Index: tst_main_treatment_ultrafilteration/test.py =================================================================== diff -u -r327472c2822efea2c331eaefb25be61191640c34 -r5a06807f7a8fb592ff1aee40003c82655ba41c9e --- tst_main_treatment_ultrafilteration/test.py (.../test.py) (revision 327472c2822efea2c331eaefb25be61191640c34) +++ tst_main_treatment_ultrafilteration/test.py (.../test.py) (revision 5a06807f7a8fb592ff1aee40003c82655ba41c9e) @@ -140,12 +140,7 @@ @return: N/A """ test.startSection("verification of uf value on main treatment section") - #FIXME: Slider shows some unexpected behaviour during its movement. - hd_simulator.cmd_set_treatment_parameter_ranges(MIN_TREATMENT_DURATION, MAX_TREATMENT_DURATION, 0, 600, MIN_DIALYSATE_FLOW_RATE, MAX_DIALYSATE_FLOW_RATE) - - for value in SLIDER_RANGE: - test_ultrafilteration_volume_slider(names.o_volumeSlider_slider, value, SLIDER_PARAMETERS) - + #FIXME: Slider shows some unexpected behaviour during its movement. for option in range(len(config.TREATMENT_ULTRAFILTRATION_TESTING_OPTION)): if object.exists(names.o_ultrafilteration_next_btn): mouseClick(waitForObjectExists(names.o_ultrafilteration_next_btn)) @@ -321,6 +316,11 @@ reset_treatment_ultrafiltration() verify_ultrafiltration_paused_state(uf_states.UF_PAUSED_STATE) + hd_simulator.cmd_set_treatment_parameter_ranges(MIN_TREATMENT_DURATION, MAX_TREATMENT_DURATION, 0, 600, MIN_DIALYSATE_FLOW_RATE, MAX_DIALYSATE_FLOW_RATE) + + for value in SLIDER_RANGE: + test_ultrafilteration_volume_slider(names.o_volumeSlider_slider, value, SLIDER_PARAMETERS) + test_treatment_ultrafiltration(uf_states.UF_PAUSED_STATE) test.startSection("verification of ultrafiltration flow parameters based on uf values")