" BLOOD_PRIMING_VALUE_0 = "0" BLOOD_PRIMING_DEFAULT_VALUE = BLOOD_PRIMING_VALUE_0 + " " + BLOOD_PRIMING_UNIT +#main_treatment_ultrafilteration +ULTRAFILTRATION_PAUSED_TEXT = "Ultrafiltration Paused" +INCREASE_UF_RATE_TEXT = "Increase UF Rate" +INCREASE_TREATMENT_DURATION_TEXT = "Increase Treatment Duration" +RESUME_TEXT = "RESUME ULTRAFILTRATION" +EDIT_TEXT = "EDIT ULTRAFILTRATION VOLUME" +TITLE_TEXT = "ULTRAFILTRATION VOLUME (L)" +#recirculate screens +RECIRCULATE_TEXT = "Recirculate" +RECONNECT_BUTTON_TEXT = "RECONNECT" # pretreatment_patient_connection DIASTOLIC_TEXT = "diastolic" SYSTOLIC_TEXT = "systolic" @@ -43,8 +57,59 @@ 3 : "Connect arterial and venous cartridge lines to your patient access lines." } -PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] +#rinseback-setup +RINSE_BACK_SETUP_TEXT = "Rinseback Setup" +RINSEBACK_BUTTON_TEXT = "START RINSEBACK" +BACK_TO_TREATMENT_BUTTON_TEXT = "BACK TO TREATMENT" +END_IMAGE_BUTTON_TEXT = "End" +DECELERATE_BUTTON_TEXT = "Decelerate" +ACCELERATE_BUTTON_TEXT = "Accelerate" +PAUSE_BUTTON_TEXT = "Pause" +RESUME_BUTTON_TEXT = "Resume" +RINSEBACK_TEXT = "Rinseback" +RINSEBACK_COMPLETE_TEXT = "Rinseback Complete" +END_TREATMENT_BUTTON_TEXT = "END TREATMENT" +BACK_TO_RECIRCULATE_TEXT = "BACK TO RECIRCULATE" +RECONNECT_TEXT = "Reconnect" +#Blood Priming +BLOOD_PRIMING_TEXT = "Blood Priming" + +#tst_treatment_saline_data +SALINE_UNIT = "mL" +ACK_REQ_STATUS = 'Ack Req' +ACK_BAK_STATUS = 'Ack Bak' + +#standby mode +LIQUID_UNIT = "mL" + + +#main treatment pressure +PRESSURE_TITLE = "PRESSURES" +ARTERIAL_TITLE = "Arterial" +VENOUS_TITLE = "Venous" +VENOUS_UOM = "mmHg" +ARTERIAL_UOM = "mmHg" +LOW_TEXT = "LOW" +HIGH_TEXT = "HIGH" + +#main-treatment heparin +HEPARIN_TEXT = "HEPARIN" +HEPARIN_UNIT = "mL" +HEPARIN_DELIVERY_TEXT = "HEPARIN DELIVERY" +HEPARIN_DELIVERY_OFF_TEXT = "OFF" +HEPARIN_STOP_TEXT = "Delivery Stopped" +HEPARIN_RESUME_DELIVERY_TEXT = "RESUME DELIVERY" +HEPARIN_BOLUS_ACTIVE_TEXT = "Bolus Active" +HEPARIN_BOLUS_DISPENSING_TEXT = "Dispensing Active" +HEPARIN_PAUSE_DELIVERY_TEXT = "PAUSE DELIVERY" +HEPARIN_MAXIMUM_VOLUME_DELIVERED_TEXT = "Maximum Cumulative Heparin Volume Delivered" +HEPARIN_SYRINGE_EMPTY_TEXT = "Syringe Empty" +HEPARIN_DISPENSING_RATE = "Heparin Dispensing Rate" +HEPARIN_BOLUS_VOLUME = "Heparin Bolus Volume" +HEPARIN_STOP_TIME = "Heparin Stop Time" +HEPARIN_TYPE_TITLE = "Heparin Type" + # dictionary of rejection reason from application source code. # message location -> denali-> Headers-> common-> MsgDefs.h REJECTION_REASON = { @@ -96,4 +161,596 @@ 45: "REQUEST_REJECT_REASON_DG_CONCENTRATE_CAP_OPEN" # Concentrate cap open } +#Alarm List +UNDEFINED_ALARM_ID_MSG = "ALARM_ID_UNDEFINED" +ALARM_TITLE = "Alarm" +ALARM_LIST_TITLE = "Alarm list" +ALARM_PRIORITIES = {1: "LOW", 2: "MEDIUM", 3: "HIGH"} +ALARM_PRIORITY_OPTIONS = { + 0 : 'ALARM_PRIORITY_NONE' , + 1 : 'ALARM_PRIORITY_LOW' , + 2 : 'ALARM_PRIORITY_MEDIUM', + 3 : 'ALARM_PRIORITY_HIGH', +} + +#Setting +EXPORT_TEXT ="Export" +DEVICE_SETTINGS_TEXT = "Device Settings" +DEVICE_SETTINGS_SCREEN_PARAMETER = ["Information", "Volume And Brightness", "Wi-Fi", "Bluetooth Cuff", "Dialysate Generator Settings", "Services"] + +#UF treatment +LIQUID_UNIT = "mL" + +#create custom treatment +TREATMENT_TITLE = "Create a Custom Treatment" +CONFIRM_TREATMENT_TITLE = "Confirm Treatment" +PRESCRIPTION_TITLE = "PRESCRIPTION" +OPERATING_PARAMETERS_TITLE = "OPERATING PARAMETERS" +PATIENT_ID_TEXT = "Patient ID" +BLOOD_FLOW_RATE = "Blood Flow Rate" +DIALYSATE_FLOW_RATE = "Dialysate Flow Rate" +DURATION = "Duration" +SALINE_BOLUS_VOLUME = "Saline Bolus Volume" + +#main treatment screen +VITALS_TITLE = "VITALS" +HEART_RATE_TITLE = "Heart Rate" +BLOOD_PRESSURE_TITLE = "Blood Pressure" +BLOOD_PRESSURE_UNIT = "mmHg" +HEART_RATE_UNIT = "BPM" + +PATIENT_ID = "demopatientid" +SALINE_BOLUS_TITLE = "Saline Bolus" +ACID_CONCENTRATE_TITLE = "Acid Concentrate" +BICARBONATE_CONCENTRATE_TITLE = "Bicarbonate Concentrate" +DIALYZER_TYPE_TITLE = "Dialyzer Type" +DIALYSATE_TEMPERATURE = "Dialysate Temperature" +BLOOD_PRESSURE_MEASUREMENT_INTERVAL = "Blood Pressure Measurement Interval" +RINSEBACK_FLOW_RATE = "Rinseback Flow Rate" + + +CREATE_TREATMENT_PARAMETERS = ["Blood Flow Rate", "Dialysate Flow Rate", "Duration","Heparin Dispensing Rate", + "Heparin Bolus Volume","Heparin Stop Time", "Saline Bolus", "Dialysate Temperature", + "Blood Pressure Measurement Interval", "Rinseback Flow Rate"] + +OFF_TEXT = "OFF" +BACK_TEXT = "BACK" +CONTINUE_TEXT = "CONTINUE" +CREATE_TREATMENT_PARAMETERS = ["Blood Flow Rate", "Dialysate Flow Rate", "Duration","Heparin Dispensing Rate", + "Heparin Bolus Volume","Heparin Stop Time", "Saline Bolus", "Dialysate Temperature", + "Blood Pressure Measurement Interval", "Rinseback Flow Rate"] + + +PRESCRIPTION_DETAILS = { + "Blood Flow Rate": "mL/min", "Dialysate Flow Rate": "mL/min", + "Duration": "min", "Heparin Dispensing Rate": "mL/hr", + "Heparin Bolus Volume": "mL", "Heparin Stop Time": "min", + "Saline Bolus Volume": "mL", + "Dialysate Temperature": "°C", "Arterial Pressure Limit Low": "mmHg", + "Arterial Pressure Limit High": "mmHg","Blood Pressure Measure Interval":"min", + "Rinseback Rate" : "mL/min", "Venous Pressure Limit High": "mmHg", + "Venous Pressure Limit Low": "mmHg" +} + +#Time_Duration +TREATMENT_PAUSED_TEXT = "Treatment Paused" + +#pre-treatment +CONSUMABLE_SELF_TEST_TEXT = "Consumables Self Test" +BICARB_PUMP_CHECK_TEXT = "BiCarb Pump Check" +ACID_PUMP_CHECK_TEXT = "Acid Pump Check" +SELF_TEST_COMPLETE_TEXT = "Self Test Complete!" +BUSY = "Busy" +CHECK = "Check" +BEGIN_PRIME_TITLE = "Begin Prime" +PRIMING_TITLE = "Priming" +SYSTEM_SELF_TEST_TITLE = "System Self Test" +FILTER_FLUSH_TITLE = "Filter Flush" + +#Pre-Treatment Disposables +NEXT_BUTTON_TEXT = "NEXT" +BACK_BUTTON_TEXT = "BACK" +CONFIRM_BUTTON_TEXT = "CONFIRM" + +#Pre_treatment_water_sample +SAMPLE_TEXT = "Sample" +CREATE_TEXT = "Create" +WATER_SAMPLE_BUTTON_TEXT = "WATER SAMPLE" +PASS_TEXT = "PASS" +FAIL_TEXT = "FAIL" +OK_TEXT = "OK" + +#Vitals & pretreatment_patient_connection +SYSTOLIC_TEXT = "systolic" +DIASTOLIC_TEXT = "diastolic" + +#priming +TEST_DRY_STEPS = ["Used Cartridge Check", "Occlusion Sensors Check", "Pressure Sensors Check", "Syringe Pump Check"] +PRIMING_STEPS = ["Blood Circuit Priming", "Dialysate Circuit Priming", "Wet Self Tests"] +SELF_TEST_COMPLETION_MSG = "Self Test Complete!" +PRIMING_COMPLETE_MSG = "Priming Complete!" + +#main_treatment_flows +FLOW_UNIT = "mL/min" +BLOOD_FLOW_TEXT = "blood" +DIALYSATE_FLOW_TEXT = "dialysate" + +LIQUID_UNIT = "mL" + +#tst_treatment_monitoring +FLOWS_TEXT = "FLOWS" +BLOOD_UNIT = "mL/min" +DIALYSATE_TEXT = "Dialysate" +BLOOD_TEXT = "Blood" +VITALS_TEXT = "VITALS" +BLOOD_PRESSURE_TEXT = "mmHg" +HEARTRATE_UNIT = "BPM" +PRESSURE_TEXT = "PRESSURE (mmHg)" +ARTEREAL_TEXT = "Arterial" +VENOUS_TEXT = "Venous" +TIME_DURATION_TEXT = "Time Remaining" +ULTRAFILTERATION_TEXT = "ULTRAFILTRATION VOLUME (L)" +SALINE_BOLUS = "SALINE BOLUS" +VOLUME_DELIVERED = "Volume Delivered" +CUMULATIVE_DELIVERED = "Cumulative Delivered" + +#post-treatment +PATIENT_DISCONNECTION_TEXT = "Patient Disconnection" +REVIEW_TEXT = "Review" +PATIENT_DISCONNECTION_CONFIRM_BUTTON_TEXT = "CONFIRM" +DISPOSABLE_TEXT = "Disposables" +NEXT_TEXT = "NEXT" +BACK_TEXT = "BACK" +CONFIRM_TEXT = "CONFIRM" +DISPOSABLES_REMOVAL_TEXT = "Disposables Removal" + +#post treatment review +TREATMENT_REVIEW_TITLE_TEXT = "Treatment Review" +POST_TREATMENT_REVIEW_SCREEN_UNITS = { + "Patient ID" : "", + "Heparin Stop" : "min", + "Device ID" : "", + "Heparin Delivered Volume": "mL", + "Blood Flow