[Water Sample Result] Fail Pass [Blood Flow Rate Ranges] # U32 - mL/min Blood_Flow_Rate_Min = +100 # PRS 30, SRSUI 619, 639, 765 Blood_Flow_Rate_Max = +500 # PRS 30, SRSUI 619, 639, 765 Blood_Flow_Rate_Res = 25 # PRS 30, SRSUI 619, 639, 765 Blood_Flow_Rate_Def = 100 # PRS 30, SRSUI 619, 639, 765 [Dialysate Flow Rate Ranges] # U32 - mL/min Dialysate_Flow_Rate_Min = +100 # PRS 345, SRSUI 620, 640, 766 Dialysate_Flow_Rate_Max = +600 # PRS 345, SRSUI 620, 640, 766 Dialysate_Flow_Rate_Res = 50 # PRS 345, SRSUI 620, 640, 766 Dialysate_Flow_Rate_Def = 100 # PRS 345, SRSUI 620, 640, 766 [Treatment Duration Ranges] # U32 - min Treatment_Duration_Min = +60 # PRS 346, SRSUI 569, 767 Treatment_Duration_Max = +480 # PRS 346, SRSUI 569, 767 Treatment_Duration_Res = 15 # PRS 346, SRSUI 569, 767 Treatment_Duration_Def = 60 # PRS 346, SRSUI 569, 767 [Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) Heparin_Dispensing_Rate_Min = 0.2 # PRS 347, SRSUI 768 Heparin_Dispensing_Rate_Max = 1.0 # PRS 347, SRSUI 768 Heparin_Dispensing_Rate_Res = 0.1 # PRS 347, SRSUI 768 Heparin_Dispensing_Rate_Def = 0.0 # PRS 347, SRSUI 768 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. [Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) Heparin_Bolus_Volume_Min = 0.2 # PRS 348, SRSUI 769 Heparin_Bolus_Volume_Max = 2.0 # PRS 348, SRSUI 769 Heparin_Bolus_Volume_Res = 0.1 # PRS 348, SRSUI 769 Heparin_Bolus_Volume_Def = 0.0 # PRS 348, SRSUI 769 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. [Heparin Stop Time Ranges] #U32 - min Heparin_Stop_Time_Min = 0 # PRS 464. SRSUI 771 Heparin_Stop_Time_Max = 480 # PRS 464. SRSUI 771 Heparin_Stop_Time_Res = 10 # PRS 464. SRSUI 771 Heparin_Stop_Time_Def = 0 # PRS 464. SRSUI 771 [Saline Bolus Volume Ranges] # U32 - mL Saline_Bolus_Volume_Min = 100 # PRS 350 Saline_Bolus_Volume_Max = 300 # PRS 350 Saline_Bolus_Volume_Res = 100 # PRS 350 Saline_Bolus_Volume_Def = 100 # PRS 350 [Heparin Type Options] # U32 - PRS ???, SRSUI ??? UFH 1,000 IU/mL [Acid Concentrate Options] # U32 - PRS 35, SRSUI 772 08-1251-1 08-2251-0 08-3251-9 [Bicarbonate Concentrate Options] # U32 - PRS 351, SRSUI 773 Fres. Centrisol [Dialyzer Type Options] # U32 - PRS 352, SRSUI 774 BB Diacap Pro 13H BB Diacap Pro 16H BB Diacap Pro 19H F Optiflux F160NRe F Optiflux F180NRe F Optiflux F200NRe [Dialysate Temperature] # Celsius Dialysate_Temperature_Min = 35 # PRS 353, SRSUI 775 Dialysate_Temperature_Max = 37 # PRS 353, SRSUI 775 Dialysate_Temperature_Res = 0.5 # PRS 353, SRSUI 775 Dialysate_Temperature_Def = 37 # PRS 353, SRSUI 775 [Arterial Blood Pressure Ranges] # Adjustment, Arterial Low Arterial_Adjust_Low_Min = -300 # PRS 354, SRSUI 776, 726 Arterial_Adjust_Low_Max = +70 # PRS 354, SRSUI 776, 726 Arterial_Adjust_Low_Res = 10 # PRS 354, SRSUI 776, 726 # this has to be the same for all since they are on the same slider with the same scale. Arterial_Adjust_Low_Def = -300 # PRS 354, SRSUI 776, 726 Arterial_Adjust_Low_Gap = 30 # PRS 354, SRSUI 728 # Adjustment, Arterial High Arterial_Adjust_High_Min = -270 # PRS 355, SRSUI 780, 727 Arterial_Adjust_High_Max = +100 # PRS 355, SRSUI 780, 727 Arterial_Adjust_High_Res = 10 # PRS 355, SRSUI 780, 727 # this has to be the same for all since they are on the same slider with the same scale. Arterial_Adjust_High_Def = +100 # PRS 355, SRSUI 780, 727 Arterial_Adjust_High_Gap = 30 # PRS 355, SRSUI 728 #Monitoring Arterial_Monitor_Min = -400 # PRS 362, SRS 617 Arterial_Monitor_Max = +100 # PRS 362, SRS 617 Arterial_Monitor_Res = 1 # PRS 362, SRS 617 [Venous Blood Pressure Ranges] # Adjustment, Venous Low Venous_Adjust_Low_Min = +20 # PRS 356, SRSUI 777, 731 Venous_Adjust_Low_Max = +370 # PRS 356, SRSUI 777, 731 Venous_Adjust_Low_Res = 10 # PRS 356, SRSUI 777, 731 # this has to be the same for all since they are on the same slider with the same scale. Venous_Adjust_Low_Def = +20 # PRS 356, SRSUI 777, 731 Venous_Adjust_Low_Gap = 30 # PRS 356, SRSUI 1084 # Adjustment, Venous High Venous_Adjust_High_Min = +50 # PRS 357, SRSUI 781, 732 Venous_Adjust_High_Max = +300 # PRS 357, SRSUI 781, 732 Venous_Adjust_High_Res = 10 # PRS 357, SRSUI 781, 732 # this has to be the same for all since they are on the same slider with the same scale. Venous_Adjust_High_Def = +300 # PRS 357, SRSUI 781, 732 Venous_Adjust_High_Gap = 30 # PRS 357, SRSUI 1084 ##Monitoring Venous_Monitor_Min = -100 # PRS 361, SRS 616 Venous_Monitor_Max = +500 # PRS 361, SRS 616 Venous_Monitor_Res = 1 # PRS 361, SRS 616 [Blood Pressure Measure Interval Ranges] # min (0 = OFF) Blood_Pressure_Measure_Interval_Min = 5 # PRS 358, SRSUI 778 Blood_Pressure_Measure_Interval_Max = 60 # PRS 358, SRSUI 778 Blood_Pressure_Measure_Interval_Res = 5 # PRS 358, SRSUI 778 Blood_Pressure_Measure_Interval_Def = 0 # PRS 358, SRSUI 778 [Rinseback Flow Rate Ranges] # mL/min Rinseback_Flow_Rate_Min = 50 # PRS 359, SRSUI 330, 770 Rinseback_Flow_Rate_Max = 150 # PRS 359, SRSUI 330, 770 Rinseback_Flow_Rate_Res = 25 # PRS 359, SRSUI 330, 770 Rinseback_Flow_Rate_Def = 50 # PRS 359, SRSUI 330, 770