[0]
Title = Indication for when no alarms is a possible situation
ListTitle =
[1]
Title = Service Required: Hemodialysis Device
ListTitle = TD Software fault. Software found itself in an unexpected state
[2]
Title = Service Required: Dialysate Device
ListTitle = DD software fault. Software found itself in an unexpected state
[3]
Title = Service Required: Dialysate Device
ListTitle = DD software fault. Software found itself in an unexpected state
[4]
Title = Service Required: Hemodialysis Device
ListTitle = TD FPGA POST failure
[5]
Title = Service Required: Hemodialysis Device
ListTitle = TD Watchdog POST failure
[6]
Title = Service Required: Hemodialysis Device
ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up
[7]
Title = Service Required: Hemodialysis Device
ListTitle = TD processor clock speed checks against FPGA clock failure
[8]
Title = Service Required: Hemodialysis Device
ListTitle = TD FPGA communication down for too long
[9]
Title = Service Required: Hemodialysis Device
ListTitle = User interface communication timeout
[10]
Title = Communication Error
ListTitle = Dialysate generator communication timeout
[11]
Title = Service Required: Hemodialysis Device
ListTitle = Too many bad CRCs detected on received system messages
[12]
Title = Service Required: Hemodialysis Device
ListTitle = System message that required acknowledgment was not acknowledged
[13]
Title = Communication Error
ListTitle = TD to DD comm error.
[14]
Title = Service Required: Hemodialysis Device
ListTitle = TD critical data integrity check failed
[15]
Title = Service Required: Dialysate Device
ListTitle = DD critical data integrity check failed
[16]
Title = Service Required: Hemodialysis Device
ListTitle = TD watchdog expired
[17]
Title = Service Required: Dialysate Device
ListTitle = DD watchdog expired
[18]
Title = Service Required: Dialysate Device
ListTitle = Too many bad CRCs detected on received system messages
[19]
Title = Service Required: Dialysate Device
ListTitle = DD to RO comm error
[20]
Title = Service Required: Dialysate Device
ListTitle = DD to TD comm error
[21]
Title = Service Required: Dialysate Device
ListTitle = DD FPGA POST failure
[22]
Title = Service Required: Dialysate Device
ListTitle = FPGA clock speed failure
[23]
Title = Service Required: Dialysate Device
ListTitle = DD FPGA communication down for too long
[24]
Title = Service Required: Hemodialysis Device
ListTitle = TD alarm audio failed POST
[25]
Title = Service Required: Dialysate Device
ListTitle = DD FPGA not accepting commanded valve states
[26]
Title = Service Required: Hemodialysis Device
ListTitle = TD arterial pressure sensor reporting error or not being read successfully
[27]
Title = Service Required: Hemodialysis Device
ListTitle = TD venous pressure sensor reporting error or not being read successfully
[28]
Title = Low TMP Pressure
ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully
[29]
Title = High TMP Pressure
ListTitle = DD bibag pressure sensor reporting error or not being read successfully
[30]
Title = Self Test: Venous Air Detector
ListTitle = TD venous air bubble detector self-test failure
[31]
Title = MAX Blood Prime Pause Time
ListTitle = Blood Prime pause approaching timeout (> 4 min).
[32]
Title = MAX Blood Prime Pause Time Reached
ListTitle = Blood Prime pause timeout reached (> 5 min).
[33]
Title = Service Required: Hemodialysis Device
ListTitle = TD monitored voltage is out of range
[34]
Title = Power Loss
ListTitle = TD AC power lost alarm
[35]
Title = Power Loss
ListTitle = TD AC power lost alarm
[36]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully
[37]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully
[38]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully
[39]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully
[40]
Title = Venous Air Detected
ListTitle = TD venous air bubble detected alarm
[41]
Title = Blood Pump Stopped: Door Open
ListTitle = TD cartridge door opened alarm
[42]
Title = Stuck Button
ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up
[43]
Title = Low Arterial Pressure
ListTitle = Arterial pressure too low during treatment
[44]
Title = High Arterial Pressure
ListTitle = Arterial pressure too high during treatment
[45]
Title = Service Required: Hemodialysis Device
ListTitle = TD arterial pressure sensor is reading out of range
[46]
Title = Self Test: Cartridge
ListTitle = TD arterial pressure self-test failure alarm
[47]
Title = Low Venous Pressure
ListTitle = Venous pressure too low during treatment
[48]
Title = High Venous Pressure
ListTitle = Venous pressure too high during treatment
[49]
Title = Service Required: Hemodialysis Device
ListTitle = TD venous pressure sensor is reading out of range
[50]
Title = Self Test: Cartridge
ListTitle = TD venous pressure self-test failure alarm
[51]
Title = Self Test: Cartridge Check
ListTitle = TD pre-treatment mode dry self-tests failure
[52]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump D11_PUMP speed control error
[53]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump D10_PUMP speed control error
[54]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump hall sensor out of range.
