[0] Title = Indication for when no alarms is a possible situation ListTitle = [1] Title = Service Required: Hemodialysis Device ListTitle = TD Software fault. Software found itself in an unexpected state [2] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state [3] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state [4] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA POST failure [5] Title = Service Required: Hemodialysis Device ListTitle = TD Watchdog POST failure [6] Title = Service Required: Hemodialysis Device ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up [7] Title = Service Required: Hemodialysis Device ListTitle = TD processor clock speed checks against FPGA clock failure [8] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA communication down for too long [9] Title = Service Required: Hemodialysis Device ListTitle = User interface communication timeout [10] Title = Communication Error ListTitle = Dialysate generator communication timeout [11] Title = Service Required: Hemodialysis Device ListTitle = Too many bad CRCs detected on received system messages [12] Title = Service Required: Hemodialysis Device ListTitle = System message that required acknowledgment was not acknowledged [13] Title = Communication Error ListTitle = TD to DD comm error. [14] Title = Service Required: Hemodialysis Device ListTitle = TD critical data integrity check failed [15] Title = Service Required: Dialysate Device ListTitle = DD critical data integrity check failed [16] Title = Service Required: Hemodialysis Device ListTitle = TD watchdog expired [17] Title = Service Required: Dialysate Device ListTitle = DD watchdog expired [18] Title = Service Required: Dialysate Device ListTitle = Too many bad CRCs detected on received system messages [19] Title = Service Required: Dialysate Device ListTitle = DD to RO comm error [20] Title = Service Required: Dialysate Device ListTitle = DD to TD comm error [21] Title = Service Required: Dialysate Device ListTitle = DD FPGA POST failure [22] Title = Service Required: Dialysate Device ListTitle = FPGA clock speed failure [23] Title = Service Required: Dialysate Device ListTitle = DD FPGA communication down for too long [24] Title = Service Required: Hemodialysis Device ListTitle = TD alarm audio failed POST [25] Title = Service Required: Dialysate Device ListTitle = DD FPGA not accepting commanded valve states [26] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor reporting error or not being read successfully [27] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor reporting error or not being read successfully [28] Title = Low TMP Pressure ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully [29] Title = High TMP Pressure ListTitle = DD bibag pressure sensor reporting error or not being read successfully [30] Title = Self Test: Venous Air Detector ListTitle = TD venous air bubble detector self-test failure [31] Title = MAX Blood Prime Pause Time ListTitle = Blood Prime pause approaching timeout (> 4 min). [32] Title = MAX Blood Prime Pause Time Reached ListTitle = Blood Prime pause timeout reached (> 5 min). [33] Title = Service Required: Hemodialysis Device ListTitle = TD monitored voltage is out of range [34] Title = Power Loss ListTitle = TD AC power lost alarm [35] Title = Power Loss ListTitle = TD AC power lost alarm [36] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully [37] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully [38] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully [39] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully [40] Title = Venous Air Detected ListTitle = TD venous air bubble detected alarm [41] Title = Blood Pump Stopped: Door Open ListTitle = TD cartridge door opened alarm [42] Title = Stuck Button ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up [43] Title = Low Arterial Pressure ListTitle = Arterial pressure too low during treatment [44] Title = High Arterial Pressure ListTitle = Arterial pressure too high during treatment [45] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor is reading out of range [46] Title = Self Test: Cartridge ListTitle = TD arterial pressure self-test failure alarm [47] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment Ensure venous access lines are secured. Check needle placement and fix access issues, if needed. Ensure transducers are secured and no moisture is present. Replace if needed. Inspect dialyzer for clots. Follow facility policy to end treatment if clots are present. [48] Title = High Venous Pressure ListTitle = Venous pressure too high during treatment [49] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor is reading out of range [50] Title = Self Test: Cartridge ListTitle = TD venous pressure self-test failure alarm [51] Title = Self Test: Cartridge Check ListTitle = TD pre-treatment mode dry self-tests failure [52] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D11_PUMP speed control error [53] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D10_PUMP speed control error [54] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump hall sensor out of range. [55] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump fault [56] Title = Service Required: Dialysate Device ListTitle = DD temperature sensor out of range [57] Title = Service Required: Dialysate Device ListTitle = DD temperature sensors not being read successfully [58] Title = Service Required: Dialysate Device ListTitle = DD Balancing chamber fill timeout error [59] Title = Service Required: Dialysate Device ListTitle = DD barometric pressure sensor bad coefficients CRC [60] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump RPM out of range [61] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump off fault [62] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump current out of range [63] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump invalid direction [64] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump direction FPGA fault [65] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump RPM out of range [66] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump off fault [67] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump current out of range [68] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump invalid direction [69] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump direction FPGA fault [70] Title = Service Required: Dialysate Device ListTitle = DD D5 heater is On with low fluid level [71] Title = Service Required: Dialysate Device ListTitle = DD D45 heater is On with low fluid level [72] Title = Service Required: Dialysate Device ListTitle = TD barometric pressure sensor bad coefficients CRC [73] Title = Service Required: Dialysate Device ListTitle = Barometric pressure is out of range [74] Title = Service Required: Dialysate Device ListTitle = DD UF pump D76_PUMP speed control error [75] Title = Service Required: Dialysate Device ListTitle = DD Spent chamber fill state 1 start pressure is not in range [76] Title = Service Required: Dialysate Device ListTitle = DD Spent chamber fill state 2 start pressure is not in range [77] Title = Service Required: Dialysate Device ListTitle = DD Spent chamber fill timeout error [78] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap fill timeout during treatment [79] Title = Service Required: Hemodialysis Device ListTitle = Air trap level sensors reporting illegal combination of air/fluid [80] Title = Service Required: Hemodialysis Device ListTitle = Pinch valve reporting a fault [81] Title = Service Required: Hemodialysis Device ListTitle = TD valve homing failed [82] Title = Service Required: Hemodialysis Device ListTitle = TD valve transition time out [83] Title = Communication Error ListTitle = DD to RO comm error. [84] Title = Service Required: Dialysate Device ListTitle = RO FPGA communication down for too long [85] Title = Service Required: Hemodialysis Device ListTitle = RO Too many bad CRCs detected on received system messages [86] Title = Service Required: Filtration Device ListTitle = RO FPGA not accepting commanded valve states [87] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill start pressure is not in range [88] Title = Service Required: Dialysate Device ListTitle = AVAILABLE [89] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill start pressure is not in range [90] Title = Service Required: Dialysate Device ListTitle = AVAILABLE [91] Title = Service Required: Dialysate Device ListTitle = DD hydraulics chamber degassing pressure is not in range [92] Title = Service Required: Dialysate Device ListTitle = DD Bicarb chamber fill timeout error [93] Title = Service Required: Dialysate Device ListTitle = DD D79 rinse pump commanded vs measured speed out of tolerance [94] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range [95] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range [96] Title = Blood Pump Stopped: Treatment Paused ListTitle = TD treatment stopped by user action - pressed stop button Treatment will end without blood return if not resumed within five minutes. [97] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully [98] Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood sitting too long warning (> 4 min) [99] Title = Blood Pump Stop Time Exceeded ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min) [100] Title = Service Required: Hemodialysis Device ListTitle = TD communication timeout [101] Title = Service Required: Hemodialysis Device ListTitle = UI POST OS Version compatibility failure [102] Title = Service Required: Hemodialysis Device ListTitle = UI POST Application Integrity (Sha256Sum) failure [103] Title = Service Required: Hemodialysis Device ListTitle = UI POST CANBus failure [104] Title = Service Required: Hemodialysis Device ListTitle = If the UI POST reports display failure. [105] Title = Service Required: Hemodialysis Device ListTitle = UI POST Touch failure [106] Title = Service Required: Hemodialysis Device ListTitle = UI POST SD-Card failure [107] Title = Service Required: Hemodialysis Device ListTitle = UI POST RTC failure [108] Title = Service Required: Hemodialysis Device ListTitle = UI POST WiFi failure [109] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bluetooth failure [110] Title = Service Required: Hemodialysis Device ListTitle = UI POST Ethernet