Index: shared/scripts/configuration/config.py =================================================================== diff -u -r7f5d92792f49814d405ae6266af82b19cd999d4b -ra4f1e264727015d92baa065dde52c33c0bc33ef3 --- shared/scripts/configuration/config.py (.../config.py) (revision 7f5d92792f49814d405ae6266af82b19cd999d4b) +++ shared/scripts/configuration/config.py (.../config.py) (revision a4f1e264727015d92baa065dde52c33c0bc33ef3) @@ -19,7 +19,6 @@ AUT_NAME = "denaliSquish" COMMON_PATH = os.environ['HOME']+"/Projects" - #create custom treatment HIGHER_VALUE_OF_BLOOD_FLOW_RATE = 500 @@ -37,6 +36,491 @@ POSITIVE_BUFFER = 10 NEGATIVE_BUFFER = -10 +ALARM_ID_AND_MESSAGES = { + 1: "Stuck button POST failure.\nStop or Off button detected to be pressed for at least 1 second during test shortly after power up.", + 2: "HD FPGA POST failure.", + 3: "DG FPGA POST failure.", + 4: "HD Watchdog POST failure.", + 5: "DG Watchdog POST failure.", + 6: "UI communication POST failure.\nUI failed to communicate within a reasonable time after power up.", + 7: "Real-time clock configuration error.", + 8: "HD accelerometer failed POST.", + 9: "DG accelerometer failed POST." , + 10: "RTC or MCU timer inaccurate.", + 11: "DG heaters failed POST.", + 12: "HD firmware image integrity POST test failed.", + 13: "DG firmware image integrity POST test failed.", + 14: "HD blood flow invalid calibration.", + 15: "HD dialysate flow invalid calibration.", + 16: "HD alarm audio failed POST.", + 17: "HD UI POST failed.", + 18: "DG load cells invalid calibration record.", + 19: "DG flow sensors invalid calibration record.", + 20: "DG conductivity sensors invalid calibration record.", + 21: "DG drain line volume invalid calibration record.", + 22: "DG reservoirs invalid calibration record.", + 23: "DG acid concentrate calibration record.", + 24: "DG bicarb concentrate calibration record.", + 25: "DG accelerometer invalid calibration record.", + 26: "HD accelerometer invalid calibration record.", + 27: "HD blood flow invalid calibration record.", + 28: "HD dialyzer flow invalid calibration record.", + 29: "HD heparin force sensor invalid calibration record.", + 30: "HD Software fault.\nSoftware found itself in an unexpected state.", + 31: "Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 32: "Blood pump failed motor off check.\nMeasured speed while commanded off.", + 33: "Blood pump failed motor direction check.\nMeasured vs commanded.", + 34: "Blood pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 35: "Dialysis inlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 36: "Dialysis inlet pump failed motor off check.\nMeasured speed while commanded off.", + 37: "Dialysis inlet pump failed motor direction check.\nMeasured vs commanded.", + 38: "Dialysis inlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 39: "Dialysis outlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 40: "Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off.", + 41: "Dialysis outlet pump failed motor direction check.\nMeasured vs commanded.", + 42: "Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 43: "Watchdog expired error.\nWatchdog was not \"pet\" within the appropriate time.", + 44: "User interface communication timeout.", + 45: "Too many bad CRCs detected on received system messages.", + 46: "System message that required acknowledgment was not acknowledged.", + 47: "Ultrafiltration rate is too high error during treatment.", + 48: "Ultrafiltration volume accuracy error during treatment.", + 49: "HD FPGA communication down for too long.", + 50: "FPGA not accepting commanded valve states.", + 51: "Blood pump failed motor speed check.\nMeasured vs commanded.", + 52: "Dialysate inlet pump failed motor speed check.\nMeasured vs commanded.", + 53: "Dialysate outlet pump failed motor speed check.\nMeasured vs commanded.", + 54: "HD critical data integrity check failed.", + 55: "DG critical data integrity check failed.", + 56: "HD accelerometer error (no readings or FPGA reports error).", + 57: "DG accelerometer error (no readings or FPGA reports error).", + 58: "HD valve homing failed.", + 59: "HD valve transition time out.", + 60: "HD valve not functional.", + 61: "HD valve current out of range.", + 62: "HD valve position out of target.", + 63: "FPGA reports failure reading arterial pressure sensor.", + 64: "FPGA reports failure reading venous pressure sensor.", + 65: "HD requests DG command with invalid parameter fault.", + 66: "HD blood leak sensor set point set failure.", + 67: "HD blood pump occlusion self-test failure alarm.", + 68: "HD active reservoir recirculation out of range.", + 69: "HD blood leak sensor invalid calibration record.", + 70: "HD arterial pressure self-test failure alarm." , + 71: "HD venous pressure self-test failure alarm.", + 72: "HD blood flow meter status check self-test failure alarm.", + 73: "HD dialysate flow meter status check self-test failure alarm.", + 74: "HD blood leak detector self-test failure alarm.", + 75: "HD syringe pump self-test failure alarm.", + 76: "HD monitored voltage is out of range.", + 77: "DG monitored voltage is out of range.", + 78: "HD syringe pump direction (from encoder) error.", + 79: "HD syringe pump direction (from controller) error.", + 80: "HD syringe pump fault reported by FPGA.", + 81: "HD syringe pump over travel error.", + 82: "HD syringe pump DAC write failure.", + 83: "HD syringe pump is running while the BP is off.", + 84: "HD venous pressure sensor not being read.", + 85: "HD pump track latch opened alarm.", + 86: "HD BP occlusion sensor not being read.", + 87: "DG heating invalid calibration record.", + 88: "DG concentrate pumps hall sensor out of range.", + 89: "HD BP occlusion sensor error.", + 90: "This alarm ID is available for use.", + 91: "HD No dialysate flow data receive in the last 3 seconds.", + 92: "DG conductivity sensor bad status.", + 93: "This alarm ID is available for use.", + 94: "This alarm ID is available for use.", + 95: "This alarm ID is available for use.", + 96: "This alarm ID is available for use.", + 97: "This alarm ID is available for use.", + 98: "HD battery communication fault.", + 99: "HD syringe pump stall alarm.", + 100: "HD no cartridge self-test timeout.", + 101: "HD dry self-test timeout.", + 102: "Real-time clock communication error.", + 103: "HD non-volatile calibration group invalid record CRC.", + 104: "Air trap level sensors reporting illegal combination of air/fluid.", + 105: "DG non-volatile system group invalid record CRC.", + 106: "This alarm ID is available for use.", + 107: "HD reports DG restarted fault.", + 108: "HD syringe pump ADC error.", + 109: "HD syringe pump volume check error.", + 110: "HD syringe pump speed check error.", + 111: "HD syringe pump not stopped in off state error.", + 112: "HD blood leak detector fault.", + 113: "HD arterial air bubble detector self-test failure.", + 114: "HD venous air bubble detector self-test failure.", + 115: "DG temperature sensor out of range.", + 116: "DG temperature sensor ADC out of range.", + 117: "This alarm ID is available for use.", + 118: "This alarm ID is available for use.", + 119: "This alarm ID is available for use.", + 120: "This alarm ID is available for use.", + 121: "DG main primary heater voltage out of range.", + 122: "DG small primary heater voltage out of range.", + 123: "DG trimmer heater voltage out of range.", + 124: "HD end of treatment alarm (high priority - user still not acting to end treatment).", + 125: "Blood sitting too long after treatment stopped by user alarm (>5 min).", + 126: "HD blood leak detected alarm.", + 127: "Venous pressure too low during treatment." , + 128: "HD venous air bubble detected alarm.", + 129: "HD venous air bubble detected rinseback alarm.", + 130: "Venous pressure too high during