Index: Alarms/Alarms.conf =================================================================== diff -u -r8b9e523bba4d7064f94af468d21538b2847c3ad8 -r801eb9280285a442d21be69da25d961638fbdf1f --- Alarms/Alarms.conf (.../Alarms.conf) (revision 8b9e523bba4d7064f94af468d21538b2847c3ad8) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 801eb9280285a442d21be69da25d961638fbdf1f) @@ -3,1075 +3,1075 @@ Message = [1] -Title = HD stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up -Message = HD +Title = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up +Message = Stop or Off button are pressed for more than 1 sec during POST. [2] Title = HD FPGA POST failure -Message = HD +Message = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. [3] Title = DG FPGA POST failure -Message = DG +Message = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = HD Watchdog POST failure -Message = HD +Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA [5] Title = DG Watchdog POST failure -Message = DG +Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V SRSDG 53 = [6] -Title = HD UI communication POST failure. UI failed to communicate within a reasonable time after power up -Message = HD +Title = UI communication POST failure. UI failed to communicate within a reasonable time after power up +Message = If the UI failed to checkin via the CAN bus within 40 seconds [7] -Title = HD RTC battery low -Message = HD +Title = Real-time clock configuration error +Message = If the RTC is not functional as indicated by the bits in the first 3 control registers. [8] Title = HD accelerometer failed POST -Message = HD +Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. [9] Title = DG accelerometer failed POST -Message = DG +Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. SRSDG 323 = [10] -Title = HD blood leak sensor setting embedded mode failure. -Message = HD +Title = HD blood leak sensor embedded mode communication failed +Message = If the blood leak sensor embedded mode is unresponsive. [11] -Title = HD dialysate temperature below target temperature or below safety temp -Message = HD +Title = Dialysate temperature is below the target temperature +Message = If the dialysate temperature is below the target temperature for a certain period of time. [12] Title = HD firmware image integrity POST test failed -Message = HD +Message = If the calculated CRC of the HD image does not match the expected CRC. [13] Title = DG firmware image integrity POST test failed -Message = DG +Message = If the calculated CRC of the DG firmware image does not matched the expected CRC SRSDG 813 = [14] -Title = DG invalid usage record CRC -Message = DG +Title = DG non-volatile usage information group invalid record CRC +Message = If the DG usage information record's CRC does not match the calculated CRC in POST. SRSDG 957 = [15] -Title = HD dialysate temperature above high safety. -Message = HD +Title = Dialysate temperature is out of high safety range +Message = If the dialysate temperature is > 46 C for a certain period of time. [16] Title = HD alarm audio failed POST -Message = HD +Message = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output [17] Title = HD UI POST failed -Message = HD +Message = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. [18] -Title = HD didn't get ACK on message to DG that required acknowledgment -Message = HD +Title = HD to DG comm error. +Message = HD message to DG that requires acknowledge was not acknowledged within a certain time. [19] -Title = HD dialysate temperature above target temperature. -Message = HD +Title = Dialysate temperature is above the target temperature +Message = If the dialysate temperature is above the target temperature for a certain period of time. [20] Title = DG conductivity sensors invalid calibration record -Message = DG +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 16 = [21] Title = DG drain line volume invalid calibration record -Message = DG +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 936 = [22] Title = DG reservoirs invalid calibration record -Message = DG +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 935 = [23] Title = DG acid concentrate calibration record -Message = DG +Message = If the non-volatile data is valid and if not raise the alarm SRSDG 934 = [24] Title = DG bicarb concentrate calibration record -Message = DG +Message = If the non-volatile data is valid and if not raise the alarm SRSDG 933 = [25] Title = DG accelerometer invalid calibration record -Message = DG +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 817 = [26] Title = HD accelerometer invalid calibration record -Message = HD +Message = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. [27] -Title = HD blood leak sensor zero and self test sequence failed -Message = HD +Title = HD BLD sensor failed zero sequence +Message = BLD fails to complete zero sequence. [28] Title = DG two wire sensors FPGA fault -Message = DG +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 864 = [29] Title = HD heparin force sensor invalid calibration record -Message = HD +Message = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. [30] Title = HD Software fault. Software found itself in an unexpected state -Message = HD +Message = If the specific software faults listed in alarm management SW faults has been detected [31] -Title = HD blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = HD +Title = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Message = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. [32] -Title = HD blood pump failed motor off check. Measured speed while commanded off -Message = HD +Title = Blood pump failed motor off check. Measured speed while commanded off +Message = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM [33] -Title = HD blood pump failed motor direction check. Measured vs. commanded -Message = HD +Title = Blood pump failed motor direction check. Measured vs. commanded +Message = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. [34] -Title = HD blood pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = HD +Title = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds +Message = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. [35] -Title = HD dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = HD +Title = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Message = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. [36] -Title = HD dialysis inlet pump failed motor off check. Measured speed while commanded off -Message = HD +Title = Dialysis inlet pump failed