failure [111] Title = Service Required: Hemodialysis Device ListTitle = UI POST Sound failure [112] Title = Service Required: Hemodialysis Device ListTitle = UI POST cloud sync POST failure [113] Title = Service Required: Hemodialysis Device ListTitle = UI POST Invalid Year [114] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bad Settings [115] Title = Service Required: Hemodialysis Device ListTitle = SD card is non-operational [116] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [117] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [118] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [119] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [120] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [121] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [122] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [123] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [124] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [125] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [126] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [127] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [128] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [129] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [130] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [131] Title = Battery Capacity Low ListTitle = Insufficient battery charge to support 20 min without A/C power [132] Title = High Dialysate Temperature ListTitle = Dialysate temperature is out of high safety range [133] Title = High Dialysate Temperature ListTitle = Dialysate temperature is above the target temperature [134] Title = Low Dialysate Temperature ListTitle = Dialysate temperature is below the target temperature [135] Title = Service Required: FP Device ListTitle = FP pressure relief is too low [136] Title = Service Required: FP Device ListTitle = FP pressure relief is too High [137] Title = Service Required: FP Device ListTitle = FP device is using too much power [138] Title = Service Required: FP Device ListTitle = FP RO pump flow out of range [139] Title = Low Inlet Water Temperature ListTitle = Inlet Water temperature is too low [140] Title = High Inlet Water Temperature ListTitle = Inlet Water temperature is too high [141] Title = Low Inlet Water Pressure ListTitle = Inlet Water pressure is too low [142] Title = High Inlet Water Pressure ListTitle = Inlet Water pressure is too high [143] Title = High Inlet Water Conductivity ListTitle = Inlet Water conductivity is too high [144] Title = Low Inlet Water Conductivity ListTitle = Inlet Water conductivity is too low [145] Title = Service Required: FP Device ListTitle = FP RO pump pressure out of range [146] Title = Service Required: FP Device ListTitle = FP outlet primary conductivity sensor out of range [147] Title = Service Required: FP Device ListTitle = FP inlet primary conductivity sensor out of range [148] Title = Service Required: FP Device ListTitle = FP RO rejection ratio out of range [149] Title = Service Required: FP Device ListTitle = FP pressure regulator is out of bounds [150] Title = Service Required: FP Device ListTitle = FP inlet temperature too high during filters flush [151] Title = FP: Filter Health ListTitle = FP conductivity after flushing the system is too high [152] Title = FP: Filter Health ListTitle = FP took more than 60 seconds to fill the permeate tank [153] Title = FP: Permeate Overfill ListTitle = FP tank level is too high [154] Title = Service Required: Dialysate Device ListTitle = DD dialysate temperature sensors drift timeout. [155] Title = Blood Leak ListTitle = DD blood leak detected alarm [156] Title = Service Required: Hemodialysis Device ListTitle = DD blood leak sensor invalid calibration record [157] Title = Service Required: Hemodialysis Device ListTitle = DD blood leak sensor embedded mode communication failed [158] Title = Service Required: Hemodialysis Device ListTitle = DD blood leak sensor set point set failure [159] Title = Blood Leak: Bypass ListTitle = DD blood leak detected and it is in the process of attempting to remove the bubbles/foam [160] Title = FP: Permeate Underfill ListTitle = FP tank level is too low [161] Title = High Inlet Water Pressure Fault ListTitle = Inlet Water pressure is too high [162] Title = High Permeate flow ListTitle = Permeate flow is too high from set target flow rate [163] Title = low Permeate flow ListTitle = Permeate flow is too low from set target flow rate [164] Title = low inlet pressure ListTitle = Inlet pressure is too low [165] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap lowering timeout during treatment [166] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump self-test failure alarm [167] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump is running while the BP is off [168] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump ADC error [169] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump fault reported by FPGA [170] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump DAC write failure [171] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump speed check error [172] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump volume