[55]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump fault
[56]
Title = Service Required: Dialysate Device
ListTitle = DD temperature sensor out of range
[57]
Title = Service Required: Dialysate Device
ListTitle = DD temperature sensors not being read successfully
[58]
Title = Service Required: Dialysate Device
ListTitle = DD Balancing chamber fill timeout error
[59]
Title = Service Required: Dialysate Device
ListTitle = DD barometric pressure sensor bad coefficients CRC
[60]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump RPM out of range
[61]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump off fault
[62]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump current out of range
[63]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump invalid direction
[64]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump direction FPGA fault
[65]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump RPM out of range
[66]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump off fault
[67]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump current out of range
[68]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump invalid direction
[69]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump direction FPGA fault
[70]
Title = Service Required: Dialysate Device
ListTitle = DD D5 heater is On with low fluid level
[71]
Title = Service Required: Dialysate Device
ListTitle = DD D45 heater is On with low fluid level
[72]
Title = Service Required: Dialysate Device
ListTitle = TD barometric pressure sensor bad coefficients CRC
[73]
Title = Service Required: Dialysate Device
ListTitle = Barometric pressure is out of range
[74]
Title = Service Required: Dialysate Device
ListTitle = DD UF pump D76_PUMP speed control error
[75]
Title = Service Required: Dialysate Device
ListTitle = DD Spent chamber fill state 1 start pressure is not in range
[76]
Title = Service Required: Dialysate Device
ListTitle = DD Spent chamber fill state 2 start pressure is not in range
[77]
Title = Service Required: Dialysate Device
ListTitle = DD Spent chamber fill timeout error
[78]
Title = Blood Pump Stopped: Venous Air Detector
ListTitle = Air trap fill timeout during treatment
[79]
Title = Service Required: Hemodialysis Device
ListTitle = Air trap level sensors reporting illegal combination of air/fluid
[80]
Title = Service Required: Hemodialysis Device
ListTitle = Pinch valve reporting a fault
[81]
Title = Service Required: Hemodialysis Device
ListTitle = TD valve homing failed
[82]
Title = Service Required: Hemodialysis Device
ListTitle = TD valve transition time out
[83]
Title = Communication Error
ListTitle = DD to RO comm error.
[84]
Title = Service Required: Dialysate Device
ListTitle = RO FPGA communication down for too long
[85]
Title = Service Required: Hemodialysis Device
ListTitle = RO Too many bad CRCs detected on received system messages
[86]
Title = Service Required: Filtration Device
ListTitle = RO FPGA not accepting commanded valve states
[87]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 1 fill start pressure is not in range
[88]
Title = Service Required: Dialysate Device
ListTitle = AVAILABLE
[89]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 2 fill start pressure is not in range
[90]
Title = Service Required: Dialysate Device
ListTitle = AVAILABLE
[91]
Title = Service Required: Dialysate Device
ListTitle = DD hydraulics chamber degassing pressure is not in range
[92]
Title = Service Required: Dialysate Device
ListTitle = DD Bicarb chamber fill timeout error
[93]
Title = Service Required: Dialysate Device
ListTitle = DD D79 rinse pump commanded vs measured speed out of tolerance
[94]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range
[95]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range
[96]
Title = Blood Pump Stopped: Treatment Paused
ListTitle = TD treatment stopped by user action - pressed stop button
[97]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully
[98]
Title = Approaching MAX Blood Pump Stop Time
ListTitle = Blood sitting too long warning (> 4 min)
[99]
Title = Blood Pump Stop Time Exceeded
ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min)
[100]
Title = Service Required: Hemodialysis Device
ListTitle = TD communication timeout
[101]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST OS Version compatibility failure
[102]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Application Integrity (Sha256Sum) failure
[103]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST CANBus failure
[104]
Title = Service Required: Hemodialysis Device
ListTitle = If the UI POST reports display failure.
