[0] Title = ListTitle = Message = [1] Title = Stuck Button ListTitle = HD POST: Stuck Button Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. [2] Title = Service Required: Hemodialysis Device ListTitle = HD POST: FPGA Compatibility Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [3] Title = Service Required: Dialysate Device ListTitle = DG POST: FPGA Compatibility Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = Service Required: Hemodialysis Device ListTitle = HD POST: WD Safestate Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [5] Title = Service Required: Dialysate Device ListTitle = DG POST: WD Safestate Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 53 = [6] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Proc- UI Comm Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [7] Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): RTC Bat Low Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [8] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Acc Vector Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [9] Title = Service Required: Dialysate Device ListTitle = DG POST: Acc Vector Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 323 = [10] Title = Service Required: Hemodialysis Device ListTitle = HD POST: BLD Config Emb Mode Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [11] Title = Low Dialysate Temperature ListTitle = Dial. Gen: Dialysate Temp Low Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [12] Title = Service Required: Hemodialysis Device ListTitle = HD POST: HD Image Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [13] Title = Service Required: Dialysate Device ListTitle = DG POST: DG Image Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 813 = [14] Title = Service Required: Dialysate Device ListTitle = DG POST: Usage Rec NVRAM CRC Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 957 = [15] Title = High Dialysate Temperature ListTitle = Dial. Gen: Dialysate Temp High Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [16] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Primary Alarm Current Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [17] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI POST Not Complete Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [18] Title = Communication Error ListTitle = Loss of DG Comm (DG ACK) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [19] Title = High Dialysate Temperature ListTitle = Dial. Gen: Dialysate Temp High Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [20] Title = Service Required: Dialysate Device ListTitle = DG POST: Conductivity Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 16 = [21] Title = Service Required: Dialysate Device ListTitle = DG POST: Drain Line Vol Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 936 = [22] Title = Service Required: Dialysate Device ListTitle = DG POST: Reservoir Vol Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 935 = [23] Title = Service Required: Dialysate Device ListTitle = DG POST: Acid Conc Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 934 = [24] Title = Service Required: Dialysate Device ListTitle = DG POST: Bicarb Conc Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 933 = [25] Title = Service Required: Dialysate Device ListTitle = DG POST: Acc Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 817 = [26] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Acc Cal Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [27] Title = Self Test: Blood Leak ListTitle = Process: Pre- Tx Wet Self- Test Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [28] Title = Service Required: Dialysate Device ListTitle = DG Fault: Conductivity RTD Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 864 = [29] Title = Service Required: Hemodialysis Device ListTitle = HD POST: SP DAC Cal Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [30] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: SW Invalid State Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [31] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Current Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [32] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [33] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Direction Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [34] Title = Blood Pump Error ListTitle = HD Fault: BP Rotor Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [35] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Current Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [36] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [37] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Direction Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [38] Title = Dialysate Pump Error ListTitle = HD Fault: DPi Rotor Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [39] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Current Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [40] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [41] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Direction Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [42] Title = Dialysate Pump Error ListTitle = HD Fault: DPo Rotor Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [43] Title = Service Required: Dialysate Device ListTitle = DG Fault: Iso ADC Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 954 = [44] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: UI Comm Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [45] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: CAN CRC Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [46] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: CAN ACK UI Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [47] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Pump High Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [48] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: UF Rate Tare Err Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [49] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: FPGA Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [50] Title = Service Required: Dialysate Device ListTitle = DG Fault: Valve Control Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 26 = [51] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [52] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [53] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [54] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Critical Data Err Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [55] Title = Service Required: Dialysate Device ListTitle = DG Fault: Critical Data Err Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 921 = [56] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Acc Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [57] Title = Service Required: Dialysate Device ListTitle = DG Fault: Acc Com Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 869 SRSDG 870 = [58] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Home Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [59] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Transition Err Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [60] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Control Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [61] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Current Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [62] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Position Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [63] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Pre- Tx Hep. Pump Timeout. Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [64] Title = Service Required: Dialysate Device ListTitle = DG Fault: Baro Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 874 = [65] Title = Communication Error ListTitle = Loss of DG Comm (Reject) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [66] Title = Service Required: Hemodialysis Device ListTitle = HD POST: BLD Config Set Point Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [67] Title = Self Test: Cartridge ListTitle = Process: Pre- Tx Remove Cartridge Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [68] Title = Dialysate Generation Error ListTitle = Dial. Gen: Dialysate not ready Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. [69] Title = Service Required: Hemodialysis Device ListTitle = HD POST: BLD Cal Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [70] Title = Self Test: Cartridge ListTitle = Process: Pre- Tx Remove Cartridge Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [71] Title = Self Test: Cartridge ListTitle = Process: Pre- Tx Remove Cartridge Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [72] Title = Communication Error ListTitle = Loss of DG Comm (LC) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [73] Title = Communication Error ListTitle = Loss of DG Comm (TDI/TRO) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [74] Title = Service Required: Dialysate Device ListTitle = DG Fault: UVi On No Flow Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 600 = [75] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe DAC Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [76] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Voltage Monitor Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [77] Title = Service Required: Dialysate Device ListTitle = DG Fault: Voltage Monitor Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Enc Dir Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [79] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Drv Dir Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [80] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Driver Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [81] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump OT Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [82] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump DAC Err Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [83] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump On Err Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [84] Title = Service Required: Dialysate Device ListTitle = DG POST: DG RTC Time Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 815 = [85] Title = Cartridge Lever is Open ListTitle = Blood Stop: Cartridge Lever Open Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. [86] Title = Service Required: Hemodialysis Device ListTitle = HD POST: HD RTC Time Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [87] Title = Service Required: Dialysate Device ListTitle = DG POST: Heat Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 929 = [88] Title = Service Required: Dialysate Device ListTitle = DG Fault: CP1 CP2 Speed Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [89] Title = Service Required: Dialysate Device ListTitle = DG Fault: UVo On No Flow Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 599 = [90] Title = Service Required: Dialysate Device ListTitle = DG Fault: LC A1/B1 Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 604 = [91] Title = Communication Error ListTitle = Loss of DG Comm (FMD) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [92] Title = ListTitle = Message = [93] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Permeate Cond High Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 430 = [94] Title = Service Required: Dialysate Device ListTitle = DG Fault: Pressure Sensor Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 629 = [95] Title = Service Required: Dialysate Device ListTitle = DG Fault: WD Error Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 907 = [96] Title = Incoming Water Temperature High ListTitle = Dial. Gen: Inlet Water Temp High Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 410 = [97] Title = Conductivity Out of Range ListTitle = Dial. Gen: Dialysate Cond. Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 397 = [98] Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): Battery Com Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [99] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Stall Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [100] Title = Service Required: Dialysate Device ListTitle = DG POST: Conductivity Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [101] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI- Proc Comm Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [102] Title = Disinfection Interrupted ListTitle = Cleaning: Tpo THD Temp Delta Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Cal NVRAM CRC Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [104] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Air Trap Sensor Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [105] Title = Service Required: Dialysate Device ListTitle = DG POST: Cal NVRAM CRC Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 608 = [106] Title = Service Required: Dialysate Device ListTitle = DG POST: Flow Sensor Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = Service Required: Dialysate Device ListTitle = DG Fault: DG Restarted Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [108] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump ADC Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [109] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Vol Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [110] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [111] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [112] Title = Service Required: Dialysate Device ListTitle = DG Fault: LC A2/B2 Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 953 = [113] Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Current Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 890 = [114] Title = Self Test: Venous Air Detector ListTitle = Process: Pre- Tx Wet Self- Test Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [115] Title = Service Required: Dialysate Device ListTitle = DG Fault: Temp Sensor Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 224 = [116] Title = Service Required: Dialysate Device ListTitle = HD Alarm (Low): SD- Card Error Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [117] Title = Communication Error ListTitle = Loss of HD Comm (HD ACK) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 909 = [118] Title = Service Required: Dialysate Device ListTitle = DG Fault: RTC Config Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 43 = [119] Title = Service Required: Dialysate Device ListTitle = DG Alarm (Low): RTC Bat Low Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. SRSDG 908 = [120] Title = Self Test: Venous Air Detector ListTitle = Process: Pre- Tx Wet Self- Test Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [121] Title = Service Required: Dialysate Device ListTitle = DG Fault: Prim Heater Lg Control Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 611 = [122] Title = Service Required: Dialysate Device ListTitle = DG Fault: Prim Heater Sm Control Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 612 = [123] Title = Service Required: Dialysate Device ListTitle = DG Fault: Trimmer Heater Control Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 613 = [124] Title = Blood Pump Stop Time Exceeded ListTitle = Blood Stop: Blood Clot Risk Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [125] Title = Blood Pump Stop Time Exceeded ListTitle = Blood Stop: Blood Clot Risk Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [126] Title = Blood Leak ListTitle = Blood Stop: Dialyzer Blood Leak Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. [127] Title = Low Venous Pressure ListTitle = Blood Stop: Venous Pres. Low Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. [128] Title = Venous Air Detected ListTitle = Blood Stop: Venous Air Bubble Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. [129] Title = Blood Leak: Bypass ListTitle = Blood Stop: Dialyzer Blood Leak Bypass Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. [130] Title = High Venous Pressure ListTitle = Blood Stop: Venous Pres. High Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. [131] Title = Low Arterial Pressure ListTitle = Blood Stop: Arterial Pres. Low Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. [132] Title = High Arterial Pressure ListTitle = Blood Stop: Arterial Pres. High Message = Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. [133] Title = Dialysate Device Fluid Leak ListTitle = DG Fault: Fluid Leak in DG Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 328 = [134] Title = Fluid Leak ListTitle = Blood Stop: Fluid Leak in HD Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. [135] Title = Blood Pump Stopped: Excessive Movement ListTitle = Blood Stop: HD Shock Detected Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. [136] Title = Dialysate Device: Excessive Movement ListTitle = Dial. Gen: Shock- Vibration Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. SRSDG 327 = [137] Title = Blood Pump Stopped: Excessive Tilt ListTitle = Blood Stop: HD Tilted Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. [138] Title = Dialysate Device: Excessive Tilt ListTitle = Dial. Gen: Tilt Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 326 = [139] Title = Power Loss ListTitle = Loss of AC Power Message = AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. [140] Title = Communication Error ListTitle = Loss of DG Comm (Timeout) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [141] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Blood Stop: Air Trap Fill Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. [142] Title = Blood Pump Stopped: Occlusion ListTitle = Blood Stop: Blood Pump Occlusion Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [143] Title = Service Required: Dialysate Device ListTitle = DG Fault: TRo TDi Delta Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 912 = [144] Title = Disinfection: Water Conductivity High ListTitle = Cleaning: Inlet Water Cond High Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] Title = Conductivity Error During Prime ListTitle = Dial. Gen: Acid Conc. Cond. Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 846 = [146] Title = Service Required: Dialysate Device ListTitle = DG POST: DG RTC Timer Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 312 = [147] Title = Dialysate Generation Error ListTitle = Dial. Gen: Dialysate not ready Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. SRSDG 398 = [148] Title = Disinfection: Water Temperature High ListTitle = Cleaning: Inlet Water Temp High Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: WD Error Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [150] Title = Incoming Water Conductivity High ListTitle = Dial. Gen: Inlet Water Cond High Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 391 = [151] Title = Incoming Water Conductivity Low ListTitle = Dial. Gen: Inlet Water Cond Low Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 826 = [152] Title = Incoming Water Pressure Low ListTitle = Dial. Gen: Inlet Water Pres Low Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 417 = [153] Title = Pre-Treatment Process Terminated ListTitle = Process: Pre- Tx Terminated Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. [154] Title = Disinfection: Water Conductivity Low ListTitle = Cleaning: Inlet Water Cond Low Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] Title = Communication Error ListTitle = Loss of DG Comm (Reservoir) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [156] Title = Communication Error ListTitle = Loss of DG Comm (Op Mode) Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [157] Title = Chemical Disinfectant Not Detected ListTitle = Cleaning: Chem Disinf PAA Prime Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 988 = [158] Title = Incoming Water Temperature Low ListTitle = Dial. Gen: Inlet Water Temp Low Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 408 = [159] Title = Remove Chemical Disinfectant ListTitle = Cleaning: Chem Disinf Connect Cap Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. SRSDG 1044 = [160] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Rotor Max Speed Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [161] Title = Service Required: Dialysate Device ListTitle = DG Fault: HD- DG Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 122 = [162] Title = Service Required: Dialysate Device ListTitle = DG Fault: Proc- FPGA Clock Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 867 = [163] Title = Ultrafiltration Error ListTitle = Dial. Gen: UF Rate Check Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. [164] Title = Ultrafiltration Error ListTitle = Dial. Gen: UF Rate Check Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. [165] Title = Blood Pump Stop Time Exceeded ListTitle = Blood Stop: Blood Clot Risk Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [166] Title = Test Alarm - Alarm 166 ListTitle = Message = [167] Title = Blood Pump Stopped: Door Open ListTitle = Blood Stop: Door Open Message = The device door is open.\n Close the device door. \n- Press Resume/OK to continue. [168] Title = Test Alarm - Alarm 168 ListTitle = Message = [169] Title = Service Required: Dialysate Device ListTitle = DG Fault: Dialysate Flow Sens Rg Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 643 = [170] Title = Heparin Syringe Empty ListTitle = Heparin Syringe Empty Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. [171] Title = Heparin Pump Occlusion ListTitle = Process: Pre- Tx Hep. Pump Occ. Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. [172] Title = Not Enough Heparin ListTitle = Process: Pre- Tx Not Enough Hep. Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. [173] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: RTC Config Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [174] Title = Service Required: Hemodialysis Device ListTitle = HD POST: HD RTC Timer Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [175] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Pump Commutation Err Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [176] Title = Test Alarm - Alarm 176 ListTitle = Message = [177] Title = Service Required: Dialysate Device ListTitle = DG Fault: Software Invalid State Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 919 = [178] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: UI- Proc Comm Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1066 [179] Title = Service Required: Dialysate Device ListTitle = DG Fault: FPGA Com Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 605 = [180] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Flow Sensor Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 883 = [181] Title = Service Required: Dialysate Device ListTitle = DG Fault: Tare Out of Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 849 SRSDG 838 = [182] Title = Service Required: Dialysate Device ListTitle = DG POST: LC Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 162 = [183] Title = Disinfection Interrupted ListTitle = Cleaning: Invalid Resv Transfer Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 851 = [184] Title = Service Required: Dialysate Device ListTitle = DG Fault: UVi Error Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 559 = [185] Title = Service Required: Dialysate Device ListTitle = DG Alarm (Low): Fan Speed Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. SRSDG 628 = [186] Title = Service Required: Dialysate Device ListTitle = DG Fault: CP1/CP2 Fault Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 891 = [187] Title = Service Required: Dialysate Device ListTitle = DG Fault: CP1 Speed Fault Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 905 = [188] Title = Service Required: Dialysate Device ListTitle = DG Fault: CP2 Speed Fault Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 906 = [189] Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Speed Fault Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 448 = [190] Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Speed Not