Index: Alarms/Alarms.conf =================================================================== diff -u -rd9dc3d70cd64985cc56e8a6d38f61ebfc95dabbb -r8faf38d02db36b4fde0683da7f1b47eb377d45c5 --- Alarms/Alarms.conf (.../Alarms.conf) (revision d9dc3d70cd64985cc56e8a6d38f61ebfc95dabbb) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 8faf38d02db36b4fde0683da7f1b47eb377d45c5) @@ -57,7 +57,7 @@ Message = [10] -Title = +Title = Self Test Error Message = = ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE @@ -79,7 +79,7 @@ [14] Title = Message = - = ALARM_ID_DG_INVALID_USAGE_RECORD + = ALARM_ID_DG_INVALID_USAGE_RECORD_CRC [15] Title = @@ -142,23 +142,15 @@ HD accelerometer invalid calibration record = ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD [27] -Title = -Message = +Title = BLD Sensor Error +Message = Call service to schedule a repair. +HD BLD sensor failed zero sequence = ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED [28] Title = Message = + = ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT -[27] -Title = -Message = - = ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED - -[28] -Title = -Message = - = ALARM_ID_HD___AVAILABLE_4 - [29] Title = Calibration Error Message = Call service to schedule a repair. @@ -232,7 +224,7 @@ [43] Title = Machine Error Message = Call service to schedule a repair. -BLD failed its internal self-test after cartridge prime = ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAULT +BLD failed its internal self-test after cartridge prime = ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT [44] Title = Machine Error @@ -332,12 +324,12 @@ [63] Title = Arterial Pressure Error Message = Call service to schedule a repair. -FPGA reports failure reading arterial pressure sensor = ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT +FPGA reports failure reading arterial pressure sensor = ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT [64] Title = Venous Pressure Error Message = Call service to schedule a repair. -FPGA reports failure reading venous pressure sensor = ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT +FPGA reports failure reading venous pressure sensor = ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE [65] Title = Machine Error @@ -385,19 +377,19 @@ HD No dialysate temperature data message receive at least once every 2 seconds = [72] -Title = -Message = +Title = Self Test Error +Message = Call service to schedule a repair. = ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE [73] -Title = -Message = +Title = Self Test Error +Message = Call service to schedule a repair. = ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE [74] Title = Self Test Error Message = Call service to schedule a repair. -HD blood leak detector self-test failure alarm = ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_SEND_TIMEOUT +HD blood leak detector self-test failure alarm = ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW [75] Title = Self Test Error @@ -447,7 +439,7 @@ [84] Title = Venous Pressure Error Message = Call service to schedule a repair. -HD venous pressure sensor not being read = ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR +HD venous pressure sensor not being read = ALARM_ID_DG_SET_RTC_YEAR_INVALID [85] Title = Pump Track Lever Not Fully Latched @@ -457,41 +449,41 @@ [86] Title = Occlusion Sensor Error Message = Call service to schedule a repair. -HD BP occlusion sensor not being read = ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR +HD BP occlusion sensor not being read = ALARM_ID_HD_SET_RTC_YEAR_INVALID [87] -Title = -Message = +Title = Calibration Error +Message = Call service to schedule a repair. = ALARM_ID_DG_HEATING_INVALID_CAL_RECORD [88] -Title = -Message = +Title = Self Test Error +Message = Call service to schedule a repair. = ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE [89] Title = Occlusion Sensor Error Message = Call service to schedule a repair. -HD BP occlusion sensor error = ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR +HD BP occlusion sensor error = ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW [90] Title = Message = - = ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR + = ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT [91] Title = Dialysate Measurement Error Message = Call service to schedule a repair. HD No dialysate flow data receive in the last 3 seconds = ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE [92] -Title = -Message = +Title = Self Test Error +Message = Call service to schedule a repair. = ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS [93] -Title = -Message = +Title = RO Rejection Ratio Error +Message = Call service to schedule a repair. = ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE [94] @@ -500,18 +492,18 @@ = ALARM_ID_DG_PRESSURE_OUT_OF_RANGE [95] -Title = -Message = +Title = Self Test Error +Message = Call service to schedule a repair. = ALARM_ID_DG_WATCHDOG_EXPIRED [96] -Title = -Message = +Title = Inlet Water Temperature High +Message = Inlet water temperature is high.