[0]
Title = Indication for when no alarms is a possible situation
ListTitle =
[1]
Title = Service Required: Hemodialysis Device
ListTitle = TD Software fault. Software found itself in an unexpected state
[2]
Title = Service Required: Dialysate Device
ListTitle = DD software fault. Software found itself in an unexpected state
[3]
Title = Service Required: Dialysate Device
ListTitle = DD software fault. Software found itself in an unexpected state
[4]
Title = Service Required: Hemodialysis Device
ListTitle = TD FPGA POST failure
[5]
Title = Service Required: Hemodialysis Device
ListTitle = TD Watchdog POST failure
[6]
Title = Service Required: Hemodialysis Device
ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up
[7]
Title = Service Required: Hemodialysis Device
ListTitle = TD processor clock speed checks against FPGA clock failure
[8]
Title = Service Required: Hemodialysis Device
ListTitle = TD FPGA communication down for too long
[9]
Title = Service Required: Hemodialysis Device
ListTitle = User interface communication timeout
[10]
Title = Communication Error
ListTitle = Dialysate generator communication timeout
[11]
Title = Service Required: Hemodialysis Device
ListTitle = Too many bad CRCs detected on received system messages
[12]
Title = Service Required: Hemodialysis Device
ListTitle = System message that required acknowledgment was not acknowledged
[13]
Title = Communication Error
ListTitle = TD to DD comm error.
[14]
Title = Service Required: Hemodialysis Device
ListTitle = TD critical data integrity check failed
[15]
Title = Service Required: Dialysate Device
ListTitle = DD critical data integrity check failed
[16]
Title = Service Required: Hemodialysis Device
ListTitle = TD watchdog expired
[17]
Title = Service Required: Dialysate Device
ListTitle = DD watchdog expired
[18]
Title = Service Required: Dialysate Device
ListTitle = Too many bad CRCs detected on received system messages
[19]
Title = Service Required: Dialysate Device
ListTitle = DD to RO comm error
[20]
Title = Service Required: Dialysate Device
ListTitle = DD to TD comm error
[21]
Title = Service Required: Dialysate Device
ListTitle = DD FPGA POST failure
[22]
Title = Service Required: Dialysate Device
ListTitle = FPGA clock speed failure
[23]
Title = Service Required: Dialysate Device
ListTitle = DD FPGA communication down for too long
[24]
Title = Service Required: Hemodialysis Device
ListTitle = TD alarm audio failed POST
[25]
Title = Service Required: Dialysate Device
ListTitle = DD FPGA not accepting commanded valve states
[26]
Title = Service Required: Hemodialysis Device
ListTitle = TD arterial pressure sensor reporting error or not being read successfully
[27]
Title = Service Required: Hemodialysis Device
ListTitle = TD venous pressure sensor reporting error or not being read successfully
[28]
Title = Low TMP Pressure
ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully
[29]
Title = High TMP Pressure
ListTitle = DD bibag pressure sensor reporting error or not being read successfully
[30]
Title = Self Test: Venous Air Detector
ListTitle = TD venous air bubble detector self-test failure
[31]
Title = MAX Blood Prime Pause Time
ListTitle = Blood Prime pause approaching timeout (> 4 min).
[32]
Title = MAX Blood Prime Pause Time Reached
ListTitle = Blood Prime pause timeout reached (> 5 min).
[33]
Title = Service Required: Hemodialysis Device
ListTitle = TD monitored voltage is out of range
[34]
Title = Power Loss
ListTitle = TD AC power lost alarm
[35]
Title = Power Loss
ListTitle = TD AC power lost alarm
[36]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully
[37]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully
[38]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully
[39]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully
[40]
Title = Venous Air Detected
ListTitle = TD venous air bubble detected alarm
[41]
Title = Blood Pump Stopped: Door Open
ListTitle = TD cartridge door opened alarm
[42]
Title = Stuck Button
ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up
[43]
Title = Low Arterial Pressure
ListTitle = Arterial pressure too low during treatment
[44]
Title = High Arterial Pressure
ListTitle = Arterial pressure too high during treatment
[45]
Title = Service Required: Hemodialysis Device
ListTitle = TD arterial pressure sensor is reading out of range
[46]
Title = Self Test: Cartridge
ListTitle = TD arterial pressure self-test failure alarm
[47]
Title = Low Venous Pressure
ListTitle = Venous pressure too low during treatment
Ensure venous access lines are secured.
Check needle placement and fix access issues, if needed.
Ensure transducers are secured and no moisture is present. Replace if needed.
Inspect dialyzer for clots. Follow facility policy to end treatment if clots are present.
