[0] Title = Indication for when no alarms is a possible situation ListTitle = [1] Title = Service Required: Hemodialysis Device ListTitle = TD Software fault. Software found itself in an unexpected state TD Fault: SW Invalid State Step 1.2 Step 1.3 [2] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state DD Fault: Software Invalid State Step 2.2 = moda-fix.jpg Step 2.3 [3] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state DD Fault: Software Invalid State [4] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA POST failure TD POST: FPGA Compatibility [5] Title = Service Required: Hemodialysis Device ListTitle = TD Watchdog POST failure TD POST: WD Safestate [6] Title = Service Required: Hemodialysis Device ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up TD POST: Proc- UI Comm [7] Title = Service Required: Hemodialysis Device ListTitle = TD processor clock speed checks against FPGA clock failure TD Fault: Proc- FPGA Clock [8] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA communication down for too long TD Fault: FPGA Com [9] Title = Service Required: Hemodialysis Device ListTitle = User interface communication timeout TD Fault: UI Comm [10] Title = Communication Error ListTitle = Dialysate generator communication timeout Loss of DD Comm (Timeout) [11] Title = Service Required: Hemodialysis Device ListTitle = Too many bad CRCs detected on received system messages TD Fault: CAN CRC Com [12] Title = Service Required: Hemodialysis Device ListTitle = System message that required acknowledgment was not acknowledged TD Fault: CAN ACK UI Com [13] Title = Communication Error ListTitle = TD to DD comm error. Loss of DD Comm (DD ACK) [14] Title = Service Required: Hemodialysis Device ListTitle = TD critical data integrity check failed TD Fault: Critical Data Err [15] Title = Service Required: Dialysate Device ListTitle = DD critical data integrity check failed DD Fault: Critical Data Err [16] Title = Service Required: Hemodialysis Device ListTitle = TD watchdog expired TD Fault: WD Error [17] Title = Service Required: Dialysate Device ListTitle = DD watchdog expired DD Fault: WD Error [18] Title = Service Required: Dialysate Device ListTitle = Too many bad CRCs detected on received system messages DD Fault: CAN CRC Com [19] Title = Service Required: Dialysate Device ListTitle = DD to RO comm error Loss of RO Comm (RO ACK) [20] Title = Service Required: Dialysate Device ListTitle = DD to TD comm error Loss of TD Comm (TD ACK) [21] Title = Service Required: Dialysate Device ListTitle = DD FPGA POST failure DD POST: FPGA Compatibility [22] Title = Service Required: Dialysate Device ListTitle = FPGA clock speed failure DD Fault: Proc- FPGA Clock [23] Title = Service Required: Dialysate Device ListTitle = DD FPGA communication down for too long DD Fault: FPGA Com [24] Title = Service Required: Hemodialysis Device ListTitle = TD alarm audio failed POST TD POST: Primary Alarm Current [25] Title = Service Required: Dialysate Device ListTitle = DD FPGA not accepting commanded valve states DD Fault: Valve Control [26] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor reporting error or not being read successfully TD Arterial Pressure Sensor Error [27] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor reporting error or not being read successfully TD Venous Pressure Sensor Error [28] Title = Service Required: Dialysate Device ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully DD Hyd outlet Pressure Sensor Error [29] Title = Service Required: Dialysate Device ListTitle = DD bibag pressure sensor reporting error or not being read successfully DD Bibag Pressure Sensor Error [30] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pressure sensor reporting error or not being read successfully DD Spent Dial Pressure Sensor Error [31] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully DD Fresh Dial Pressure Sensor Error [32] Title = Service Required: Dialysate Device ListTitle = DD transmembrane sensor reporting error or not being read successfully DD Transmemb Pressure Sensor Error [33] Title = Service Required: Hemodialysis Device ListTitle = TD monitored voltage is out of range TD Fault: Voltage Monitor [34] Title = Power Loss ListTitle = TD AC power lost alarm Loss of AC Power [35] Title = Power Loss ListTitle = TD AC power lost alarm Blood Stop: Loss of AC Power [36] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully DD Conductivity Sensor Error [37] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully DD Conductivity Sensor Error [38] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully DD Conductivity Sensor Error [39] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully DD Conductivity Sensor Error [40] Title = Venous Air Detected ListTitle = TD venous air bubble detected alarm Blood Stop: Venous Air Bubble [41] Title = Blood Pump Stopped: Door Open ListTitle = TD cartridge door opened alarm Blood Stop: Door Open [42] Title = Stuck Button ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up TD POST: Stuck Button [43] Title = Low Arterial Pressure ListTitle = Arterial pressure too low during treatment Blood Stop: Arterial Pres. Low [44] Title = High Arterial Pressure ListTitle = Arterial pressure too high during treatment Blood Stop: Arterial Pres. High [45] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor is reading out of range TD Fault: Arterial Press Range [46] Title = Self Test: Cartridge ListTitle = TD arterial pressure self-test failure alarm Process: Pre- Tx Remove Cartridge [47] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment Blood Stop: Venous Pres. Low [48] Title = High Venous Pressure ListTitle = Venous pressure too high during treatment Blood Stop: Venous Pres. High [49] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor is reading out of range TD Fault: Venous Press Range [50] Title = Self Test: Cartridge ListTitle = TD venous pressure self-test failure alarm Process: Pre- Tx Remove Cartridge [51] Title = Self Test: Cartridge Check ListTitle = TD pre-treatment mode dry self-tests failure Process: Pre- Tx Dry Self- Test [52] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D11_PUMP speed control error DD Fault: Pump Speed Fault [53] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D10_PUMP speed control error DD Fault: Pump Speed Fault [54] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump hall sensor out of range. DD Fault: Pump Speed Fault [55] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump fault DD Fault: Pump Fault [56] Title = Service Required: Dialysate Device ListTitle = DD temperature sensor out of range DD Fault: Temp Sensor Range [57] Title = Service Required: Dialysate Device ListTitle = DD temperature sensors not being read successfully DD Fault: Temperature RTD Comm [58] Title = Service Required: Dialysate Device ListTitle = DD barometric sensor not being read successfully DD Fault: Baro temp Comm [59] Title = Service Required: Dialysate Device ListTitle = DD barometric pressure sensor bad coefficients CRC DD Fault: Baro temp Cal CRC [60] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump RPM out of range DD Fault: Fresh Dial Pump Speed Fault [61] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump off fault DD Fault: Fresh Dial Pump Speed Not Off [62] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump current out of range DD Fault: Fresh Dial Pump Current [63] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump invalid direction DD Fault: Fresh Dial Pump Dir Fault [64] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump direction FPGA fault DD Fault: Fresh Dial pump Dir Err Cnt [65] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump RPM out of range DD Fault: Spent Dial Pump Speed Fault [66] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump off fault DD Fault: Spent Dial Pump Speed Not Off [67] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump current out of range DD Fault: Spent Dial Pump Current [68] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump invalid direction DD Fault: Spent Dial Pump Dir Fault [69] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump direction FPGA fault DD Fault: Spent Dial pump Dir Err Cnt [70] Title = Service Required: Dialysate Device ListTitle = DD D5 heater is On with low fluid level DD Fault: Heater On With No Fluid [71] Title = Service Required: Dialysate Device ListTitle = DD D45 heater is On with low fluid level DD Fault: Heater On With No Fluid [72] Title = Service Required: Dialysate Device ListTitle = DD water inlet input pressure sensor reporting error or not being read successfully DD Water Inlet Input Press Sensor Error [73] Title = Service Required: Dialysate Device ListTitle = DD water inlet output pressure sensor reporting error or not being read successfully DD Water