[N/A] Title = Message = [Stop or Off button are pressed for more than 1 sec during POST.] Title = Stuck Button Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. [Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 53 = [If the UI failed to checkin via the CAN bus within 40 seconds] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the RTC is not functional as indicated by the bits in the first 3 control registers.] Title = Service Required: Hemodialysis Device Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [If the calculated vector length of the accelerometer x, y, z is not the nominal length.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 323 = [If the blood leak sensor embedded mode is unresponsive.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the dialysate temperature is below the target temperature for a certain period of time.] Title = Low Dialysate Temperature Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [If the calculated CRC of the HD image does not match the expected CRC.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the calculated CRC of the DG firmware image does not matched the expected CRC] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 813 = [If the DG usage information record's CRC does not match the calculated CRC in POST.] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 957 = [If the dialysate temperature is > 46 C for a certain period of time.] Title = High Dialysate Temperature Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [HD message to DG that requires acknowledge was not acknowledged within a certain time.] Title = Communication Error Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the dialysate temperature is above the target temperature for a certain period of time.] Title = High Dialysate Temperature Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [If the calibration data that was received from NVDataMgmt is NOT legitimate] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 16 = SRSDG 936 = SRSDG 935 = [If the non-volatile data is valid and if not raise the alarm] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 934 = SRSDG 933 = [If the calibration data that was received from NVDataMgmt is NOT legitimate] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 817 = [If the accel calibration data that was received from NVDataMgmt is NOT legitimate.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [BLD fails to complete zero sequence.] Title = Self Test: Blood Leak Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [If the FPGA does not report fresh data for a certain period of time] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 864 = [If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the specific software faults listed in alarm management SW faults has been detected] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the delta of the measured motor speed and the commanded motor speed is not within tolerance.] Title = Blood Pump Error Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If during command off state, the DPi pump motor speed is measured > 100.0 RMP.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds.] Title = Dialysate Pump Error Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed] Title = Dialysate Pump Error Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [If the FPGA does not report fresh data for a certain period of time] Title = Service Required: Dialysate Device Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 954 = [If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If too many bad messages CRCs are received within a set period of time] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [If no retries left in the pending acknowlegement list.] Title = Service Required: Hemodialysis Device Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.