\n- Check the incoming water for issues with cold water supply. = ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE [97] -Title = -Message = +Title = Dialysate Fill Error +Message = Dialysate Conductivity out of range. = ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE [98] @@ -557,7 +549,7 @@ [106] Title = Message = -This alarm ID is available for use = ALARM_ID_DG_DIALYSATE_FLOW_SENSOR_INVALID_CAL_RECORD +This alarm ID is available for use = ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD [107] Title = Machine Error @@ -567,7 +559,7 @@ [108] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump ADC error = ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR +HD syringe pump ADC error = ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT [109] Title = Syringe Pump Error @@ -587,12 +579,12 @@ [112] Title = Blood Leak Detector Error Message = Call service to schedule a repair. -HD blood leak detector fault = ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_SEND_TIMEOUT +HD blood leak detector fault = ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT [113] Title = Blood Leak Detector Error Message = Call service to schedule a repair. -HD blood leak zero command fault = ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_FAULT +HD blood leak zero command fault = ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE [114] Title = Air Bubble Detector Error @@ -661,82 +653,82 @@ [127] Title = Venous Pressure Low -Message = Venous pressure is low. Check to be sure your blue venous connections are secure and that your venous access is intact. +Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured properly. Venous pressure too low during treatment = ALARM_ID_VENOUS_PRESSURE_LOW [128] Title = Venous Air Bubble Detected -Message = Venous air bubble detected. Check for air bubbles in the venous line. If bubbles are present, tap the tubing to move the bubbles up to the air trap. If bubbles cannot be removed, treatment must end. +Message = Venous air bubble detected.\n- Check for air bubbles in the venous line.\n- If bubbles are present, tap the tubing to move the bubbles up to the air trap.\n- If bubbles cannot be removed, treatment must end. HD venous air bubble detected alarm = ALARM_ID_HD_VENOUS_BUBBLE_DETECTED [129] Title = Venous Air Bubble Detected During Rinseback -Message = Venous air bubble detected. Check for air bubbles in the venous line. If bubbles are present, tap the tubing to move the bubbles up to the air trap. If bubbles cannot be removed, treatment must end. +Message = Venous air bubble detected.\n- Check for air bubbles in the venous line.\n- If bubbles are present, tap the tubing to move the bubbles up to the air trap.\n- If bubbles cannot be removed, treatment must end. HD venous air bubble detected rinseback alarm = ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK [130] Title = Venous Pressure High -Message = Venous pressure is high. Check to be sure your blue venous connections are secure and that there are no kinks or clots in the tubing. +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present." Venous pressure too high during treatment = ALARM_ID_VENOUS_PRESSURE_HIGH [131] Title = Arterial Pressure Low -Message = Arterial pressure is low. Check to be sure your red arterial connections are secure and that there are no kinks or clots in the tubing. +Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. Arterial pressure too low during treatment = ALARM_ID_ARTERIAL_PRESSURE_LOW [132] Title = Arterial Pressure High -Message = Arterial pressure is low. Check to be sure your red arterial connections are secure and that your venous access is intact. +Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. Arterial pressure too high during treatment = ALARM_ID_ARTERIAL_PRESSURE_HIGH [133] -Title = Fluid Leak in Dialysate Generator -Message = Fluid leak detected in Dialysate Generation +Title = Fluid Leak in DG +Message = Fluid leak detected in Dialysate Generation Delivery Device. DG fluid leak detected alarm = ALARM_ID_DG_FLUID_LEAK_DETECTED [134] -Title = Fluid Leak in Hemodialysis Delivery Device -Message = Fluid leak detected in Hemodialysis Delivery device. Clean up the spill before continuing. +Title = Fluid Leak in HD +Message = Fluid leak detected in Hemodialysis Delivery device.\n- Identify and correct the cause of the spill.\n- Clean up the spill with a clean paper towel.\n- Resume treatment. HD fluid leak detected alarm = ALARM_ID_HD_FLUID_LEAK_DETECTED [135] -Title = Excessive Acceleration Detected in the Hemodialysis Device -Message = Excessive movement detected with Hemodialysis Delivery device. Confirm the device is stationary and level before continuing. +Title = Excessive Acceleration Detected in the HD +Message = Excessive movement detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. HD experienced large acceleration = ALARM_ID_HD_SHOCK [136] -Title = Excessive Acceleration Detected in the Dialysate Generation Device -Message = Excessive movement detected with Dialysis Generation device. Confirm the device is stationary and level before continuing. +Title = Excessive Acceleration Detected in the DG +Message = Excessive movement detected with Dialysis Generation device.