[48]
Title = High Venous Pressure
ListTitle = Venous pressure too high during treatment
[49]
Title = Service Required: Hemodialysis Device
ListTitle = TD venous pressure sensor is reading out of range
[50]
Title = Self Test: Cartridge
ListTitle = TD venous pressure self-test failure alarm
[51]
Title = Self Test: Cartridge Check
ListTitle = TD pre-treatment mode dry self-tests failure
[52]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump D11_PUMP speed control error
[53]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump D10_PUMP speed control error
[54]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump hall sensor out of range.
[55]
Title = Service Required: Dialysate Device
ListTitle = DD concentrate pump fault
[56]
Title = Service Required: Dialysate Device
ListTitle = DD temperature sensor out of range
[57]
Title = Service Required: Dialysate Device
ListTitle = DD temperature sensors not being read successfully
[58]
Title = Service Required: Dialysate Device
ListTitle = DD Balancing chamber fill timeout error
[59]
Title = Service Required: Dialysate Device
ListTitle = DD barometric pressure sensor bad coefficients CRC
[60]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump RPM out of range
[61]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump off fault
[62]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump current out of range
[63]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump invalid direction
[64]
Title = Service Required: Dialysate Device
ListTitle = DD fresh dialysate pump direction FPGA fault
[65]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump RPM out of range
[66]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump off fault
[67]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump current out of range
[68]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump invalid direction
[69]
Title = Service Required: Dialysate Device
ListTitle = DD spent dialysate pump direction FPGA fault
[70]
Title = Service Required: Dialysate Device
ListTitle = DD D5 heater is On with low fluid level
[71]
Title = Service Required: Dialysate Device
ListTitle = DD D45 heater is On with low fluid level
[72]
Title = Service Required: Dialysate Device
ListTitle = TD barometric pressure sensor bad coefficients CRC
[73]
Title = Service Required: Dialysate Device
ListTitle = Barometric pressure is out of range
[74]
Title = Service Required: Dialysate Device
ListTitle = DD UF pump D76_PUMP speed control error
[75]
Title = Service Required: Dialysate Device
ListTitle = DD Spent chamber fill state 1 start pressure is not in range
[76]
Title = Service Required: Dialysate Device
ListTitle = DD Spent chamber fill state 2 start pressure is not in range
[77]
Title = Service Required: Dialysate Device
ListTitle = DD Spent chamber fill timeout error
[78]
Title = Blood Pump Stopped: Venous Air Detector
ListTitle = Air trap fill timeout during treatment
[79]
Title = Service Required: Hemodialysis Device
ListTitle = Air trap level sensors reporting illegal combination of air/fluid
[80]
Title = Service Required: Hemodialysis Device
ListTitle = Pinch valve reporting a fault
[81]
Title = Service Required: Hemodialysis Device
ListTitle = TD valve homing failed
[82]
Title = Service Required: Hemodialysis Device
ListTitle = TD valve transition time out
[83]
Title = Communication Error
ListTitle = DD to RO comm error.
[84]
Title = Service Required: Dialysate Device
ListTitle = RO FPGA communication down for too long
[85]
Title = Service Required: Hemodialysis Device
ListTitle = RO Too many bad CRCs detected on received system messages
[86]
Title = Service Required: Filtration Device
ListTitle = RO FPGA not accepting commanded valve states
[87]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 1 fill start pressure is not in range
[88]
Title = Service Required: Dialysate Device
ListTitle = AVAILABLE
[89]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 2 fill start pressure is not in range
[90]
Title = Service Required: Dialysate Device
ListTitle = AVAILABLE
[91]
Title = Service Required: Dialysate Device
ListTitle = DD hydraulics chamber degassing pressure is not in range
[92]
Title = Service Required: Dialysate Device
ListTitle = DD Bicarb chamber fill timeout error
[93]
Title = Service Required: Dialysate Device
ListTitle = DD D79 rinse pump commanded vs measured speed out of tolerance
[94]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range
[95]
Title = Service Required: Dialysate Device
ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range
[96]
Title = Blood Pump Stopped: Treatment Paused
ListTitle = TD treatment stopped by user action - pressed stop button
Treatment will end without blood return if not resumed within five minutes.
[97]
Title = Service Required: Dialysate Device
ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully
[98]
Title = Approaching MAX Blood Pump Stop Time
ListTitle = Blood sitting too long warning (> 4 min)
[99]
Title = Blood Pump Stop Time Exceeded
ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min)
[100]
Title = Service Required: Hemodialysis Device
ListTitle = TD communication timeout
[101]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST OS Version compatibility failure
[102]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Application Integrity (Sha256Sum) failure
[103]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST CANBus failure
[104]
Title = Service Required: Hemodialysis Device
ListTitle = If the UI POST reports display failure.