treatment.", + 131: "Arterial pressure too low during treatment.", + 132: "Arterial pressure too high during treatment.", + 133: "DG fluid leak detected alarm.", + 134: "HD fluid leak detected alarm." , + 135: "HD experienced large acceleration.", + 136: "DG experienced large acceleration.", + 137: "HD tilt exceeds maximum.", + 138: "DG tilt exceeds maximum.", + 139: "HD AC power lost alarm.", + 140: "Dialysate generator communication timeout.", + 141: "Air trap fill timeout during treatment.", + 142: "Blood pump occlusion detected.", + 143: "DG dialysate temperature sensors drift timeout.", + 144: "DG software configuration record invalid CRC.", + 145: "Concentrate conductivity after adding acid out of range alarm.", + 146: "Dialysate conductivity out of range.", + 147: "DG dialysate generation conductivity fault alarm.", + 148: "Inlet water temperature too high.", + 149: "Inlet water temperature too low.", + 150: "Inlet water conductivity too high.", + 151: "Inlet water conductivity too low.", + 152: "Inlet water pressure too low.", + 153: "HD prime completed high priority alarm.", + 154: "EEPROM operations (read, write, erase) failure.", + 155: "DG software configuration record invalid CRC.", + 156: "HW usage data (treatment time in HD and total consumed water in DG) failure.", + 157: "DG chemical disinfect prime acid line timeout.", + 158: "This alarm ID is available for use.", + 159: "Dialysate inlet pump failed flow vs motor speed check.", + 160: "Blood pump rotor speed too high.", + 161: "This alarm ID is available for use.", + 162: "This alarm ID is available for use.", + 163: "HD sees primary load cell for reservoir 1 change too much too fast.", + 164: "HD sees primary load cell for reservoir 2 change too much too fast.", + 165: "HD in treatment rinseback operation timeout.", + 166: "HD in treatment re-circ sub-mode for too long.", + 167: "HD cartridge door opened alarm.", + 168: "HD active reservoir depletion timeout.", + 169: "Dialysate flow rate out of range.", + 170: "HD syringe empty alarm.", + 171: "HD syringe pump occlusion alarm.", + 172: "HD syringe pump not enough Heparin alarm.", + 173: "HD arterial air bubble detected alarm.", + 174: "HD arterial air bubble detected rinseback alarm.", + 175: "HD pump direction status error.", + 176: "HD active reservoir is full during treatment and DG not ready to switch reservoirs.", + 177: "DG software fault.\nSoftware found itself in an unexpected state.", + 178: "HD communication timeout.", + 179: "DG FPGA communication down for too long.", + 180: "DG load cell ADC error.", + 181: "DG load cells weight out of range for tare.", + 182: "DG load cells invalid calibration.", + 183: "DG invalid load cell value.", + 184: "Inlet or Outlet UV reactor not healthy.", + 185: "DG fan RPM out of range.", + 186: "DG concentrate pump fault.", + 187: "DG concentrate pump CP1 speed control error.", + 188: "DG concentrate pump CP2 speed control error.", + 189: "DG drain pump RPM out of range.", + 190: "DG drain pump off fault.", + 191: "DG flow rate out of upper range.", + 192: "DG flow rate out of lower range.", + 193: "RO pump flow rate out of range.", + 194: "DG RO pump duty cycle out of range.", + 195: "DG RO pump pressure out of range.", + 196: "DG temperature sensor error flag fault.", + 197: "DG temperature sensors ADC fault.", + 198: "DG heaters cold junction temperature out of range.", + 199: "DG heaters fault.", + 200: "DG thermistors/sensors temperature out of range.", + 201: "Inlet water quality is bad (temperature, conductivity, pressure).", + 202: "Inlet water pressure fault.", + 203: "Pressure sensor fault.", + 204: "RO rejection ratio out of range.", + 205: "Conductivity sensor fault.", + 206: "DG dialysate fill runs out of time.", + 207: "DG flow meter check failure alarm.", + 208: "DG primary heater on with no flow time out.", + 209: "DG drain circulation line timeout.", + 210: "HD battery pack detected an error.", + 211: "Blood sitting too long warning (>4 min).", + 212: "HD end of treatment alarm (user not acting to end treatment).", + 213: "HD prime completed medium priority alarm.", + 214: "HD Saline bolus volume check failure.", + 215: "HD rinseback volume check failure.", + 216: "HD end treatment sub-mode timeout alarm.", + 217: "HD blood prime volume check failure.", + 218: "HD syringe detected alarm.", + 219: "HD syringe pump syringe removed alarm.", + 220: "HD battery pack SOC is too low.", + 221: "HD Saline bag is empty.", + 222: "HD dialysate temperature too high alarm.", + 223: "HD dialysate temperature too low alarm.", + 224: "Treatment stopped by user action - pressed stop button.", + 225: "HD end of treatment warning.", + 226: "HD prime completed low priority alarm.", + 227: "HD prime out of time alarm.", + 228: "HD time out on prime sub-mode purge air state.", + 229: "HD prime dialysate dialyzer time out alarm.", + 230: "HD prime dialysate bypass time out alarm.", + 231: "HD pre-treatment mode dry self-tests failure.", + 232: "HD pre-treatment mode wet self-tests failure.", + 233: "RTC battery low.", + 234: "RTC (i.e read time) or RAM operations failure (read or write).", + 235: "HD in treatment stopped sub-mode after rinseback completed (no escalation).", + 236: "HD needs new cartridge to be installed.", + 237: "HD prime saline dialyzer time out alarm.", + 238: "HD no cartridge loaded or installed improperly alarm.", + 239: "HD fail to remove cartridge alarm.", + 240: "Bicarb conductivity out of range during bicarb pump check alarm.", + 241: "DG reservoir drain time out.", + 242: "DG reservoir fill time out.", + 243: "DG reservoir leak time out.", + 244: "DG temperature sensors difference out of range.", + 245: "DG heat disinfect target temperature time out (could not reach to temperature).", + 246: "DG heat disinfect inlet pressure and temperature sensors out of range.", + 247: "DG heat disinfect inlet conductivity and temperature out of range.", + 248: "DG chemical disinfect could not reach to target temperature.", + 249: "DG chemical disinfect inlet pressure and temperature sensors out of range.", + 250: "DG chemical disinfect inlet conductivity and temperature out of range.", + 251: "HD invalid system record.", + 252: "HD invalid service record.", + 253: "DG invalid system record.", + 254: "DG invalid service record.", + 255: "HD and UI software builds are not compatible.", + 256: "HD and DG software builds are not compatible.", + 257: "DG FPGA power out timeout.", + 258: "HD FPGA power out timeout.", + 259: "HD temperatures out of range.", + 260: "UI POST Application Integrity (Sha256Sum) failure.", + 261: "UI POST CANBus failure.", + 262: "UI POST Display failure.", + 263: "UI POST Touch failure.", + 264: "UI POST SD-Card failure.", + 265: "UI POST RTC failure.", + 266: "UI POST WiFi failure.", + 267: "UI POST Bluetooth failure.", + 268: "UI POST Ethernet failure.", + 269: "UI POST Sound failure.", + 270: "HD POST Safety Shutdown failure.", + 271: "DG POST Safety Shutdown failure.", + 272: "HD Fan RPM out of range.", + 273: "HD measured blood pump flow rate is out of range.", + 274: "HD measured dialysate inlet pump flow rate is out of range.", + 275: "HD arterial pressure sensor is reading out of range.", + 276: "HD venous pressure sensor is reading out of range.", + 277: "HD BP occlusion sensor is reading out of range.", + 278: "HD active reservoir weight out of range.", + 279: "DG dialysate drain time out.", + 280: "HD arterial pressure sensor read timeout error.", + 281: "Acid concentration bottle low volume alarm.", + 282: "Bicarbonate concentration bottle low volume alarm.", + 283: "DG load cells weight out of range.", + 284: "DG load cells primary/back up drift out of range.", + 285: "HD treatment recirculate timeout warning.", + 286: "HD treatment rinseback complete timeout