\n- Confirm the device is stationary and level before continuing. DG experienced large acceleration = ALARM_ID_DG_SHOCK [137] Title = Excessive Tilt Detected in the Hemodialysis Device -Message = Excessive tilt detected with Hemodialysis Delivery device. Confirm the device is stationary and level before continuing. +Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing HD tilt exceeds maximum = ALARM_ID_HD_EXCESSIVE_TILT [138] Title = Excessive Tilt Detected in the Dialysate Generation Device -Message = Excessive tilt detected with Dialysis Generation device. Confirm the device is stationary and level before continuing. +Message = Excessive tilt detected with Dialysis Generation device.\n-Confirm the device is stationary and level before continuing. DG tilt exceeds maximum = ALARM_ID_DG_EXCESSIVE_TILT [139] Title = Power Lost -Message = Power has been lost. Check all electrical connections. The display and CPU will run off of the battery until power returns. +Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until power returns. HD AC power lost alarm = ALARM_ID_HD_AC_POWER_LOST [140] Title = Communication Failure with Dialysate Generator -Message = Communication failure with Dialysis Generator. Ensure all cords are securely connected. +Message = Communication failure with Dialysis Generator.\n- Ensure all cords are securely connected. Dialysate generator communication timeout = ALARM_ID_DG_COMM_TIMEOUT [141] Title = Air Trap Unable to Fill -Message = Failure to fill the Air Trap. Check the air trap vent line and the transducer protector. If the transducer protector is wet, replace it. +Message = Failure to fill the Air Trap.\n- Check the air trap vent line and the transducer protector.\n- If the transducer protector is wet, replace it. Air trap fill timeout during treatment = ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT [142] Title = Blood Pump Occlusion Detected -Message = Occlusion detected after the blood pump. Check for clots or kinks around the dialyzer and in the venous line. +Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the venous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. Blood pump occlusion detected = ALARM_ID_OCCLUSION_BLOOD_PUMP [143] @@ -751,7 +743,7 @@ [145] Title = Acid Concentrate Conductivity Out of Range -Message = Dialysate conductivity is out of range. Check the acid dialysate concentrates and make sure the correct concentrate is connected. +Message = Dialysate conductivity is out of range.\n- Check correct acid concentrate is connected. DG Concentrate conductivity after adding acid out of range alarm = ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE [146] @@ -766,27 +758,27 @@ [148] Title = Inlet Water Temperature High -Message = Inlet water temperature is high. Check the incoming water for issues with cold water supply. +Message = Inlet water temperature is high.\n- Check the incoming water for issues with cold water supply. Inlet water temperature too high = ALARM_ID_INLET_WATER_HIGH_TEMPERATURE [149] Title = Inlet Water Temperature Low -Message = Inlet water temperature is low. Check the incoming water for issues with hot water supply. +Message = Inlet water temperature is low.\n- Check the incoming water for issues with hot water supply. Inlet water temperature too low = ALARM_ID_HD_WATCHDOG_EXPIRED [150] Title = Inlet Water Conductivity High -Message = Inlet water conductivity is high. Check the incoming water for issues. +Message = Inlet water conductivity is high.\n- Check the incoming water for issues. Inlet water conductivity too high = ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE [151] Title = Inlet Water Conductivity Low -Message = Inlet water conductivity is low. Check the incoming water for issues. +Message = Inlet water conductivity is low.\n- Check the incoming water for issues. Inlet water conductivity too low = ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE [152] Title = Inlet Water Pressure Low -Message = Inlet water pressure is low. Check the incoming water supply. +Message = Inlet water pressure is low.\n- Check the incoming water supply. Inlet water pressure too low = ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE [153] @@ -821,12 +813,12 @@ [159] Title = Dialysate Pump Failure -Message = Dialysate pump failure. Check there are no kinks in the dialysate tubing. +Message = Dialysate pump failure.\n- Check there are no kinks in the dialysate tubing. Dialysate inlet pump failed flow vs. motor speed check. = ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK [160] Title = Blood Pump Speed High -Message = Blood pump failure. Check the pump rotor for secure connection. +Message = Blood pump failure.\n- Check the pump rotor for secure connection. Blood pump rotor speed too high = ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH [161] @@ -872,11 +864,11 @@ [169] Title = Dialysate Generation Error Message = Call service to schedule a repair. -Dialysate flow rate out of range = ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE +Dialysate flow rate out of range = ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE [170] Title = Syringe Empty -Message = Heparin syringe empty. Option to resume treatment without heparin. +Message = Heparin syringe empty.