[105]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Touch failure
[106]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST SD-Card failure
[107]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST RTC failure
[108]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST WiFi failure
[109]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Bluetooth failure
[110]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Ethernet failure
[111]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Sound failure
[112]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST cloud sync POST failure
[113]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Invalid Year
[114]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Bad Settings
[115]
Title = Service Required: Hemodialysis Device
ListTitle = SD card is non-operational
[116]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[117]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[118]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[119]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[120]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[121]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[122]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[123]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[124]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[125]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[126]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[127]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[128]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[129]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[130]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[131]
Title = Battery Capacity Low
ListTitle = Insufficient battery charge to support 20 min without A/C power
[132]
Title = High Dialysate Temperature
ListTitle = Dialysate temperature is out of high safety range
[133]
Title = High Dialysate Temperature
ListTitle = Dialysate temperature is above the target temperature
[134]
Title = Low Dialysate Temperature
ListTitle = Dialysate temperature is below the target temperature
[135]
Title = Service Required: FP Device
ListTitle = FP pressure relief is too low
[136]
Title = Service Required: FP Device
ListTitle = FP pressure relief is too High
[137]
Title = Service Required: FP Device
ListTitle = FP device is using too much power
[138]
Title = Service Required: FP Device
ListTitle = FP RO pump flow out of range
[139]
Title = Low Inlet Water Temperature
ListTitle = Inlet Water temperature is too low
[140]
Title = High Inlet Water Temperature
ListTitle = Inlet Water temperature is too high
[141]
Title = Low Inlet Water Pressure
ListTitle = Inlet Water pressure is too low
[142]
Title = High Inlet Water Pressure
ListTitle = Inlet Water pressure is too high
[143]
Title = High Inlet Water Conductivity
ListTitle = Inlet Water conductivity is too high
[144]
Title = Low Inlet Water Conductivity
ListTitle = Inlet Water conductivity is too low
[145]
Title = Service Required: FP Device
ListTitle = FP RO pump pressure out of range
[146]
Title = Service Required: FP Device
ListTitle = FP outlet primary conductivity sensor out of range
[147]
Title = Service Required: FP Device
ListTitle = FP inlet primary conductivity sensor out of range
[148]
Title = Service Required: FP Device
ListTitle = FP RO rejection ratio out of range
[149]
Title = Service Required: FP Device
ListTitle = FP pressure regulator is out of bounds
[150]
Title = Service Required: FP Device
ListTitle = FP inlet temperature too high during filters flush
[151]
Title = FP: Filter Health
ListTitle = FP conductivity after flushing the system is too high
[152]
Title = FP: Filter Health
ListTitle = FP took more than 60 seconds to fill the permeate tank
[153]
Title = FP: Permeate Overfill
ListTitle = FP tank level is too high
[154]
Title = Service Required: Dialysate Device
ListTitle = DD dialysate temperature sensors drift timeout.
[155]
Title = Blood Leak
ListTitle = DD blood leak detected alarm
[156]
Title = Service Required: Hemodialysis Device
ListTitle = DD blood leak sensor invalid calibration record
[157]
Title = Service Required: Hemodialysis Device
ListTitle = DD blood leak sensor embedded mode communication failed
[158]
Title = Service Required: Hemodialysis Device
ListTitle = DD blood leak sensor set point set failure
[159]
Title = Blood Leak: Bypass
ListTitle = DD blood leak detected and it is in the process of attempting to remove the bubbles/foam
[160]
Title = FP: Permeate Underfill
ListTitle = FP tank level is too low
[161]
Title = High Inlet Water Pressure Fault
ListTitle = Inlet Water pressure is too high
[162]
Title = High Permeate flow
ListTitle = Permeate flow is too high from set target flow rate
[163]
Title = low Permeate flow
ListTitle = Permeate flow is too low from set target flow rate
[164]
Title = low inlet pressure
ListTitle = Inlet pressure is too low
[165]
Title = Blood Pump Stopped: Venous Air Detector
ListTitle = Air trap lowering timeout during treatment
[166]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump self-test failure alarm