motor off check. Measured speed while commanded off +Message = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. [37] -Title = HD dialysis inlet pump failed motor direction check. Measured vs. commanded -Message = HD +Title = Dialysis inlet pump failed motor direction check. Measured vs. commanded +Message = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. [38] -Title = HD dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = HD +Title = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds +Message = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. [39] -Title = HD dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = HD +Title = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Message = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. [40] -Title = HD dialysis outlet pump failed motor off check. Measured speed while commanded off -Message = HD +Title = Dialysis outlet pump failed motor off check. Measured speed while commanded off +Message = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. [41] -Title = HD dialysis outlet pump failed motor direction check. Measured vs. commanded -Message = HD +Title = Dialysis outlet pump failed motor direction check. Measured vs. commanded +Message = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. [42] -Title = HD dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = HD +Title = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds +Message = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed [43] -Title = DG main primary heater FPGA fault -Message = DG +Title = Main primary heater FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 954 = [44] -Title = HD UI communication timeout -Message = HD +Title = User interface communication timeout +Message = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. [45] -Title = HD too many bad communications CRC -Message = HD +Title = Too many bad CRCs detected on received system messages +Message = If too many bad messages CRCs are received within a set period of time [46] -Title = HD didn't get ACK on message to UI that required acknowledgment -Message = HD +Title = System message that required acknowledgment was not acknowledged +Message = If no retries left in the pending acknowlegement list. [47] -Title = Available for use +Title = Message = [48] -Title = HD ultrafiltration volume accuracy error during treatment -Message = HD +Title = Ultrafiltration volume accuracy error during treatment +Message = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. [49] Title = HD FPGA communication down for too long -Message = HD +Message = If there are more than 3 comm errors in a 1 min period. [50] Title = DG FPGA not accepting commanded valve states -Message = DG +Message = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. SRSDG 26 = [51] -Title = HD blood pump failed motor speed check. Measured vs. commanded -Message = HD +Title = Blood pump failed motor speed check. Measured vs. commanded +Message = If measured motor speed versus commanded rotor speed exceeded limits. [52] -Title = HD dialysate inlet pump failed motor speed check. Measured vs. commanded -Message = HD +Title = Dialysate inlet pump failed motor speed check. Measured vs. commanded +Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. [53] -Title = HD dialysate outlet pump failed motor speed check. Measured vs. commanded -Message = HD +Title = Dialysate outlet pump failed motor speed check. Measured vs. commanded +Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. [54] Title = HD critical data integrity check failed -Message = HD +Message = If the critical data failed range check. [55] Title = DG critical data integrity check failed -Message = DG +Message = If the critical data failed range check. SRSDG 921 = [56] Title = HD accelerometer error (no readings or FPGA reports error) -Message = HD +Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. [57] Title = DG accelerometer error (no readings or FPGA reports error) -Message = DG +Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. SRSDG 869 SRSDG 870 = [58] Title = HD valve homing failed -Message = HD +Message = If the number of failed homing > = 3 counts. [59] Title = HD valve transition time out -Message = HD +Message = if the valve's transition time has timed out (3000msec) [60] Title = HD valve not functional -Message = HD +Message = If the valve bit masks are not as expected in different valve control mode. [61] Title = HD valve current out of range -Message = HD +Message = If the valve current is over the threshold of 1.0 A for 1000 msec. [62] Title = HD valve position out of target -Message = HD +Message = If the current position has deviated from the expected position for more than 1 second. [63] -Title = HD blood leak FPGA fault -Message = HD +Title = FPGA reports error +Message = If the FPGA error counter increments for a certain period of time. [64] -Title = DG barometric pressure sensor out of range -Message = DG +Title = Barometric pressure is out of range +Message = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. SRSDG 874 = [65] -Title = HD DG requests DG command with invalid parameter fault -Message = HD +Title = HD requests DG command with invalid parameter fault +Message = If the latest DG command response indicated invalid parameter [66] Title = HD blood leak sensor set point set failure -Message = HD +Message = If the set point of the blood leak sensor does not match the set point in the calibration record. [67] Title = HD blood pump occlusion self-test failure alarm -Message = HD +Message = If the blood pump pressure > 5000 [68] Title = HD active reservoir recirculation out of range -Message = HD +Message = If the recirculation level has exceeded the max allowed [69] Title = HD blood leak sensor invalid calibration record -Message = HD +Message = If the BLD data that was received from NVDataMgmt is NOT legitimate. [70] Title = HD arterial pressure self-test failure alarm -Message = HD +Message = If arterial pressure < -300 or > 100 mmHg [71] Title = HD venous pressure self-test failure alarm -Message = HD +Message = If the venous pressure < -100 or > 600 mmHg [72] Title = HD No load cell data message receive at least once every 2 seconds -Message = HD +Message = No new load cell data for 2 seconds. [73] -Title = HD No dialysate temperature data message receive at least once every 2 seconds -Message = HD +Title = HD No dialysate temperature data message received at least once every 2 seconds +Message = No new dialysate temperature data for 2 seconds. [74] -Title = DG inlet UV reactor is on with no flow -Message = DG +Title = DG inlet UV reactor is on while there is no flow +Message = If the inlet UV reactor is on while VPi is closed for a certain period of time. SRSDG 600 = [75] Title = HD syringe pump self-test failure alarm -Message = HD +Message = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. [76] Title = HD monitored voltage is out of range -Message = HD +Message = If the HD monitored voltages are out of range for a certain period of time. [77] Title = DG monitored voltage is out of range -Message = DG +Message = If the DG monitored voltages are out of range for a certain period of time. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = HD syringe pump direction (from encoder) error -Message = HD +Message = If the measured encoder direction of the syringe pump does not agree with the expected direction [79] Title = HD syringe pump direction (from controller) error -Message = HD +Message = If the measured direction of the syringe pump motor controller does not agree with the expected direction [80] Title = HD syringe pump fault reported by FPGA -Message = HD +Message = If the FPGA reported the syringe pump fault [81] Title = HD syringe pump over travel error -Message = HD +Message = If the syringe pump position is deemed to have overtraveled [82] Title = HD syringe pump DAC write failure -Message = HD +Message = If the syringe pump DAC write error is reported by the FPGA [83] Title = HD syringe pump is running while the BP is off -Message = HD +Message = If the syringe pump is on while the blood pump is off [84] -Title = DG set RTC year is invalid -Message = DG +Title = DG RTC set year is invalid +Message = If DG RTC set year is < 2021. SRSDG 815 = [85] -Title = HD pump track latch opened alarm -Message = HD +Title = HD pump track lever is not fully latched +Message = If the pump track lever is detected as unlatched when it should be. [86] -Title = HD set RTC year is invalid -Message = HD +Title = HD RTC set year is invalid +Message = If HD RTC set year is < 2021. [87] -Title = DG heating invalid calibration record -Message = DG +Title = DG heating parameters invalid calibration record +Message = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. SRSDG 929 = [88] -Title = DG concentrate pumps hall sensor out of range -Message = DG +Title = DG concentrate pump hall sensor out of range. +Message = Hall sensor pulse to pulse timing is < 1ms (speed too fast). [89] -Title = DG outlet UV reactor on with no flow -Message = DG +Title = DG outlet UV reactor is on while there is no flow +Message = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. SRSDG 599 = [90] Title = DG load cells A1/B1 FPGA fault -Message = DG +Message = If the FPGA does not report fresh data for a certain period of time. SRSDG 604 = [91] -Title = HD No dialysate flow data receive in the last 3 seconds -Message = HD +Title = HD No dialysate flow data received in the last 3 seconds +Message = If not receiving new dialysate flow readings in expected interval (3 seconds) [92] Title = Message = [93] -Title = DG outlet primary conductivity out of range -Message = DG +Title = DG outlet primary conductivity sensor out of range +Message = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. SRSDG 430 = [94] -Title = DG pressure out of range -Message = DG +Title = DG pressure sensor out of range +Message = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. SRSDG 629 = [95] Title = DG watchdog expired -Message = DG +Message = If the watchdog signal is low while not in POST. SRSDG 907 = [96] -Title = DG inlet water temperature in high range -Message = DG +Title = DG inlet water temperature out of range +Message = If the inlet water temperature is > 37 C. SRSDG 410 = [97] Title = DG fill conductivity out of range -Message = DG +Message = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. SRSDG 397 = [98] -Title = HD battery communication fault -Message = HD +Title = HD battery or battery charger communication fault +Message = If battery I2C communication fault persistence count > 5. [99] Title = HD syringe pump stall alarm -Message = HD +Message = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. [100] Title = HD no cartridge self-test timeout -Message = HD +Message = If during the no cartridge self-tests state, the elapsed self-test time > maximum no cartridge self test time (30secs) [101] -Title = UI POST HD communication failure -Message = UI +Title = UI POST FW Communication failure +Message = If the HD is not communicating with the UI during POST. [102] -Title = DG heat disinfect temperature gradient out of range -Message = DG +Title = DG heat disinfection temperature gradient out of range +Message = If temperature gradient between Tpo and Thd exceeds maximum 15°C. SRSDG 856 SRSDG 930 = [103] -Title = HD invalid calibration CRC -Message = HD +Title = HD non-volatile calibration group invalid record CRC +Message = If the computed CRC of the calibration group record is not equal to the record CRC. [104] -Title = HD air trap level sensors reporting illegal combination of air/fluid -Message = HD +Title = Air trap level sensors reporting illegal combination of air/fluid +Message = If the air trap level readings are not valid for more than 2 seconds. [105] -Title = DG invalid calibration CRC -Message = DG +Title = DG non-volatile system group invalid record CRC +Message = If the DG NVDATAMGMT CAL group calculated CRC does not mached the expected CRC. SRSDG 608 = [106] -Title = DG dialysate flow sensor invalid calibration record -Message = DG +Title = DG flow sensors calibration error +Message = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = HD reports DG restarted fault -Message = HD +Message = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. [108] -Title = HD syringe pump ADC FPGA fault -Message = HD +Title = HD syringe pump ADC error +Message = If the ADC read is not fresh after POST init and the ADC read counter has not changed. [109] Title = HD syringe pump volume check error -Message = HD +Message = If the volume delivered is off by more than 5% or 0.1mL. [110] Title = HD syringe pump speed check error -Message = HD +Message = If the syringe measured rate is off by more than 5% or 0.1mL. [111] Title = HD syringe pump not stopped in off state error -Message = HD +Message = If the syringe pump position is not equal to its last position. [112] -Title = DG load cells A1/B1 FPGA fault -Message = DG +Title = DG load cells A2/B2 FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time. SRSDG 953 = [113] -Title = DG Drain pump current out of