[105]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Touch failure
[106]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST SD-Card failure
[107]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST RTC failure
[108]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST WiFi failure
[109]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Bluetooth failure
[110]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Ethernet failure
[111]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Sound failure
[112]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST cloud sync POST failure
[113]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Invalid Year
[114]
Title = Service Required: Hemodialysis Device
ListTitle = UI POST Bad Settings
[115]
Title = Service Required: Hemodialysis Device
ListTitle = SD card is non-operational
[116]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[117]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[118]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[119]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[120]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[121]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[122]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[123]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[124]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[125]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[126]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[127]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[128]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[129]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[130]
Title = Service Required: Hemodialysis Device
ListTitle = RESERVED
[131]
Title = Battery Capacity Low
ListTitle = Insufficient battery charge to support 20 min without A/C power
[132]
Title = High Dialysate Temperature
ListTitle = Dialysate temperature is out of high safety range
[133]
Title = High Dialysate Temperature
ListTitle = Dialysate temperature is above the target temperature
[134]
Title = Low Dialysate Temperature
ListTitle = Dialysate temperature is below the target temperature
[135]
Title = Service Required: FP Device
ListTitle = FP pressure relief is too low
[136]
Title = Service Required: FP Device
ListTitle = FP pressure relief is too High
[137]
Title = Service Required: FP Device
ListTitle = FP device is using too much power
[138]
Title = Service Required: FP Device
ListTitle = FP RO pump flow out of range
[139]
Title = Low Inlet Water Temperature
ListTitle = Inlet Water temperature is too low
[140]
Title = High Inlet Water Temperature
ListTitle = Inlet Water temperature is too high
[141]
Title = Low Inlet Water Pressure
ListTitle = Inlet Water pressure is too low
[142]
Title = High Inlet Water Pressure
ListTitle = Inlet Water pressure is too high
[143]
Title = High Inlet Water Conductivity
ListTitle = Inlet Water conductivity is too high
[144]
Title = Low Inlet Water Conductivity
ListTitle = Inlet Water conductivity is too low
[145]
Title = Service Required: FP Device
ListTitle = FP RO pump pressure out of range
[146]
Title = Service Required: FP Device
ListTitle = FP outlet primary conductivity sensor out of range
[147]
Title = Service Required: FP Device
ListTitle = FP inlet primary conductivity sensor out of range
[148]
Title = Service Required: FP Device
ListTitle = FP RO rejection ratio out of range
[149]
Title = Service Required: FP Device
ListTitle = FP pressure regulator is out of bounds
[150]
Title = Service Required: FP Device
ListTitle = FP inlet temperature too high during filters flush
[151]
Title = FP: Filter Health
ListTitle = FP conductivity after flushing the system is too high
[152]
Title = FP: Filter Health
ListTitle = FP took more than 60 seconds to fill the permeate tank
[153]
Title = FP: Permeate Overfill
ListTitle = FP tank level is too high
[154]
Title = Service Required: Dialysate Device
ListTitle = DD dialysate temperature sensors drift timeout.
[155]
Title = Blood Leak
ListTitle = DD blood leak detected alarm
[156]
Title = Service Required: Hemodialysis Device
ListTitle = DD blood leak sensor invalid calibration record
[157]
Title = Service Required: Hemodialysis Device
ListTitle = DD blood leak sensor embedded mode communication failed
[158]
Title = Service Required: Hemodialysis Device
ListTitle = DD blood leak sensor set point set failure
[159]
Title = Blood Leak: Bypass
ListTitle = DD blood leak detected and it is in the process of attempting to remove the bubbles/foam
[160]
Title = FP: Permeate Underfill
ListTitle = FP tank level is too low
[161]
Title = High Inlet Water Pressure Fault
ListTitle = Inlet Water pressure is too high
[162]
Title = High Permeate flow
ListTitle = Permeate flow is too high from set target flow rate
[163]
Title = low Permeate flow
ListTitle = Permeate flow is too low from set target flow rate
[164]
Title = low inlet pressure
ListTitle = Inlet pressure is too low
[165]
Title = Blood Pump Stopped: Venous Air Detector
ListTitle = Air trap lowering timeout during treatment
[166]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump self-test failure alarm
[167]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump is running while the BP is off
[168]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump ADC error
[169]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump fault reported by FPGA