Rate" : "mL/min", + "Treatment Start DateTime" : "", + "Dialysate Flow Rate" : "mL/min", + "Treatment End DateTime" : "", + "Treatment Duration" : "min", + "Water Sample Test Result": "", + "Actual Treatment Duration":"min", + "Dialysate Volume Used" : "L", + "Acid ConcentrateType" : "", + "Prescribed UF Volume" : "L", + "Bicarbonate Concentrate Type" : "", + "Target UF Volume" : "L", + "Potassium Concentration" : "mEg/L", + "Actual UF Volume" : "L", + "Calcium Concentration" :"mEg/L", + "Prescribed UF Rate" : "mL/min", + "Bicarbonate Concentration" : "mEg/L", + "Target UF Rate": "mL/min", + "Sodium Concentration" : "mEg/L", + "Actual UF Rate": "mL/min", + "Dialysate Temperature" :"°C", + "Saline Bolus Volume" : "mL", + "Dialyzer Type" : "", + "Average Blood Flow" : "mL/min", + "Heparin Type" : "IU/mL", + "Average Dialysate Flow" : "mL/min", + "Heparin Concentration" : "IU/mL", + "Average Dialysate Temp" : "°C", + "Heparin Bolus Volume": "mL", + "Average Arterial Pressure" : "mmHg", + "Heparin Dispense Rate" : "mL/hr", + "Average Venous Pressure" : "mmHg", + } + +#Information +INFORMATION_TITLES = ["Information", "Versions"] +INFORMATION_PARAMETERS = ["Information", "UI Version", "HD Version", "HD FPGA Version", "HD Serial Number", "DG Version", "DG FPGA Version", "DG Serial Number"] +SERIVCES_TITLE = "Services" +SERVICES_PARAMETERS = ["HD Last Service Date", "HD Next Service Date", "DG Last Service Date", "DG Next Service Date"] + + +#volume and brightness +VOLUME_AND_BRIGHTNESS_TEXT = "Volume And Brightness" +BRIGHTNESS_TEXT = "Brightness" +ALARM_VOLUME_TEXT = "Alarm Volume" +UNIT_OF_VOLUME_AND_BRIGHTNESS = "%" + +#services +SHUTDOWN_TEXT = "SHUTDOWN" + +# pretreatment_patient_connection & ui logs +ACK_REQ_STATUS = 'Ack Req' +ACK_BAK_STATUS = 'Ack Bak' +PATIENT_CONNECTION_TEXT = "Patient Connection" +UF_VOLUME_TEXT = "Ultrafiltration Volume (L)" +UF_TITLE_TEXT = "Ultrafiltration Setup" +BP_HR_TEXT = "BP/HR" +SKIP_TEXT = "SKIP" +TUTORIAL_TEXT = { + 0 : "Wash your hands and establish vascular access per your clinic's instructions.", + 1 : "Pinch arterial clamp and venous clamp on located on the cartridge patient lines.", + 2 : "Remove arterial and venous shunt.", + 3 : "Connect arterial and venous cartridge lines to your patient access lines." +} + +#End treatment +RINSEBACK_BUTTON_TEXT = "START RINSEBACK" +END_TREATMENT_COMPLETE_STATE_TEXT = "Treatment Complete" +END_TREATMENT_COMPLETE_PAUSED_STATE_TEXT = "Treatment Complete Paused" + +#rinseback-setup +RINSE_BACK_SETUP_TEXT = "Rinseback Setup" +RINSEBACK_BUTTON_TEXT = "START RINSEBACK" +END_IMAGE_BUTTON_TEXT = "End" +DECELERATE_BUTTON_TEXT = "Decelerate" +ACCELERATE_BUTTON_TEXT = "Accelerate" +PAUSE_BUTTON_TEXT = "Pause" +RESUME_BUTTON_TEXT = "Resume" +RINSEBACK_TEXT = "Rinseback" +RINSEBACK_COMPLETE_TEXT = "Rinseback Complete" +END_TREATMENT_BUTTON_TEXT = "END TREATMENT" +ADDITIONAL_BUTTON_TEXT = "ADDITIONAL" +ADDITIONAL_RINSEBACK_TEXT = "Additional Rinseback" +RINSEBACK_UNIT = " mL" +RINSEBACK_UNIT_MIN = " mL/min" + + + +ALARM_ID_AND_MESSAGES = { + 1: "Stuck button POST failure.\nStop or Off button detected to be pressed for at least 1 second during test shortly after power up.", + 2: "HD FPGA POST failure.", + 3: "DG FPGA POST failure.", + 4: "HD Watchdog POST failure.", + 5: "DG Watchdog POST failure.", + 6: "UI communication POST failure.