Off Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [191] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Flow High Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 885 = [192] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Flow Low Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 945 = [193] Title = Blood Pump Stopped: Partial Occlusion ListTitle = Blood Stop: Blood Pump Partial Occlusion Message = Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment. [194] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Driver Feedback Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 881 = [195] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Pump High Pressure Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 925 = [196] Title = Service Required: Dialysate Device ListTitle = DG Fault: CPi/CPo Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 955 = [197] Title = Service Required: Dialysate Device ListTitle = DG Fault: CD1/CD2 Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 956 = [198] Title = Service Required: Dialysate Device ListTitle = DG Fault: HP On With No Flow Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 872 = [199] Title = Service Required: Dialysate Device ListTitle = DG Fault: HR On With No Flow Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 871 = [200] Title = Service Required: Dialysate Device ListTitle = DG Fault: Thermistor Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 178 = [201] Title = Self Test: Dialysate Flow Rate ListTitle = Process: Pre- Tx Wet Self- Test Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [202] Title = Self Test: Pressure ListTitle = Process: Pre- Tx Dry Self- Test Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [203] Title = Service Required: Dialysate Device ListTitle = DG POST: Fill Cond Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [204] Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Rejection Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 429 = [205] Title = Service Required: Dialysate Device ListTitle = DG Fault: Cond Sensors Error Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 913 = [206] Title = Service Required: Dialysate Device ListTitle = DG Fault: Reservoir Fill Timeout Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 437 = [207] Title = Service Required: Dialysate Device ListTitle = DG Fault: FMP Flow Rate Error Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 240 = [208] Title = High Venous Pressure ListTitle = Blood Stop: Venous Pres. High Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [209] Title = Disinfection Interrupted ListTitle = Cleaning: Flush Circ Error Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 980 = [210] Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): Battery Fault Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [211] Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood Stopped: Warning Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. [212] Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood Stopped: Warning Message = The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. [213] Title = Prime Complete - 30 MIN WARNING ListTitle = Process: Pre- Tx Prime Complete Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. [214] Title = Service Required: Dialysate Device ListTitle = DG POST: Conc Pump Cal Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [215] Title = ListTitle = Message = [216] Title = Treatment Complete ListTitle = Process: End of Dialysis Tx Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. [217] Title = ListTitle = Message = [218] Title = Syringe Detected ListTitle = Process: Remove Syringe Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. [219] Title = Blood Stopped: Syringe Not Detected ListTitle = Blood Stop: Syringe Not Detected Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. [220] Title = Service Required: Dialysate Device ListTitle = DG Fault: THD RTD Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 947 = [221] Title = Blood Pump Stopped: Saline Bag Empty ListTitle = Blood Stop: Saline Bag Empty Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. [222] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Occ Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [223] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Art. Pres. Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [224] Title = Blood Pump Stopped: Treatment Paused ListTitle = Blood Stopped: Treatment Pause Message = Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. [225] Title = Treatment Complete ListTitle = Process: End of Treatment Message = Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. [226] Title = Priming Complete ListTitle = Process: Pre- Tx Priming Complete Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. [227] Title = ListTitle = Message = [228] Title = Priming Error ListTitle = Process: Pre- Tx Prime Timeout Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [229] Title = Priming Error ListTitle = Process: Pre- Tx Prime Timeout Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [230] Title = Priming Error ListTitle = Process: Pre- Tx Prime Timeout Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [231] Title = Self Test: Cartridge Check ListTitle = Process: Pre- Tx Dry Self- Test Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [232] Title = Self Test: Dialysate Flow Rate ListTitle = Process: Pre- Tx Wet Self- Test Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [233] Title = Service Required: Dialysate Device ListTitle = DG Fault: Baro Cal CRC Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] Title = Incoming Water Pressure High ListTitle = Dial. Gen: Inlet Water Pres High Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 1037 = [235] Title = Rinseback Complete ListTitle = Process: Rinseback Complete Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. [236] Title = Self Test: Fluid Present ListTitle = Process: Pre- Tx Dry Self- Test Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. [237] Title = ListTitle = Message = [238] Title = Cartridge Not Detected ListTitle = Process: Pre- Tx Dry Self- Test Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. [239] Title = Cartridge Still Detected ListTitle = Process: Post- Tx Remove Cartridge Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. [240] Title = Conductivity Error During Prime ListTitle = Dial. Gen: Bicarb Conc. Cond. Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 482 = [241] Title = Disinfection Interrupted ListTitle = Cleaning: Resv Drain Timeout Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 SRSDG 577 SRSDG 590 SRSDG 591 SRSDG 336 SRSDG 879 SRSDG 340 SRSDG 496 SRSDG 501 SRSDG 505 SRSDG 532 SRSDG 535 SRSDG 556 SRSDG 557 SRSDG 1048 SRSDG 1049 SRSDG 651 SRSDG 655 SRSDG 744 SRSDG 745 SRSDG 1004 SRSDG 1008 SRSDG 1019 SRSDG 1023 SRSDG 1033 SRSDG 1034 = [242] Title = Disinfection Interrupted ListTitle = Cleaning: Resv Fill Timeout Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 SRSDG 579 SRSDG 580 SRSDG 492 SRSDG 493 SRSDG 498 SRSDG 499 SRSDG 508 SRSDG 509 SRSDG 519 SRSDG 690 SRSDG 691 SRSDG 703 SRSDG 711 SRSDG 1020 SRSDG 1024 = [243] Title = Disinfection Interrupted ListTitle = Cleaning: Fluid Loss Detected Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 853 = [244] Title = Disinfection Interrupted ListTitle = Cleaning: Temp Sensor Check Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 852 = [245] Title = Disinfection Interrupted ListTitle = Cleaning: Heat Disinfect Temp TO Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 854 = [246] Title = Disinfection Interrupted ListTitle = Cleaning: RO Water Cond Error Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. [247] Title = ListTitle = Message = [248] Title = Disinfection Interrupted ListTitle = Cleaning: Chem Disinfect Proc Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. [249] Title = Chemical Disinfectant Not Detected ListTitle = Cleaning: Chem Disinf PAA Cond Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 992 = [250] Title = Connect Chemical Disinfectant ListTitle = Cleaning: Chem Disinf Connect PAA Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. SRSDG 1046 = [251] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Sys Rec NVRAM CRC Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [252] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Serv Rec NVRAM CRC Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [253] Title = Service Required: Dialysate Device ListTitle = DG POST: Sys Rec NVRAM CRC Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 609 = [254] Title = Service Required: Dialysate Device ListTitle = DG POST: Serv Rec NVRAM CRC Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 610 = [255] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Compatibility Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [256] Title = Chemical Flush in Progress ListTitle = Cleaning: Chem Flush In Progress Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. = SRSUI 497 [257] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Institutional Rec NVRAM CRC Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [258] Title = Service Required: Hemodialysis Device ListTitle = HD POST: OS Version Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [259] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Temp Sensors Error Message = A problem was detected with the hemodialysis device.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [260] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Image Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1067 [261] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI CAN Driver Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1068 [262] Title = ListTitle = Message = [263] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Touch Driver Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1070 [264] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI SD- Card Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1071 [265] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI RTC Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1072 [266] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI WiFi Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1073 [267] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Bluetooth Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1074 [268] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Ethernet Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1075 [269] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Sound Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1076 [270] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Proc Safestate Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [271] Title = Service Required: Dialysate Device ListTitle = DG POST: Proc Safestate Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 822 = [272] Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): Fan Speed Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [273] Title = ListTitle = Message = [274] Title = Service Required: Dialysate Device ListTitle = DG Fault: Reservoir Fault Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [275] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Arterial Press Range Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [276] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Venous Press Range Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [277] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Occlusion Sens Range Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [278] Title = Dialysate Reservoir Error ListTitle = Dial. Gen: Reservoir Overfill Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. [279] Title = Service Required: Dialysate Device ListTitle = DG Fault: Reservoir Drain Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 752 = [280] Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Direction Fault Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 888 = [281] Title = Acid Concentrate Low ListTitle = Dial. Gen: Acid Conc. Low Message = The acid concentrate is low.