Index: Alarms/Alarms.conf =================================================================== diff -u -r0e5ea69da0c7b82ee3485d8198307a318a19a3e4 -raaa0ce2a11504f70a62b0aa2c3057a04857216ae --- Alarms/Alarms.conf (.../Alarms.conf) (revision 0e5ea69da0c7b82ee3485d8198307a318a19a3e4) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision aaa0ce2a11504f70a62b0aa2c3057a04857216ae) @@ -1,1077 +1,1077 @@ [0] -Title = ` -Message = +Title = +Message = \n\n [1] -Title = HD POST: Stuck Button -Message = Stop or On button pressed during startup\n-Release button and power cycle\n-Call service to schedule a repair if problem persists +Title = Stuck Button +Message = A button was pressed during start up and appears stuck. \n-Press and release the power and stop buttons on the front of the device. \n-Turn device off, then back on. \n-If problem continues, call service to schedule a repair. \n\nHD POST: Stuck Button [2] -Title = HD POST: FPGA Compatibility -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: FPGA Compatibility [3] -Title = DG POST: FPGA Compatibility -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: FPGA Compatibility SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = HD POST: WD Safestate -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: WD Safestate [5] -Title = DG POST: WD Safestate -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: WD Safestate SRSDG 53 = [6] -Title = HD POST: Proc-UI Comm -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Proc-UI Comm [7] -Title = HD Alarm (Low): RTC Bat Low -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): RTC Bat Low [8] -Title = HD POST: Acc Vector -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Acc Vector [9] -Title = DG POST: Acc Vector -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acc Vector SRSDG 323 = [10] -Title = HD POST: BLD Config Emb Mode -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Emb Mode [11] -Title = Dial. Gen: Dialysate Temp Low -Message = Dialysate temperature is low\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. +Title = Low Dialysate Temperature +Message = Dialysate temperature is low. \n-Treatment is currently paused.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Dialysate Temp Low [12] -Title = HD POST: HD Image -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: HD Image [13] -Title = DG POST: DG Image -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: DG Image SRSDG 813 = [14] -Title = DG POST: Usage Rec NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Usage Rec NVRAM CRC SRSDG 957 = [15] -Title = Dial. Gen: Dialysate Temp High -Message = Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. +Title = High Dialysate Temperature +Message = Dialysate temperature is high. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed.\n\nDial. Gen: Dialysate Temp High [16] -Title = HD POST: Primary Alarm Current -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Primary Alarm Current [17] -Title = HD POST: UI POST Not Complete -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI POST Not Complete [18] -Title = Loss of DG Comm (DG ACK) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (DG ACK) [19] -Title = Dial. Gen: Dialysate Temp High -Message = Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. +Title = High Dialysate Temperature +Message = Dialysate temperature is high. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed. \n\nDial. Gen: Dialysate Temp High [20] -Title = DG POST: Conductivity Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Conductivity Cal SRSDG 16 = [21] -Title = DG POST: Drain Line Vol Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Drain Line Vol Cal SRSDG 936 = [22] -Title = DG POST: Reservoir Vol Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Reservoir Vol Cal SRSDG 935 = [23] -Title = DG POST: Acid Conc Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acid Conc Cal SRSDG 934 = [24] -Title = DG POST: Bicarb Conc Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Bicarb Conc Cal SRSDG 933 = [25] -Title = DG POST: Acc Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acc Cal SRSDG 817 = [26] -Title = HD POST: Acc Cal -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Acc Cal [27] -Title = Process: Pre-Tx Wet Self-Test -Message = Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. +Title = Self Test: Blood Leak +Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test [28] -Title = DG Fault: Conductivity RTD Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Conductivity RTD Comm SRSDG 864 = [29] -Title = HD POST: SP DAC Cal -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: SP DAC Cal [30] -Title = HD Fault: SW Invalid State -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: SW Invalid State [31] -Title = HD Fault: BP Current -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Current [32] -Title = HD Fault: BP Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed [33] -Title = HD Fault: BP Direction -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Direction [34] -Title = HD Fault: BP Rotor -Message = Blood pump rotor failure.\n- Check the pump rotor for secure connection. +Title = Blood Pump Error +Message = A problem has been detected with the blood pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: BP Rotor [35] -Title = HD Fault: DPi Current -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Current [36] -Title = HD Fault: DPi Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed [37] -Title = HD Fault: DPi Direction -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Direction [38] -Title = HD Fault: DPi Rotor -Message = Dialysate in pump rotor failure.\n- Check the pump rotor for secure connection. +Title = Dialysate Pump Error +Message = A problem has been detected with the dialysate pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: DPi Rotor [39] -Title = HD Fault: DPo Current -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Current [40] -Title = HD Fault: DPo Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed [41] -Title = HD Fault: DPo Direction -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Direction [42] -Title = HD Fault: DPo Rotor -Message = Dialysate out pump rotor failure.