Inlet Output Press Sensor Error [74] Title = Incoming Water Pressure Input Low ListTitle = DD Inlet water pressure input too low DD: Inlet Water Pres Input Low [75] Title = Incoming Water Pressure Input High ListTitle = DD inlet water pressure input is too high DD: Inlet Water Pres Input High [76] Title = Incoming Water Pressure Output Low ListTitle = DD Inlet water pressure output too low DD: Inlet Water Pres Output Low [77] Title = Incoming Water Pressure Output High ListTitle = DD inlet water pressure output is too high DD: Inlet Water Pres Output High [78] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap fill timeout during treatment Blood Stop: Air Trap Fill [79] Title = Service Required: Hemodialysis Device ListTitle = Air trap level sensors reporting illegal combination of air/fluid TD Fault: Air Trap Sensor [80] Title = Service Required: Hemodialysis Device ListTitle = Pinch valve reporting a fault TD Fault: Pinch Valve [81] Title = Service Required: Hemodialysis Device ListTitle = TD valve homing failed TD Fault: Valve Home [82] Title = Service Required: Hemodialysis Device ListTitle = TD valve transition time out TD Fault: Valve Transition Err [83] Title = Communication Error ListTitle = DD to RO comm error. RO Loss of DD Comm (DD ACK) [84] Title = Service Required: Dialysate Device ListTitle = RO FPGA communication down for too long RO Fault: FPGA Comm [85] Title = Service Required: Hemodialysis Device ListTitle = RO Too many bad CRCs detected on received system messages RO Fault: CAN CRC Comm [86] Title = Service Required: Filtration Device ListTitle = RO FPGA not accepting commanded valve states RO Fault: Valve Control [87] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill start pressure is not in range DD Fault: BC Pressure not in range [88] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill end pressure is not in range DD Fault: BC Pressure not in range [89] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill start pressure is not in range DD Fault: BC Pressure not in range [90] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill end pressure is not in range DD Fault: BC Pressure not in range [91] Title = Service Required: Dialysate Device ListTitle = DD hydraulics chamber degassing pressure is not in range DD Fault: Hyd Cham Neg Pres not in range [92] Title = ListTitle = [93] Title = ListTitle = [94] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range DD Fault: BC Pressure not in range [95] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range DD Fault: BC Pressure not in range [96] Title = Blood Pump Stopped: Treatment Paused ListTitle = TD treatment stopped by user action - pressed stop button Blood Stopped: Treatment Pause [97] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully DD Conductivity Sensor Error [98] Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood sitting too long warning (> 4 min) Blood Stopped: Warning [99] Title = Blood Pump Stop Time Exceeded ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min) Blood Stop: Blood Clot Risk [100] Title = Service Required: Hemodialysis Device ListTitle = TD communication timeout TD Fault: UI- Proc Comm [101] Title = Service Required: Hemodialysis Device ListTitle = UI POST OS Version compatibility failure TD POST: OS Version [102] Title = Service Required: Hemodialysis Device ListTitle = UI POST Application Integrity (Sha256Sum) failure TD POST: UI Image [103] Title = Service Required: Hemodialysis Device ListTitle = UI POST CANBus failure TD POST: UI CAN Driver [104] Title = Service Required: Hemodialysis Device ListTitle = If the UI POST reports display failure. UI POST Display failure [105] Title = Service Required: Hemodialysis Device ListTitle = UI POST Touch failure TD POST: UI Touch Driver [106] Title = Service Required: Hemodialysis Device ListTitle = UI POST SD-Card failure TD POST: UI SD- Card [107] Title = Service Required: Hemodialysis Device ListTitle = UI POST RTC failure TD POST: UI RTC [108] Title = Service Required: Hemodialysis Device ListTitle = UI POST WiFi failure TD POST: UI WiFi [109] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bluetooth failure TD POST: UI Bluetooth [110] Title = Service Required: Hemodialysis Device ListTitle = UI POST Ethernet failure TD POST: UI Ethernet [111] Title = Service Required: Hemodialysis Device ListTitle = UI POST Sound failure TD POST: UI Sound [112] Title = Service Required: Hemodialysis Device ListTitle = UI POST cloud sync POST failure TD POST: UI CloudSync [113] Title = Service Required: Hemodialysis Device ListTitle = UI POST Invalid Year TD POST Alarm: UI Invalid Year [114] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bad Settings TD POST Alarm: UI Bad Settings [115] Title = Service Required: Hemodialysis Device ListTitle = SD card is non-operational TD Fault: UI- SD- Card Fault [116] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [117] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [118] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [119] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [120] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [121] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [122] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [123] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [124] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [125] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [126] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [127] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [128] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [129] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [130] Title = Service Required: Hemodialysis Device ListTitle = RESERVED TD Fault: UI- RESERVED [131] Title = Battery Capacity Low ListTitle = Insufficient battery charge to support 20 min without A/C power TD Alarm (Low): Battery Low [132] Title = High Dialysate Temperature ListTitle = Dialysate temperature is out of high safety range Dial. Gen: Dialysate Temp High [133] Title = High Dialysate Temperature ListTitle = Dialysate temperature is above the target temperature Dial. Gen: Dialysate Temp High [134] Title = Low Dialysate Temperature ListTitle = Dialysate temperature is below the target temperature Dial. Gen: Dialysate Temp Low [135] Title = Service Required: FP Device ListTitle = FP pressure relief is too low FP Fault: Pressure Relief Valve Low [136] Title = Service Required: FP Device ListTitle = FP pressure relief is too High FP Fault: Pressure Relief Valve High [137] Title = Service Required: FP Device ListTitle = FP device is using too much power FP Fault: Power usage has exceeded budget [138] Title = Service Required: FP Device ListTitle = FP RO pump flow out of range FP Fault: RO Pump Flow Low [139] Title = Low Inlet Water Temperature ListTitle = Inlet Water temperature is too low FP Water Gen: Inlet Water Temp Low [140] Title = High Inlet Water Temperature ListTitle = Inlet Water temperature is too high FP Water Gen: Inlet Water Temp High [141] Title = Low Inlet Water Pressure ListTitle = Inlet Water pressure is too low FP Water Gen: Inlet Pressure Low [142] Title = High Inlet Water Pressure ListTitle = Inlet Water pressure is too high FP Water Gen: Inlet Pressure High [143] Title = High Inlet Water Conductivity ListTitle = Inlet Water conductivity is too high FP Water Gen: Inlet Conductivity High [144] Title = Low Inlet Water Conductivity ListTitle = Inlet Water conductivity is too low FP Water Gen: Inlet Conductivity Low [145] Title = Service Required: FP Device ListTitle = FP RO pump pressure out of range FP Fault: RO Pump High Pressure [146] Title = Service Required: FP Device ListTitle = FP outlet primary conductivity sensor out of range FP Fault: RO Permeate Cond High [147] Title = Service Required: FP Device ListTitle = FP inlet primary conductivity sensor out of range FP Fault: RO Permeate Cond Low [148] Title = Service Required: FP Device ListTitle = FP RO rejection ratio out of range FP Fault: RO Rejection [149] Title = Service Required: FP Device ListTitle = FP pressure regulator is out of bounds FP Fault: Pressure regulator not in range [150] Title = FP: Filter Health ListTitle = FP pressure difference between the filters is too high FP Filters: Pressure difference too high [151] Title = FP: Filter Health ListTitle = FP conductivity after flushing the system is too high FP Filters: Permeate conductivity high [152] Title = FP: Filter Health ListTitle = FP volume difference between integrated flow and expected volume FP Filters: Volume difference too high [153] Title = FP: Permeate Overfill ListTitle = FP tank level is too high FP Tank: Tank is too full [154] Title = Service Required: Dialysate Device ListTitle = DD dialysate temperature sensors drift timeout. DD Fault: D28 and D30 Delta