\nUI failed to communicate within a reasonable time after power up.", + 7: "Real-time clock configuration error.", + 8: "HD accelerometer failed POST.", + 9: "DG accelerometer failed POST." , + 10: "RTC or MCU timer inaccurate.", + 11: "DG heaters failed POST.", + 12: "HD firmware image integrity POST test failed.", + 13: "DG firmware image integrity POST test failed.", + 14: "HD blood flow invalid calibration.", + 15: "HD dialysate flow invalid calibration.", + 16: "HD alarm audio failed POST.", + 17: "HD UI POST failed.", + 18: "DG load cells invalid calibration record.", + 19: "DG flow sensors invalid calibration record.", + 20: "DG conductivity sensors invalid calibration record.", + 21: "DG drain line volume invalid calibration record.", + 22: "DG reservoirs invalid calibration record.", + 23: "DG acid concentrate calibration record.", + 24: "DG bicarb concentrate calibration record.", + 25: "DG accelerometer invalid calibration record.", + 26: "HD accelerometer invalid calibration record.", + 27: "HD blood flow invalid calibration record.", + 28: "HD dialyzer flow invalid calibration record.", + 29: "HD heparin force sensor invalid calibration record.", + 30: "HD Software fault.\nSoftware found itself in an unexpected state.", + 31: "Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 32: "Blood pump failed motor off check.\nMeasured speed while commanded off.", + 33: "Blood pump failed motor direction check.\nMeasured vs commanded.", + 34: "Blood pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 35: "Dialysis inlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 36: "Dialysis inlet pump failed motor off check.\nMeasured speed while commanded off.", + 37: "Dialysis inlet pump failed motor direction check.\nMeasured vs commanded.", + 38: "Dialysis inlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 39: "Dialysis outlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 40: "Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off.", + 41: "Dialysis outlet pump failed motor direction check.\nMeasured vs commanded.", + 42: "Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 43: "Watchdog expired error.\nWatchdog was not \"pet\" within the appropriate time.", + 44: "User interface communication timeout.", + 45: "Too many bad CRCs detected on received system messages.", + 46: "System message that required acknowledgment was not acknowledged.", + 47: "Ultrafiltration rate is too high error during treatment.", + 48: "Ultrafiltration volume accuracy error during treatment.", + 49: "HD FPGA communication down for too long.", + 50: "FPGA not accepting commanded valve states.", + 51: "Blood pump failed motor speed check.\nMeasured vs commanded.", + 52: "Dialysate inlet pump failed motor speed check.\nMeasured vs commanded.", + 53: "Dialysate outlet pump failed motor speed check.\nMeasured vs commanded.", + 54: "HD critical data integrity check failed.", + 55: "DG critical data integrity check failed.", + 56: "HD accelerometer error (no readings or FPGA reports error).", + 57: "DG accelerometer error (no readings or FPGA reports error).", + 58: "HD valve homing failed.", + 59: "HD valve transition time out.", + 60: "HD valve not functional.", + 61: "HD valve current out of range.", + 62: "HD valve position out of target.", + 63: "FPGA reports failure reading arterial pressure sensor.", + 64: "FPGA reports failure reading venous pressure sensor.", + 65: "HD requests DG command with invalid parameter fault.", + 66: "HD blood leak sensor set point set failure.", + 67: "HD blood pump occlusion self-test failure alarm.", + 68: "HD active reservoir recirculation out of range.", + 69: "HD blood leak sensor invalid calibration record.", + 70: "HD arterial pressure self-test failure alarm." , + 71: "HD venous pressure self-test failure alarm.", + 72: "HD blood flow meter status check self-test failure alarm.", + 73: "HD dialysate flow meter status check self-test failure alarm.", + 74: "HD blood leak detector self-test failure alarm.", + 75: "HD syringe pump self-test failure alarm.", + 76: "HD monitored voltage is out of range.", + 77: "DG monitored voltage is out of range.", + 78: "HD syringe pump direction (from encoder) error.", + 79: "HD syringe pump direction (from