\n- Check the pump rotor for secure connection. +Title = Dialysate Pump Error +Message = A problem has been detected with the dialysate pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: DPo Rotor [43] -Title = DG Fault: Iso ADC Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Iso ADC Comm SRSDG 954 = [44] -Title = HD Fault: UI Comm -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UI Comm [45] -Title = HD Fault: CAN CRC Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: CAN CRC Com [46] -Title = HD Fault: CAN ACK UI Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: CAN ACK UI Com [47] Title = -Message = +Message = \n\n [48] -Title = HD Fault: UF Rate Tare Err -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UF Rate Tare Err [49] -Title = HD Fault: FPGA Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: FPGA Com [50] -Title = DG Fault: Valve Control -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Valve Control SRSDG 26 = [51] -Title = HD Fault: BP Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed [52] -Title = HD Fault: DPi Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed [53] -Title = HD Fault: DPo Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed [54] -Title = HD Fault: Critical Data Err -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Critical Data Err [55] -Title = DG Fault: Critical Data Err -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Critical Data Err SRSDG 921 = [56] -Title = HD Fault: Acc Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Acc Com [57] -Title = DG Fault: Acc Com -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Acc Com SRSDG 869 SRSDG 870 = [58] -Title = HD Fault: Valve Home -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Home [59] -Title = HD Fault: Valve Transition Err -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Transition Err [60] -Title = HD Fault: Valve Control -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Control [61] -Title = HD Fault: Valve Current -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Current [62] -Title = HD Fault: Valve Position -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Position [63] -Title = HD Fault: BLD Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BLD Com [64] -Title = DG Fault: Baro Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Range SRSDG 874 = [65] -Title = Loss of DG Comm (Reject) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reject) [66] -Title = HD POST: BLD Config Set Point -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Set Point [67] -Title = Process: Pre-Tx Remove Cartridge -Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume +Title = Self Test: Cartridge +Message = A cartridge has been detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge [68] -Title = Dial. Gen: Dialysate not ready -Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. +Title = Dialysate Generation Error +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. \n\nDial. Gen: Dialysate not ready [69] -Title = HD POST: BLD Cal -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: BLD Cal [70] -Title = Process: Pre-Tx Remove Cartridge -Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume +Title = Self Test: Cartridge +Message = A cartridge has been detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge [71] -Title = Process: Pre-Tx Remove Cartridge -Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume +Title = Self Test: Cartridge +Message = A cartridge was detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge [72] -Title = Loss of DG Comm (LC) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (LC) [73] -Title = Loss of DG Comm (TDI/TRO) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (TDI/TRO) [74] -Title = DG Fault: UVi On No Flow -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVi On No Flow SRSDG 600 = [75] -Title = HD Fault: Syringe DAC -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe DAC [76] -Title = HD Fault: Voltage Monitor -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Voltage Monitor [77] -Title = DG Fault: Voltage Monitor -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Voltage Monitor SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = HD Fault: Syringe Pump Enc Dir -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Enc Dir [79] -Title = HD Fault: Syringe Pump Drv Dir -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Drv Dir [80] -Title = HD Fault: Syringe Pump Driver -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Driver [81] -Title = HD Fault: Syringe Pump OT -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump OT [82] -Title = HD Fault: Syringe Pump DAC Err -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump DAC Err [83] -Title = HD Fault: Syringe Pump On Err -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump On Err [84] -Title = DG POST: HD RTC Time -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: HD RTC Time SRSDG 815 = [85] -Title = Blood Stop: Cartridge Lever Open -Message = Cartridge Lock Lever is not closed.\n- Open front door and push lever all the way down.\n- Close door and press RESUME button to continue. +Title = Cartridge Lever is Open +Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press RESUME button to continue.\n\nBlood Stop: Cartridge Lever Open [86] -Title = HD POST: HD RTC Time -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: HD RTC Time [87] -Title = DG POST: Heat Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Heat Cal SRSDG 929 = [88] -Title = DG Fault: CP1 CP2 Speed Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 CP2 Speed Range [89] -Title = DG Fault: UVo On No Flow -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVo On No Flow SRSDG 599 = [90] -Title = DG Fault: LC A1/B1 Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC A1/B1 Comm SRSDG 604 = [91] -Title = Loss of DG Comm (FMD) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nLoss of DG Comm (FMD) [92] Title = Message = [93] -Title = DG Fault: RO Permeate Cond High -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Permeate Cond High SRSDG 430 = [94] -Title = DG Fault: Pressure Sensor Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Pressure Sensor Range SRSDG 629 = [95] -Title = DG Fault: WD Error -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: WD Error SRSDG 907 = [96] -Title = Dial. Gen: Inlet Water Temp High -Message = Inlet water temperature is high.\n- Treatment is paused\n- Check the incoming water supply. +Title = Incoming Water Temperature High +Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp High SRSDG 410 = [97] -Title = Dial. Gen: Dialysate Cond. -Message = Dialysate conductivity is out of range.\n- Check acid and bicarb concentrate is correct and not empty. +Title = Conductivity Out of Range +Message = Treatment is currently paused.\n-Confirm that acid and bicarbonate concentrates are securely connected. \n- If empty, replace. \n-Treatment will need to be resumed when conductivity is within target range.