\n- Option to resume treatment without heparin. HD syringe empty alarm = ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY [171] @@ -901,7 +893,7 @@ [175] Title = Blood Pump Error -Message = Blood pump eailure. Check the pump rotor for secure connection. +Message = Blood pump failure. \n- Check the pump rotor for secure connection. HD pump direction status error = ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR [176] @@ -927,7 +919,7 @@ [180] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. -DG load cell ADC error = ALARM_ID_DG_LOAD_CELL_ADC_ERROR +DG load cell ADC error = ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE [181] Title = Ultrafiltration Measurement Error @@ -951,7 +943,7 @@ [185] Title = Fan Error -Message = Fan failure. Check that the fans inlets are not blocked or obstructed. +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. DG fan RPM out of range = ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE [186] @@ -1007,12 +999,12 @@ [196] Title = Dialysate Temperature Error Message = Call service to schedule a repair. -DG temperature sensor error flag fault. = ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT +DG temperature sensor error flag fault. = ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT [197] Title = Dialysate Temperature Error Message = Call service to schedule a repair. -DG temperature sensors ADC fault = ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT +DG temperature sensors ADC fault = ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT [198] Title = Dialysate Heater Error @@ -1096,12 +1088,12 @@ [214] Title = Saline Bolus Error -Message = Inability to deliver saline bolus. Check for kinks or clamped lines in the saline line. +Message = Inability to deliver saline bolus.\n- Check for kinks or clamped lines in the saline line. HD Saline bolus volume check failure = ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE [215] Title = Rinseback Error -Message = Unable to complete Rinseback. Check for kinks or clamped blood lines. +Message = Unable to complete Rinseback.\n- Check for kinks or clamped blood lines. HD rinseback volume check failure = ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE [216] @@ -1127,22 +1119,22 @@ [220] Title = Battery Error Message = Battery pack error. -HD battery pack SOC is too low. = ALARM_ID_HD_BATTERY_PACK_CHARGE_TOO_LOW +HD battery pack SOC is too low. = ALARM_ID_DG_THD_SENSORS_FPGA_FAULT [221] Title = Saline Bag Empty -Message = Saline back appears empty. Please check saline bag and line and replace if saline bag is empty. +Message = Saline back appears empty.\n- Check saline bag and line and replace if saline bag is empty. HD Saline bag is empty = ALARM_ID_EMPTY_SALINE_BAG [222] Title = Dialysate Temperature High Message = Dialysate temperature high. -HD dialysate temperature too high alarm = ALARM_ID_DIALYSATE_TEMPERATURE_HIGH +HD dialysate temperature too high alarm = ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT [223] Title = Dialysate Temperature Low Message = Dialysate temperature low. -HD dialysate temperature too low alarm = ALARM_ID_DIALYSATE_TEMPERATURE_LOW +HD dialysate temperature too low alarm = ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT [224] Title = Treatment Stopped @@ -1166,17 +1158,17 @@ [228] Title = Maximum Prime Time Exceeded -Message = Priming failure. Check blood and dialysate lines for clamps and kinks. +Message = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. HD time out on prime sub-mode purge air state. = ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT [229] Title = Maximum Prime Time Exceeded -Message = Priming failure. Check blood and dialysate lines for clamps and kinks. +Message = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. HD prime dialysate dialyzer time out alarm = ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT [230] Title = Maximum Prime Time Exceeded -Message = Priming failure. Check blood and dialysate lines for clamps and kinks. +Message = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. HD prime dialysate bypass time out alarm = ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT [231] @@ -1196,7 +1188,7 @@ [234] Title = Communication Failure -Message = Communication failure. Ensure all cords are securely connected. +Message = Communication failure.\n- Ensure all cords are securely connected. RTC (i.e read time) or RAM operations failure (read or write) = ALARM_ID_RTC_RAM_OPS_ERROR [235] @@ -1216,17 +1208,17 @@ [238] Title = Cartridge Needed -Message = Cartridge not loaded or loaded improperly. Check that a cartridge is properly installed. +Message = Cartridge not loaded or loaded improperly.\n- Check that a cartridge is properly installed. HD no cartridge loaded or installed improperly alarm = ALARM_ID_NO_CARTRIDGE_LOADED [239] Title = Cartridge Removal Error -Message = Cartridge not unloaded, or unloaded improperly. Check that the cartridge has been removed. +Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. HD fail to remove cartridge alarm = ALARM_ID_CARTRIDGE_REMOVAL_FAILURE [240] Title = Dialysate Conductivity Error -Message = Dialysate conductivity out of range. Check the bicarbonate dialysate concentrates and make sure the correct concentrate is connected. +Message = Dialysate conductivity out of range.