Index: Alarms/Alarms.conf =================================================================== diff -u -r16112cb383ed0ff50c88df5723e204adbe6dab88 -r3c1be356c12564e0ca2849acbc866020b49dcce6 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 16112cb383ed0ff50c88df5723e204adbe6dab88) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 3c1be356c12564e0ca2849acbc866020b49dcce6) @@ -1053,9 +1053,9 @@ SRSDG 945 = [193] -Title = -ListTitle = -Message = +Title = Blood Pump Stopped: Partial Occlusion +ListTitle = Blood Stop: Blood Pump Partial Occlusion +Message = Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [194] Title = Service Required: Dialysate Device