\n\nDial. Gen: Dialysate Cond. SRSDG 397 = [98] -Title = HD Alarm (Low): Battery Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Com [99] -Title = HD Fault: Syringe Pump Stall -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Stall [100] -Title = HD Fault: Timeout No Cart ST -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Timeout No Cart ST [101] -Title = HD POST: UI-Proc Comm -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI-Proc Comm [102] -Title = Cleaning: Tpo THD Temp Delta -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Disinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Tpo THD Temp Delta SRSDG 856 SRSDG 930 = [103] -Title = HD POST: Cal NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: Cal NVRAM CRC [104] -Title = HD Fault: Air Trap Sensor -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Air Trap Sensor [105] -Title = DG POST: Cal NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Cal NVRAM CRC SRSDG 608 = [106] -Title = DG POST: Flow Sensor Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Flow Sensor Cal SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = DG Fault: DG Restarted -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DG Restarted [108] -Title = HD Fault: Syringe Pump ADC Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump ADC Com [109] -Title = HD Fault: Syringe Pump Vol -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Vol [110] -Title = HD Fault: Syringe Pump Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed [111] -Title = HD Fault: Syringe Pump Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed [112] -Title = DG Fault: LC A2/B2 Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC A2/B2 Comm SRSDG 953 = [113] -Title = DG Fault: DRP Current -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Current SRSDG 890 = [114] -Title = Process: Pre-Tx Wet Self-Test -Message = Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. +Title = Self Test: Venous Air Detector +Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test [115] -Title = DG Fault: Temp Sensor Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Temp Sensor Range SRSDG 224 = [116] -Title = HD Alarm (Low): SD-Card Error -Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. +Title = Service Required: Dialysate Device +Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): SD-Card Error [117] -Title = Loss of HD Comm (HD ACK) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of HD Comm (HD ACK) SRSDG 909 = [118] -Title = DG Fault: RTC Config -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RTC Config SRSDG 43 = [119] -Title = DG Alarm (Low): RTC Bat Low -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A non-critical problem was detected with the dialysate device.\n-Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): RTC Bat Low SRSDG 908 = [120] -Title = Process: Pre-Tx Wet Self-Test -Message = Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. +Title = Self Test: Venous Air Detector +Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test [121] -Title = DG Fault: Prim Heater Lg Control -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Lg Control SRSDG 611 = [122] -Title = DG Fault: Prim Heater Sm Control -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Sm Control SRSDG 612 = [123] -Title = DG Fault: Trimmer Heater Control -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Trimmer Heater Control SRSDG 613 = [124] -Title = Blood Stop: Blood Clot Risk -Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. +Title = Blood Pump Stop Time Exceeded +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk [125] -Title = Blood Stop: Blood Clot Risk -Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. +Title = Blood Pump Stop Time Exceeded +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk [126] -Title = Blood Stop: Dialyzer Blood Leak -Message = A blood leak has been detected. Treatment must end. +Title = Blood Leak +Message = A blood leak has been detected. \n-Treatment must be terminated without rinseback.\n\nBlood Stop: Dialyzer Blood Leak [127] -Title = Blood Stop: Venous Pres. Low -Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured \nproperly. +Title = Low Venous Pressure +Message = Venous pressure is low.\n- Ensure venous access connections are properly secured. \n-If there has been an access disconnection or needle dislodgement, \nfollow facility policy to address and terminate treatment without rinseback. \n-If access connections are properly secured, \npress RESUME to continue treatment.\n\nBlood Stop: Venous Pres. Low [128] -Title = Blood Stop: Venous Air Bubble -Message = Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. +Title = Venous Air Detected +Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm connections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Venous Air Bubble [129] Title = -Message = +Message = \n\n [130] -Title = Blood Stop: Venous Pres. High -Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. +Title = High Venous Pressure +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n-Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Venous Pres. High [131] -Title = Blood Stop: Arterial Pres. Low -Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. +Title = Low Arterial Pressure +Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. Low [132] -Title = Blood Stop: Arterial Pres. High -Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. +Title = High Arterial Pressure +Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n-Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. High [133] -Title = DG Fault: Fluid Leak in DG -Message = Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and \nelectrical connections from system.\n- Call service to schedule a repair. +Title = Dialysate Device Fluid Leak +Message = Fluid leak detected in the dialysate device.\n-Treatment must be terminated.\n- Turn off water source. \n-Disconnect water and electrical connections from the device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Fluid Leak in DG SRSDG 328 = [134] -Title = Blood Stop: Fluid Leak in HD -Message = Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment, BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment, resume treatment. +Title = Fluid Leak +Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n-Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, resume treatment.\n\nBlood Stop: Fluid Leak in HD [135] -Title = Blood Stop: HD Shock Detected -Message = Excessive movement detected with Hemodialysis Delivery \ndevice.\n- Confirm the device is stationary and level before continuing. +Title = Blood Pump Stopped: Excessive Movement +Message = Excessive movement of the hemodialysis device has been detected. \n-Treatment is currently paused. \n-When system movement has stablized, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nBlood Stop: HD Shock Detected [136] -Title = Dial. Gen: Shock-Vibration -Message = Excessive movement detected with Dialysis Generation \ndevice.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. +Title = Dialysate Device: Excessive Movement +Message = Excessive movement of the dialysate device has been detected.\n-Treatment is currently paused. \n-Treatment can be resumed when system movement has stablized.