\n- Check to be sure the correct bicarbonate dialysate concentrate is connected. Bicarb conductivity out of range during bicarb pump check alarm = ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE [241] @@ -1346,7 +1338,7 @@ [264] Title = Machine Error -Message = Power cycle the system.\nCall service to schedule a repair if problem persist. +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persist. UI POST SD-Card failure = ALARM_ID_UI_POST_FAILURE_SDCARD [265] @@ -1386,7 +1378,7 @@ [272] Title = -Message = Fan failure. Check that the fans inlets are not blocked or obstructed. +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. HD Fan RPM out of range = ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE [273] @@ -1421,22 +1413,22 @@ [279] Title = Dialysate Drain Pump Error -Message = +Message = Call service to schedule a repair. DG dialysate drain time out = ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT [280] Title = Arterial Pressure Error Message = Call service to schedule a repair. -HD arterial pressure sensor read timeout error = ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR +HD arterial pressure sensor read timeout error = ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID [281] Title = Acid Concentrate Low -Message = Acid concentrate low. Replace acid concentrate bottle. +Message = Acid concentrate low.\n- Replace acid concentrate bottle. DG Acid concentration bottle low volume alarm = ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME [282] Title = Bicarbonate Concentrate Low -Message = Bicarbonate concentrate low. Replace bicarbonate concentrate bottle. +Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. DG Bicarbonate concentration bottle low volume alarm = ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME [283] @@ -1477,21 +1469,21 @@ [290] Title = RO Flow Low Message = -DG RO flow too low while heating = ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON +DG RO flow too low while heating = ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT [291] -Title = -Message = +Title = DG Flush Failure +Message = Dialsyate Generation Device flush failure. = ALARM_ID_HD_DISINFECT_FLUSH [292] -Title = -Message = +Title = DG Heat Disinfection Failure +Message = DG Heat Disinfection Failure. = ALARM_ID_HD_DISINFECT_HEAT [293] -Title = -Message = +Title = DG Chemical Disinfection Failure +Message = DG Chemical Disinfection Failure = ALARM_ID_HD_DISINFECT_CHEM [294] @@ -1500,7 +1492,62 @@ = ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1 [295] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_UI_POST_FAILURE_CLOUDSYNC + +[296] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT + +[297] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT + +[298] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT + +[299] Title = Message = - = ALARM_ID_UI_POST_FAILURE_CLOUDSYNC + = ALARM_ID_DG_INVALID_SERIAL_NUMBER +[300] +Title = +Message = + = ALARM_ID_HD_INVALID_SERIAL_NUMBER + +[301] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE + +[302] +Title = Processore RAM error +Message = Call service to schedule a repair. +HD_CPU RAM error, or DG_CPU RAM error = ALARM_ID_DG_CPU_RAM_ERROR + +[303] +Title = +Message = + = ALARM_ID__AVAILABLE_4 + +[304] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_HD_INVALID_USAGE_RECORD_CRC + +[305] +Title = Self-Test Error +Message = Call service to schedule a repair. + = ALARM_ID_HD_CPU_RAM_ERROR + +[306] +Title = Air Trap Error +Message = Air Trap Error.\n- Check air trap fill level. + = ALARM_ID_HD_AIR_TRAP_OVERFILL + Index: Confirm/Confirm.conf =================================================================== diff -u --- Confirm/Confirm.conf (revision 0) +++ Confirm/Confirm.conf (revision 8faf38d02db36b4fde0683da7f1b47eb377d45c5) @@ -0,0 +1,18 @@ +[1] +Title = Water Flush +Message = Are sure you want to stop water flush? +Confirm = CONFIRM +Cancel = CANCEL + +[2] +Title = Heat Disinfect +Message = Are sure you want to stop heat disinfection? +Confirm = CONFIRM +Cancel = CANCEL + +[3] +Title = Chemical Disinfect +Message = Are sure you want to stop chemical disinfection? +Confirm = CONFIRM +Cancel = CANCEL + Index: Settings/System.conf =================================================================== diff -u -r84cce4c71892afc693c98cfa2603890191fb1f69 -r8faf38d02db36b4fde0683da7f1b47eb377d45c5 --- Settings/System.conf (.../System.conf) (revision 84cce4c71892afc693c98cfa2603890191fb1f69) +++ Settings/System.conf (.../System.conf) (revision 8faf38d02db36b4fde0683da7f1b47eb377d45c5) @@ -1,12 +1,12 @@ -[Service] -Password = 123 - [Alarm] Volume = 100 [Date] Format = MM/dd/yyyy +[Service] +Password = 123 + [Time] Format = HH:mm:ss Fisheye: Tag 8faf38d02db36b4fde0683da7f1b47eb377d45c5 refers to a dead (removed) revision in file `ui.config'. Fisheye: No comparison available. Pass `N' to diff?