Index: Alarms/Alarms.conf =================================================================== diff -u -r6459f898c32307d53d0fb4eb5c38534810bef3d2 -r9252ffd4bb94f989ff7c3b222d19dbea72467f5f --- Alarms/Alarms.conf (.../Alarms.conf) (revision 6459f898c32307d53d0fb4eb5c38534810bef3d2) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 9252ffd4bb94f989ff7c3b222d19dbea72467f5f) @@ -1053,9 +1053,9 @@ SRSDG 945 = [193] -Title = -ListTitle = -Message = +Title = Blood Pump Stopped: Partial Occlusion +ListTitle = Blood Stop: Blood Pump Partial Occlusion +Message = Partial Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [194] Title = Service Required: Dialysate Device