[170]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump DAC write failure
[171]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump speed check error
[172]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump volume check error
[173]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump over travel error
[174]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump not stopped in off state error
[175]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump direction (from encoder) error
[176]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump direction (from controller) error
[177]
Title = Heparin Syringe Empty
ListTitle = TD syringe empty alarm
[178]
Title = Blood Stopped: Syringe Not Detected
ListTitle = TD syringe pump syringe removed alarm
[179]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump stall alarm
[180]
Title = Service Required: Hemodialysis Device
ListTitle = TD heparin force sensor invalid calibration record
[181]
Title = Syringe Detected
ListTitle = TD syringe detected alarm
[182]
Title = Not Enough Heparin
ListTitle = TD syringe pump not enough Heparin alarm
[183]
Title = Heparin Pump Occlusion
ListTitle = TD syringe pump occlusion alarm
[184]
Title = Service Required: Hemodialysis Device
ListTitle = TD syringe pump timeout alarm
[185]
Title = Service Required: FP Device
ListTitle = FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully
[186]
Title = Service Required: FP Device
ListTitle = FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully
[187]
Title = Service Required: Hemodialysis Device
ListTitle = Blood pump failed to complete homing or one full revolution during auto-load or auto-eject
[188]
Title = Service Required: Hemodialysis Device
ListTitle = Ejector failed to reach extended or retracted position during auto-eject
[189]
Title = Service Required: Hemodialysis Device
ListTitle = TD ejector did not reach the fully retracted position within the allowed time during power-on self-test
[190]
Title = Blood Pump Stopped: Tubing Set Jam
ListTitle = Blood pump torque exceeded limit during Auto-Load/Eject one-revolution and tubing set jam suspected
[191]
Title = Air Pump Stall Detected
ListTitle = Air pump stall detected during treatment
[192]
Title = Service Required: Dialysate Device
ListTitle = TD CPU RAM check failure
[193]
Title = Service Required: Dialysate Device
ListTitle = DD CPU RAM check failure
[194]
Title = Service Required: Hemodialysis Device
ListTitle = TD firmware image integrity POST test failed
[195]
Title = Service Required: Hemodialysis Device
ListTitle = DD firmware image integrity POST test failed
[196]
Title = Blood Pump Stopped: Saline Bag Empty
ListTitle = TD Saline bag is empty
[197]
Title = Service Required: Dialysate Device
ListTitle = DD substitution pump RPM out of range
[198]
Title = Prime Complete - 30 MIN WARNING
ListTitle = TD prime completed medium priority alarm
[199]
Title = Blood Pump Stopped: Venous Air Detector
ListTitle = Repeated air trap lower events detected during treatment
[200]
Title = Wrong Tube Set Installed
ListTitle = TD installed tube set type does not match the selected treatment modality
[201]
Title = Load Syringe
ListTitle = Load Syringe
Fill syringe with <$FD8A06>{140:2:mL:1}$> of heparin = heparin_1.png
Connect heparin filled syringe to infusion port = heparin_2.png
Insert syringe into the syringe pump = heparin_3.png
Unclamp heparin pinch clamp = heparin_4.png
Lock heparin syringe in place = heparin_5.png
[210]
Title = Prime Completed
ListTitle = TD Prime Completed
System priming is completed. Press ‘OK’ to continue.
[244]
Title = Low Dry Bicarbonate Cartridge
ListTitle = DD Low dry bicarb level
Press ‘Resume’ and follow steps to disconnect dry bicarbonate cartridge. Dialysate flow rate is in bypass=dry_bicarb.png
[246]
Title = Connect Online Fluid Line
ListTitle = TD HDF submode change alarm
Use aseptic technique to cleanse the dilution port per facility policies
Connect online fluid line to the <$FD8A06>post-dialyzer dilution port$> = connect_online_fluid_2.png
Ensure all connection points are secure
Ensure online fluid line is properly routed and not kinked, clamped, compressed, or under tension
[247]
Title = Treatment Mode Changed
ListTitle = TD treatment modality change alarm
Treatment mode has been changed.
[248]
Title = Replace Dry Bicarbonate Cartridge
ListTitle = DD replace bicarb cartridge alarm
Pull <$2268FF>blue$> bicarbonate knob upwards and remove cartridge =dry_bicarb.png
Insert the new dry bicarbonate cartridge
Press ‘Resume’ to resume treatment
[249]
Title = Replace Acid Concentrate
ListTitle = DD replace acid concentrate alarm
Remove <$B53333>red$> acid connector from acid concentrate = acid_concentrate.png
Insert <$B53333>red$> acid connector into new acid concentrate
Press ‘Resume’ button to resume treatment
[250]
Title = Dialysate Proportioning in Progress
ListTitle = Dialysate Proportioning in Progress
Machine is in bypass and treatment will automatically resume once conductivity is within range.
Review pH and conductivity based on facility policies once treatment is resumed.
[251]
Title = Dialysis Treatment Complete
ListTitle = Dialysis Treatment Complete
Press ‘OK’ to continue to end of treatment options
[252]
Title = End of Treatment Timeout
ListTitle = End of Treatment Timeout
The maximum time allowed on this screen has been reached.
Press 'Return Blood & End Treatment' to begin the blood return process