check error [173] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump over travel error [174] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump not stopped in off state error [175] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump direction (from encoder) error [176] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump direction (from controller) error [177] Title = Heparin Syringe Empty ListTitle = TD syringe empty alarm [178] Title = Blood Stopped: Syringe Not Detected ListTitle = TD syringe pump syringe removed alarm [179] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump stall alarm [180] Title = Service Required: Hemodialysis Device ListTitle = TD heparin force sensor invalid calibration record [181] Title = Syringe Detected ListTitle = TD syringe detected alarm [182] Title = Not Enough Heparin ListTitle = TD syringe pump not enough Heparin alarm [183] Title = Heparin Pump Occlusion ListTitle = TD syringe pump occlusion alarm [184] Title = Service Required: Hemodialysis Device ListTitle = TD syringe pump timeout alarm [185] Title = Service Required: FP Device ListTitle = FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully [186] Title = Service Required: FP Device ListTitle = FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully [187] Title = Service Required: Hemodialysis Device ListTitle = Blood pump failed to complete homing or one full revolution during auto-load or auto-eject [188] Title = Service Required: Hemodialysis Device ListTitle = Ejector failed to reach extended or retracted position during auto-eject [189] Title = Service Required: Hemodialysis Device ListTitle = TD ejector did not reach the fully retracted position within the allowed time during power-on self-test [190] Title = Blood Pump Stopped: Tubing Set Jam ListTitle = Blood pump torque exceeded limit during Auto-Load/Eject one-revolution and tubing set jam suspected [191] Title = Air Pump Stall Detected ListTitle = Air pump stall detected during treatment [192] Title = Service Required: Dialysate Device ListTitle = TD CPU RAM check failure [193] Title = Service Required: Dialysate Device ListTitle = DD CPU RAM check failure [194] Title = Service Required: Hemodialysis Device ListTitle = TD firmware image integrity POST test failed [195] Title = Service Required: Hemodialysis Device ListTitle = DD firmware image integrity POST test failed [196] Title = Blood Pump Stopped: Saline Bag Empty ListTitle = TD Saline bag is empty [197] Title = Service Required: Dialysate Device ListTitle = DD substitution pump RPM out of range [198] Title = Prime Complete - 30 MIN WARNING ListTitle = TD prime completed medium priority alarm [199] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Repeated air trap lower events detected during treatment [200] Title = Wrong Tube Set Installed ListTitle = TD installed tube set type does not match the selected treatment modality [201] Title = Load Syringe ListTitle = Load Syringe Fill syringe with <$FD8A06>{140:2:mL:1} of heparin = heparin_1.png Connect heparin filled syringe to infusion port = heparin_2.png Insert syringe into the syringe pump = heparin_3.png Unclamp heparin pinch clamp = heparin_4.png Lock heparin syringe in place = heparin_5.png [210] Title = Prime Completed ListTitle = TD Prime Completed System priming is completed. Press ‘OK’ to continue. [244] Title = Low Dry Bicarbonate Cartridge ListTitle = DD Low dry bicarb level Press ‘Resume’ and follow steps to disconnect dry bicarbonate cartridge. Dialysate flow rate is in bypass=dry_bicarb.png [246] Title = Connect Online Fluid Line ListTitle = TD HDF submode change alarm Use aseptic technique to cleanse the dilution port per facility policies Connect online fluid line to the <$FD8A06>post-dialyzer dilution port = connect_online_fluid_2.png Ensure all connection points are secure Ensure online fluid line is properly routed and not kinked, clamped, compressed, or under tension [247] Title = Treatment Mode Changed ListTitle = TD treatment modality change alarm Treatment mode has been changed. [248] Title = Replace Dry Bicarbonate Cartridge ListTitle = DD replace bicarb cartridge alarm Pull <$2268FF>blue bicarbonate knob upwards and remove cartridge =dry_bicarb.png Insert the new dry bicarbonate cartridge Press ‘Resume’ to resume treatment [249] Title = Replace Acid Concentrate ListTitle = DD replace acid concentrate alarm Remove <$B53333>red acid connector from acid concentrate = acid_concentrate.png Insert <$B53333>red acid connector into new acid concentrate Press ‘Resume’ button to resume treatment [250] Title = Dialysate Proportioning in Progress ListTitle = Dialysate Proportioning in Progress Machine is in bypass and treatment will automatically resume once conductivity is within range. Review pH and conductivity based on facility policies once treatment is resumed. [251] Title = Dialysis Treatment Complete ListTitle = Dialysis Treatment Complete Press ‘OK’ to continue to end of treatment options [252] Title = End of Treatment Timeout ListTitle = End of Treatment Timeout The maximum time allowed on this screen has been reached. Press 'Return Blood & End Treatment' to begin the blood return process