warning.", + 287: "HD processor clock speed checks against FPGA clock failure.", + 288: "DG trimmer heater on with flow timeout.", + 289: "DG dialysate or concentrate caps not closed." + } + +ALARM_PRIORITIES = {1: "LOW", 2: "MEDIUM", 3: "HIGH"} +NUM_OF_ALARM_PRIORITIES = 4 # Total number of alarm priorities +ACCEPTED = True +REJECTED = False +ALARM_LIST_TITLE = "Alarm list" + + +NUM_OF_REQUEST_REJECT_REASONS = 43 +NUM_OF_MSG_ID = 289 +UNDEFINED_ALARM_ID_MSG = "ALARM_ID_UNDEFINED" +ALARM_TITLE = "Alarm" + + +ALARM_ID_NO_ALARM = 0 +ALARM_ID_OUT_OF_RANGE = 290 + +ALARM_PRIORITY_OPTIONS = { + 0 : 'ALARM_PRIORITY_NONE' , + 1 : 'ALARM_PRIORITY_LOW' , + 2 : 'ALARM_PRIORITY_MEDIUM', + 3 : 'ALARM_PRIORITY_HIGH', +} + +# alarm colors +ALARMS_COLORS_HEADER = { + "ALARM_PRIORITY_NONE" : "#438feb", + "ALARM_PRIORITY_LOW" : "#db8f00", + "ALARM_PRIORITY_MEDIUM" : "#db8f00", + "ALARM_PRIORITY_HIGH" : "#831913" +} + +ALARMS_COLORS_BODY = { + "ALARM_PRIORITY_NONE" : "#254670", + "ALARM_PRIORITY_LOW" : "#f5a623", + "ALARM_PRIORITY_MEDIUM" : "#f5a623", + "ALARM_PRIORITY_HIGH" : "#c53b33" +} + +ALARM_MUTE_FLAG = 0B0000001000000000 +ALARM_UNMUTE_FLAG = 0B0000000000000000 + +ALARM_BUTTONS_OK = 0B0000000010111000 + +ALARM_BUTTONS_RESUME = 0B0000000000110000 +ALARM_BUTTONS_RESUME_RINSEBACK = 0B0000000000100000 +ALARM_BUTTONS_RESUME_END = 0B0000000000010000 + +ALARM_BUTTONS_RINSEBACK = 0B0000000000101000 +ALARM_BUTTONS_RINSEBACK_RESUME = ALARM_BUTTONS_RESUME_RINSEBACK +ALARM_BUTTONS_RINSEBACK_END = 0B0000000000001000 + +ALARM_BUTTONS_END = 0B0000000000011000 +ALARM_BUTTONS_END_RESUME = ALARM_BUTTONS_RESUME_END +ALARM_BUTTONS_END_RINSEBACK = ALARM_BUTTONS_RINSEBACK_END + +ALARM_BUTTONS_RESUME_RINSEBACK_END = 0B0000000000000000 +ALARM_BUTTONS_RINSEBACK_RESUME_END = ALARM_BUTTONS_RESUME_RINSEBACK_END +ALARM_BUTTONS_END_RESUME_RINSEBACK = ALARM_BUTTONS_RESUME_RINSEBACK_END + +CURRENT_COLOR = '#000000' +COMPLETE_COLOR= '#4290ec' +ENABLED_COLOR = '#fcfcfc' +INCOMPLETE_COLOR = '#607a91' +OUT_OF_RANGE_COLOR = "#c53b33" + +#standby mode +GOODMORNING_START_TIME_SEC = 0 +GOODEVENING_START_TIME_SEC = 43200 +BLOOD_PRIMING_TEXT = "Blood Priming" +LIQUID_UNIT = "mL" +BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" +SALINE_BOLUS_CUMULATIVE = 300 +SALINE_UNIT = "mL" + +UF_MINIMUM_VALUE = 0 +UF_MAXIMUM_VALUE = 600 +ULTRAFILTRATION_PAUSED_TEXT = "Ultrafiltration Paused" +RESUME_TEXT = "RESUME ULTRAFILTRATION" +EDIT_TEXT = "EDIT ULTRAFILTRATION VOLUME" +TITLE_TEXT = "ULTRAFILTRATION VOLUME (L)" + +#main treatment pressure +PRESSURE_TITLE = "PRESSURES" +ARTERIAL_TITLE = "Arterial" +VENOUS_TITLE = "Venous" +VENOUS_UOM = "mmHg" +ARTERIAL_UOM = "mmHg" +LOW_TEXT = "LOW" +HIGH_TEXT = "HIGH" +ACCEPTED = True +REJECTED = False +NUM_OF_REQUEST_REJECT_REASONS = 43 + +#pressure +PRESSURE_STEPS = 10 +ARTERIAL_PRESSURE_MINIMUM = -400 +ARTERIAL_PRESSURE_MAXIMUM = +600 +ARTERIAL_PRESSURE_LOW_MIN = -300 +ARTERIAL_PRESSURE_LOW_DEF = -300 +ARTERIAL_PRESSURE_LOW_MAX = +200 +ARTERIAL_PRESSURE_HIGH_MIN = -300 +ARTERIAL_PRESSURE_HIGH_DEF = +100 +ARTERIAL_PRESSURE_HIGH_MAX = +200 + +VENOUS_PRESSURE_MINIMUM = -100 +VENOUS_PRESSURE_MAXIMUM = +700 +VENOUS_PRESSURE_LOW_MIN = -100 +VENOUS_PRESSURE_LOW_DEF = -100 +VENOUS_PRESSURE_LOW_MAX = +600 +VENOUS_PRESSURE_HIGH_MIN = +100 +VENOUS_PRESSURE_HIGH_DEF = +400 +VENOUS_PRESSURE_HIGH_MAX = +600 + +#main_treatment_flows +BLOOD_FLOW_MINIMUM_VALUE = 100 +BLOOD_FLOW_BUFFER = 25 +FLOW_SLIDER_WIDTH_BUFFER = 37 +BLOOD_FLOW_MAXIMUM_VALUE = 500 +DIALYSATE_FLOW_MINIMUM_VALUE = 0 +DIALYSATE_FLOW_MAXIMUM_VALUE = 0 # Seems to be bug. need change +FLOW_UNIT = "mL/min" +BLOOD_FLOW_TEXT = "blood" +DIALYSATE_FLOW_TEXT = "dialysate" + +#main treatment screen +VITALS_TITLE = "VITALS" +OUT_OF_RANGE_COLOR = "#c53b33" +IN_RANGE_COLOR = "#fcfcfc" +SYSTOLIC_LOWER_LIMIT = 60 +SYSTOLIC_UPPER_LIMIT = 250 +DIASTOLIC_LOWER_LIMIT = 40 +DIASTOLIC_UPPER_LIMIT = 200 +HEART_RATE_LOWER_LIMIT = 40 +HEART_RATE_UPPER_LIMIT = 180 +HEART_RATE_TITLE = "Heart Rate" +BLOOD_PRESSURE_TITLE = "Blood Pressure" +BLOOD_PRESSURE_UNIT = "mmHg" +HEART_RATE_UNIT = "BPM" +BLOOD_PRIMING_RANGE = 300 +BLOOD_PRIMING_TARGET_MAXIMUM = 300 + + +#tst_treatment_monitoring +FLOWS_TEXT = "FLOWS" +BLOOD_UNIT = "mL/min" +DIALYSATE_TEXT = "Dialysate" +BLOOD_TEXT = "Blood" +VITALS_TEXT = "VITALS" +BLOOD_PRESSURE_TEXT = "mmHg" +HEARTRATE_UNIT = "BPM" +PRESSURE_TEXT = "PRESSURE (mmHg)" +ARTEREAL_TEXT = "Arterial" +VENOUS_TEXT = "Venous" +TIME_DURATION_TEXT = "Time Remaining" +ULTRAFILTERATION_TEXT = "ULTRAFILTRATION VOLUME (L)" +SALINE_BOLUS = "SALINE BOLUS" +VOLUME_DELIVERED = "Volume Delivered" +CUMULATIVE_DELIVERED = "Cumulative Delivered" +HEPARIN_TEXT = "HEPARIN" +VOLUME_DELIVERED = "Volume Delivered" + +#patient id +PATIENT_ID = { + "qwerty" : "qwerty", + "sampleid" : "sampleid", + "abcdefghijklmnop" : "abcdefghijklmnop", + "patientid" : "patientid", + "username" : "username", + "sA@MpLE#123" : "sAMpLE123", + "A@BH&321" : "ABH321", + "5432@qwerty" : "5432qwerty" +} + +KEYBOARD_SWITCH = ["ABC", "&123"] + + +#main-treatment heparin +HEPARIN_UNIT = "mL" +HEPARIN_DELIVERY_TEXT = "HEPARIN DELIVERY" +HEPARIN_DELIVERY_OFF_TEXT = "OFF" +HEPARIN_STOP_TEXT = "Delivery Stopped" +HEPARIN_RESUME_DELIVERY_TEXT = "RESUME DELIVERY" +HEPARIN_BOLUS_ACTIVE_TEXT = "Bolus Active" +HEPARIN_PAUSE_DELIVERY_TEXT = "PAUSE DELIVERY" +HEPARIN_MAXIMUM_VOLUME_DELIVERED_TEXT = "Maximum Cumulative Heparin Volume Delivered" +HEPARIN_SYRINGE_EMPTY_TEXT = "Syringe Empty" + +#Time_Duration +CREATE_TREATMENT_PARAMETERS = ["Blood Flow Rate", "Dialysate Flow Rate", "Duration","Heparin Dispensing Rate", + "Heparin Bolus Volume","Heparin Stop Time", "Saline Bolus","Dialysate Temperature","Blood Pressure Measurement Interval", + "Rinseback Flow Rate"] +TREATMENT_PAUSED_TEXT = "Treatment Paused" + #dictionary consist of custom treatment values CREATE_CUSTOM_TREATMENT = { "Blood Flow Rate": 250, @@ -88,4 +572,140 @@ "rinseback_flow_rate" : [50, 25, 187], } +#ultrafilteration volume based on flow parameters. Python list contains Rotational speed, Motor speed, +# MC speed, MC current and PWM. +TREATMENT_ULTRAFILTRATION_FLOW_PARAMETERS = { + 0 : [100, 1, 1, 1, 1, 1, 1], + 1 : [550, 24, 32, 425, 85, 18, 36], + 2 : [80, 125, 187, 257, 530, 458, 985], + 3 : [100, 0, 0, 0, 0, 0, 0], + 4 : [220, 1.45, 1.45, 1.45, 1.45, 1.45, 1.45], + } +#dictionary contains uf maximum and minimum value's to set. +TREATMENT_ULTRAFILTRATION_TESTING_OPTION = { + 0 : {"uf_maximum" : 600, "uf_minimum" : 0}, + 1 : {"uf_maximum" : 500, "uf_minimum" : 0}, + 2 : {"uf_maximum" : 550, "uf_minimum" : 0}, + 3 : {"uf_maximum" : 300, "uf_minimum" : 0}, + 4 : {"uf_maximum" : 280, "uf_minimum" : 0}, + 5 : {"uf_maximum" : 765, "uf_minimum" : 500}, + 6 : {"uf_maximum" : 680, "uf_minimum" : 450}, + 7 : {"uf_maximum" : 480, "uf_minimum" : 380}, + 8 : {"uf_maximum" : 390, "uf_minimum" : 675}, + 9 : {"uf_maximum" : 290, "uf_minimum" : 125}, + } + +#tst_pretreatment_screens +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] +NUM_OF_PRETREATMENT_BULLETS = len(PRE_TREATMENT_SCREENS) + +#tst_pretreatment_screens +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] + + +#tst_pretreatment_screens color palettes +CURRENT_COLOR = '#000000' +COMPLETE_COLOR= '#4290ec' +ENABLED_COLOR = '#fcfcfc' +INCOMPLETE_COLOR = '#607a91' + +CONSUMABLE_SELF_TEST_TEXT = "Consumables Self Test" +BICARB_PUMP_CHECK_TEXT = "BiCarb Pump Check" +ACID_PUMP_CHECK_TEXT = "Acid Pump Check" +SELF_TEST_COMPLETE_TEXT = "Self Test Complete!" + + +#pre-treatment_priming +NOT_VISIBLE = False +VISIBLE = True +FOCUS = True +NOT_IN_FOCUS = False +BUSY = "Busy" +CHECK = "Check" +BEGIN_PRIME_TITLE = "Begin Prime" +PRIMING_TITLE = "Priming" +SYSTEM_SELF_TEST_TITLE = "System Self Test" +FILTER_FLUSH_TITLE = "Filter Flush" +MAXIMUM_COUNTDOWN_TIME = 300 +COUNT_DOWN_TIME_100 = 100 +MINIMUM_COUNTDOWN_TIME = 0 +ENABLED = True +DISABLED = False +CONSUMABLE_SELF_TEST_BICARB_PUMP_CHECK_STATE = 3 + +LOOD_PRIMING_TEXT = "Blood Priming" +SALINE_UNIT = "mL" +BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" + +#Pre-Treatment Disposables +NEXT_BUTTON_TEXT = "NEXT" +BACK_BUTTON_TEXT = "BACK" +ENABLED = True +DISABLED = False +CONFIRM_BUTTON_TEXT = "CONFIRM" + +#Filter Flush + +#Pre_treatment_water_sample +SAMPLE_TEXT = "Sample" +CREATE_TEXT = "Create" +NEXT_BUTTON_TEXT = "NEXT" +WATER_SAMPLE_BUTTON_TEXT = "WATER SAMPLE" +BACK_BUTTON_TEXT = "BACK" +PASS_TEXT = "PASS" +FAIL_TEXT = "FAIL" +OK_TEXT = "OK" + +# pretreatment_patient_connection +SYSTOLIC_PRESSSURE_120 = 120 +SYSTOLIC_PRESSSURE_113 = 113 +SYSTOLIC_PRESSSURE_150 = 150 +SYSTOLIC_PRESSSURE_175 = 175 +SYSTOLIC_PRESSSURE_200 = 200 +SYSTOLIC_PRESSSURE_200 = 200 +SYSTOLIC_PRESSSURE_251 = 251 +SYSTOLIC_PRESSSURE_59 = 59 +SYSTOLIC_PRESSSURE_15 = 15 +DIASTOLIC_PRESSSURE_25 = 25 +DIASTOLIC_PRESSSURE_39 = 39 +DIASTOLIC_PRESSSURE_74 = 74 +DIASTOLIC_PRESSSURE_80 = 80 +DIASTOLIC_PRESSSURE_85 = 85 +DIASTOLIC_PRESSSURE_60 = 60 +DIASTOLIC_PRESSSURE_150 = 150 +DIASTOLIC_PRESSSURE_200 = 200 +DIASTOLIC_PRESSSURE_201 = 201 +HEART_RATE_VAL_39 = 39 +HEART_RATE_VAL_60 = 60 +HEART_RATE_VAL_70 = 70 +HEART_RATE_VAL_85 = 85 +HEART_RATE_VAL_100 = 100 +HEART_RATE_VAL_101 = 101 +HEART_RATE_VAL_110 = 110 +HEART_RATE_VAL_181 = 181 +HEART_RATE_VAL_200 = 200 +SYSTOLIC_LOWER_LIMIT = 60 +SYSTOLIC_UPPER_LIMIT = 250 +DIASTOLIC_LOWER_LIMIT = 40 +DIASTOLIC_UPPER_LIMIT = 200 +HEART_RATE_LOWER_LIMIT = 40 +HEART_RATE_UPPER_LIMIT = 180 +UF_MINIMUM_SLIDER_WIDTH = -2 +UF_MAXIMUM_SLIDER_WIDTH = 638.00 +UF_VALID_RANGE = 700 +TRAINING_INDICATOR_SCREEN = 4 +TREATMENT_ULTRAFILTERATION_TESTING_OPTION = { + "OPTION_1" : {"uf_maximum" : 800, "uf_minimum" : 0}, + "OPTION_2" : {"uf_maximum" : 500, "uf_minimum" : 0}, + "OPTION_3" : {"uf_maximum" : 550, "uf_minimum" : 0}, + "OPTION_4" : {"uf_maximum" : 300, "uf_minimum" : 0}, + "OPTION_5" : {"uf_maximum" : 280, "uf_minimum" : 0}, + "OPTION_6" : {"uf_maximum" : 765, "uf_minimum" : 0}, + "OPTION_7" : {"uf_maximum" : 680, "uf_minimum" : 0}, + "OPTION_8" : {"uf_maximum" : 1500, "uf_minimum" : 0}, + "OPTION_9" : {"uf_maximum" : 390, "uf_minimum" : 0}, + "OPTION_10" : {"uf_maximum" : 800, "uf_minimum" : 0}, + "OPTION_11" : {"uf_maximum" : 1000, "uf_minimum" : 0}, + "OPTION_12" : {"uf_maximum" : 1250, "uf_minimum" : 0}, + } Index: shared/scripts/configuration/strings.py =================================================================== diff -u -rcea724e0b33dcd446f230ab23d6aa3729e6f86fd -ra4f1e264727015d92baa065dde52c33c0bc33ef3 --- shared/scripts/configuration/strings.py (.../strings.py) (revision cea724e0b33dcd446f230ab23d6aa3729e6f86fd) +++ shared/scripts/configuration/strings.py (.../strings.py) (revision a4f1e264727015d92baa065dde52c33c0bc33ef3) @@ -8,7 +8,7 @@ # # @file strings.py # -# @author (last) LTTS +# @author (last) LTTS # @date (last) 18-Jan-2022 # ############################################################################ @@ -23,49 +23,77 @@ # dictionary of rejection reason from application source code. # message location -> denali-> Headers-> common-> MsgDefs.h -REJECTION_REASON = { - 0 : "REQUEST_REJECT_REASON_NONE" , # Used when there is no rejection - 1 : "REQUEST_REJECT_REASON_NOT_ALLOWED_IN_CURRENT_MODE" , # "REQuest is not allowed in the current operating mode - 2 : "REQUEST_REJECT_REASON_TIMEOUT_WAITING_FOR_USER_CONFIRM" , # Validated "REQuest was not confirmed by user in reasonable time - 3 : "REQUEST_REJECT_REASON_NOT_IN_TREATMENT_MODE" , # "REQuest is not allowed if not in treatment mode - 4 : "REQUEST_REJECT_REASON_INVALID_TREATMENT_STATE" , # "REQuest is not allowed in current treatment state - 5 : "REQUEST_REJECT_REASON_TREATMENT_TOO_CLOSE_TO_FINISHED" , # "REQuest is not allowed so near end of treatment - 6 : "REQUEST_REJECT_REASON_TREATMENT_TIME_OUT_OF_RANGE" , # Treatment duration is out of range - 7 : "REQUEST_REJECT_REASON_TREATMENT_TIME_LESS_THAN_CURRENT" , # Treatment time change is less than currently elapsed treatment time - 8 : "REQUEST_REJECT_REASON_BLOOD_FLOW_OUT_OF_RANGE" , # Blood flow is out of range - 9 : "REQUEST_REJECT_REASON_DIAL_FLOW_OUT_OF_RANGE" , # Dialysate flow is out of range - 10 : "REQUEST_REJECT_REASON_DIAL_VOLUME_OUT_OF_RANGE" , # Dialysate flow rate or treatment duration causes dialysate volume to exceed limit - 11 : "REQUEST_REJECT_REASON_UF_VOLUME_OUT_OF_RANGE" , # Ultrafiltration volume is out of range - 12 : "REQUEST_REJECT_REASON_UF_RATE_OUT_OF_RANGE" , # Ultrafiltration rate is out of range - 13 : "REQUEST_REJECT_REASON_TREATMENT_TIME_LESS_THAN_MINIMUM" , # Treatment time change is less than minimum treatment time - 14 : "REQUEST_REJECT_REASON_UF_NOT_IN_PROGESS" , # Ultrafiltration is not currently in progress - 15 : "REQUEST_REJECT_REASON_UF_NOT_PAUSED" , # Ultrafiltration is not currently paused - 16 : "REQUEST_REJECT_REASON_SALINE_BOLUS_IN_PROGRESS" , # A saline bolus is in progress - 17 : "REQUEST_REJECT_REASON_PARAM_OUT_OF_RANGE" , # A treatment parameter is out of range - 18 : "REQUEST_REJECT_REASON_HEPARIN_PRESTOP_EXCEEDS_DURATION" , # The Heparin pre-stop setting is greater than the treatment duration - 19 : "REQUEST_REJECT_REASON_ARTERIAL_PRESSURE_LOW_VS_HIGH" , # Arterial pressure low and high alarm limits are not inconsistent - 20 : "REQUEST_REJECT_REASON_VENOUS_PRESSURE_LOW_VS_HIGH" , # Venous pressure low and high alarm limits are inconsistent - 21 : "REQUEST_REJECT_REASON_SALINE_MAX_VOLUME_REACHED" , # Saline bolus volume maximum has been reached - no more saline allowed - 22 : "REQUEST_REJECT_REASON_SALINE_BOLUS_NOT_IN_PROGRESS" , # A saline bolus is not in progress - 23 : "REQUEST_REJECT_REASON_ACTION_DISABLED_IN_CURRENT_STATE" , # "REQuested user action is disabled in current state - 24 : "REQUEST_REJECT_REASON_ALARM_IS_ACTIVE" , # "REQuested user action not allowed while alarm is active - 25 : "REQUEST_REJECT_REASON_INVALID_COMMAND" , # "REQuested user action invalid - 26 : "REQUEST_REJECT_REASON_TREATMENT_IS_COMPLETED" , # The treatment has been completed - 27 : "REQUEST_REJECT_REASON_ADDL_RINSEBACK_MAX_VOLUME_REACHED" , # Rinseback additional volume maximum has been reached - no more additional rinsebacks allowed - 28 : "REQUEST_REJECT_REASON_UF_VOLUME_NOT_SET" , # Ultrafiltration volume is not set yet - 29 : "REQUEST_REJECT_REASON_NO_PATIENT_CONNECTION_CONFIRM" , # The user has not confirmed patient connection - 30 : "REQUEST_REJECT_REASON_HEPARIN_PAUSE_INVALID_IN_THIS_STATE" , # Heparin cannot be paused if not currently deliverying Heparin - 31 : "REQUEST_REJECT_REASON_HEPARIN_NOT_PAUSED" , # Heparin cannot be resumed if not paused - 32 : "REQUEST_REJECT_REASON_DG_COMM_LOST" , # Treatment cannot initiate if DG comm is lost - 33 : "REQUEST_REJECT_REASON_DRAIN_NOT_COMPLETE" , # Post-treatment reservoirs drain not complete - 34 : "REQUEST_REJECT_REASON_DG_NOT_IN_STANDBY_IDLE_STATE" , # Treatment cannot initiate if DG is not in standby idle state - 35 : "REQUEST_REJECT_REASON_INVALID_REQUEST_FORMAT" , # "REQuest message not formatted properly - 36 : "REQUEST_REJECT_REASON_INVALID_DATE_OR_TIME" , # Given date/time is invalid - 37 : "REQUEST_REJECT_REASON_NO_NEW_TREATMENT_ALARM_TRIGGERED" , # Treatment cannot initiate if an alarm with no new treatment property has triggered before - 38 : "REQUEST_REJECT_REASON_BATTERY_IS_NOT_CHARGED" , # Battery does not have enough charge to start treatment - 39 : "REQUEST_REJECT_REASON_RINSEBACK_NOT_COMPLETED" , # Cannot move on to recirculate without completing full rinseback +# pretreatment_patient_connection +DIASTOLIC_TEXT = "diastolic" +SYSTOLIC_TEXT = "systolic" +OUT_OF_RANGE_COLOR = "#c53b33" +IN_RANGE_COLOR = "#fcfcfc" +HEART_RATE_TITLE = "Heart Rate" +BLOOD_PRESSURE_TITLE = "Blood Pressure" +BLOOD_PRESSURE_UNIT = "mmHg" +HEART_RATE_UNIT = "BPM" +PATIENT_CONNECTION_TEXT = "Patient Connection" +UF_VOLUME_TEXT = "Ultrafiltration Volume (L)" +UF_TITLE_TEXT = "Ultrafiltration Setup" +BP_HR_TEXT = "BP/HR" +SKIP_TEXT = "SKIP" +TUTORIAL_TEXT = { + 0 : "Wash your hands and establish vascular access per your clinic's instructions.", + 1 : "Pinch arterial clamp and venous clamp on located on the cartridge patient lines.", + 2 : "Remove arterial and venous shunt.", + 3 : "Connect arterial and venous cartridge lines to your patient access lines." } +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] +# dictionary of rejection reason from application source code. +# message location -> denali-> Headers-> common-> MsgDefs.h +REJECTION_REASON = { + 0: "No Active Alarm List", + 1: "REQUEST_REJECT_REASON_NOT_ALLOWED_IN_CURRENT_MODE", # Request is not allowed in the current operating mode + 2: "REQUEST_REJECT_REASON_TIMEOUT_WAITING_FOR_USER_CONFIRM", # Validated request was not confirmed by user in reasonable time + 3: "REQUEST_REJECT_REASON_NOT_IN_TREATMENT_MODE", # Request is not allowed if not in treatment mode + 4: "REQUEST_REJECT_REASON_INVALID_TREATMENT_STATE", # Request is not allowed in current treatment state + 5: "REQUEST_REJECT_REASON_TREATMENT_TOO_CLOSE_TO_FINISHED", # Request is not allowed so near end of treatment + 6: "REQUEST_REJECT_REASON_TREATMENT_TIME_OUT_OF_RANGE", # Treatment duration is out of range + 7: "REQUEST_REJECT_REASON_TREATMENT_TIME_LESS_THAN_CURRENT", # Treatment time change is less than currently elapsed treatment time + 8: "REQUEST_REJECT_REASON_BLOOD_FLOW_OUT_OF_RANGE", # Blood flow is out of range + 9: "REQUEST_REJECT_REASON_DIAL_FLOW_OUT_OF_RANGE", # Dialysate flow is out of range + 10: "REQUEST_REJECT_REASON_DIAL_VOLUME_OUT_OF_RANGE", # Dialysate flow rate or treatment duration causes dialysate volume to exceed limit + 11: "REQUEST_REJECT_REASON_UF_VOLUME_OUT_OF_RANGE", # Ultrafiltration volume is out of range + 12: "REQUEST_REJECT_REASON_UF_RATE_OUT_OF_RANGE", # Ultrafiltration rate is out of range + 13: "REQUEST_REJECT_REASON_TREATMENT_TIME_LESS_THAN_MINIMUM", # Treatment time change is less than minimum treatment time + 14: "REQUEST_REJECT_REASON_UF_NOT_IN_PROGESS", # Ultrafiltration is not currently in progress + 15: "REQUEST_REJECT_REASON_UF_NOT_PAUSED", # Ultrafiltration is not currently paused + 16: "REQUEST_REJECT_REASON_SALINE_BOLUS_IN_PROGRESS", # A saline bolus is in progress + 17: "REQUEST_REJECT_REASON_PARAM_OUT_OF_RANGE", # A treatment parameter is out of range + 18: "REQUEST_REJECT_REASON_HEPARIN_PRESTOP_EXCEEDS_DURATION", # The Heparin pre-stop setting is greater than the treatment duration + 19: "REQUEST_REJECT_REASON_ARTERIAL_PRESSURE_LOW_VS_HIGH", # Arterial pressure low and high alarm limits are not inconsistent + 20: "REQUEST_REJECT_REASON_VENOUS_PRESSURE_LOW_VS_HIGH", # Venous pressure low and high alarm limits are inconsistent + 21: "REQUEST_REJECT_REASON_SALINE_MAX_VOLUME_REACHED", # Saline bolus volume maximum has been reached - no more saline allowed + 22: "REQUEST_REJECT_REASON_SALINE_BOLUS_NOT_IN_PROGRESS", # A saline bolus is not in progress + 23: "REQUEST_REJECT_REASON_ACTION_DISABLED_IN_CURRENT_STATE", # Requested user action is disabled in current state + 24: "REQUEST_REJECT_REASON_ALARM_IS_ACTIVE", # Requested user action not allowed while alarm is active + 25: "REQUEST_REJECT_REASON_INVALID_COMMAND", # Requested user action invalid + 26: "REQUEST_REJECT_REASON_TREATMENT_IS_COMPLETED", # The treatment has been completed + 27: "REQUEST_REJECT_REASON_ADDL_RINSEBACK_MAX_VOLUME_REACHED", # Rinseback additional volume maximum has been reached - no more additional rinsebacks allowed + 28: "REQUEST_REJECT_REASON_UF_VOLUME_NOT_SET", # Ultrafiltration volume is not set yet + 29: "REQUEST_REJECT_REASON_NO_PATIENT_CONNECTION_CONFIRM", # The user has not confirmed patient connection + 30: "REQUEST_REJECT_REASON_HEPARIN_PAUSE_INVALID_IN_THIS_STATE", # Heparin cannot be paused if not currently deliverying Heparin + 31: "REQUEST_REJECT_REASON_HEPARIN_NOT_PAUSED", # Heparin cannot be resumed if not paused + 32: "REQUEST_REJECT_REASON_DG_COMM_LOST", # Treatment cannot initiate if DG comm is lost + 33: "REQUEST_REJECT_REASON_DRAIN_NOT_COMPLETE", # Post-treatment reservoirs drain not complete + 34: "REQUEST_REJECT_REASON_DG_NOT_IN_STANDBY_IDLE_STATE", # Treatment cannot initiate if DG is not in standby idle state + 35: "REQUEST_REJECT_REASON_INVALID_REQUEST_FORMAT", # Request message not formatted properly + 36: "REQUEST_REJECT_REASON_INVALID_DATE_OR_TIME", # Given date/time is invalid + 37: "REQUEST_REJECT_REASON____AVAILABLE_1", # Not used - available for use + 38: "REQUEST_REJECT_REASON_BATTERY_IS_NOT_CHARGED", # Battery does not have enough charge to start treatment + 39: "REQUEST_REJECT_REASON_RINSEBACK_NOT_COMPLETED", # Cannot move on to recirculate without completing full rinseback. + 40: "REQUEST_REJECT_REASON_RESERVOIR_ONE_IS_NOT_READY", # Reservoir one fill is not complete + 41: "REQUEST_REJECT_REASON_PUMP_TRACK_NOT_CLOSED", # Pump track is not ready / closed + 42: "REQUEST_REJECT_REASON_DOOR_NOT_CLOSED", # Door is not closed + 43: "REQUEST_REJECT_REASON_SYRINGE_NOT_PRESENT", # Syringe is not present + } + #create custom treatment PATIENT_ID = "demopatientid" TREATMENT_TITLE = "Create a Custom Treatment" @@ -125,4 +153,4 @@ OPERATIONAL_PARAMETERS = {"Heparin Type": HEPARIN_TYPE, "Acid Concentrate": "Fres. Naturalyte", "Bicarbonate Concentrate": BICARBONATE_CONCENTRATE, "Dialyzer Type": "BB Diacap Pro 16H"} - + \ No newline at end of file Index: shared/scripts/configuration/utility.py =================================================================== diff -u -rcea724e0b33dcd446f230ab23d6aa3729e6f86fd -ra4f1e264727015d92baa065dde52c33c0bc33ef3 --- shared/scripts/configuration/utility.py (.../utility.py) (revision cea724e0b33dcd446f230ab23d6aa3729e6f86fd) +++ shared/scripts/configuration/utility.py (.../utility.py) (revision a4f1e264727015d92baa065dde52c33c0bc33ef3) @@ -6,28 +6,48 @@ # WHOLE, WITHOUT THE EXPLICIT PERMISSION OF THE COPYRIGHT OWNER. # # @file utils.py -# -# @author (last) LTTS +# @author (last) LTTS # @date (last) 15-Jan-2022 # ############################################################################ - -import names -import test import object +import names +import sys import squish -from configuration import config +import time +import test + +from builtins import str as pyStr from builtins import int as pyInt +from configuration import config +from dialin.ui import utils +from datetime import datetime +from dialin.ui import utils +from dialin.common.hd_defs import HDOpModes +from dialin.ui.hd_simulator import HDSimulator +from dialin.ui.dg_simulator import DGSimulator +hd_simulator = HDSimulator() +dg_simulator = DGSimulator() - +def color_verification(exp_val = "Red", act_val = "#c53b33"): + """ + Function to verify item color verification + Argument: + exp_val - Expected required value + act_val - Color displayed on UI + Return: + handle the application for log + """ + test.compare(config.COLOR_CODES[color_name],(act_val.color[name])) + def check_if_object_is_within_the_container(obj=None, container=None): """ check if an object is inside a container - @param obj - child UI object - @param container - container UI object - @return boolean + @param obj - child UI object + @param container - container UI object + @return boolean true and false """ container = squish.findObject(container) containerPos = container.mapToGlobal(squish.QPoint(0, 0)) @@ -45,10 +65,9 @@ return False - def scroll_to_zone(zone=None, screen_object=None, direction = None): """ - scroll to the UI, if object is hidden + scroll to the UI, if object is hidden @param zone - object to