[167]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump is running while the BP is off
[168]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump ADC error
[169]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump fault reported by FPGA
[170]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump DAC write failure
[171]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump speed check error
[172]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump volume check error
[173]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump over travel error
[174]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump not stopped in off state error
[175]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump direction (from encoder) error
[176]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump direction (from controller) error
[177]
Title = Heparin Syringe Empty
ListTitle = TD syringe empty alarm
[178]
Title = Blood Stopped: Syringe Not Detected
ListTitle = TD syringe pump syringe removed alarm
[179]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump stall alarm
[180]
Title = Service Required: Hemodialysis Device
ListTitle = TD heparin force sensor invalid calibration record
[181]
Title = Syringe Detected
ListTitle = TD syringe detected alarm
[182]
Title = Not Enough Heparin
ListTitle = TD syringe pump not enough Heparin alarm
[183]
Title = Heparin Pump Occlusion
ListTitle = TD syringe pump occlusion alarm
[184]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump timeout alarm
[185]
Title = Service Required: FP Device
ListTitle = FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully
[186]
Title = Service Required: FP Device
ListTitle = FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully
[187]
Title = Service Required: Hemodialysis Device
ListTitle = Blood pump failed to complete homing or one full revolution during auto-load or auto-eject
[188]
Title = Service Required: Hemodialysis Device
ListTitle = Ejector failed to reach extended or retracted position during auto-eject
[189]
Title = Service Required: Hemodialysis Device
ListTitle = TD ejector did not reach the fully retracted position within the allowed time during power-on self-test
[190]
Title = Blood Pump Stopped: Tubing Set Jam
ListTitle = Blood pump torque exceeded limit during Auto-Load/Eject one-revolution and tubing set jam suspected
[191]
Title = Air Pump Stall Detected
ListTitle = Air pump stall detected during treatment
[192]
Title = Service Required: Dialysate Device
ListTitle = TD CPU RAM check failure
[193]
Title = Service Required: Dialysate Device
ListTitle = DD CPU RAM check failure
[194]
Title = Service Required: Hemodialysis Device
ListTitle = TD firmware image integrity POST test failed
[195]
Title = Service Required: Hemodialysis Device
ListTitle = DD firmware image integrity POST test failed
[196]
Title = Blood Pump Stopped: Saline Bag Empty
ListTitle = TD Saline bag is empty
[197]
Title = Service Required: Dialysate Device
ListTitle = DD substitution pump RPM out of range
[198]
Title = Prime Complete - 30 MIN WARNING
ListTitle = TD prime completed medium priority alarm
[199]
Title = Blood Pump Stopped: Venous Air Detector
ListTitle = Repeated air trap lower events detected during treatment
[200]
Title = Wrong Tube Set Installed
ListTitle = TD installed tube set type does not match the selected treatment modality
[201]
Title = Load Syringe
ListTitle = Load Syringe
Fill Syring with <$FD8A06>{140:2:mL:1}$> of heparin = heparin_1.png
Connect heparin filled syring to infusion port = heparin_2.png
Insert syringe into the syring pump = heparin_3.png
Unclamp heparin pinch clamp = heparin_4.png
Lock heparin syring in place = heparin_5.png
[244]
Title = Low Dry Bicarbonate Cartridge
ListTitle = DD Low dry bicarb level
Press ‘OK’ and follow steps to disconnect dry bicarbonate cartridge. Dialysate flow rate is in bypass=dry_bicarb.png
[246]
Title = Connect Online Fluid Line
ListTitle = TD HDF submode change alarm
Use aseptic technique to cleanse the dilution port per facility policies
Connect online fluid line to the <$FD8A06>post-dialyzer dilution port$> = connect_online_fluid_2.png
Ensure all connection points are secure
Ensure online fluid line is properly routed and not kinked, clamped, compressed, or under tension
[247]
Title = Treatment Mode Changed
ListTitle = TD treatment modality change alarm
Treatment mode has been changed.
[248]
Title = Replace Dry Bicarbonate Cartridge
ListTitle = DD replace bicarb cartridge alarm
Pull <$2268FF>blue$> bicarbonate knob upwards and remove cartridge =dry_bicarb.png
Insert the new dry bicarbonate cartridge
Press ‘Resume’ to resume treatment
[249]
Title = Replace Acid Concentrate
ListTitle = DD replace acid concentrate alarm
Remove <$B53333>red$> acid connector from acid concentrate = acid_concentrate.png
Insert <$B53333>red$> acid connector into new acid concentrate
Press ‘Resume’ to resume treatment
[250]
Title = Dialysate Proportioning in Progress
ListTitle = Dialysate Proportioning in Progress
Machine is in bypass and treatment will automatically resume once conductivity is within range.
Review pH and conductivity based on facility policies once treatment is resumed.