controller) error.", + 80: "HD syringe pump fault reported by FPGA.", + 81: "HD syringe pump over travel error.", + 82: "HD syringe pump DAC write failure.", + 83: "HD syringe pump is running while the BP is off.", + 84: "HD venous pressure sensor not being read.", + 85: "HD pump track latch opened alarm.", + 86: "HD BP occlusion sensor not being read.", + 87: "DG heating invalid calibration record.", + 88: "DG concentrate pumps hall sensor out of range.", + 89: "HD BP occlusion sensor error.", + 90: "This alarm ID is available for use.", + 91: "HD No dialysate flow data receive in the last 3 seconds.", + 92: "DG conductivity sensor bad status.", + 93: "This alarm ID is available for use.", + 94: "This alarm ID is available for use.", + 95: "This alarm ID is available for use.", + 96: "Inlet water temperature in high range.", + 97: "Fill conductivity out of range.", + 98: "HD battery communication fault.", + 99: "HD syringe pump stall alarm.", + 100: "HD no cartridge self-test timeout.", + 101: "HD dry self-test timeout.", + 102: "Real-time clock communication error.", + 103: "HD non-volatile calibration group invalid record CRC.", + 104: "Air trap level sensors reporting illegal combination of air/fluid.", + 105: "DG non-volatile system group invalid record CRC.", + 106: "DG inlet water conductivity is greater than threshold.", + 107: "HD reports DG restarted fault.", + 108: "HD syringe pump ADC error.", + 109: "HD syringe pump volume check error.", + 110: "HD syringe pump speed check error.", + 111: "HD syringe pump not stopped in off state error.", + 112: "HD blood leak detector fault.", + 113: "HD arterial air bubble detector self-test failure.", + 114: "HD venous air bubble detector self-test failure.", + 115: "DG temperature sensor out of range.", + 116: "DG temperature sensor ADC out of range.", + 117: "DG primary heater internal temperature out of range.", + 118: "DG primary heater cold junction temperature out of range.", + 119: "DG trimmer heater internal temperature out of range.", + 120: "DG trimmer heater cold junction temperature out of range.", + 121: "DG main primary heater voltage out of range.", + 122: "DG small primary heater voltage out of range.", + 123: "DG trimmer heater voltage out of range.", + 124: "HD end of treatment alarm (high priority).", + 125: "Blood sitting too long after treatment stopped by user alarm (>5 min).", + 126: "HD blood leak detected alarm.", + 127: "Venous pressure too low during treatment." , + 128: "HD venous air bubble detected alarm.", + 129: "HD venous air bubble detected rinseback alarm.", + 130: "Venous pressure too high during treatment.", + 131: "Arterial pressure too low during treatment.", + 132: "Arterial pressure too high during treatment.", + 133: "DG fluid leak detected alarm.", + 134: "HD fluid leak detected alarm." , + 135: "HD experienced large acceleration.", + 136: "DG experienced large acceleration.", + 137: "HD tilt exceeds maximum.", + 138: "DG tilt exceeds maximum.", + 139: "HD AC power lost alarm.", + 140: "Dialysate generator communication timeout.", + 141: "Air trap fill timeout during treatment.", + 142: "Blood pump occlusion detected.", + 143: "DG dialysate temperature sensors drift timeout.", + 144: "DG software configuration record invalid CRC.", + 145: "Concentrate conductivity after adding acid out of range alarm.", + 146: "Dialysate conductivity out of range.", + 147: "Wait for the DG to produce dialysate.", + 148: "Inlet water temperature too high.", + 149: "Alarm ID available for use.", + 150: "Inlet water conductivity in high range.", + 151: "Inlet water conductivity in low range.", + 152: "Inlet water pressure in low range.", + 153: "HD prime completed high priority alarm.", + 154: "EEPROM operations (read, write, erase) failure.", + 155: "DG software