be find out. @param screen_object - object of the screen. @return boolean @@ -64,6 +83,85 @@ else: raise RuntimeError except RuntimeError: + ScreenObj = squish.waitForObject(screen_object) + screenHeight = pyInt(ScreenObj.height) + screenWidth = pyInt(ScreenObj.width) + squish.mouseWheel(ScreenObj, screenWidth//2, + screenHeight//2, 0, -50, squish.Qt.NoModifier) + + raise LookupError("value object is not in view to the user after trying 100 times") + + +def pressure_pop_up_text_obj(text): + names.o_pop_up_pressure_text_obj["text"] = text + return names.o_pop_up_pressure_text_obj + +def pressure_text_obj(text): + names.o_pressure_text_obj["text"] = text + return names.o_pressure_text_obj + +def get_current_date_and_time(): + + date_format='%Y/%b/%d - %H:%M' + date = datetime.now() + return str(date.strftime(date_format)) + +def enter_keypad_value(entry): + """ + Method to enter user desired + value using keypad + @param entry: (str) User expected value + """ + test.startSection("Entering {}".format(entry)) + for value in entry: + value = pyInt(value) + key_val = squish.waitForObject(keypad_input(value)) + squish.mouseClick(key_val) + utils.waitForGUI(1) + test.endSection() + +def erase_entered_value(input_field): + """ + Method to erase the entered value + @param input_field - (obj) object of input field + """ + test.startSection("Erasing value") + input_field= squish.waitForObject(input_field) + entered_value = str(input_field.text) + for value in range(len(entered_value)+1): + utils.waitForGUI(1) + squish.mouseClick(squish.waitForObjectExists(names.o_back_space_key)) + + test.compare(str(input_field.text), "", "Input field should be empty") + test.endSection() + +def vitals_reading_obj(reading): + names.o_vitals_reading["text"] = reading + return names.o_vitals_reading + +def keypad_input(key_value): + names.o_keypad_input["text"] = key_value + return names.o_keypad_input + + +def scroll_to_zone(zone=None, screen_object=None, direction = None): + """ + scroll to the to the value if object is hidden + @param value - (obj) value object + @param container - (obj) Container of the value + @return boolean true and false + """ + counter = 0 + while counter <= 100: + try: + counter += 1 + squish.findObject(zone) + squish.snooze(0.5) + if check_if_object_is_within_the_container(obj=zone, container=screen_object): + return True + else: + raise RuntimeError + except RuntimeError: ScreenObj = squish.findObject(screen_object) screenHeight = pyInt(ScreenObj.height) screenWidth = pyInt(ScreenObj.width) @@ -74,9 +172,20 @@ squish.mouseWheel(ScreenObj, (screenWidth-100), -(screenHeight-700), 0, 200, squish.Qt.NoModifier) - raise LookupError("zone object is not in view to the user after " + \ - "trying 100 times") + raise LookupError("zone object is not in view to the user after trying 100 times") + +def get_alarm_id_obj(id): + names.o_alarm_id["text"] = id + return names.o_alarm_id +def get_alarm_msg_obj(msg): + names.o_alarm_message["text"] = msg + return names.o_alarm_message + +def rejection_msg(text): + names.o_rejection_msg["text"] = text + return names.o_rejection_msg + def verify_page_step_indicator(screen_obj, pre_treatment_step): """ @@ -116,8 +225,6 @@ test.compare(bullet_border_color,config.INCOMPLETE_COLOR, " the bullet border color must be " + str(config.INCOMPLETE_COLOR)) test.endSection() - - #Methods for create custom treatment def set_venous_ranges_min_val(ven_low): """ @@ -219,11 +326,73 @@ test.compare(high_handler_parent.maxValue, ven_max, "Venous range maximum value cannot be moved beyond {}".format(ven_max)) else: - test.compare(ven_max, ven_high, - "Actual Venous range maximum value: {} is equal to Expected value: {}".format(ven_max, ven_high)) - test.endSection() + x_value = 0 + counter = 0 + while True: + squish.mouseDrag(slider_object, x_value, 0, slider_value, 0, squish.Qt.NoModifier, + squish.Qt.LeftButton) + if counter == 50: + test.fail("User unable to fix slider slider_value") + return False + if bidirectional is True: + if slider_range == "Low": + if actual_value == pyFloat(slider_object.minValue): + return True + if slider_value >= pyFloat(slider_object.minValue): + slider_value = slider_value + 30 + if slider_value <= pyFloat(slider_object.minValue): + slider_value = slider_value - 30 + if slider_range == "Max": + if actual_value == pyFloat(slider_object.maxValue): + return True + if slider_value >= pyFloat(slider_object.maxValue): + slider_value = slider_value + 30 + if slider_value <= pyFloat(slider_object.maxValue): + slider_value = slider_value - 30 + counter = counter + 1 + + +def scroll_to_value_on_pop_up(value=None, container=None): + """ + scroll to the to the value if object is hidden + @param value - (obj) value object + @param container - (obj) Container of the value + @return boolean true and false + """ + counter = 0 + while counter <= 100: + try: + counter += 1 + squish.findObject(value) + squish.snooze(0.5) + if check_if_object_is_within_the_container(obj=value, container=container): + return True + else: + raise RuntimeError + except RuntimeError: + ScreenObj = squish.waitForObject(container) + screenHeight = pyInt(ScreenObj.height) + screenWidth = pyInt(ScreenObj.width) + squish.mouseWheel(ScreenObj, screenWidth//2, + screenHeight//2, 0, -50, squish.Qt.NoModifier) + + raise LookupError("value object is not in view to the user after " + \ + "trying 100 times") + +def expected_heparin_value(val): + names.o_heparin_value["text"] = val + return names.o_heparin_value +def msg(string: str): + """ + added ### at the right side of the string to make sure that it is a message. + @param string: (str) the string to add trailing ### to + @return: (str) padded string + """ + pad_str = "###"+string + return pad_str + def set_arterial_ranges_min_val(art_low): """ Method to set the Arterial range maximum value to user expected value @@ -325,18 +494,211 @@ test.compare(arterial_max, art_high, "Actual Arterial range maximum value: {} is equal to Expected value: {}".format(arterial_max, art_high)) test.endSection() + +def navigate_to_pretreatment_screen(mode): + """ + Method to navigate to sub mode under pre-treatment screen + @param mode - (int) pre treatment state + """ + hd_simulator.cmd_set_hd_operation_mode_data(HDOpModes.MODE_PRET.value,0) + hd_simulator.cmd_send_pre_treatment_state_data(sub_mode=mode, water_sample_state=0, + consumables_self_test_state=0, no_cartridge_self_test_state=0, + installation_state=0, dry_self_test_state=0, prime_state=0, + recirculate_state=0, patient_connection_state=0) + +def convert_seconds_into_min_and_sec(seconds): + min_and_sec = time.strftime("%M:%S", time.gmtime(seconds)) + return min_and_sec + + +def self_test_dry_check_list_text(text): + names.o_self_test_dry_check_list_text["text"] = text + return names.o_self_test_dry_check_list_text + +def navigate_to_pretreatment_screen(mode): + """ + Method to navigate to sub mode under pre-treatment screen + @param mode - (int) pre treatment state + """ + hd_simulator.cmd_set_hd_operation_mode_data(HDOpModes.MODE_PRET.value,0) + hd_simulator.cmd_send_pre_treatment_state_data(sub_mode=mode, water_sample_state=0, + consumables_self_test_state=0, no_cartridge_self_test_state=0, + installation_state=0, dry_self_test_state=0, prime_state=0, + recirculate_state=0, patient_connection_state=0) + +def convert_seconds_into_min_and_sec(seconds): + min_and_sec = time.strftime("%M:%S", time.gmtime(seconds)) + return min_and_sec + +def get_time(screen_title): + """ + Method to return the current count down + in the application + @param screen_title - (str) current title of the screen + @return time_text - (str) count down in the application + """ + if screen_title == config.BEGIN_PRIME_TITLE or screen_title == config.PRIMING_TITLE: + parent_object = object.parent(squish.waitForObjectExists(self_test_dry_check_list_text(screen_title))) + elif screen_title == config.SYSTEM_SELF_TEST_TITLE: + parent_object = object.parent(squish.waitForObjectExists(names.o_system_self_test)) + else: + parent_object = object.parent(squish.waitForObjectExists(names.o_PreTreatmentBase_Filter_Flush_Text)) + time_parent_children = object.children(parent_object) + progress_circle_parent = time_parent_children[4] + progress_circle_parent = object.children(progress_circle_parent) + progress_circle_parent = progress_circle_parent[0] + progress_circle_parent = object.children(progress_circle_parent) + progress_circle_children = object.children(progress_circle_parent[0]) + time_text = progress_circle_children[1] + return time_text.time + + +def verify_countdown(screen_title): + """ + Method to verify the count down + time in application + @param screen_title - (str) current title of the screen + """ + test.startSection("Verify the count down time in application") + for count_down in range(config.COUNT_DOWN_TIME_100, config.MINIMUM_COUNTDOWN_TIME-1, -1): + if screen_title == config.BEGIN_PRIME_TITLE: + hd_simulator.cmd_send_pre_treatment_self_test_dry_progress_data(300, count_down) + elif screen_title == config.PRIMING_TITLE: + hd_simulator.cmd_send_pre_treatment_disposables_prime_progress_data(300, count_down) + elif screen_title == config.SYSTEM_SELF_TEST_TITLE: + hd_simulator.cmd_send_pre_treatment_self_test_no_cartridge_progress_data(300, count_down) + else: + dg_simulator.cmd_send_dg_pre_treatment_filter_flush_progress_data(300, count_down) + + +def verify_countdown(screen_title): + """ + Method to verify the count down + time in application + @param screen_title - (str) current title of the screen + """ + test.startSection("Verify the count down time in application") + for count_down in range(config.COUNT_DOWN_TIME_100, config.MINIMUM_COUNTDOWN_TIME-1, -1): + if screen_title == config.BEGIN_PRIME_TITLE: + hd_simulator.cmd_send_pre_treatment_self_test_dry_progress_data(total=300, countdown=count_down) + elif screen_title == config.PRIMING_TITLE: + hd_simulator.cmd_send_pre_treatment_disposables_prime_progress_data(timeout=300, countdown=count_down) + elif screen_title == config.SYSTEM_SELF_TEST_TITLE: + hd_simulator.cmd_send_pre_treatment_self_test_no_cartridge_progress_data(total=300, countdown=count_down) + else: + dg_simulator.cmd_send_dg_pre_treatment_filter_flush_progress_data(total=300, countdown=count_down) + actual_time = get_time(screen_title) + expected_time = convert_seconds_into_min_and_sec(count_down) + test.compare(actual_time, expected_time, "Actual count down time: {} should be equal to expected count down time {}".format(actual_time, expected_time)) + verify_the_progress(count_down, screen_title) + test.endSection() + +def verify_the_progress(count_down, screen_title): + """ + Method to verify the current progress + @param count_down - (int) current count down time, screen_title: (str) current title of the screen + """ + test.startSection("Verifying the current progress") + if screen_title == config.BEGIN_PRIME_TITLE or screen_title == config.PRIMING_TITLE: + current_progress = (squish.waitForObjectExists(names.o_self_test_dry_progress)).progressValue + elif screen_title == config.SYSTEM_SELF_TEST_TITLE: + current_progress = (squish.waitForObjectExists(names.o_system_self_test_progress)).progressValue + else: + current_progress = (squish.waitForObjectExists(names.o_filter_flush_progress)).progressValue + #Since progress value is equal maximum count down value - current count down value + expected_progress = config.MAXIMUM_COUNTDOWN_TIME - count_down + test.compare(current_progress, expected_progress, "{} should be the current progress".format(expected_progress)) + test.endSection() + + +def verify_page_step_indicator(screen_obj, pre_treatment_step): + """ + Method to verify the Page Step indicators [the object on top of the screen which indicates the steps passed, current, remained] + @param pre_treatment_step : (int) indicates the Current pre-treatment step + """ + test.startSection("verification of page step indicators") + for page in range(len(config.PRE_TREATMENT_SCREENS)): + bullet_children = object.children(squish.waitForObjectExists(get_bullet_object(screen_obj, page))) + bullet_circle_color = bullet_children[0].color.name + bullet_border_color = bullet_children[0].border.color.name + step_title = squish.waitForObjectExists(get_text_object(screen_obj, config.PRE_TREATMENT_SCREENS[page])) + #To verify the step indicators of the completed pre treatment screens + if page < pre_treatment_step: + test.verify(squish.waitForObjectExists(get_bullet_object(screen_obj, page)).complete) + test.verify(not squish.waitForObjectExists(get_bullet_object(screen_obj, page)).current) + test.compare(bullet_circle_color, config.COMPLETE_COLOR) + test.compare(bullet_border_color, config.COMPLETE_COLOR) + test.compare(step_title.color.name, config.ENABLED_COLOR) + #To verify the step indicators of the current pre treatment screen + elif page == pre_treatment_step: + test.verify(squish.waitForObjectExists(get_bullet_object(screen_obj, page)).current,) + test.verify(not squish.waitForObjectExists(get_bullet_object(screen_obj, page)).complete) + test.compare(bullet_circle_color, config.CURRENT_COLOR) + test.compare(bullet_border_color, config.COMPLETE_COLOR) + test.compare(step_title.color.name, config.ENABLED_COLOR) + test.verify(step_title.font.bold) + #To verify the step indicators of the remaining pre-treatment screens + else: + test.verify(not squish.waitForObjectExists(get_bullet_object(screen_obj, page)).current,) + test.verify(not squish.waitForObjectExists(get_bullet_object(screen_obj, page)).complete,) + test.compare(step_title.color.name, config.INCOMPLETE_COLOR) + test.compare(bullet_circle_color, config.CURRENT_COLOR) + test.compare(bullet_border_color, config.INCOMPLETE_COLOR) + test.endSection() + +def verify_missing_object(object_to_check): + """ + Method to verify the given object is invisible or is not present on the screen + @param object_to_check: the object whose invisibility must be verified + """ + try: + squish.testSettings.objectNotFoundDebugging = False + squish.waitForObject(object_to_check,3000) + test.fail("Given object should not be present initially") + except LookupError as _: + test.passes("object is not present as expected") + + squish.testSettings.objectNotFoundDebugging = True + def get_text_object(screen_obj, txt): """ To obtain a text object based on text provided @param screen_obj: provides the container on which the txt must be present @returns a real name object - """ + """ names.o_text_object["container"] = screen_obj names.o_text_object["text"] = txt return names.o_text_object - +def verify_color_of_entry(entry, vital_parameter, input_field): + """ + Method to verify the color of entry + of systolic, diastolic and heart rate + @param entry: (int) user user entered value + @param vital_parameter - (str) parameter name under which user is entering value (sys/dia/heart rate) + @param input_field - (obj) object of input field + """ + test.startSection("Verify the color of {} value {}".format(vital_parameter, entry)) + input_field_color = input_field.color.name + entry = pyInt(entry) + if vital_parameter is config.SYSTOLIC_TEXT: + if (entry < config.SYSTOLIC_LOWER_LIMIT) or (entry > config.SYSTOLIC_UPPER_LIMIT): + test.compare(input_field_color, config.OUT_OF_RANGE_COLOR, "systolic value {} is out of range, systolic value should be in range of {} and {}".format(entry, config.SYSTOLIC_LOWER_LIMIT, config.SYSTOLIC_UPPER_LIMIT)) + elif (entry >= config.SYSTOLIC_LOWER_LIMIT) and (entry <= config.SYSTOLIC_UPPER_LIMIT): + test.compare(input_field_color, config.IN_RANGE_COLOR, "systolic value {} is in range of {} and {}".format(entry, config.SYSTOLIC_LOWER_LIMIT, config.SYSTOLIC_UPPER_LIMIT)) + elif vital_parameter is config.DIASTOLIC_TEXT: + if (entry < config.DIASTOLIC_LOWER_LIMIT) or (entry > config.DIASTOLIC_UPPER_LIMIT): + test.compare(input_field_color, config.OUT_OF_RANGE_COLOR, "diastolic value {} is out of range, diastolic value should be in range of {} and {}".format(entry, config.DIASTOLIC_LOWER_LIMIT, config.DIASTOLIC_UPPER_LIMIT)) + elif (entry >= config.DIASTOLIC_LOWER_LIMIT) and (entry <= config.DIASTOLIC_UPPER_LIMIT): + test.compare(input_field_color, config.IN_RANGE_COLOR, "diastolic value {} is in range of {} and {}".format(entry, config.DIASTOLIC_LOWER_LIMIT, config.DIASTOLIC_UPPER_LIMIT)) + elif vital_parameter is config.HEART_RATE_TITLE: + if (entry < config.HEART_RATE_LOWER_LIMIT) or (entry > config.HEART_RATE_UPPER_LIMIT): + test.compare(input_field_color, config.OUT_OF_RANGE_COLOR, "Heart Rate value {} is out of range, Heart Rate value should be in range of {} and {}".format(entry, config.HEART_RATE_LOWER_LIMIT, config.HEART_RATE_UPPER_LIMIT)) + elif (entry >= config.HEART_RATE_LOWER_LIMIT) and (entry <= config.HEART_RATE_UPPER_LIMIT): + test.compare(input_field_color, config.IN_RANGE_COLOR, "Heart Rate value {} is in range of {} and {}".format(entry, config.HEART_RATE_LOWER_LIMIT, config.HEART_RATE_UPPER_LIMIT)) + test.endSection() + def get_bullet_object(screen_obj, num): """ To obtain a bullet object based on occurrence provided. @@ -347,4 +709,45 @@ names.o_bullet_object["container"] = screen_obj names.o_bullet_object["occurrence"] = num + 1 return names.o_bullet_object + +def keypad_input(key_value): + """ + Method to enter values using application UI keyboard + @param key_value: (str) User expected value + """ + if key_value is not None: + names.o_keypad_input["text"] = key_value + return names.o_keypad_input + else: + test.log("Invalid text for object.") + names.o_keypad_input["text"] = None + + +def enter_keypad_value(entry): + """ + Method to enter user desired + value using keypad + @param entry: (str) User expected value + """ + test.startSection("Entering {}".format(entry)) + entry = pyStr(entry) #type casted into string format + for value in entry: + squish.mouseClick(squish.waitForObject(keypad_input(value))) + test.endSection() + +def erase_entered_value(input_field): + """ + Method to erase the entered value + @param input_field - (obj) object of input field + """ + test.startSection("Erasing value") + input_field= squish.waitForObject(input_field) + entered_value = str(input_field.text) + for value in range(len(entered_value)+1): + utils.waitForGUI(1) + squish.mouseClick(squish.waitForObjectExists(names.o_back_space_key)) + + test.compare(str(input_field.text), "", "Input field should be empty") + test.endSection() + Index: shared/scripts/names.py =================================================================== diff -u -rf36146a660d814576b22fbefaed163ba46058ec6 -ra4f1e264727015d92baa065dde52c33c0bc33ef3 --- shared/scripts/names.py (.../names.py) (revision f36146a660d814576b22fbefaed163ba46058ec6) +++ shared/scripts/names.py (.../names.py) (revision a4f1e264727015d92baa065dde52c33c0bc33ef3) @@ -22,6 +22,7 @@ o_minimizeButton_UpDownButton = {"container": o_Overlay, "gradient": 0, "id": "_minimizeButton", "type": "UpDownButton", "unnamed": 1, "visible": True} o_shape_Shape = {"container": o_Gui_MainView, "id": "_shape", "type": "Shape", "unnamed": 1, "visible": True} o_PreTreatmentCreateStack_PreTreatmentCreateStack = {"container": o_Gui_MainView, "objectName": "_PreTreatmentCreateStack", "type": "PreTreatmentCreateStack", "visible": True} +o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate = {"container": o_PreTreatmentCreateStack_PreTreatmentCreateStack, "objectName": "_PreTreatmentCreate", "type": "PreTreatmentCreate", "visible": True} o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase = {"container": o_PreTreatmentCreateStack_PreTreatmentCreateStack, "objectName": "_PreTreatmentBase", "type": "TreatmentFlowBase", "visible": True} o_PreTreatmentBase_titleBar_StepNavigationTitleBar = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "gradient": 0, "id": "_titleBar", "type": "StepNavigationTitleBar", "unnamed": 1, "visible": True} o_titleBar_Rectangle = {"container": o_Overlay, "gradient": 0, "id": "_titleBar", "type": "Rectangle", "unnamed": 1, "visible": True} @@ -44,8 +45,10 @@ o_PreTreatmentBase_backgroundRect_Rectangle = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "id": "_backgroundRect", "type": "Rectangle", "unnamed": 1, "visible": True} t_Text = {"container": o_Gui_MainView, "text": "t", "type": "Text", "unnamed": 1, "visible": True} o_PreTreatmentBase_confirmButton_TouchRect = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "objectName": "_confirmButton", "type": "TouchRect", "visible": True} +#page step indicator dynamic object property +o_text_object = {"type": "Text", "unnamed": 1, "visible": True} +o_bullet_object = {"type": "StepBullet", "unnamed": 1, "visible": True} - #standbymode o_MainHome_MainHome = {"container": o_Gui_MainView, "objectName": "_MainHome", "type": "MainHome", "visible": True} o_standby_page = {"container": o_MainHome_MainHome, "type": "Column", "unnamed": 1, "visible": True} @@ -445,8 +448,6 @@ #create_treatment #create_custom treatment testcase -o_text_object = {"type": "Text", "unnamed": 1, "visible": True} -o_bullet_object = {"type": "StepBullet", "unnamed": 1, "visible": True} o_PreTreatmentBase_confirmButton_TouchRect = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "text": "CONFIRM", "type": "Text", "unnamed": 1, "visible": True} o_PreTreatmentCreate_bloodFlowRate_SliderCreateTreatment = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "objectName": "_bloodFlowRate", "type": "SliderCreateTreatment", "visible": True} o_PreTreatmentBase_confirmButton_TouchRect_2 = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "objectName": "_confirmButton", "type": "TouchRect", "visible": True} @@ -471,6 +472,8 @@ o_MainHome_MainHome = {"container": o_Gui_MainView, "objectName": "_MainHome", "type": "MainHome", "visible": True} o_standby_page = {"container": o_MainHome_MainHome, "type": "Column", "unnamed": 1, "visible":True} o_PreTreatmentCreate_progressRect_ProgressRect = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "id": "_progressRect", "type": "ProgressRect", "unnamed": 1, "visible": True} +#page step indicator verification +o_PreTreatmentCreate_gridSteps_Grid = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "id": "_gridSteps", "type": "Grid", "unnamed": 1, "visible": True} #create custom treatment slider object property's #prescription object propert's o_PreTreatmentCreate_bloodFlowRateSlider_Slider = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "objectName": "_bloodFlowRateSlider", "type": "Slider", "visible": True} @@ -579,6 +582,9 @@ #text and bullet object o_text_object = {"type": "Text", "unnamed": 1, "visible": True} o_bullet_object = {"type": "StepBullet", "unnamed": 1, "visible": True} +o_PreTreatmentBase_Patient_ID_Text = {"container": o_PreTreatmentCreateStack_PreTreatmentBase_TreatmentFlowBase, "text": "Patient ID", "type": "Text", "unnamed": 1, "visible": True} +o_PreTreatmentCreate_PRESCRIPTION_Text = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "text": "PRESCRIPTION", "type": "Text", "unnamed": 1, "visible": True} +o_PreTreatmentCreate_OPERATING_PARAMETERS_Text = {"container": o_PreTreatmentCreateStack_PreTreatmentCreate_PreTreatmentCreate, "text": "OPERATING PARAMETERS", "type": "Text", "unnamed": 1, "visible": True} o_PreTreatmentBase_Indicators = {"container": o_PreTreatmentConnectionStack_PreTreatmentBase_TreatmentFlowBase_2, "type": "StepBullet", "unnamed": 1, "visible": True} #systolic, diastolic and heart rate input field