\n\nDial. Gen: Shock-Vibration SRSDG 327 = [137] -Title = Blood Stop: HD Tilted -Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. +Title = Blood Pump Stopped: Excessive Tilt +Message = Excessive tilt of the hemodialysis device has been detected. \n-Treatment is currently paused. \n-When the device is stationary and level, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nBlood Stop: HD Tilted [138] -Title = Dial. Gen: Tilt -Message = Excessive tilt detected with Dialysis Generation device.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing. +Title = Dialysate Device: Excessive Tilt +Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing.\n\nDial. Gen: Tilt SRSDG 326 = [139] -Title = Loss of AC Power -Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns. +Title = Power Loss +Message = AC Power has been lost.\n-Check all electrical connections.\n- Press OK to return to welcome screen. \n\nLoss of AC Power [140] -Title = Loss of DG Comm (Timeout) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Timeout) [141] -Title = Blood Stop: Air Trap Fill -Message = Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. +Title = Blood Pump Stopped: Venous Air Detector +Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure connections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Air Trap Fill [142] -Title = Blood Stop: Blood Pump Occlusion -Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the \nvenous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. +Title = Blood Pump Stopped: Occulusion +Message = Occlusion detected after the blood pump.\n-Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Blood Pump Occlusion [143] -Title = DG Fault: TRo TDi Delta -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TRo TDi Delta SRSDG 912 = [144] -Title = Cleaning: Inlet Water Cond High -Message = Cleaning Mode Inlet water conductivity too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. +Title = Disinfection: Water Conductivity High +Message = Incoming water conductivity is too high.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond High SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = Dial. Gen: Acid Conc. Cond. -Message = Acid concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. +Title = Conductivity Error During Prime +Message = Conductivity out of range during prime.\n-Confirm that acid and bicarbonate concentrates are securely connected. \n- If empty, replace. \n-Treatment will need to be resumed when conductivity is within target range.\n\nDial. Gen: Acid Conc. Cond. SRSDG 846 = [146] -Title = DG POST: DG RTC Timer -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: DG RTC Timer SRSDG 312 = [147] -Title = Dial. Gen: Dialysate not ready -Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. +Title = Dialysate Generation Error +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \nthe dialysate is ready for use. \n\nDial. Gen: Dialysate not ready SRSDG 398 = [148] -Title = Cleaning: Inlet Water Temp High -Message = Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. +Title = Disinfection: Water Temperature High +Message = Incoming water temperature is too high.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp High SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = HD Fault: WD Error -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: WD Error [150] -Title = Dial. Gen: Inlet Water Cond High -Message = Inlet water conductivity is high.\n- Treatment is paused.\n- Check the incoming water supply. +Title = Incoming Water Conductivity High +Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the conductivity is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond High SRSDG 391 = [151] -Title = Dial. Gen: Inlet Water Cond Low -Message = Inlet water conductivity is low.\n- Treatment is paused.\n- Check the incoming water supply. +Title = Incoming Water Conductivity Low +Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the conductivity is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond Low SRSDG 826 = [152] -Title = Dial. Gen: Inlet Water Pres Low -Message = Inlet water pressure is low.\n- Treatment is paused.\n- Check the incoming water supply. +Title = Incoming Water Pressure Low +Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the pressure is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres Low SRSDG 417 = [153] -Title = Process: Pre-Tx Terminated -Message = Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system. +Title = Pre-Treatment Process Terminated +Message = The pre-treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n-Press OK for continue with next steps to remove cartridge from the device.\n\nProcess: Pre-Tx Terminated [154] -Title = Cleaning: Inlet Water Cond Low -Message = Cleaning Mode Inlet water conductivity too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. +Title = Disinfection: Water Conductivity Low +Message = Incoming water conductivity is too low.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond Low SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = Loss of DG Comm (Reservoir) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reservoir) [156] -Title = Loss of DG Comm (Op Mode) -Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. +Title = Communication Error +Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Op Mode) [157] -Title = Cleaning: Chem Disinf PAA Prime -Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. +Title = Chemical Disinfectant Not Detected +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the concentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Prime SRSDG 988 = [158] -Title = Dial. Gen: Inlet Water Temp Low -Message = Inlet water temperature is low.\n- Treatment is paused.\n- Check the incoming water supply. +Title = Incoming Water Temperature Low +Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp Low SRSDG 408 = [159] -Title = Cleaning: Chem Disinf Connect Cap -Message = Remove chemical disinfectant connection and close cap. +Title = Remove Chemical Disinfectant +Message = The chemical disinfectant is no longer needed. \n-Disconnect chemical disinfectant from the device. \n-Place the concentrate connection cap on the device.\n-Keep dialysate cap on the device.\n\nCleaning: Chem Disinf Connect Cap SRSDG 1044 = [160] -Title = HD Fault: BP Rotor Max Speed -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Rotor Max Speed [161] -Title = DG Fault: HD-DG Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HD-DG Comm SRSDG 122 = [162] -Title = DG Fault: Proc-FPGA Clock -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Proc-FPGA Clock SRSDG 867 = [163] -Title = Dial. Gen: UF Rate Check -Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. +Title = Ultrafiltration Error +Message = A problem was detected with the dialysate device. \n-Confirm the device is level with minimal movement. \n-Resume treatment. \n-If problem persists, call service to report the issue and schedule a repair. \n\nDial. Gen: UF Rate Check [164] -Title = Dial. Gen: UF Rate Check -Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. +Title = Ultrafiltration Error +Message = A problem was detected with the dialysate device. \n-Confirm the device is level with minimal movement. \n-Resume treatment. \n-If problem persists, call service to report the issue and schedule a repair. \n\nDial. Gen: UF Rate Check [165] -Title = Blood Stop: Blood Clot Risk -Message = Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. +Title = Blood Pump Stop Time Exceeded +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk [166] Title = -Message = +Message = \n\n [167] -Title = Blood Stop: Door Open -Message = Device door is open.