Index: Alarms/Alarms.conf =================================================================== diff -u -r5d0b098f15d890e6ae97bb3590344d1c12a55eb8 -rde1c5d548f0121f78c48484d439fe4dd4e526101 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 5d0b098f15d890e6ae97bb3590344d1c12a55eb8) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision de1c5d548f0121f78c48484d439fe4dd4e526101) @@ -1176,7 +1176,7 @@ [215] Title = Air Trap Overfill -ListTitle = Air Trap: Air Trap Upper Fluid +ListTitle = Air Trap: Air Trap Overfill Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [216]