[Water Sample Result] Fail Pass [Blood Flow Rate Ranges] # U32 - mL/min Blood_Flow_Rate_Def = +50 # LEAHI-SRS-287 Blood_Flow_Rate_Min = +50 # LEAHI-SRS-287 Blood_Flow_Rate_Max = +500 # LEAHI-SRS-287 Blood_Flow_Rate_Res = 10 # LEAHI-SRS-287 [Dialysate Flow Rate Ranges] # U32 - mL/min Dialysate_Flow_Rate_Def = +50 # LEAHI-SRS-294 Dialysate_Flow_Rate_Min = +50 # LEAHI-SRS-294 Dialysate_Flow_Rate_Max = +600 # LEAHI-SRS-294 Dialysate_Flow_Rate_Res = +25 # LEAHI-SRS-294 [Treatment Duration Ranges] # U32 - minutes Treatment_Duration_Def = +60 # LEAHI-SRS-295 Treatment_Duration_Min = +60 # LEAHI-SRS-295 Treatment_Duration_Max = +480 # LEAHI-SRS-295 Treatment_Duration_Res = +15 # LEAHI-SRS-295 [Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) Heparin_Dispensing_Rate_Def = +0.0 # LEAHI-SRS-296 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. Heparin_Dispensing_Rate_Min = +0.2 # LEAHI-SRS-296 Heparin_Dispensing_Rate_Max = +1.0 # LEAHI-SRS-296 Heparin_Dispensing_Rate_Res = +0.1 # LEAHI-SRS-296 [Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) Heparin_Bolus_Volume_Def = +0.0 # LEAHI-SRS-298 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. Heparin_Bolus_Volume_Min = +0.2 # LEAHI-SRS-298 Heparin_Bolus_Volume_Max = +2.0 # LEAHI-SRS-298 Heparin_Bolus_Volume_Res = +0.1 # LEAHI-SRS-298 [Heparin Stop Time Ranges] # U32 - min # Missing parameters are dynamically update in run time) Heparin_Stop_Time_Min = +0 # LEAHI-SRS-300 [Saline Bolus Volume Ranges] # U32 - mL Saline_Bolus_Volume_Def = +100 # LEAHI-SRS-302 Saline_Bolus_Volume_Min = +100 # LEAHI-SRS-302 Saline_Bolus_Volume_Max = +300 # LEAHI-SRS-302 Saline_Bolus_Volume_Res = +100 # LEAHI-SRS-302 [Dialysate Temperature] # Celsius Dialysate_Temperature_Def = +37 # LEAHI-SRS-303 Dialysate_Temperature_Min = +35 # LEAHI-SRS-303 Dialysate_Temperature_Max = +38 # LEAHI-SRS-303 Dialysate_Temperature_Res = +0.5 # LEAHI-SRS-303 [Potassium] # K Potassium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 Potassium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 Potassium_Max = +4.0 # LEAHI-PRS-660, LEAHI-PRS-241 Potassium_Res = +1.0 # LEAHI-PRS-660, LEAHI-PRS-241 [Calcium] # Ca Calcium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 Calcium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 Calcium_Max = +3.5 # LEAHI-PRS-660, LEAHI-PRS-241 Calcium_Res = +0.25 # LEAHI-PRS-660, LEAHI-PRS-241 [Acid Concentrate Options] # U32 1.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 2.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 3.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 [Bicarbonate Concentrate Options] # U32 Sodium Bicarbonate # LEAHI-SRS-289 [Dialyzer Type Options] # U32 Diacap Pro 13H # LEAHI-SRS-290 Diacap Pro 16H # LEAHI-SRS-290 Diacap Pro 19H # LEAHI-SRS-290 Optiflux F160NRe # LEAHI-SRS-290 Optiflux F180NRe # LEAHI-SRS-290 Optiflux F200NRe # LEAHI-SRS-290 Optiflux F250NRe # LEAHI-SRS-290 [Heparin Type Options] # U32 Unfractionated 1,000 IU/mL # LEAHI-SRS-291 [Trancememberance Pressure Ranges] # Adjustment, Arterial Window Trancememberance_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? Trancememberance_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? Trancememberance_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? Trancememberance_Adjust_Window_Res = +20 # LEAHI-PRS-255 ? #Monitoring Trancememberance_Monitor_Res = 1 # LEAHI-PRS-269 LEAHI-SRS-222 Trancememberance_Monitor_Min = -650 # LEAHI-PRS-269 LEAHI-SRS-222 Trancememberance_Monitor_Max = +250 # LEAHI-PRS-269 LEAHI-SRS-222 [Arterial Blood Pressure Ranges] # Adjustment, Arterial Window Arterial_Adjust_Window_Def = +120 # LEAHI-PRS-215 ? Arterial_Adjust_Window_Min = +120 # LEAHI-PRS-245 ? Arterial_Adjust_Window_Max = +200 # LEAHI-PRS-245 ? Arterial_Adjust_Window_Res = +40 # LEAHI-PRS-245 ? #Monitoring Arterial_Monitor_Res = 1 # LEAHI-PRS-252 LEAHI-SRS-218 Arterial_Monitor_Min = -400 # LEAHI-PRS-252 LEAHI-SRS-218 Arterial_Monitor_Max = +200 # LEAHI-PRS-252 LEAHI-SRS-218 [Venous Blood Pressure Ranges] # Adjustment, Venous Window Venous_Adjust_Window_Def = +100 # LEAHI-PRS-215 ? Venous_Adjust_Window_Min = +100 # LEAHI-PRS-246 ? Venous_Adjust_Window_Max = +200 # LEAHI-PRS-246 ? Venous_Adjust_Window_Res = +20 # LEAHI-PRS-246 ? # Adjustment, Venous Asymmetric Venous_Adjust_Asymmetric_Def = +20 # LEAHI-PRS-215 ? Venous_Adjust_Asymmetric_Min = +20 # LEAHI-PRS-247 ? Venous_Adjust_Asymmetric_Max = +35 # LEAHI-PRS-247 ? Venous_Adjust_Asymmetric_Res = +5 # LEAHI-PRS-247 ? ##Monitoring Venous_Monitor_Res = 1 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Min = -100 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Max = +500 # LEAHI-PRS-253 LEAHI-SRS-220 [Blood Pressure Measure Interval Ranges] # min (0 = OFF) Blood_Pressure_Measure_Interval_Def = +0 # LEAHI-SRS-304 Blood_Pressure_Measure_Interval_Min = +5 # LEAHI-SRS-304 Blood_Pressure_Measure_Interval_Max = +60 # LEAHI-SRS-304 Blood_Pressure_Measure_Interval_Res = +5 # LEAHI-SRS-304 [Rinseback Volume Ranges] # mL/min Rinseback_Volume_Def = 300 # LEAHI-PRS-559 Rinseback_Volume_Min = 0 # LEAHI-? Rinseback_Volume_Max = 0 # LEAHI-? Rinseback_Volume_Res = 0 # LEAHI-? # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. [Rinseback Flow Rate Ranges] # mL/min Rinseback_Flow_Rate_Def = +200 # LEAHI-PRS-353 Rinseback_Flow_Rate_Min = +100 # LEAHI-PRS-354 Rinseback_Flow_Rate_Max = +300 # LEAHI-PRS-354 Rinseback_Flow_Rate_Res = +10 # LEAHI-PRS-354 # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore.