[0] Title = ` Message = [1] Title = HD POST: Stuck Button Message = Stop or On button pressed during startup\n-Release button and power cycle\n-Call service to schedule a repair if problem persists [2] Title = HD POST: FPGA Compatibility Message = Call service to schedule a repair. [3] Title = DG POST: FPGA Compatibility Message = Call service to schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = HD POST: WD Safestate Message = Call service to schedule a repair. [5] Title = DG POST: WD Safestate Message = Call service to schedule a repair. SRSDG 53 = [6] Title = HD POST: Proc-UI Comm Message = Call service to schedule a repair. [7] Title = HD Alarm (Low): RTC Bat Low Message = Call service to schedule a repair. [8] Title = HD POST: Acc Vector Message = Call service to schedule a repair. [9] Title = DG POST: Acc Vector Message = Call service to schedule a repair. SRSDG 323 = [10] Title = HD POST: BLD Config Emb Mode Message = Call service to schedule a repair. [11] Title = Dial. Gen: Dialysate Temp Low Message = Dialysate temperature is low\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [12] Title = HD POST: HD Image Message = Call service to schedule a repair. [13] Title = DG POST: DG Image Message = Call service to schedule a repair. SRSDG 813 = [14] Title = DG POST: Usage Rec NVRAM CRC Message = Call service to schedule a repair. SRSDG 957 = [15] Title = Dial. Gen: Dialysate Temp High Message = Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. [16] Title = HD POST: Primary Alarm Current Message = Call service to schedule a repair. [17] Title = HD POST: UI POST Not Complete Message = Call service to schedule a repair. [18] Title = Loss of DG Comm (DG ACK) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [19] Title = Dial. Gen: Dialysate Temp High Message = Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [20] Title = DG POST: Conductivity Cal Message = Call service to schedule a repair. SRSDG 16 = [21] Title = DG POST: Drain Line Vol Cal Message = Call service to schedule a repair. SRSDG 936 = [22] Title = DG POST: Reservoir Vol Cal Message = Call service to schedule a repair. SRSDG 935 = [23] Title = DG POST: Acid Conc Cal Message = Call service to schedule a repair. SRSDG 934 = [24] Title = DG POST: Bicarb Conc Cal Message = Call service to schedule a repair. SRSDG 933 = [25] Title = DG POST: Acc Cal Message = Call service to schedule a repair. SRSDG 817 = [26] Title = HD POST: Acc Cal Message = Call service to schedule a repair. [27] Title = Process: Pre-Tx Wet Self-Test Message = Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [28] Title = DG Fault: Conductivity RTD Comm Message = Call service to schedule a repair. SRSDG 864 = [29] Title = HD POST: SP DAC Cal Message = Call service to schedule a repair. [30] Title = HD Fault: SW Invalid State Message = Call service to schedule a repair. [31] Title = HD Fault: BP Current Message = Call service to schedule a repair. [32] Title = HD Fault: BP Speed Message = Call service to schedule a repair. [33] Title = HD Fault: BP Direction Message = Call service to schedule a repair. [34] Title = HD Fault: BP Rotor Message = Blood pump rotor failure.\n- Check the pump rotor for secure connection. [35] Title = HD Fault: DPi Current Message = Call service to schedule a repair. [36] Title = HD Fault: DPi Speed Message = Call service to schedule a repair. [37] Title = HD Fault: DPi Direction Message = Call service to schedule a repair. [38] Title = HD Fault: DPi Rotor Message = Dialysate in pump rotor failure.\n- Check the pump rotor for secure connection. [39] Title = HD Fault: DPo Current Message = Call service to schedule a repair. [40] Title = HD Fault: DPo Speed Message = Call service to schedule a repair. [41] Title = HD Fault: DPo Direction Message = Call service to schedule a repair. [42] Title = HD Fault: DPo Rotor Message = Dialysate out pump rotor failure.