configuration record invalid CRC.", + 156: "HW usage data (treatment time in HD and total consumed water in DG) failure.", + 157: "DG chemical disinfect prime acid line timeout.", + 158: "Inlet water temperature is in the low range.", + 159: "Dialysate inlet pump failed flow vs motor speed check.", + 160: "Blood pump rotor speed too high.", + 161: "This alarm ID is available for use.", + 162: "This alarm ID is available for use.", + 163: "HD sees primary load cell for reservoir 1 change too much too fast.", + 164: "HD sees primary load cell for reservoir 2 change too much too fast.", + 165: "HD in treatment rinseback operation timeout.", + 166: "HD in treatment re-circ sub-mode for too long.", + 167: "HD cartridge door opened alarm.", + 168: "HD active reservoir depletion timeout.", + 169: "Dialysate flow rate out of range.", + 170: "HD syringe empty alarm.", + 171: "HD syringe pump occlusion alarm.", + 172: "HD syringe pump not enough Heparin alarm.", + 173: "HD arterial air bubble detected alarm.", + 174: "HD arterial air bubble detected rinseback alarm.", + 175: "HD pump direction status error.", + 176: "HD active reservoir is full during treatment and DG not ready to switch reservoirs.", + 177: "DG software fault.\nSoftware found itself in an unexpected state.", + 178: "HD communication timeout.", + 179: "DG FPGA communication down for too long.", + 180: "DG load cell ADC error.", + 181: "DG load cells weight out of range for tare.", + 182: "DG load cells invalid calibration.", + 183: "DG invalid load cell value.", + 184: "Inlet or Outlet UV reactor not healthy.", + 185: "DG fan RPM out of range.", + 186: "DG concentrate pump fault.", + 187: "DG concentrate pump CP1 speed control error.", + 188: "DG concentrate pump CP2 speed control error.", + 189: "DG drain pump RPM out of range.", + 190: "DG drain pump off fault.", + 191: "DG flow rate out of upper range.", + 192: "DG flow rate out of lower range.", + 193: "RO pump flow rate out of range.", + 194: "DG RO pump duty cycle out of range.", + 195: "DG RO pump pressure out of range.", + 196: "DG temperature sensor error flag fault.", + 197: "DG temperature sensors ADC fault.", + 198: "DG heaters cold junction temperature out of range.", + 199: "DG heaters fault.", + 200: "DG thermistors/sensors temperature out of range.", + 201: "Inlet water quality is bad (temperature, conductivity, pressure).", + 202: "Inlet water pressure fault.", + 203: "Pressure sensor fault.", + 204: "RO rejection ratio out of range.", + 205: "Conductivity sensor fault.", + 206: "DG dialysate fill runs out of time.", + 207: "DG flow meter check failure alarm.", + 208: "DG primary heater on with no flow time out.", + 209: "DG drain circulation line timeout.", + 210: "HD battery pack detected an error.", + 211: "Blood sitting too long warning (>4 min).", + 212: "HD end of treatment alarm (user not acting to end treatment).", + 213: "HD prime completed medium priority alarm.", + 214: "HD Saline bolus volume check failure.", + 215: "HD rinseback volume check failure.", + 216: "HD end treatment sub-mode timeout alarm.", + 217: "HD blood prime volume check failure.", + 218: "HD syringe detected alarm.", + 219: "HD syringe pump syringe removed alarm.", + 220: "HD battery pack SOC is too low.", + 221: "HD Saline bag is empty.", + 222: "HD dialysate temperature too high alarm.", + 223: "HD dialysate temperature too low alarm.", + 224: "Treatment stopped by user action.", + 225: "HD end of treatment warning.", + 226: "HD prime completed low priority alarm.", + 227: "HD prime out of time alarm.", + 228: "HD time out on prime sub-mode purge air state.", + 229: "HD prime dialysate dialyzer time out alarm.", + 230: "HD prime dialysate bypass time out alarm.", + 231: "HD pre-treatment mode dry self-tests failure.", + 232: "HD