range -Message = DG +Title = DG drain pump current out of range +Message = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. SRSDG 890 = [114] Title = HD venous air bubble detector self-test failure -Message = HD +Message = If the venous air bubble self test did not detect bubble for more than 500 msec. [115] Title = DG temperature sensor out of range -Message = DG +Message = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. SRSDG 224 = [116] -Title = HD UI SD card failure -Message = UI HD +Title = SD card is non-operational +Message = SD card is non-operational [117] -Title = DG CAN message not acked -Message = DG +Title = System message that required acknowledgment was not acknowledged +Message = If no retries left in the pending acknowlegement list. SRSDG 909 = [118] -Title = DG RTC config error -Message = DG +Title = DG RTC configuration error +Message = If the RTC configuration registers report an error. SRSDG 43 = [119] -Title = DG RTC battery low error -Message = DG +Title = DG RTC battery low +Message = If the RTC battery bit report a low alarm. SRSDG 908 = [120] -Title = HD pre-treatment mode wet self-test prime check failure -Message = HD +Title = HD pre-treatment wet prime test failure +Message = If the HD pre-treatment wet self test failed. [121] Title = DG main primary heater voltage out of range -Message = DG +Message = If the main primary heater's voltage is out of range. SRSDG 611 = [122] Title = DG small primary heater voltage out of range -Message = DG +Message = If the small primary heater's voltage is out of range. SRSDG 612 = [123] Title = DG trimmer heater voltage out of range -Message = DG +Message = If the trimmer voltage is out of range. SRSDG 613 = [124] Title = HD end of treatment alarm (high priority - user still not acting to end treatment) -Message = HD +Message = In treatment end paused state for > 5 minutes. [125] -Title = HD blood sitting too long after treatment stopped by user alarm (>5 min) -Message = HD +Title = Blood sitting too long after treatment stopped by user alarm (> 5 min) +Message = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. [126] Title = HD blood leak detected alarm -Message = HD +Message = If the blood leak is detected based on the blood leak status reported by the sensor. [127] -Title = HD venous pressure too low during treatment -Message = HD +Title = Venous pressure too low during treatment +Message = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. [128] Title = HD venous air bubble detected alarm -Message = HD +Message = If bubbles are detected during the treatment. [129] -Title = Available for use +Title = Message = [130] -Title = HD venous pressure too high during treatment -Message = HD +Title = Venous pressure too high during treatment +Message = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. [131] -Title = HD arterial pressure too low during treatment -Message = HD +Title = Arterial pressure too low during treatment +Message = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. [132] -Title = HD arterial pressure too high during treatment -Message = HD +Title = Arterial pressure too high during treatment +Message = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. [133] Title = DG fluid leak detected alarm -Message = DG +Message = If the fluid leak is detected as reported by the fluid leak detector. SRSDG 328 = [134] Title = HD fluid leak detected alarm -Message = HD +Message = If the fluid leak is detected as reported by the fluid leak detector. [135] Title = HD experienced large acceleration -Message = HD +Message = If the system experienced excessive shock > 2.5 g. [136] Title = DG experienced large acceleration -Message = DG +Message = If the system experienced excessive shock > 2.5 g. SRSDG 327 = [137] Title = HD tilt exceeds maximum -Message = HD +Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. [138] Title = DG tilt exceeds maximum -Message = DG +Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. SRSDG 326 = [139] Title = HD AC power lost alarm -Message = HD +Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. [140] -Title = HD DG communication timeout -Message = HD +Title = Dialysate generator communication timeout +Message = If DG communication timeout is detected for more than 2 seconds. [141] -Title = HD air trap fill timeout during treatment -Message = HD +Title = Air trap fill timeout during treatment +Message = If the air trap fill timeout during treatment. [142] -Title = HD blood pump occlusion detected -Message = HD +Title = Blood pump occlusion detected +Message = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall [143] -Title = DG dialysate temperature sensors out of range -Message = DG +Title = DG dialysate temperature sensors drift timeout. +Message = If TDi and TRo have drifted for more 1 C for a certain period of time. SRSDG 912 = [144] -Title = DG cleaning mode inlet water conductivity out of high range. -Message = DG +Title = DG cleaning mode inlet water conductivity out of high range +Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = DG concentrate conductivity after adding acid out of range alarm -Message = DG +Title = DG Concentrate conductivity after adding acid out of range alarm +Message = If the conductivity value for the acid concentrate is out of range for more than 45 secs. SRSDG 846 = [146] -Title = DG RTC or timer accuracy failure. -Message = DG +Title = DG RTC vs. Timer accuracy error +Message = If RTC and timer are not aligned on 1 second elapsed time. SRSDG 312 = [147] -Title = DG wait for the DG to produce dialysate -Message = DG +Title = DG dialysate generation conductivity fault alarm +Message = If dialysate conductivity is not with 5% of the TCD SRSDG 398 = [148] -Title = DG Cleaning mode inlet water temperature too high -Message = DG +Title = Cleaning Mode Inlet water temperature too high +Message = Inlet temperature > 37 C. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = HD watchdog expired -Message = HD +Message = If the watchdog signal is low while not in POST. [150] -Title = DG inlet water conductivity in high range -Message = DG +Title = Inlet water conductivity too high +Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). SRSDG 391 = [151] -Title = DG inlet water conductivity in low range -Message = DG +Title = Inlet water conductivity too low +Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). SRSDG 826 = [152] -Title = DG inlet water pressure in low range -Message = DG +Title = Inlet water pressure too low +Message = Inlet water pressure < 25 PSI. SRSDG 417 = [153] Title = HD prime completed high priority alarm -Message = HD +Message = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) [154] -Title = DG cleaning mode inlet water conductivity out of low range. -Message = DG +Title = Cleaning mode inlet water conductivity too low +Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = HD No reservoirs data message receive at least once every 2 seconds -Message = HD +Title = HD No reservoir data message receive at least once every 2 seconds +Message = No new reservoir data for 2 seconds. [156] -Title = HD No DG operation mode message receive at least once every 2 seconds -Message = HD +Title = HD No DG op mode message receive at least once every 2 seconds +Message = No DG op mode for 2 seconds. [157] -Title = DG chemical disinfect prime acid line timeout. -Message = DG +Title = Chemical disinfect priming the acid line timeout +Message = If the disinfectant acid line is not primed within a certain period of time. SRSDG 988 = [158] -Title = DG inlet water temperature is in the low range -Message = DG +Title = Inlet water temperature low +Message = Inlet temperature < 24 °C for 5 seconds. SRSDG 408 = [159] -Title = DG chemical disinfect flush remove acid and close the concentrate cap. -Message = DG +Title = Close concentrate line be able to flush it +Message = Start of chemical disinfect flush. SRSDG 1044 = [160] -Title = HD blood pump rotor speed too high -Message = HD +Title = Blood pump rotor speed too high +Message = If blood pump motor speed exceeded 100.0 RPM. [161] -Title = DG comm too many bad CRCs -Message = DG +Title = Too many bad CRCs detected on received system messages +Message = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. SRSDG 122 = [162] -Title = DG FPGA clock speed check failure. -Message = DG +Title = FPGA clock speed failure +Message = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. SRSDG 867 = [163] Title = HD sees primary load cell for reservoir 1 change too much too fast -Message = HD +Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). [164] Title = HD sees primary load cell for reservoir 2 change too much too fast -Message = HD +Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). [165] Title = HD in treatment rinseback operation timeout -Message = HD +Message = If the maximum time inrinseback operation has been exceeded. [166] -Title = Available for use +Title = Message = [167] Title = HD cartridge door opened alarm -Message = HD +Message = If the cartridge door is opened as reported by the FPGA. [168] -Title = Available for use +Title = Message = [169] -Title = DG Dialysate flow rate out of maximum range -Message = DG +Title = Dialysate flow rate out of range +Message = If the dialysate flow rate > 2.00 L/min. SRSDG 643 = [170] Title = HD syringe empty alarm -Message = HD +Message = If the syringe pump position is near empty position. [171] Title = HD syringe pump occlusion alarm -Message = HD +Message = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. [172] Title = HD syringe pump not enough Heparin alarm -Message = HD +Message = If the estimated syringe volume needed for treatment is not sufficient [173] Title = HD RTC configuration error -Message = HD +Message = If the RTC configuration registers report an error. [174] -Title = HD RTC or timer accuracy failure -Message = HD +Title = HD RTC vs. Timer accuracy error +Message = If RTC and timer are not aligned on 1 second elapsed time. [175] -Title = HD pump direction status error -Message = HD +Title = HD pump commutation error +Message = If the set pump direction does not agree with the hall sensor direction. [176] -Title = HD dialysate temperature above high safety range +Title = Message = [177] Title = DG software fault. Software found itself in an unexpected state -Message = DG +Message = If the specific software faults listed in alarm management SW faults has been detected SRSDG 919 = [178] Title = HD communication timeout -Message = UI +Message = If the HD stops communicating with the UI. = SRSUI 1066 [179] Title = DG FPGA communication down for too long -Message = DG +Message = If retries for commands exceeds limit or FPGA reports comm error. SRSDG 605 = [180] -Title = DG RO flow out of maximum range -Message = DG +Title = RO flow rate out of range +Message = If the RO flow rate > 2.00 L/min. SRSDG 883 = [181] Title = DG load cells weight out of range for tare -Message = DG +Message = If the load cell weight is out of range for tare. SRSDG 849 SRSDG 838 = [182] Title = DG load cells invalid calibration -Message = DG +Message = If the CRC of load cells calibration data does not match the calculated CRC. SRSDG 162 = [183] Title = DG invalid load cell value -Message = DG +Message = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. SRSDG 851 = [184] -Title = DG inlet UV reactor not healthy -Message = DG +Title = Inlet reactor not healthy +Message = If the inlet reactor are not healthy. SRSDG 559 = [185] Title = DG fan RPM out of range -Message = DG +Message = If the fan RPM is out of Min/Max range. SRSDG 628 = [186] Title = DG concentrate pump fault -Message = DG +Message = If the concentrate pumps fault is reported by the FPGA. SRSDG 891 = [187] Title = DG concentrate pump CP1 speed control error -Message = DG +Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. SRSDG 905 = [188] Title = DG concentrate pump CP2 speed control error -Message = DG +Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. SRSDG 906 = [189] Title = DG drain pump RPM out of range -Message = DG +Message = If the drain pump RPM > 100.0 of the target RPM. SRSDG 448 = [190] Title = DG drain pump off fault -Message = DG +Message = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. [191] Title = DG flow rate out of upper range -Message = DG +Message = If the RO flow rate > 10% from the target flow rate. SRSDG 885 = [192] Title = DG flow rate out of lower range -Message = DG +Message = If the RO flow rate < 10% from the target flow rate. SRSDG 945 = [193] -Title = HD dialysate temperature out of low safety range. +Title = Message = [194] Title = DG RO pump duty cycle out of range -Message = DG +Message = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. SRSDG 881 = [195] Title = DG RO pump pressure out of range -Message = DG +Message = If the DG RO pump pressure > 135 psi for a certain period of time. SRSDG 925 = [196] -Title = DG CPi/CPo sensors FPGA fault -Message = DG +Title = DG CPi/CPo conductivity sensors FPGA fault +Message = If the FPGA reports error for a certain period of time. SRSDG 955 = [197] -Title = DG CD1/CD2 sensors FPGA fault -Message = DG +Title = DG CD1/CD2 conductivity sensors FPGA fault +Message = If the FPGA reports error for a certain period of time. SRSDG 956 = [198] -Title = DG RO flow too low while primary heater is on -Message = DG +Title = DG primary heater while RO flow is low +Message = If the RO flow is < 0.2 L/min for a certain period of time. SRSDG 872 = [199] -Title = DG Dialysate flow too low while trimmer heater is on -Message = DG +Title = DG trimmer heater while dialysate flow is low +Message = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. SRSDG 871 = [200] -Title = DG thermistors/sensors temperature out of range -Message = DG +Title = DG thermistor or temperature sensor out of range +Message = If thermistor > 80.0 C or < 0.0 C. SRSDG 178 = [201] -Title = HD pre-treatment mode wet self-test lc vs FMD failure -Message = HD +Title = HD pre-treatment wet flow test failure +Message = If the pre-treatment wet flow test has failed. [202] -Title = HD pre-treatment mode dry pressure normal self-test failure -Message = HD +Title = HD pre-treatment dry pressure normal test failure +Message = If the pre-treatment dry pressure normal test failed. [203] -Title = Available for use +Title = Message = [204] -Title = DG RO rejection ratio out of range -Message = DG +Title = RO rejection ratio out of range +Message = If RO rejection ratio > 0.1 SRSDG 429 = [205] -Title = DG conductivity sensor fault -Message = DG +Title = Conductivity sensor fault +Message = If conductivity sensor reports 0x11 for more than 2 consecutive reads. SRSDG 913 = [206] Title = DG dialysate fill runs out of time -Message = DG +Message = If DG dialysate fill > 5 minutes. SRSDG 437 = [207] Title = DG flow meter check failure alarm -Message = DG +Message = If integrated Volume To Load Cell Reading Percent > 0.1. SRSDG 240 = [208] -Title = HD venous pressure occlusion alarm -Message = HD +Title = HD venous line occlusion fault +Message = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. [209] Title = DG drain circulation line timeout -Message = DG +Message = If the integrated volume < 0.1 for more than 60 secs. SRSDG 980 = [210] Title = HD battery pack detected an error -Message = HD +Message = If the battery pack status error is detected from the slave device I2C address 0x0B. [211] -Title = HD blood sitting too long warning (>4 min) -Message = HD +Title = Blood sitting too long warning (> 4 min) +Message = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. [212] -Title = HD end of treatment alarm (user not acting to end treatment) -Message = HD +Title = Blood sitting too long warning (> 4 min) +Message = During treatment end paused state, if the blood is in the line for more than 4 minutes. [213] Title = HD prime completed medium priority alarm -Message = HD +Message = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) [214] -Title = Available for use +Title = Message = [215] -Title = Available for use +Title = Message = [216] Title = HD end treatment sub-mode timeout alarm -Message = HD +Message = If the treatment ended successfully. [217] -Title = Available for use +Title = Message = [218] Title = HD syringe detected alarm -Message = HD +Message = If the syringe detection switch > 2.0 volts. [219] Title = HD syringe pump syringe removed alarm -Message = HD +Message = If the syringe detection switch < 2.0 volts. [220] -Title = DG THd sensors FPGA fault -Message = DG +Title = DG THd sensor FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 947 = [221] Title = HD Saline bag is empty -Message = HD +Message = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. [222] Title = HD occlusion sensor FPGA fault -Message = HD +Message = If the FPGA does not report fresh data or indicates an error for a certain period of time. [223] Title = HD arterial sensor FPGA fault -Message = HD +Message = If the FPGA does not report fresh data for a certain period of time [224] -Title = HD treatment stopped by user action - pressed stop button -Message = HD +Title = Treatment stopped by user action - pressed stop button +Message = If the user has pressed the stop button. [225] Title = HD end of treatment warning -Message = HD +Message = If the treatment duration has exceeded the prescribed treatment time entered by the user. [226] Title = HD prime completed low priority alarm -Message = HD +Message = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection [227] -Title = Available for use +Title = Message = [228] -Title = HD time out on prime saline purge air state. -Message = HD +Title = HD time out on prime sub-mode purge air state. +Message = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. [229] Title = HD prime dialysate dialyzer time out alarm -Message = HD +Message = If the priming of the dialyzer dialysate flow path has exceeded the time limit. [230] Title = HD prime dialysate bypass time out alarm -Message = HD +Message = If the priming of the dialyzer bypass flow path has exceeded the time limit. [231] -Title = HD pre-treatment mode dry pressure self-test failure -Message = HD +Title = HD pre-treatment mode dry self-tests failure +Message = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState [232] -Title = HD pre-treatment mode wet self-test LC vs. LC failure -Message = HD +Title = HD pre-treatment mode wet self-tests failure +Message = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState [233] -Title = DG barometric sensor coefficients bad CRC -Message = DG +Title = DG barometric pressure sensor bad coefficients CRC +Message = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = DG inlet water pressure in high range. -Message = DG +Title = DG inlet water pressure is too high +Message = If the inlet water pressure is > 80 psig. SRSDG 1037 = [235] Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) -Message = HD +Message = If there is no active alarm and rinseback has completed. [236] Title = HD needs new cartridge to be installed -Message = HD +Message = If the air trap sensor reported AIR and bubble sensor detected bubbles. [237] Title = HD prime saline dialyzer time out alarm -Message = HD +Message = If timeout, just transition to HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE [238] Title = HD no cartridge loaded or installed improperly alarm -Message = HD +Message = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. [239] Title = HD fail to remove cartridge alarm -Message = HD +Message = In the post treatment state, if the cartridge has not been removed. [240] -Title = DG bicarb conductivity out of range during bicarb pump check alarm -Message = DG +Title = Bicarb conductivity out of range during bicarb pump check alarm +Message = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. SRSDG 482 = [241] Title = DG reservoir drain time out -Message = DG +Message = If the DG reservoir drain is not complete in 120 secs. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1103,7 +1103,7 @@ [242] Title = DG reservoir fill time out -Message = DG +Message = If the DG reservoir fill is not complete in 120 secs. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1125,372 +1125,372 @@ [243] Title = DG reservoir leak time out -Message = DG +Message = If the reservoir level is drifted by 100 mL for more than 30 seconds. SRSDG 853 = [244] -Title = DG cleaning mode temperature sensors difference out of range -Message = DG +Title = DG temperature sensors difference out of range +Message = If any of the temperature sensors are out of the average of the sensors by more than 3 C. SRSDG 852 = [245] Title = DG heat disinfect target temperature time out (could not reach to temperature) -Message = DG +Message = If heating up to minimum temperature for heat disinfect failed after 4 hours. SRSDG 854 = [246] -Title = DG cleaning mode conductivity sensors out of range -Message = DG +Title = DG chemical disinfect inlet conducitivty out of range +Message = If CPo and CD2 conductivity values are above 100 uS/cm. [247] -Title = Alarm Id available +Title = Message = [248] -Title = DG chemical disinfect could not reach to target temperature out of range -Message = DG +Title = DG chemical disinfect out of target temperature +Message = If TPo > 50 C after 1 second. [249] -Title = DG chemical disinfect target conductivity out of range -Message = DG +Title = DG chemical disinfect out of target conductivity +Message = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. SRSDG 992 = [250] -Title = DG chemical disinfect insert acid and remove the concentrate cap -Message = DG +Title = Insert the acid to start chemical disinfect +Message = Start of chemical disinfect SRSDG 1046 = [251] Title = HD invalid system record -Message = HD +Message = If the calculated CRC of the HD system record image does not match the expected CRC. [252] Title = HD invalid service record -Message = HD +Message = If the calculated CRC of the HD service record image does not match the expected CRC. [253] Title = DG invalid system record -Message = DG +Message = If the calculated CRC of the DG system record image does not match the expected CRC. SRSDG 609 = [254] Title = DG invalid service record -Message = DG +Message = If the calculated CRC of the DG service record image does not match the expected CRC. SRSDG 610 = [255] Title = HD and UI software builds are not compatible -Message = HD +Message = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. [256] -Title = HD processor is in Disinfect Chemical Flush mode -Message = HD +Title = DG flush started +Message = If DG flush successfully started from the UI. = SRSUI 497 [257] -Title = Alarm ID available for use +Title = Message = [258] -Title = Alarm Id available +Title = Message = [259] Title = HD temperatures out of range -Message = HD +Message = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. [260] -Title = HD UI POST Application Integrity (Sha256Sum) failure -Message = UI HD +Title = UI POST Application Integrity (Sha256Sum) failure +Message = If the UI POST reports a shasum faillure. = SRSUI 1067 [261] -Title = HD UI POST CANBus failure -Message = UI HD +Title = UI POST CANBus failure +Message = If the UI POST reports a can bus failure. = SRSUI 1068 [262] -Title = HD UI POST Display failure -Message = UI HD +Title = UI POST Display failure +Message = If the UI POST reports a display failure. = SRSUI 1069 [263] -Title = HD UI POST Touch failure -Message = UI HD +Title = UI POST Touch failure +Message = If the UI POST reports a touch failure. = SRSUI 1070 [264] -Title = HD UI POST SD-Card failure -Message = UI HD +Title = UI POST SD-Card failure +Message = If the UI POST reports SD card failure. = SRSUI 1071 [265] -Title = HD UI POST RTC failure -Message = UI HD +Title = UI POST RTC failure +Message = If the UI POST reports RTC failure. = SRSUI 1072 [266] -Title = HD UI POST WiFi failure -Message = UI HD +Title = UI POST WiFi failure +Message = If the UI POST reports WiFi failure. = SRSUI 1073 [267] -Title = HD UI POST Bluetooth failure -Message = UI HD +Title = UI POST Bluetooth failure +Message = If the UI POST reports bluetooth failure. = SRSUI 1074 [268] -Title = HD UI POST Ethernet failure -Message = UI HD +Title = UI POST Ethernet failure +Message = If the UI POST reports ethernet failure. = SRSUI 1075 [269] -Title = HD UI POST Sound failure -Message = UI HD +Title = UI POST Sound failure +Message = If the UI POST reports sound failure. = SRSUI 1076 [270] Title = HD POST Safety Shutdown failure -Message = HD +Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. [271] Title = DG POST Safety Shutdown failure -Message = DG +Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. SRSDG 822 = [272] Title = HD Fan RPM out of range -Message = HD +Message = If the fan RPM is out of Min/Max range. [273] -Title = Available for use +Title = Message = [274] -Title = Available for use +Title = Message = [275] Title = HD arterial pressure sensor is reading out of range -Message = HD +Message = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. [276] Title = HD venous pressure sensor is reading out of range -Message = HD +Message = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. [277] Title = HD BP occlusion sensor is reading out of range -Message = HD +Message = If the blood pump occlusion count is < 2000. [278] Title = HD active reservoir weight out of range. -Message = HD +Message = If the active reservoir weight is > 1900.0 g. [279] Title = DG dialysate drain time out -Message = DG +Message = If the DG dialysate drain exceeded 120 seconds. SRSDG 752 = [280] -Title = DG drain pump direction invalid -Message = DG +Title = DG drain pump invalid direction +Message = If the drain pump's direction is not 1 when draining. SRSDG 888 = [281] -Title = Acid concentration bottle low volume alarm -Message = DG +Title = DG Acid concentration bottle low volume alarm +Message = If the acid concentration bottle volume > 3240 (3340 - 100). SRSDG 836 = [282] -Title = Bicarbonate concentration bottle low volume alarm -Message = DG +Title = DG Bicarbonate concentration bottle low volume alarm +Message = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). SRSDG 837 = [283] Title = DG load cells weight out of range -Message = DG +Message = If the weight of the DG load cells > 4500.0 g or < 0.0 g. SRSDG 615 = [284] Title = DG load cells primary/back up drift out of range -Message = DG +Message = If redundant load cells deviate by more than 60g after tare for 5 seconds. SRSDG 383 = [285] -Title = Available for use +Title = Message = [286] -Title = DG concentrate cap not in proper position -Message = DG +Title = DG cap in wrong state +Message = If the concentrate cap is not in the right position in the cleaning modes. [287] Title = HD processor clock speed checks against FPGA clock failure -Message = HD +Message = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. [288] -Title = HD load cells primary/back up drift out of range -Message = HD +Title = HD load cell inconsistency +Message = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. [289] -Title = DG dialysate cap not closed -Message = DG +Title = DG cap in wrong state +Message = If the dialysate cap is not in the right position in the cleaning modes. SRSDG 928 = [290] -Title = HD venous pressure sensor FPGA fault -Message = HD +Title = HD venous sensor FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time [291] -Title = HD processor is in Disinfect Flush mode -Message = HD +Title = DG flush started +Message = If DG flush successfully started from the UI. [292] -Title = HD processor is in Disinfect Heat mode -Message = HD +Title = DG heat disinfect started +Message = If DG heat disinfect successfully started from the UI. [293] -Title = HD processor is in Disinfect Chemical mode -Message = HD +Title = DG flush started +Message = If DG flush successfully started from the UI. [294] -Title = Available for use +Title = Message = [295] -Title = HD UI POST CloudSync failure -Message = UI HD +Title = UI cloud sync POST failure +Message = If the UI POST reports cloud sync failure. = SRSUI 1238 [296] -Title = DG TDi sensors FPGA fault -Message = DG +Title = DG TDi sensor FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 951 = [297] -Title = DG TRo sensors FPGA fault -Message = DG +Title = DG TRo sensor FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 950 = [298] -Title = DG baro sensor FPGA fault -Message = DG +Title = DG barometric sensor FPGA fault +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 948 = [299] Title = DG invalid serial number -Message = DG +Message = If the DG serial number is empty and does not have any characters. SRSDG 952 = [300] Title = HD invalid serial number -Message = HD +Message = If the HD serial number is empty and does not have any characters. [301] -Title = Alarm ID available +Title = Message = [302] -Title = DG processor RAM error -Message = DG +Title = DG_CPU RAM error +Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. SRSDG 126 = [303] Title = DG drain pump direction FPGA fault -Message = DG +Message = If the FPGA indicates an error for a certain period of time. SRSDG 958 = [304] -Title = HD invalid usage record -Message = HD +Title = HD invalid usage reocord +Message = If the CRC of the HD usage record does not match the calculated CRC. [305] -Title = HD processor RAM error -Message = HD +Title = HD_CPU RAM error +Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. [306] -Title = DG CD1 CD2 check fault -Message = DG +Title = +Message = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other [307] -Title = HD AC power lost second alarm -Message = HD +Title = HD AC power lost alarm +Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. [308] -Title = DG CPi conductivity sensor invalid character received -Message = DG +Title = DG CPi conductivity sensor transmits invalid characters +Message = If the CPi conductivity sensor transmits invalid characters for a certain period of time. SRSDG 959 = [309] -Title = DG CPo conductivity sensor invalid character received -Message = DG +Title = DG CPo conductivity sensor transmits invalid characters +Message = If the CPo conductivity sensor transmits invalid characters for a certain period of time. SRSDG 960 = [310] -Title = DG CD1 conductivity sensor invalid character received -Message = DG +Title = DG CD1 conductivity sensor transmits invalid characters +Message = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. SRSDG 962 = [311] -Title = DG CD2 conductivity sensor invalid character received -Message = DG +Title = DG CD2 conductivity sensor transmits invalid characters +Message = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. SRSDG 961 = [312] -Title = HD Dialysate In flow rate to too low. -Message = HD +Title = The dialysate inlet flow is too low. +Message = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. [313] -Title = DG cleaning mode inlet water temperature too low. -Message = DG +Title = DG cleaning mode inlet water temperature too low +Message = If inlet water temperature is < 24 C. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = DG cleaning mode inlet water pressure too high. -Message = DG +Title = DG cleaning mode inlet water pressure too high +Message = If inlet water temperature is > 80 psig. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = DG cleaning mode inlet water pressure too low. -Message = DG +Title = DG cleaning mode inlet water pressure too low +Message = If inlet water pressure is < 14 psig. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = DG acid concentrate pump park fault. -Message = DG +Title = Acid concentrate pump did not park successfully +Message = If the park took more than 1 second. SRSDG 983 = [317] -Title = DG bicarb concentrate pump part fault. -Message = DG +Title = Bicarb concentrate pump did not park successfully +Message = If the park took more than 1 second. SRSDG 984 = [318] -Title = HD Air Pump timeout alarm -Message = HD +Title = HD Air Pump operation timeout +Message = If air pump has been turned on for greater than 10s. [319] -Title = DG chemical disinfect flush flush sample. -Message = DG +Title = Collect sample to check the chemical disinfectant acid +Message = Chemical disinfect flush sample alarm is prompted. SRSDG 1040 = [320] -Title = DG chemical disinfect flush sample timeout. -Message = DG +Title = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes +Message = Chemical disinfect flush sample was not collected form more than 30 minutes. SRSDG 1043 = [321] -Title = DG outlet UV reactor not healthy -Message = DG +Title = Outlet reactor not healthy +Message = If the outlet reactor are not healthy. SRSDG 1069 = [322] -Title = HD UI POST Year check failure -Message = UI HD +Title = UI POST Invalid Year +Message = If the UI POST reports a invalid year.