[0] Title = Used when there is no rejection [1] Title = Request is not allowed in the current operating mode [2] Title = Validated request was not confirmed by user in reasonable time [3] Title = Request is not allowed if not in treatment mode [4] Title = Request is not allowed in current treatment state [5] Title = Request is not allowed so near end of treatment [6] Title = Treatment duration is out of range [7] Title = Treatment time change is less than currently elapsed treatment time [8] Title = Blood flow is out of range [9] Title = Dialysate flow is out of range [10] Title = Dialysate flow rate or treatment duration causes dialysate volume to exceed limit [11] Title = Ultrafiltration volume is out of range [12] Title = Ultrafiltration rate is out of range [13] Title = Treatment time change is less than minimum treatment time [14] Title = Ultrafiltration is not currently in progress [15] Title = Ultrafiltration is not currently paused [16] Title = A saline bolus is in progress [17] Title = A treatment parameter is out of range [18] Title = The Heparin pre-stop setting is greater than the treatment duration [19] Title = DG filter flush time has been expired [20] Title = Not used [21] Title = Saline bolus volume maximum has been reached - no more saline allowed [22] Title = A saline bolus is not in progress [23] Title = Requested user action is disabled in current state [24] Title = Requested user action not allowed while alarm is active [25] Title = Requested user action invalid [26] Title = The treatment has been completed [27] Title = Rinseback additional volume maximum has been reached - no more additional rinsebacks allowed [28] Title = Dialyzer not inverted yet [29] Title = The user has not confirmed patient connection [30] Title = Heparin cannot be paused if not currently deliverying Heparin [31] Title = Heparin cannot be resumed if not paused [32] Title = Treatment cannot initiate if DG comm is lost [33] Title = Post-treatment reservoirs drain not complete [34] Title = Treatment cannot initiate if DG is not in standby idle state [35] Title = Request message not formatted properly [36] Title = Given date/time is invalid [37] Title = Treatment is in progress [38] Title = Battery does not have enough charge to start treatment [39] Title = Cannot move on to recirculate without completing full rinseback. [40] Title = Reservoir one fill is not complete [41] Title = Pump track is not ready / closed [42] Title = Door is not closed [43] Title = Syringe is not present [44] Title = Dialysate cap open [45] Title = Concentrate cap open [46] Title = DG disinfect has been expired [47] Title = DG service is due [48] Title = HD service is due [49] Title = DG chemical flush is not complete [50] Title = DG RO filter temperature out of range [51] Title = DG firmware version is not compatible with HD firmware [52] Title = Heparin settings are zero but pre-stop setting is non-zero [53] Title = Saline bolus is not allowed during dialyzer reprime [54] Title = DG RO only mode DG is busy [55] Title = DG RO only mode invalid parameter [56] Title = DG RO only mode invalid payload length [57] Title = Cannot resume treatment per alarm [58] Title = Cannot continue due to maximum heparin volume [59] Title = Cannot start chemical disinfect because it is not enabled in institutional record.