pre-treatment mode wet self-tests failure.", + 233: "RTC battery low.", + 234: "RTC (i.e read time) or RAM operations failure (read or write).", + 235: "HD in treatment stopped sub-mode after rinseback completed (no escalation).", + 236: "HD needs new cartridge to be installed.", + 237: "HD prime saline dialyzer time out alarm.", + 238: "HD no cartridge loaded or installed improperly alarm.", + 239: "HD fail to remove cartridge alarm.", + 240: "Bicarb conductivity out of range during bicarb pump check alarm.", + 241: "DG reservoir drain time out.", + 242: "DG reservoir fill time out.", + 243: "DG reservoir leak time out.", + 244: "DG temperature sensors difference out of range.", + 245: "DG heat disinfect target temperature time out (could not reach to temperature).", + 246: "DG heat disinfect inlet pressure and temperature sensors out of range.", + 247: "DG heat disinfect inlet conductivity and temperature out of range.", + 248: "DG chemical disinfect could not reach to target temperature.", + 249: "DG chemical disinfect inlet pressure and temperature sensors out of range.", + 250: "DG chemical disinfect inlet conductivity and temperature out of range.", + 251: "HD invalid system record.", + 252: "HD invalid service record.", + 253: "DG invalid system record.", + 254: "DG invalid service record.", + 255: "HD and UI software builds are not compatible.", + 256: "HD and DG software builds are not compatible.", + 257: "DG FPGA power out timeout.", + 258: "HD FPGA power out timeout.", + 259: "HD temperatures out of range.", + 260: "UI POST Application Integrity (Sha256Sum) failure.", + 261: "UI POST CANBus failure.", + 262: "UI POST Display failure.", + 263: "UI POST Touch failure.", + 264: "UI POST SD-Card failure.", + 265: "UI POST RTC failure.", + 266: "UI POST WiFi failure.", + 267: "UI POST Bluetooth failure.", + 268: "UI POST Ethernet failure.", + 269: "UI POST Sound failure.", + 270: "HD POST Safety Shutdown failure.", + 271: "DG POST Safety Shutdown failure.", + 272: "HD Fan RPM out of range.", + 273: "HD measured blood pump flow rate is out of range.", + 274: "HD measured dialysate inlet pump flow rate is out of range.", + 275: "HD arterial pressure sensor is reading out of range.", + 276: "HD venous pressure sensor is reading out of range.", + 277: "HD BP occlusion sensor is reading out of range.", + 278: "HD active reservoir weight out of range.", + 279: "DG dialysate drain time out.", + 280: "HD arterial pressure sensor read timeout error.", + 281: "Acid concentration bottle low volume alarm.", + 282: "Bicarbonate concentration bottle low volume alarm.", + 283: "DG load cells weight out of range.", + 284: "DG load cells primary/back up drift out of range.", + 285: "HD treatment recirculate timeout warning.", + 286: "HD treatment rinseback complete timeout warning.", + 287: "HD processor clock speed checks against FPGA clock failure.", + 288: "DG trimmer heater on with flow timeout.", + 289: "DG dialysate or concentrate caps not closed." + } + +ALARM_PRIORITIES = {1: "LOW", 2: "MEDIUM", 3: "HIGH"} +ALARM_LIST_TITLE = "Alarm list" +UNDEFINED_ALARM_ID_MSG = "ALARM_ID_UNDEFINED" +ALARM_TITLE = "Alarm" + +ALARM_PRIORITY_OPTIONS = { + 0 : 'ALARM_PRIORITY_NONE' , + 1 : 'ALARM_PRIORITY_LOW' , + 2 : 'ALARM_PRIORITY_MEDIUM', + 3 : 'ALARM_PRIORITY_HIGH', +} + + +PRESSURE_TITLE = "PRESSURES" +ARTERIAL_TITLE = "Arterial" +VENOUS_TITLE = "Venous" +VENOUS_UOM = "mmHg" +ARTERIAL_UOM = "mmHg" +LOW_TEXT = "LOW" +HIGH_TEXT = "HIGH" + +FLOW_UNIT = "mL/min" +BLOOD_FLOW_TEXT = "blood" +DIALYSATE_FLOW_TEXT = "dialysate" + +VITALS_TITLE = "VITALS" +HEART_RATE_TITLE = "Heart Rate" +BLOOD_PRESSURE_TITLE = "Blood Pressure" +BLOOD_PRESSURE_UNIT = "mmHg" +HEART_RATE_UNIT = "BPM" + +HEPARIN_TEXT = "HEPARIN" +VOLUME_DELIVERED = "Volume Delivered" + + +#main-treatment