\n- Check the pump rotor for secure connection. [43] Title = DG Fault: Iso ADC Comm Message = Call service to schedule a repair. SRSDG 954 = [44] Title = HD Fault: UI Comm Message = Call service to schedule a repair. [45] Title = HD Fault: CAN CRC Com Message = Call service to schedule a repair. [46] Title = HD Fault: CAN ACK UI Com Message = Call service to schedule a repair. [47] Title = Message = [48] Title = HD Fault: UF Rate Tare Err Message = Call service to schedule a repair. [49] Title = HD Fault: FPGA Com Message = Call service to schedule a repair. [50] Title = DG Fault: Valve Control Message = Call service to schedule a repair. SRSDG 26 = [51] Title = HD Fault: BP Speed Message = Call service to schedule a repair. [52] Title = HD Fault: DPi Speed Message = Call service to schedule a repair. [53] Title = HD Fault: DPo Speed Message = Call service to schedule a repair. [54] Title = HD Fault: Critical Data Err Message = Call service to schedule a repair. [55] Title = DG Fault: Critical Data Err Message = Call service to schedule a repair. SRSDG 921 = [56] Title = HD Fault: Acc Com Message = Call service to schedule a repair. [57] Title = DG Fault: Acc Com Message = Call service to schedule a repair. SRSDG 869 SRSDG 870 = [58] Title = HD Fault: Valve Home Message = Call service to schedule a repair. [59] Title = HD Fault: Valve Transition Err Message = Call service to schedule a repair. [60] Title = HD Fault: Valve Control Message = Call service to schedule a repair. [61] Title = HD Fault: Valve Current Message = Call service to schedule a repair. [62] Title = HD Fault: Valve Position Message = Call service to schedule a repair. [63] Title = HD Fault: BLD Com Message = Call service to schedule a repair. [64] Title = DG Fault: Baro Range Message = Call service to schedule a repair. SRSDG 874 = [65] Title = Loss of DG Comm (Reject) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [66] Title = HD POST: BLD Config Set Point Message = Call service to schedule a repair. [67] Title = Process: Pre-Tx Remove Cartridge Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [68] Title = Dial. Gen: Dialysate not ready Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. [69] Title = HD POST: BLD Cal Message = Call service to schedule a repair. [70] Title = Process: Pre-Tx Remove Cartridge Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [71] Title = Process: Pre-Tx Remove Cartridge Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [72] Title = Loss of DG Comm (LC) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [73] Title = Loss of DG Comm (TDI/TRO) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [74] Title = DG Fault: UVi On No Flow Message = Call service to schedule a repair. SRSDG 600 = [75] Title = HD Fault: Syringe DAC Message = Call service to schedule a repair. [76] Title = HD Fault: Voltage Monitor Message = Call service to schedule a repair. [77] Title = DG Fault: Voltage Monitor Message = Call service to schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = HD Fault: Syringe Pump Enc Dir Message = Call service to schedule a repair. [79] Title = HD Fault: Syringe Pump Drv Dir Message = Call service to schedule a repair. [80] Title = HD Fault: Syringe Pump Driver Message = Call service to schedule a repair. [81] Title = HD Fault: Syringe Pump OT Message = Call service to schedule a repair. [82] Title = HD Fault: Syringe Pump DAC Err Message = Call service to schedule a repair. [83] Title = HD Fault: Syringe Pump On Err Message = Call service to schedule a repair. [84] Title = DG POST: HD RTC Time Message = Call service to schedule a repair. SRSDG 815 = [85] Title = Blood Stop: Cartridge Lever Open Message = Cartridge Lock Lever is not closed.\n- Open front door and push lever all the way down.\n- Close door and press RESUME button to continue. [86] Title = HD POST: HD RTC Time Message = Call service to schedule a repair. [87] Title = DG POST: Heat Cal Message = Call service to schedule a repair. SRSDG 929 = [88] Title = DG Fault: CP1 CP2 Speed Range Message = Call service to schedule a repair. [89] Title = DG Fault: UVo On No Flow Message = Call service to schedule a repair. SRSDG 599 = [90] Title = DG Fault: LC A1/B1 Comm Message = Call service to schedule a repair. SRSDG 604 = [91] Title = Loss of DG Comm (FMD) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [92] Title = Message = [93] Title = DG Fault: RO Permeate Cond High Message = Call service to schedule a repair. SRSDG 430 = [94] Title = DG Fault: Pressure Sensor Range Message = Call service to schedule a repair. SRSDG 629 = [95] Title = DG Fault: WD Error Message = Call service to schedule a repair. SRSDG 907 = [96] Title = Dial. Gen: Inlet Water Temp High Message = Inlet water temperature is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 410 = [97] Title = Dial. Gen: Dialysate Cond. Message = Dialysate conductivity is out of range.\n- Check acid and bicarb concentrate is correct and not empty. SRSDG 397 = [98] Title = HD Alarm (Low): Battery Com Message = Call service to schedule a repair. [99] Title = HD Fault: Syringe Pump Stall Message = Call service to schedule a repair. [100] Title = HD Fault: Timeout No Cart ST Message = Call service to schedule a repair. [101] Title = HD POST: UI-Proc Comm Message = Call service to schedule a repair. [102] Title = Cleaning: Tpo THD Temp Delta Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] Title = HD POST: Cal NVRAM CRC Message = Call service to schedule a repair. [104] Title = HD Fault: Air Trap Sensor Message = Call service to schedule a repair. [105] Title = DG POST: Cal NVRAM CRC Message = Call service to schedule a repair. SRSDG 608 = [106] Title = DG POST: Flow Sensor Cal Message = Call service to schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = DG Fault: DG Restarted Message = Call service to schedule a repair. [108] Title = HD Fault: Syringe Pump ADC Com Message = Call service to schedule a repair. [109] Title = HD Fault: Syringe Pump Vol Message = Call service to schedule a repair. [110] Title = HD Fault: Syringe Pump Speed Message = Call service to schedule a repair. [111] Title = HD Fault: Syringe Pump Speed Message = Call service to schedule a repair. [112] Title = DG Fault: LC A2/B2 Comm Message = Call service to schedule a repair. SRSDG 953 = [113] Title = DG Fault: DRP Current Message = Call service to schedule a repair. SRSDG 890 = [114] Title = Process: Pre-Tx Wet Self-Test Message = Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [115] Title = DG Fault: Temp Sensor Range Message = Call service to schedule a repair. SRSDG 224 = [116] Title = HD Alarm (Low): SD-Card Error Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. [117] Title = Loss of HD Comm (HD ACK) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. SRSDG 909 = [118] Title = DG Fault: RTC Config Message = Call service to schedule a repair. SRSDG 43 = [119] Title = DG Alarm (Low): RTC Bat Low Message = Call service to schedule a repair. SRSDG 908 = [120] Title = Process: Pre-Tx Wet Self-Test Message = Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [121] Title = DG Fault: Prim Heater Lg Control Message = Call service to schedule a repair. SRSDG 611 = [122] Title = DG Fault: Prim Heater Sm Control Message = Call service to schedule a repair. SRSDG 612 = [123] Title = DG Fault: Trimmer Heater Control Message = Call service to schedule a repair. SRSDG 613 = [124] Title = Blood Stop: Blood Clot Risk Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [125] Title = Blood Stop: Blood Clot Risk Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [126] Title = Blood Stop: Dialyzer Blood Leak Message = A blood leak has been detected. Treatment must end. [127] Title = Blood Stop: Venous Pres. Low Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured \nproperly. [128] Title = Blood Stop: Venous Air Bubble Message = Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [129] Title = Message = [130] Title = Blood Stop: Venous Pres. High Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [131] Title = Blood Stop: Arterial Pres. Low Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [132] Title = Blood Stop: Arterial Pres. High Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [133] Title = DG Fault: Fluid Leak in DG Message = Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and \nelectrical connections from system.\n- Call service to schedule a repair. SRSDG 328 = [134] Title = Blood Stop: Fluid Leak in HD Message = Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment, BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment, resume treatment. [135] Title = Blood Stop: HD Shock Detected Message = Excessive movement detected with Hemodialysis Delivery \ndevice.\n- Confirm the device is stationary and level before continuing. [136] Title = Dial. Gen: Shock-Vibration Message = Excessive movement detected with Dialysis Generation \ndevice.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 327 = [137] Title = Blood Stop: HD Tilted Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. [138] Title = Dial. Gen: Tilt Message = Excessive tilt detected with Dialysis Generation device.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing. SRSDG 326 = [139] Title = Loss of AC Power Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns. [140] Title = Loss of DG Comm (Timeout) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [141] Title = Blood Stop: Air Trap Fill Message = Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [142] Title = Blood Stop: Blood Pump Occlusion Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the \nvenous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. [143] Title = DG Fault: TRo TDi Delta Message = Call service to schedule a repair. SRSDG 912 = [144] Title = Cleaning: Inlet Water Cond High Message = Cleaning Mode Inlet water conductivity too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] Title = Dial. Gen: Acid Conc. Cond. Message = Acid concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 846 = [146] Title = DG POST: DG RTC Timer Message = Call service to schedule a repair. SRSDG 312 = [147] Title = Dial. Gen: Dialysate not ready Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. SRSDG 398 = [148] Title = Cleaning: Inlet Water Temp High Message = Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = HD Fault: WD Error Message = Call service to schedule a repair. [150] Title = Dial. Gen: Inlet Water Cond High Message = Inlet water conductivity is high.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 391 = [151] Title = Dial. Gen: Inlet Water Cond Low Message = Inlet water conductivity is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 826 = [152] Title = Dial. Gen: Inlet Water Pres Low Message = Inlet water pressure is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 417 = [153] Title = Process: Pre-Tx Terminated Message = Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system. [154] Title = Cleaning: Inlet Water Cond Low Message = Cleaning Mode Inlet water conductivity too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] Title = Loss of DG Comm (Reservoir) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [156] Title = Loss of DG Comm (Op Mode) Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [157] Title = Cleaning: Chem Disinf PAA Prime Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 988 = [158] Title = Dial. Gen: Inlet Water Temp Low Message = Inlet water temperature is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 408 = [159] Title = Cleaning: Chem Disinf Connect Cap Message = Remove chemical disinfectant connection and close cap. SRSDG 1044 = [160] Title = HD Fault: BP Rotor Max Speed Message = Call service to schedule a repair. [161] Title = DG Fault: HD-DG Comm Message = Call service to schedule a repair. SRSDG 122 = [162] Title = DG Fault: Proc-FPGA Clock Message = Call service to schedule a repair. SRSDG 867 = [163] Title = Dial. Gen: UF Rate Check Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. [164] Title = Dial. Gen: UF Rate Check Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. [165] Title = Blood Stop: Blood Clot Risk Message = Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [166] Title = Message = [167] Title = Blood Stop: Door Open Message = Device door is open.\n - Close door and press RESUME button to continue. [168] Title = Message = [169] Title = DG Fault: Dialysate Flow Sens Rg Message = Call service to schedule a repair. SRSDG 643 = [170] Title = Heparin Syringe Empty Message = Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment, \nadminister manually. [171] Title = Process: Pre-Tx Hep. Pump Occ. Message = Heparin pump occlusion detected.\n- Check heparin line is unclamped.\n- Check there are no kinks in the heparin line. [172] Title = Process: Pre-Tx Not Enough Hep. Message = Not enough heparin is available for treatment.\n- Check if syringe pump is touching syringe plunger.\n- If syringe pump is touching syringe plunger, end treatment.\n- If syringe pump is not touching syringe plunger, reload \nsyringe with enough heparin. [173] Title = HD Fault: RTC Config Message = Call service to schedule a repair. [174] Title = HD POST: HD RTC Timer Message = Call service to schedule a repair. [175] Title = HD Fault: Pump Commutation Err Message = Call service to schedule a repair. [176] Title = Message = [177] Title = DG Fault: Software Invalid State Message = Call service to schedule a repair. SRSDG 919 = [178] Title = HD Fault: UI-Proc Comm Message = Call service to schedule a repair. = SRSUI 1066 [179] Title = DG Fault: FPGA Com Message = Call service to schedule a repair. SRSDG 605 = [180] Title = DG Fault: RO Flow Sensor Range Message = Call service to schedule a repair. SRSDG 883 = [181] Title = DG Fault: Tare Out of Range Message = Call service to schedule a repair. SRSDG 849 SRSDG 838 = [182] Title = DG POST: LC Cal Message = Call service to schedule a repair. SRSDG 162 = [183] Title = Cleaning: Invalid Resv Transfer Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 851 = [184] Title = DG Fault: UVi Error Message = Call service to schedule a repair. SRSDG 559 = [185] Title = DG Alarm (Low): Fan Speed Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. SRSDG 628 = [186] Title = DG Fault: CP1/CP2 Fault Message = Call service to schedule a repair. SRSDG 891 = [187] Title = DG Fault: CP1 Speed Fault Message = Call service to schedule a repair. SRSDG 905 = [188] Title = DG Fault: CP2 Speed Fault Message = Call service to schedule a repair. SRSDG 906 = [189] Title = DG Fault: DRP Speed Fault Message = Call service to schedule a repair. SRSDG 448 = [190] Title = DG Fault: DRP Speed Not Off Message = Call service to schedule a repair. [191] Title = DG Fault: RO Flow High Message = Call service to schedule a repair. SRSDG 885 = [192] Title = DG Fault: RO Flow Low Message = Call service to schedule a repair. SRSDG 945 = [193] Title = Message = [194] Title = DG Fault: RO Driver Feedback Message = Call service to schedule a repair. SRSDG 881 = [195] Title = DG Fault: RO Pump High Pressure Message = Call service to schedule a repair. SRSDG 925 = [196] Title = DG Fault: CPi/CPo Comm Message = Call service to schedule a repair. SRSDG 955 = [197] Title = DG Fault: CD1/CD2 Comm Message = Call service to schedule a repair. SRSDG 956 = [198] Title = DG Fault: HP On With No Flow Message = Call service to schedule a repair. SRSDG 872 = [199] Title = DG Fault: HR On With No Flow Message = Call service to schedule a repair. SRSDG 871 = [200] Title = DG Fault: Thermistor Range Message = Call service to schedule a repair. SRSDG 178 = [201] Title = Process: Pre-Tx Wet Self-Test Message = Self-test has detected a problem in the dialysate flow \nmeasurement or there is a leak in the dialysate in/out lines.\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [202] Title = Process: Pre-Tx Dry Self-Test Message = Self-test has detected blood circuit pressure cannot be \nrelieved.\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists, install a new cartridge and \nPBo transducer protector. [203] Title = Message = [204] Title = DG Fault: RO Rejection Message = Call service to schedule a repair. SRSDG 429 = [205] Title = DG Fault: Cond Sensors Error Message = Call service to schedule a repair. SRSDG 913 = [206] Title = DG Fault: Reservoir Fill Timeout Message = Call service to schedule a repair. SRSDG 437 = [207] Title = DG Fault: FMP Flow Rate Error Message = Call service to schedule a repair. SRSDG 240 = [208] Title = Blood Stop: Venous Pres. High Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [209] Title = Cleaning: Flush Circ Error Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 980 = [210] Title = HD Alarm (Low): Battery Fault Message = Call service to schedule a repair. [211] Title = Blood Stopped: Warning Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [212] Title = Blood Stopped: Warning Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [213] Title = Process: Pre-Tx Prime Complete Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [214] Title = Message = [215] Title = Message = [216] Title = Process: End of Dialysis Tx Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [217] Title = Message = [218] Title = Process: Remove Syringe Message = Syringe not removed from machine.\n- Remove heparin syringe. [219] Title = Blood Stop: Syringe Not Detected Message = Heparin is needed for treatment but syringe has been removed \nfrom syringe pump.\n- Install Heparin syringe. [220] Title = DG Fault: THD RTD Comm Message = Call service to schedule a repair. SRSDG 947 = [221] Title = Blood Stop: Saline Bag Empty Message = Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is \nempty. [222] Title = HD Fault: BP Occ Com Message = Call service to schedule a repair. [223] Title = HD Fault: Art. Pres. Com Message = Call service to schedule a repair. [224] Title = Blood Stopped: Treatment Pause Message = Treatment paused by user. [225] Title = Process: End of Treatment Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [226] Title = Process: Pre-Tx Priming Complete Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [227] Title = Message = [228] Title = Process: Pre-Tx Prime Timeout Message = Blood line prime with saline timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [229] Title = Process: Pre-Tx Prime Timeout Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [230] Title = Process: Pre-Tx Prime Timeout Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [231] Title = Process: Pre-Tx Dry Self-Test Message = Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion.\n- Check clamps.