\n- Replace acid concentrate container. SRSDG 836 = [282] Title = Bicarb Concentrate Low ListTitle = Dial. Gen: Bicarb Conc. Low Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. SRSDG 837 = [283] Title = Service Required: Dialysate Device ListTitle = DG Fault: LC Range Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 615 = [284] Title = Service Required: Dialysate Device ListTitle = DG Fault: Redundant LC Check Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 383 = [285] Title = RO Permeate Sample in Progress ListTitle = Cleaning: RO Permeate Sample In Progress Message = RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [286] Title = Concentrate Connection Cap Removed ListTitle = Cleaning: Conc Cap Removed Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. [287] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Proc- FPGA Clock Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [288] Title = Service Required: Dialysate Device ListTitle = DG Fault: LC Drift Check Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [289] Title = Dialysate Connection Cap Removed ListTitle = Cleaning: Dialysate Cap Removed Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. SRSDG 928 = [290] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Ven. Pres. Com Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [291] Title = Flush In Progress ListTitle = Cleaning: Flush In Progress Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [292] Title = Heat Disinfection in Progress ListTitle = Cleaning: Heat Disinf In Progress Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [293] Title = Chemical Disinfection in Progress ListTitle = Cleaning: Chem Disinfect In Progress Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [294] Title = Active Cool in Progress ListTitle = Cleaning: Active Cool In Progress Message = Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [295] Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI CloudSync Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1238 [296] Title = Service Required: Dialysate Device ListTitle = DG Fault: TDi RTD Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 951 = [297] Title = Service Required: Dialysate Device ListTitle = DG Fault: TRo RTD Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 950 = [298] Title = Service Required: Dialysate Device ListTitle = DG Fault: Baro Pres Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 948 = [299] Title = Service Required: Dialysate Device ListTitle = DG POST: Serial Number Check Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 952 = [300] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Serial Number Check Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [301] Title = Remove Dialysate Cap ListTitle = Cleaning: RO Permeate Sample Connect Cap Message = - Remove the left dialysate cap.\n- Disinfect the bottom dialysate port using a 70% isopropyl\nalcohol wipe and allow to dry completely.\n- Using aseptic technique, connect extension tubing\nto bottom dialysate port.\n- Keep the right concentrate cap on the device. [302] Title = Service Required: Dialysate Device ListTitle = DG Fault: Processor RAM error Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 126 = [303] Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Hall Commutation Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 958 = [304] Title = Service Required: Hemodialysis Device ListTitle = HD POST: Usage Rec NVRAM CRC Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [305] Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Processor RAM error Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [306] Title = Turn Off Inlet Water Valves ListTitle = Cleaning: Turn Off Inlet Water Valves Message = Heat disinfection is in progress.\n- Turn off inlet water valves. [307] Title = Power Loss ListTitle = Blood Stop: Loss of AC Power Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. [308] Title = Service Required: Dialysate Device ListTitle = DG Fault: CPi Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 959 = [309] Title = Service Required: Dialysate Device ListTitle = DG Fault: CPo Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 960 = [310] Title = Service Required: Dialysate Device ListTitle = DG Fault: CD1 Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 962 = [311] Title = Service Required: Dialysate Device ListTitle = DG Fault: CD2 Comm Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 961 = [312] Title = Low Dialysate Flow ListTitle = Dial. Gen: Dialysate In Flow Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. [313] Title = Disinfection: Water Temperature Low ListTitle = Cleaning: Inlet Water Temp Low Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] Title = Disinfection: Water Pressure High ListTitle = Cleaning: Inlet Water Press High Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] Title = Disinfection: Water Pressure Low ListTitle = Cleaning: Inlet Water Press Low Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water pressure\n before initiating disinfection again. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] Title = ListTitle = Message = SRSDG 983 = [317] Title = ListTitle = Message = SRSDG 984 = [318] Title = ListTitle = Message = [319] Title = Residual Chemical Sampling ListTitle = Cleaning: Chem Residual Sampling Message = A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. SRSDG 1040 = [320] Title = Chemical Residual Timeout ListTitle = Cleaning: Chem Residual Timeout Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. SRSDG 1043 = [321] Title = Service Required: Dialysate Device ListTitle = DG Fault: UVo Error Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 1069 = [322] Title = Service Required: Hemodialysis Device ListTitle = HD POST Alarm: UI Invalid Year Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [323] Title = Service Required: Hemodialysis Device ListTitle = HD POST Alarm: UI Bad Settings Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.