heparin +HEPARIN_UNIT = "mL" +HEPARIN_DELIVERY_TEXT = "HEPARIN DELIVERY" +HEPARIN_DELIVERY_OFF_TEXT = "OFF" +HEPARIN_STOP_TEXT = "Delivery Stopped" +HEPARIN_RESUME_DELIVERY_TEXT = "RESUME DELIVERY" +HEPARIN_BOLUS_ACTIVE_TEXT = "Bolus Active" +HEPARIN_PAUSE_DELIVERY_TEXT = "PAUSE DELIVERY" +HEPARIN_MAXIMUM_VOLUME_DELIVERED_TEXT = "Maximum Cumulative Heparin Volume Delivered" +HEPARIN_SYRINGE_EMPTY_TEXT = "Syringe Empty" + +#Time_Duration +CREATE_TREATMENT_PARAMETERS = ["Blood Flow Rate", "Dialysate Flow Rate", "Duration","Heparin Dispensing Rate", + "Heparin Bolus Volume","Heparin Stop Time", "Saline Bolus","Dialysate Temperature","Blood Pressure Measurement Interval", + "Rinseback Flow Rate"] +TREATMENT_PAUSED_TEXT = "Treatment Paused" + +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] + + + +#Pre-Treatment Disposables +NEXT_BUTTON_TEXT = "NEXT" +BACK_BUTTON_TEXT = "BACK" +CONFIRM_BUTTON_TEXT = "CONFIRM" + + Index: shared/scripts/names.py =================================================================== diff -u -r650db9740238b22facd117b9216b28a1b619d70b -r11b2970e98eb80e8bd299d25b607b8ff00a2043b --- shared/scripts/names.py (.../names.py) (revision 650db9740238b22facd117b9216b28a1b619d70b) +++ shared/scripts/names.py (.../names.py) (revision 11b2970e98eb80e8bd299d25b607b8ff00a2043b) @@ -40,8 +40,8 @@ o_PreTreatmentCreate_dialysateFlowRate_SliderCreateTreatment = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "objectName": "_dialysateFlowRate", "type": "SliderCreateTreatment", "visible": True} o_PreTreatementCreateStack_PreTreatmentBase_TreatmentFlowBase = {"container":o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase,"objectName":"_PreTreatmentBase","type":"TreatmentFlowBase", "visible": True} o_treatmentStack_treatmentHome_TreatmentHome = {"container": o_treatmentStack_TreatmentStack, "id": "_treatmentHome", "type": "TreatmentHome", "unnamed": 1, "visible": True} -o_confirmButton_TouchRect = {"container": o_Gui_MainView, "gradient": 0, "objectName": "_confirmButton", "type": "TouchRect", "visible": False} -o_backButton_BackButton = {"container": o_Gui_MainView, "gradient": 0, "objectName": "_backButton", "type": "BackButton", "visible": False} +o_confirmButton_TouchRect = {"container": o_Gui_MainView, "objectName": "_confirmButton", "type": "TouchRect", "visible": False} +o_backButton_BackButton = {"container": o_Gui_MainView, "objectName": "_backButton", "type": "BackButton", "visible": False} o_NinePatchImage = {"container": o_Gui_MainView, "occurrence": 6, "type": "NinePatchImage", "unnamed": 1, "visible": False} o_PreTreatmentBase_backgroundRect_Rectangle = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "id": "_backgroundRect", "type": "Rectangle", "unnamed": 1, "visible": True} o_PreTreatmentBase_confirmButton_TouchRect = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "objectName": "_confirmButton", "type": "TouchRect", "visible": True} @@ -279,7 +279,7 @@ o_PreTreatmentBase_BACK_Text = {"container": o_PreTreatmentConnectionStack_PreTreatmentBase_TreatmentFlowBase_2, "text": "BACK", "type": "Text", "unnamed": 1, "visible": True} o_PreTreatmentBase_CONFIRM_Text_2 = {"container":o_PreTreatmentConnectionStack_PreTreatmentBase_TreatmentFlowBase_2, "text": "CONFIRM", "type": "Text", "unnamed": 1, "visible": True} #virtal keypad input field -o_keypad_input = {"container": o_Gui_MainView, "id": "_keyText", "type": "Text", "unnamed": 1, "visible": True} +o_keypad_input = {"container": o_Gui_MainView, "id": "_keyText","type": "Text", "unnamed": 1, "visible": True} #virtal keypad back space key o_back_space_key = {"container": o_Gui_MainView, "id": "_backspaceKeyIcon", "source": "qrc:/plugins/QtQuick/VirtualKeyboard/Styles/denali/images/iBackspace.svg", "type": "Image", "unnamed": 1, "visible": True}