\n- If problem persists, install a new cartridge. [232] Title = Process: Pre-Tx Wet Self-Test Message = Self-test has detected a problem in the load cell \nmeasurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation onto machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [233] Title = DG Fault: Baro Cal CRC Message = Call service to schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] Title = Dial. Gen: Inlet Water Pres High Message = Inlet water pressure is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 1037 = [235] Title = Process: Rinseback Complete Message = Rinseback has been completed. [236] Title = Process: Pre-Tx Dry Self-Test Message = Self-test has detected fluid in the cartridge.\n- Cartridges are single-use only\n- Install new cartridge to resume. [237] Title = Process: Pre-Tx Prime Timeout Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [238] Title = Process: Pre-Tx Dry Self-Test Message = Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. [239] Title = Process: Post-Tx Remove Cartridge Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. [240] Title = Dial. Gen: Bicarb Conc. Cond. Message = Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 482 = [241] Title = Cleaning: Resv Drain Timeout Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 SRSDG 577 SRSDG 590 SRSDG 591 SRSDG 336 SRSDG 879 SRSDG 340 SRSDG 496 SRSDG 501 SRSDG 505 SRSDG 532 SRSDG 535 SRSDG 556 SRSDG 557 SRSDG 1048 SRSDG 1049 SRSDG 651 SRSDG 655 SRSDG 744 SRSDG 745 SRSDG 1004 SRSDG 1008 SRSDG 1019 SRSDG 1023 SRSDG 1033 SRSDG 1034 = [242] Title = Cleaning: Resv Fill Timeout Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 SRSDG 579 SRSDG 580 SRSDG 492 SRSDG 493 SRSDG 498 SRSDG 499 SRSDG 508 SRSDG 509 SRSDG 519 SRSDG 690 SRSDG 691 SRSDG 703 SRSDG 711 SRSDG 1020 SRSDG 1024 = [243] Title = Cleaning: Fluid Loss Detected Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 853 = [244] Title = Cleaning: Temp Sensor Check Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 852 = [245] Title = Cleaning: Heat Disinfect Temp TO Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 854 = [246] Title = Cleaning: RO Water Cond Error Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [247] Title = Message = [248] Title = Cleaning: Chem Disinfect Proc Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [249] Title = Cleaning: Chem Disinf PAA Cond Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 992 = [250] Title = Cleaning: Chem Disinf Connect PAA Message = Connect chemical disinfectant bottle to bicarb port. SRSDG 1046 = [251] Title = HD POST: Sys Rec NVRAM CRC Message = Call service to schedule a repair. [252] Title = HD POST: Serv Rec NVRAM CRC Message = Call service to schedule a repair. [253] Title = DG POST: Sys Rec NVRAM CRC Message = Call service to schedule a repair. SRSDG 609 = [254] Title = DG POST: Serv Rec NVRAM CRC Message = Call service to schedule a repair. SRSDG 610 = [255] Title = HD POST: UI Compatibility Message = Call service to schedule a repair. [256] Title = Cleaning: Chem Flush In Progress Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. = SRSUI 497 [257] Title = Message = [258] Title = Message = [259] Title = HD Fault: Temp Sensors Error Message = Call service to schedule a repair. [260] Title = HD POST: UI Image Message = Call service to schedule a repair. = SRSUI 1067 [261] Title = HD POST: UI CAN Driver Message = Call service to schedule a repair. = SRSUI 1068 [262] Title = HD POST: UI Disp Driver Message = Call service to schedule a repair. = SRSUI 1069 [263] Title = HD POST: UI Touch Driver Message = Call service to schedule a repair. = SRSUI 1070 [264] Title = HD POST: UI SD-Card Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. = SRSUI 1071 [265] Title = HD POST: UI RTC Message = Call service to schedule a repair. = SRSUI 1072 [266] Title = HD POST: UI WiFi Message = Call service to schedule a repair. = SRSUI 1073 [267] Title = HD POST: UI Bluetooth Message = Call service to schedule a repair. = SRSUI 1074 [268] Title = HD POST: UI Ethernet Message = Call service to schedule a repair. = SRSUI 1075 [269] Title = HD POST: UI Sound Message = Call service to schedule a repair. = SRSUI 1076 [270] Title = HD POST: Proc Safestate Message = Call service to schedule a repair. [271] Title = DG POST: Proc Safestate Message = Call service to schedule a repair. SRSDG 822 = [272] Title = HD Alarm (Low): Fan Speed Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. [273] Title = Message = [274] Title = Message = [275] Title = HD Fault: Arterial Press Range Message = Call service to schedule a repair. [276] Title = HD Fault: Venous Press Range Message = Call service to schedule a repair. [277] Title = HD Fault: Occlusion Sens Range Message = Call service to schedule a repair. [278] Title = DG Fault: Reservoir Overfill Message = Call service to schedule a repair. [279] Title = DG Fault: Reservoir Drain Message = Call service to schedule a repair. SRSDG 752 = [280] Title = DG Fault: DRP Direction Fault Message = Call service to schedule a repair. SRSDG 888 = [281] Title = Dial. Gen: Acid Conc. Low Message = Acid concentrate low.