\n - Close door and press RESUME button to continue. +Title = Blood Pump Stopped: Door Open +Message = The device door is open.\n Close the device door. \n-Press RESUME to continue.\n\nBlood Stop: Door Open [168] Title = -Message = +Message = \n\n [169] -Title = DG Fault: Dialysate Flow Sens Rg -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Dialysate Flow Sens Rg SRSDG 643 = [170] Title = Heparin Syringe Empty -Message = Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment, \nadminister manually. +Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to complete treatment.\n\nHeparin Syringe Empty [171] -Title = Process: Pre-Tx Hep. Pump Occ. -Message = Heparin pump occlusion detected.\n- Check heparin line is unclamped.\n- Check there are no kinks in the heparin line. +Title = Heparin Pump Occlusion +Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped.\n\nProcess: Pre-Tx Hep. Pump Occ. [172] -Title = Process: Pre-Tx Not Enough Hep. -Message = Not enough heparin is available for treatment.\n- Check if syringe pump is touching syringe plunger.\n- If syringe pump is touching syringe plunger, end treatment.\n- If syringe pump is not touching syringe plunger, reload \nsyringe with enough heparin. +Title = Not Enough Heparin +Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger, end treatment.\n- If syringe pump is not touching the syringe plunger, \n- ensure the correct amount of heparin has been drawn and reload the syringe. \n\nProcess: Pre-Tx Not Enough Hep. [173] -Title = HD Fault: RTC Config -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: RTC Config [174] -Title = HD POST: HD RTC Timer -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: HD RTC Timer [175] -Title = HD Fault: Pump Commutation Err -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Pump Commutation Err [176] Title = -Message = +Message = \n\n [177] -Title = DG Fault: Software Invalid State -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Software Invalid State SRSDG 919 = [178] -Title = HD Fault: UI-Proc Comm -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UI-Proc Comm = SRSUI 1066 [179] -Title = DG Fault: FPGA Com -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: FPGA Com SRSDG 605 = [180] -Title = DG Fault: RO Flow Sensor Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Sensor Range SRSDG 883 = [181] -Title = DG Fault: Tare Out of Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Tare Out of Range SRSDG 849 SRSDG 838 = [182] -Title = DG POST: LC Cal -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: LC Cal SRSDG 162 = [183] -Title = Cleaning: Invalid Resv Transfer -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Invalid Resv Transfer SRSDG 851 = [184] -Title = DG Fault: UVi Error -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVi Error SRSDG 559 = [185] -Title = DG Alarm (Low): Fan Speed -Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. +Title = Service Required: Dialysate Device +Message = A non-critical problem was detected with the dialysate device.\n-Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): Fan Speed SRSDG 628 = [186] -Title = DG Fault: CP1/CP2 Fault -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1/CP2 Fault SRSDG 891 = [187] -Title = DG Fault: CP1 Speed Fault -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 Speed Fault SRSDG 905 = [188] -Title = DG Fault: CP2 Speed Fault -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP2 Speed Fault SRSDG 906 = [189] -Title = DG Fault: DRP Speed Fault -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Fault SRSDG 448 = [190] -Title = DG Fault: DRP Speed Not Off -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Not Off [191] -Title = DG Fault: RO Flow High -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow High SRSDG 885 = [192] -Title = DG Fault: RO Flow Low -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Low SRSDG 945 = [193] Title = -Message = +Message = \n\n [194] -Title = DG Fault: RO Driver Feedback -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Driver Feedback SRSDG 881 = [195] -Title = DG Fault: RO Pump High Pressure -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Pump High Pressure SRSDG 925 = [196] -Title = DG Fault: CPi/CPo Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPi/CPo Comm SRSDG 955 = [197] -Title = DG Fault: CD1/CD2 Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1/CD2 Comm SRSDG 956 = [198] -Title = DG Fault: HP On With No Flow -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HP On With No Flow SRSDG 872 = [199] -Title = DG Fault: HR On With No Flow -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HR On With No Flow SRSDG 871 = [200] -Title = DG Fault: Thermistor Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Thermistor Range SRSDG 178 = [201] -Title = Process: Pre-Tx Wet Self-Test -Message = Self-test has detected a problem in the dialysate flow \nmeasurement or there is a leak in the dialysate in/out lines.\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. +Title = Self Test: Dialysate Flow Rate +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test [202] -Title = Process: Pre-Tx Dry Self-Test -Message = Self-test has detected blood circuit pressure cannot be \nrelieved.\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists, install a new cartridge and \nPBo transducer protector. +Title = Self Test: Pressure +Message = A problem has been detected during self-test. \n-Confirm the cartridge is properly installed. \n-Ensure arterial and venous pressure sensor lines are not kinked. \n-Press OK to resume. \n-If problem persists, install a new cartridge.