Index: Alarms/Alarms.conf =================================================================== diff -u -r97f446ce70c77a1be1ce21cd28bde2876725f6de -r8d4f6fcbcf89fd5206ddb80ecb847e9a5d527dd2 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 97f446ce70c77a1be1ce21cd28bde2876725f6de) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 8d4f6fcbcf89fd5206ddb80ecb847e9a5d527dd2) @@ -532,9 +532,9 @@ Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [100] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [101] Title = Service Required: Hemodialysis Device @@ -1110,9 +1110,9 @@ Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [203] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DG POST: Fill Cond Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [204] Title = Service Required: Dialysate Device @@ -1445,9 +1445,9 @@ = SRSUI 497 [257] -Title = Service Required: Hemodialysis Device -ListTitle = HD POST: Institutional Rec NVRAM CRC -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Title = +ListTitle = +Message = [258] Title = Service Required: Hemodialysis Device @@ -1714,9 +1714,9 @@ Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [306] -Title = Service Required: Dialysate Device -ListTitle = DG Fault: CD1 CD2 Check -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Title = +ListTitle = +Message = [307] Title = Power Loss