\n- Replace acid concentrate bottle. SRSDG 836 = [282] Title = Dial. Gen: Bicarb Conc. Low Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. SRSDG 837 = [283] Title = DG Fault: LC Range Message = Call service to schedule a repair. SRSDG 615 = [284] Title = DG Fault: Redundant LC Check Message = Call service to schedule a repair. SRSDG 383 = [285] Title = Message = [286] Title = Cleaning: Conc Cap Removed Message = Concentrate cap removed from DG during cleaning.\n- Cleaning Process Terminated. [287] Title = HD Fault: Proc-FPGA Clock Message = Call service to schedule a repair. [288] Title = DG Fault: LC Drift Check Message = Call service to schedule a repair. [289] Title = Cleaning: Dialysate Cap Removed Message = Dialysate cap removed from DG during cleaning.\n- Cleaning Process Terminated. SRSDG 928 = [290] Title = HD Fault: Ven. Pres. Com Message = Call service to schedule a repair. [291] Title = Cleaning: Flush In Progress Message = System flush in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [292] Title = Cleaning: Heat Disinf In Progress Message = Heat disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [293] Title = Cleaning: Chem Disinf In Progress Message = Chemical disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [294] Title = Message = [295] Title = HD POST: UI CloudSync Message = Call service to schedule a repair. = SRSUI 1238 [296] Title = DG Fault: TDi RTD Comm Message = Call service to schedule a repair. SRSDG 951 = [297] Title = DG Fault: TRo RTD Comm Message = Call service to schedule a repair. SRSDG 950 = [298] Title = DG Fault: Baro Pres Comm Message = Call service to schedule a repair. SRSDG 948 = [299] Title = DG POST: Serial Number Check Message = Call service to schedule a repair. SRSDG 952 = [300] Title = HD POST: Serial Number Check Message = Call service to schedule a repair. [301] Title = Message = [302] Title = DG Fault: Processor RAM error Message = Call service to schedule a repair. SRSDG 126 = [303] Title = DG Fault: DRP Hall Commutation Message = Call service to schedule a repair. SRSDG 958 = [304] Title = HD POST: Usage Rec NVRAM CRC Message = Call service to schedule a repair. [305] Title = HD Fault: Processor RAM error Message = Call service to schedule a repair. [306] Title = DG Fault: CD1 CD2 Check Message = Call service to schedule a repair. [307] Title = Blood Stop: Loss of AC Power Message = HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns.\n-If power is returned prior to blood stopped countdown elapses, \nrinseback will be allowed. [308] Title = DG Fault: CPi Comm Message = Call service to schedule a repair. SRSDG 959 = [309] Title = DG Fault: CPo Comm Message = Call service to schedule a repair. SRSDG 960 = [310] Title = DG Fault: CD1 Comm Message = Call service to schedule a repair. SRSDG 962 = [311] Title = DG Fault: CD2 Comm Message = Call service to schedule a repair. SRSDG 961 = [312] Title = Dial. Gen: Dialysate In Flow Message = Dialysate in flow rate is low\n- Check dialysate line connection to the DG. [313] Title = Cleaning: Inlet Water Temp Low Message = Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] Title = Cleaning: Inlet Water Press High Message = Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] Title = Cleaning: Inlet Water Press Low Message = Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] Title = DG Fault: Acid CP Pump Park Message = Call service to schedule a repair. SRSDG 983 = [317] Title = DG Fault: Bicarb CP Pump Park Message = Call service to schedule a repair. SRSDG 984 = [318] Title = HD Fault: Air Pump Timeout Message = Unable to reduce fluid level.\n- Air trap may be overfilled.\n- Pressure transducer protector may be contaminated.\n- Treatment must be terminated without option for rinseback.\n- Call service to schedule a repair. [319] Title = Cleaning: Chem Residual Sampling Message = Collect chemical residuals water sample from drain line. SRSDG 1040 = [320] Title = Cleaning: Chem Residual Timeout Message = During Chemical Disinfection Flush\n-user residual sampling timed out. SRSDG 1043 = [321] Title = DG Fault: UVo Error Message = Call service to schedule a repair. SRSDG 1069 = [322] Title = HD POST Alarm: UI Invalid Year Message = Call service to schedule a repair.