[0] Title = Indication for when no alarms is a possible situation ListTitle = [1] Title = Service Required: Hemodialysis Device ListTitle = TD Software fault. Software found itself in an unexpected state [2] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state [3] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state [4] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA POST failure [5] Title = Service Required: Hemodialysis Device ListTitle = TD Watchdog POST failure [6] Title = Service Required: Hemodialysis Device ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up [7] Title = Service Required: Hemodialysis Device ListTitle = TD processor clock speed checks against FPGA clock failure [8] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA communication down for too long [9] Title = Service Required: Hemodialysis Device ListTitle = User interface communication timeout [10] Title = Communication Error ListTitle = Dialysate generator communication timeout [11] Title = Service Required: Hemodialysis Device ListTitle = Too many bad CRCs detected on received system messages [12] Title = Service Required: Hemodialysis Device ListTitle = System message that required acknowledgment was not acknowledged [13] Title = Communication Error ListTitle = TD to DD comm error. [14] Title = Service Required: Hemodialysis Device ListTitle = TD critical data integrity check failed [15] Title = Service Required: Dialysate Device ListTitle = DD critical data integrity check failed [16] Title = Service Required: Hemodialysis Device ListTitle = TD watchdog expired [17] Title = Service Required: Dialysate Device ListTitle = DD watchdog expired [18] Title = Service Required: Dialysate Device ListTitle = Too many bad CRCs detected on received system messages [19] Title = Service Required: Dialysate Device ListTitle = DD to RO comm error [20] Title = Service Required: Dialysate Device ListTitle = DD to TD comm error [21] Title = Service Required: Dialysate Device ListTitle = DD FPGA POST failure [22] Title = Service Required: Dialysate Device ListTitle = FPGA clock speed failure [23] Title = Service Required: Dialysate Device ListTitle = DD FPGA communication down for too long [24] Title = Service Required: Hemodialysis Device ListTitle = TD alarm audio failed POST [25] Title = Service Required: Dialysate Device ListTitle = DD FPGA not accepting commanded valve states [26] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor reporting error or not being read successfully [27] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor reporting error or not being read successfully [28] Title = Service Required: Dialysate Device ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully [29] Title = Service Required: Dialysate Device ListTitle = DD bibag pressure sensor reporting error or not being read successfully [30] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pressure sensor reporting error or not being read successfully [31] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully [32] Title = Service Required: Dialysate Device ListTitle = DD transmembrane sensor reporting error or not being read successfully [33] Title = Service Required: Hemodialysis Device ListTitle = TD monitored voltage is out of range [34] Title = Power Loss ListTitle = TD AC power lost alarm [35] Title = Power Loss ListTitle = TD AC power lost alarm [36] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully [37] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully [38] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully [39] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully [40] Title = Venous Air Detected ListTitle = TD venous air bubble detected alarm [41] Title = Blood Pump Stopped: Door Open ListTitle = TD cartridge door opened alarm [42] Title = Stuck Button ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up [43] Title = Low Arterial Pressure ListTitle = Arterial pressure too low during treatment [44] Title = High Arterial Pressure ListTitle = Arterial pressure too high during treatment [45] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor is reading out of range [46] Title = Self Test: Cartridge ListTitle = TD arterial pressure self-test failure alarm [47] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment [48] Title = High Venous Pressure ListTitle = Venous pressure too high during treatment [49] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor is reading out of range [50] Title = Self Test: Cartridge ListTitle = TD venous pressure self-test failure alarm [51] Title = Self Test: Cartridge Check ListTitle = TD pre-treatment mode dry self-tests failure [52] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D11_PUMP speed control error [53] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D10_PUMP speed control error [54] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump hall sensor out of range. [55] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump fault [56] Title = Service Required: Dialysate Device ListTitle = DD temperature sensor out of range [57] Title = Service Required: Dialysate Device ListTitle = DD temperature sensors not being read successfully [58] Title = Service Required: Dialysate Device ListTitle = DD barometric sensor not being read successfully [59] Title = Service Required: Dialysate Device ListTitle = DD barometric pressure sensor bad coefficients CRC [60] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump RPM out of range [61] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump off fault [62] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump current out of range [63] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump invalid direction [64] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump direction FPGA fault [65] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump RPM out of range [66] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump off fault [67] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump current out of range [68] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump invalid direction [69] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump direction FPGA fault [70] Title = Service Required: Dialysate Device ListTitle = DD D5 heater is On with low fluid level [71] Title = Service Required: Dialysate Device ListTitle = DD D45 heater is On with low fluid level [72] Title = Service Required: Dialysate Device ListTitle = DD water inlet input pressure sensor reporting error or not being read successfully [73] Title = Service Required: Dialysate Device ListTitle = DD water inlet output pressure sensor reporting error or not being read successfully [74] Title = Service Required: Dialysate Device ListTitle = DD UF pump D76_PUMP speed control error [75] Title = Service Required: Dialysate Device ListTitle = DD Spent chamber fill state 1 start pressure is not in range [76] Title = Service Required: Dialysate Device ListTitle = DD Spent chamber fill state 2 start pressure is not in range [77] Title = Service Required: Dialysate Device ListTitle = DD Spent chamber fill timeout error [78] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap fill timeout during treatment [79] Title = Service Required: Hemodialysis Device ListTitle = Air trap level sensors reporting illegal combination of air/fluid [80] Title = Service Required: Hemodialysis Device ListTitle = Pinch valve reporting a fault [81] Title = Service Required: Hemodialysis Device ListTitle = TD valve homing failed [82] Title = Service Required: Hemodialysis Device ListTitle = TD valve transition time out [83] Title = Communication Error ListTitle = DD to RO comm error. [84] Title = Service Required: Dialysate Device ListTitle = RO FPGA communication down for too long [85] Title = Service Required: Hemodialysis Device ListTitle = RO Too many bad CRCs detected on received system messages [86] Title = Service Required: Filtration Device ListTitle = RO FPGA not accepting commanded valve states [87] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill start pressure is not in range [88] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill end pressure is not in range [89] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill start pressure is not in range [90] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill end pressure is not in range [91] Title = Service Required: Dialysate Device ListTitle = DD hydraulics chamber degassing pressure is not in range [92] Title = Service Required: Dialysate Device ListTitle = DD Bicarb chamber fill timeout error [93] Title = ListTitle = [94] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range [95] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range [96] Title = Blood Pump Stopped: Treatment Paused ListTitle = TD treatment stopped by user action - pressed stop button [97] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully [98] Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood sitting too long warning (> 4 min) [99] Title = Blood Pump Stop Time Exceeded ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min) [100] Title = Service Required: Hemodialysis Device ListTitle = TD communication timeout [101] Title = Service Required: Hemodialysis Device ListTitle = UI POST OS Version compatibility failure [102] Title = Service Required: Hemodialysis Device ListTitle = UI POST Application Integrity (Sha256Sum) failure [103] Title = Service Required: Hemodialysis Device ListTitle = UI POST CANBus failure [104] Title = Service Required: Hemodialysis Device ListTitle = If the UI POST reports display failure. [105] Title = Service Required: Hemodialysis Device ListTitle = UI POST Touch failure [106] Title = Service Required: Hemodialysis Device ListTitle = UI POST SD-Card failure [107] Title = Service Required: Hemodialysis Device ListTitle = UI POST RTC failure [108] Title = Service Required: Hemodialysis Device ListTitle = UI POST WiFi failure [109] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bluetooth failure [110] Title = Service Required: Hemodialysis Device ListTitle = UI POST Ethernet failure [111] Title = Service Required: Hemodialysis Device ListTitle = UI POST Sound failure [112] Title = Service Required: Hemodialysis Device ListTitle = UI POST cloud sync POST failure [113] Title = Service Required: Hemodialysis Device ListTitle = UI POST Invalid Year [114] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bad Settings [115] Title = Service Required: Hemodialysis Device ListTitle = SD card is non-operational [116] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [117] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [118] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [119] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [120] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [121] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [122] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [123] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [124] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [125] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [126] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [127] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [128] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [129] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [130] Title = Service Required: Hemodialysis Device ListTitle = RESERVED [131] Title = Battery Capacity Low ListTitle = Insufficient battery charge to support 20 min without A/C power [132] Title = High Dialysate Temperature ListTitle = Dialysate temperature is out of high safety range [133] Title = High Dialysate Temperature ListTitle = Dialysate temperature is above the target temperature [134] Title = Low Dialysate Temperature ListTitle = Dialysate temperature is below the target temperature [135] Title = Service Required: FP Device ListTitle = FP pressure relief is too low [136] Title = Service Required: FP Device ListTitle = FP pressure relief is too High [137] Title = Service Required: FP Device ListTitle = FP device is using too much power [138] Title = Service Required: FP Device ListTitle = FP RO pump flow out of range [139] Title = Low Inlet Water Temperature ListTitle = Inlet Water temperature is too low [140] Title = High Inlet Water Temperature ListTitle = Inlet Water temperature is too high [141] Title = Low Inlet Water Pressure ListTitle = Inlet Water pressure is too low [142] Title = High Inlet Water Pressure ListTitle = Inlet Water pressure is too high [143] Title = High Inlet Water Conductivity ListTitle = Inlet Water conductivity is too high [144] Title = Low Inlet Water Conductivity ListTitle = Inlet Water conductivity is too low [145] Title = Service Required: FP Device ListTitle = FP RO pump pressure out of range [146] Title = Service Required: FP Device ListTitle = FP outlet primary conductivity sensor out of range [147] Title = Service Required: FP Device ListTitle = FP inlet primary conductivity sensor out of range [148] Title = Service Required: FP Device ListTitle = FP RO rejection ratio out of range [149] Title = Service Required: FP Device ListTitle = FP pressure regulator is out of bounds [150] Title = FP: Filter Health ListTitle = FP pressure difference between the filters is too high [151] Title = FP: Filter Health ListTitle = FP conductivity after flushing the system is too high [152] Title = FP: Filter Health ListTitle = FP volume difference between integrated flow and expected volume [153] Title = FP: Permeate Overfill ListTitle = FP tank level is too high [154] Title = Service Required: Dialysate Device ListTitle = DD dialysate temperature sensors drift timeout. [155] Title = Blood Leak ListTitle = DD blood leak detected alarm [156] Title = Service Required: Hemodialysis Device ListTitle = DD blood leak sensor invalid calibration record [157] Title = Service Required: Hemodialysis Device ListTitle = DD blood leak sensor embedded mode communication failed [158] Title = Service Required: Hemodialysis Device ListTitle = DD blood leak sensor set point set failure [159] Title = Blood Leak: Bypass ListTitle = DD blood leak detected and it is in the process of attempting to remove the bubbles/foam [160] Title = FP: Permeate Underfill ListTitle = FP tank level is too low