\n\nProcess: Pre-Tx Dry Self-Test [203] Title = -Message = +Message = \n\n [204] -Title = DG Fault: RO Rejection -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Rejection SRSDG 429 = [205] -Title = DG Fault: Cond Sensors Error -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Cond Sensors Error SRSDG 913 = [206] -Title = DG Fault: Reservoir Fill Timeout -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Fill Timeout SRSDG 437 = [207] -Title = DG Fault: FMP Flow Rate Error -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: FMP Flow Rate Error SRSDG 240 = [208] -Title = Blood Stop: Venous Pres. High -Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. +Title = High Venous Pressure +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n-Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. \n\nBlood Stop: Venous Pres. High [209] -Title = Cleaning: Flush Circ Error -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Flush Circ Error SRSDG 980 = [210] -Title = HD Alarm (Low): Battery Fault -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Fault [211] -Title = Blood Stopped: Warning -Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. +Title = Approaching MAX Blood Pump Stop Time +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time timer expires. \n- If treatment is not resumed before the timer expires, treatment will be terminated without rinseback. \n\nBlood Stopped: Warning [212] -Title = Blood Stopped: Warning -Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. +Title = Approaching MAX Blood Pump Stop Time +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time timer expires. \n- If treatment is not resumed before the timer expires, treatment will be terminated without rinseback.\n\nBlood Stopped: Warning [213] -Title = Process: Pre-Tx Prime Complete -Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. +Title = Prime Complete - 30 MIN WARNING +Message = The priming process completed 90 minutes ago. \n-Treatment must begin within 30 minutes or the pre-treatment process will be terminated. \n-Press OK to continue. \n\nProcess: Pre-Tx Prime Complete [214] Title = -Message = +Message = \n\n [215] Title = -Message = +Message = \n\n [216] -Title = Process: End of Dialysis Tx -Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. +Title = Treatment Complete +Message = Dialysis treatment is complete.\n-Press OK to end treatment and rinseback.\n\nProcess: End of Dialysis Tx [217] Title = -Message = +Message = \n\n [218] -Title = Process: Remove Syringe -Message = Syringe not removed from machine.\n- Remove heparin syringe. +Title = Syringe Detected +Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n-Press OK to continue.\n\nProcess: Remove Syringe [219] -Title = Blood Stop: Syringe Not Detected -Message = Heparin is needed for treatment but syringe has been removed \nfrom syringe pump.\n- Install Heparin syringe. +Title = Blood Stopped: Syringe Not Detected +Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n-Press Ok to resume treatment.\n\nBlood Stop: Syringe Not Detected [220] -Title = DG Fault: THD RTD Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: THD RTD Comm SRSDG 947 = [221] -Title = Blood Stop: Saline Bag Empty -Message = Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is \nempty. +Title = Blood Pump Stopped: Saline Bag Empty +Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. \n\nBlood Stop: Saline Bag Empty [222] -Title = HD Fault: BP Occ Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Occ Com [223] -Title = HD Fault: Art. Pres. Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Art. Pres. Com [224] -Title = Blood Stopped: Treatment Pause -Message = Treatment paused by user. +Title = Blood Pump Stopped: Treatment Paused +Message = Treatment has been stopped by used. \n- Press OK to resume treatment. \n-If treatment is not resumed within five minutes, treatment will be terminated without rinseback. \n\nBlood Stopped: Treatment Pause [225] -Title = Process: End of Treatment -Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. +Title = Treatment Complete +Message = Dialysis treatment is complete.\n-Press OK to end treatment and rinseback.\n\nProcess: End of Treatment [226] -Title = Process: Pre-Tx Priming Complete -Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. +Title = Priming Complete +Message = Priming is complete. \n-Treatment must begin within 120 minutes \nor the pre-treatment process will be terminated.\n-Press OK to continue. \n\nProcess: Pre-Tx Priming Complete [227] Title = -Message = +Message = \n\n [228] -Title = Process: Pre-Tx Prime Timeout -Message = Blood line prime with saline timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. +Title = Priming Error +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout [229] -Title = Process: Pre-Tx Prime Timeout -Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. +Title = Priming Error +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout [230] -Title = Process: Pre-Tx Prime Timeout -Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. +Title = Priming Error +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout [231] -Title = Process: Pre-Tx Dry Self-Test -Message = Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion.\n- Check clamps.\n- If problem persists, install a new cartridge. +Title = Self Test: Pressure +Message = A problem has been detected during self-test. \n- Confirm cartridge is properly installed without kinks or clamps. \n-Ensure all connections are secure. \n-Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre-Tx Dry Self-Test [232] -Title = Process: Pre-Tx Wet Self-Test -Message = Self-test has detected a problem in the load cell \nmeasurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation onto machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. +Title = Self Test: Dialysate Flow Rate +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test [233] -Title = DG Fault: Baro Cal CRC -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Cal CRC SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = Dial. Gen: Inlet Water Pres High -Message = Inlet water pressure is high.\n- Treatment is paused\n- Check the incoming water supply. +Title = Incoming Water Pressure High +Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the pressure is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres High SRSDG 1037 = [235] -Title = Process: Rinseback Complete -Message = Rinseback has been completed. +Title = Rinseback Complete +Message = Rinseback is complete. \n-Follow facililty policy prior to disconnecting patient. \n-Press OK to continue. \n\nProcess: Rinseback Complete [236] -Title = Process: Pre-Tx Dry Self-Test -Message = Self-test has detected fluid in the cartridge.\n- Cartridges are single-use only\n- Install new cartridge to resume. +Title = Self Test: Fluid Present +Message = A problem has been detected during self-test. \n-Cartridges are single-use only and fluid is present.\n- Install new cartridge to resume.\n\nProcess: Pre-Tx Dry Self-Test [237] -Title = Process: Pre-Tx Prime Timeout -Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. +Title = Priming Error +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout [238] -Title = Process: Pre-Tx Dry Self-Test -Message = Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. +Title = Cartridge Not Dectected +Message = A problem has been detected during self-test. \n-Cartridge is not detected.\n- Confirm cartridge is properly installed.\n\nProcess: Pre-Tx Dry Self-Test [239] -Title = Process: Post-Tx Remove Cartridge -Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. +Title = Cartridge Still Detected +Message = A problem has been detected during self-test. \n-Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed.\n\nProcess: Post-Tx Remove Cartridge [240] -Title = Dial. Gen: Bicarb Conc. Cond. -Message = Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. +Title = Conductivity Error During Prime +Message = Conductivity out of range during prime.\n-Confirm that acid and bicarbonate concentrates are securely connected. \n- If empty, replace. \n-Treatment will need to be resumed when conductivity is within target range.\n\nDial. Gen: Bicarb Conc. Cond. SRSDG 482 = [241] -Title = Cleaning: Resv Drain Timeout -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Resv Drain Timeout SRSDG 848 SRSDG 90 SRSDG 561 @@ -1102,8 +1102,8 @@ SRSDG 1034 = [242] -Title = Cleaning: Resv Fill Timeout -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Resv Fill Timeout SRSDG 850 SRSDG 573 SRSDG 574 @@ -1124,373 +1124,373 @@ SRSDG 1024 = [243] -Title = Cleaning: Fluid Loss Detected -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Fluid Loss Detected SRSDG 853 = [244] -Title = Cleaning: Temp Sensor Check -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Temp Sensor Check SRSDG 852 = [245] -Title = Cleaning: Heat Disinfect Temp TO -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Heat Disinfect Temp TO SRSDG 854 = [246] -Title = Cleaning: RO Water Cond Error -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: RO Water Cond Error [247] Title = -Message = +Message = \n\n [248] -Title = Cleaning: Chem Disinfect Proc -Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. +Title = Disinfection Interrupted +Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinfect Proc [249] -Title = Cleaning: Chem Disinf PAA Cond -Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. +Title = Chemical Disinfectant Not Detected +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the concentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Cond SRSDG 992 = [250] -Title = Cleaning: Chem Disinf Connect PAA -Message = Connect chemical disinfectant bottle to bicarb port. +Title = Connect Chemical Disinfectant +Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n-Keep dialysate cap on the device.\n-Place disinfectant straw in the disinfectant bottle. \n-Attach straw to concentrate connection and ensure connection is secure. \n-Press OK to continue.\n\nCleaning: Chem Disinf Connect PAA SRSDG 1046 = [251] -Title = HD POST: Sys Rec NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Sys Rec NVRAM CRC [252] -Title = HD POST: Serv Rec NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serv Rec NVRAM CRC [253] -Title = DG POST: Sys Rec NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Sys Rec NVRAM CRC SRSDG 609 = [254] -Title = DG POST: Serv Rec NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Serv Rec NVRAM CRC SRSDG 610 = [255] -Title = HD POST: UI Compatibility -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Compatibility [256] -Title = Cleaning: Chem Flush In Progress -Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. +Title = Chemical Flush in Progress +Message = Chemical flush in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Flush In Progress = SRSUI 497 [257] Title = -Message = +Message = \n\n [258] Title = -Message = +Message = \n\n [259] -Title = HD Fault: Temp Sensors Error -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Temp Sensors Error [260] -Title = HD POST: UI Image -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Image = SRSUI 1067 [261] -Title = HD POST: UI CAN Driver -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI CAN Driver = SRSUI 1068 [262] -Title = HD POST: UI Disp Driver -Message = Call service to schedule a repair. +Title = +Message = \n\n = SRSUI 1069 [263] -Title = HD POST: UI Touch Driver -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Touch Driver = SRSUI 1070 [264] -Title = HD POST: UI SD-Card -Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI SD-Card = SRSUI 1071 [265] -Title = HD POST: UI RTC -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI RTC = SRSUI 1072 [266] -Title = HD POST: UI WiFi -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI WiFi = SRSUI 1073 [267] -Title = HD POST: UI Bluetooth -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Bluetooth = SRSUI 1074 [268] -Title = HD POST: UI Ethernet -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Ethernet = SRSUI 1075 [269] -Title = HD POST: UI Sound -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Sound = SRSUI 1076 [270] -Title = HD POST: Proc Safestate -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Proc Safestate [271] -Title = DG POST: Proc Safestate -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Proc Safestate SRSDG 822 = [272] -Title = HD Alarm (Low): Fan Speed -Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. +Title = Service Required: Hemodialysis Device +Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Fan Speed [273] Title = -Message = +Message = \n\n [274] Title = -Message = +Message = \n\n [275] -Title = HD Fault: Arterial Press Range -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Arterial Press Range [276] -Title = HD Fault: Venous Press Range -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Venous Press Range [277] -Title = HD Fault: Occlusion Sens Range -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Occlusion Sens Range [278] -Title = DG Fault: Reservoir Overfill -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Overfill [279] -Title = DG Fault: Reservoir Drain -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Drain SRSDG 752 = [280] -Title = DG Fault: DRP Direction Fault -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Direction Fault SRSDG 888 = [281] -Title = Dial. Gen: Acid Conc. Low -Message = Acid concentrate low.\n- Replace acid concentrate bottle. +Title = Acid Concentrate Low +Message = The acid concentrate is low.\n- Replace acid concentrate container. \n\nDial. Gen: Acid Conc. Low SRSDG 836 = [282] -Title = Dial. Gen: Bicarb Conc. Low -Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. +Title = Bicarb Concentrate Low +Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container.\n\nDial. Gen: Bicarb Conc. Low SRSDG 837 = [283] -Title = DG Fault: LC Range -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC Range SRSDG 615 = [284] -Title = DG Fault: Redundant LC Check -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Redundant LC Check SRSDG 383 = [285] Title = -Message = +Message = \n\n [286] -Title = Cleaning: Conc Cap Removed -Message = Concentrate cap removed from DG during cleaning.\n- Cleaning Process Terminated. +Title = Concentrate Connection Cap Removed +Message = The concentrate connection cap was removed from the dialysate device during disinfection.\n- Disinfection terminated.\n-Replace cap before attempting disinfection again. \n\nCleaning: Conc Cap Removed [287] -Title = HD Fault: Proc-FPGA Clock -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Proc-FPGA Clock [288] -Title = DG Fault: LC Drift Check -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC Drift Check [289] -Title = Cleaning: Dialysate Cap Removed -Message = Dialysate cap removed from DG during cleaning.\n- Cleaning Process Terminated. +Title = Dialysate Connection Cap Removed +Message = The dialysate connection cap was removed from the dialysate device during disinfection.\n- Disinfection terminated.\n-Replace cap before attempting disinfection again. \n\nCleaning: Dialysate Cap Removed SRSDG 928 = [290] -Title = HD Fault: Ven. Pres. Com -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Ven. Pres. Com [291] -Title = Cleaning: Flush In Progress -Message = System flush in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. +Title = Flush In Progress +Message = System flush is in progress\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Flush In Progress [292] -Title = Cleaning: Heat Disinf In Progress -Message = Heat disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. +Title = Heat Disinfection in Progress +Message = Heat disinfection is in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Heat Disinf In Progress [293] -Title = Cleaning: Chem Disinf In Progress -Message = Chemical disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. +Title = Chemical Disinfection in Progress +Message = Chemical disinfection in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Disinfect In Progress [294] Title = -Message = +Message = \n\n [295] -Title = HD POST: UI CloudSync -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: UI CloudSync = SRSUI 1238 [296] -Title = DG Fault: TDi RTD Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TDi RTD Comm SRSDG 951 = [297] -Title = DG Fault: TRo RTD Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TRo RTD Comm SRSDG 950 = [298] -Title = DG Fault: Baro Pres Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Pres Comm SRSDG 948 = [299] -Title = DG POST: Serial Number Check -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Serial Number Check SRSDG 952 = [300] -Title = HD POST: Serial Number Check -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serial Number Check [301] Title = -Message = +Message = \n\n [302] -Title = DG Fault: Processor RAM error -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Processor RAM error SRSDG 126 = [303] -Title = DG Fault: DRP Hall Commutation -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Hall Commutation SRSDG 958 = [304] -Title = HD POST: Usage Rec NVRAM CRC -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: Usage Rec NVRAM CRC [305] -Title = HD Fault: Processor RAM error -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Processor RAM error [306] -Title = DG Fault: CD1 CD2 Check -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 CD2 Check [307] -Title = Blood Stop: Loss of AC Power -Message = HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns.\n-If power is returned prior to blood stopped countdown elapses, \nrinseback will be allowed. +Title = Power Loss +Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n-Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed.\n\nBlood Stop: Loss of AC Power [308] -Title = DG Fault: CPi Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPi Comm SRSDG 959 = [309] -Title = DG Fault: CPo Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPo Comm SRSDG 960 = [310] -Title = DG Fault: CD1 Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 Comm SRSDG 962 = [311] -Title = DG Fault: CD2 Comm -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD2 Comm SRSDG 961 = [312] -Title = Dial. Gen: Dialysate In Flow -Message = Dialysate in flow rate is low\n- Check dialysate line connection to the DG. +Title = Low Dialysate Flow +Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nRESUME button will be available.\n-Press RESUME to continue.\n\nDial. Gen: Dialysate In Flow [313] -Title = Cleaning: Inlet Water Temp Low -Message = Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. +Title = Disinfection: Water Temperature Low +Message = Incoming water temperature is too low.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp Low SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = Cleaning: Inlet Water Press High -Message = Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. +Title = Disinfection: Water Pressure High +Message = Incoming water pressure is too high. \n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Press High SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = Cleaning: Inlet Water Press Low -Message = Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. +Title = Disinfection: Water Pressure Low +Message = Incoming water pressure is too low. \n-Disinfection has been terminated.\n- Check incoming water initiating disinfection again. \n\nCleaning: Inlet Water Press Low SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = DG Fault: Acid CP Pump Park -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Acid CP Pump Park SRSDG 983 = [317] -Title = DG Fault: Bicarb CP Pump Park -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Bicarb CP Pump Park SRSDG 984 = [318] -Title = HD Fault: Air Pump Timeout -Message = Unable to reduce fluid level.\n- Air trap may be overfilled.\n- Pressure transducer protector may be contaminated.\n- Treatment must be terminated without option for rinseback.\n- Call service to schedule a repair. +Title = Air Trap Control Error +Message = A problem has been detected with the air trap level control system. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: Air Pump Timeout [319] -Title = Cleaning: Chem Residual Sampling -Message = Collect chemical residuals water sample from drain line. +Title = Residual Chemical Sampling +Message = A water sample is needed to confirm the \nabsence of residual chemical following disininfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant.\n\nCleaning: Chem Residual Sampling SRSDG 1040 = [320] -Title = Cleaning: Chem Residual Timeout -Message = During Chemical Disinfection Flush\n-user residual sampling timed out. +Title = Chemical Residual Timeout +Message = A chemical residual sample was not collected \n-within 30min of chemical flush.\n-To collect a sample for chemical residual, you must repeat the chemical flush.\n\nCleaning: Chem Residual Timeout SRSDG 1043 = [321] -Title = DG Fault: UVo Error -Message = Call service to schedule a repair. +Title = Service Required: Dialysate Device +Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVo Error SRSDG 1069 = [322] -Title = HD POST Alarm: UI Invalid Year -Message = Call service to schedule a repair. +Title = Service Required: Hemodialysis Device +Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST Alarm: UI Invalid Year