[Water Sample Result] Fail Pass [Treatment Modality] HD HDF [HDF Treatment Mode] Pre-Dilution Post-Dilution [Blood Flow Rate Ranges] # U32 - mL/min Blood_Flow_Rate_Def = +50 # LEAHI-SRS-287 Blood_Flow_Rate_Min = +50 # LEAHI-SRS-287 Blood_Flow_Rate_Max = +500 # LEAHI-SRS-287 Blood_Flow_Rate_Res = 10 # LEAHI-SRS-287 [Dialysate Flow Rate Ranges] # U32 - mL/min Dialysate_Flow_Rate_Def = +50 # LEAHI-SRS-294 Dialysate_Flow_Rate_Min = +50 # LEAHI-SRS-294 Dialysate_Flow_Rate_Max = +600 # LEAHI-SRS-294 Dialysate_Flow_Rate_Res = +25 # LEAHI-SRS-294 [Treatment Duration Ranges] # U32 - minutes Treatment_Duration_Def = +60 # LEAHI-SRS-295 Treatment_Duration_Min = +60 # LEAHI-SRS-295 Treatment_Duration_Max = +480 # LEAHI-SRS-295 Treatment_Duration_Res = +15 # LEAHI-SRS-295 [Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) Heparin_Dispensing_Rate_Def = +0.0 # LEAHI-SRS-296 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. Heparin_Dispensing_Rate_Min = +0.2 # LEAHI-SRS-296 Heparin_Dispensing_Rate_Max = +3.0 # LEAHI-PRS 215 && PRS 238 Heparin_Dispensing_Rate_Res = +0.1 # LEAHI-SRS-296 [Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) Heparin_Bolus_Volume_Def = +0.0 # LEAHI-SRS-298 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. Heparin_Bolus_Volume_Min = +0.2 # LEAHI-SRS-298 Heparin_Bolus_Volume_Max = +2.0 # LEAHI-SRS-298 Heparin_Bolus_Volume_Res = +0.1 # LEAHI-SRS-298 [Heparin Stop Time Ranges] # U32 - min # Missing parameters are dynamically update in run time) Heparin_Stop_Time_Min = +0 # LEAHI-SRS-300 [Saline Bolus Volume Ranges] # U32 - mL Saline_Bolus_Volume_Def = +100 # LEAHI-SRS-302 Saline_Bolus_Volume_Min = +100 # LEAHI-SRS-302 Saline_Bolus_Volume_Max = +300 # LEAHI-SRS-302 Saline_Bolus_Volume_Res = +100 # LEAHI-SRS-302 [Dialysate Temperature] # Celsius Dialysate_Temperature_Def = +37 # LEAHI-SRS-303 Dialysate_Temperature_Min = +35 # LEAHI-SRS-303 Dialysate_Temperature_Max = +38 # LEAHI-SRS-303 Dialysate_Temperature_Res = +0.5 # LEAHI-SRS-303 [Acid Concentrate Options] # U32 1.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 2.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 3.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 [Acid Concentrate Ranges] # K Acid_Concentrate_Potassium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 Acid_Concentrate_Potassium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 Acid_Concentrate_Potassium_Max = +4.0 # LEAHI-PRS-660, LEAHI-PRS-241 Acid_Concentrate_Potassium_Res = +1.0 # LEAHI-PRS-660, LEAHI-PRS-241 # Ca Acid_Concentrate_Calcium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 Acid_Concentrate_Calcium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 Acid_Concentrate_Calcium_Max = +3.5 # LEAHI-PRS-660, LEAHI-PRS-241 Acid_Concentrate_Calcium_Res = +0.25 # LEAHI-PRS-660, LEAHI-PRS-241 [Dialysate Composition Ranges] # mEq/L Dialysate_Composition_Bicarbonate_Def = +137 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Bicarbonate_Min = +130 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Bicarbonate_Max = +155 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Bicarbonate_Res = +1 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Sodium_Def = +33 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Sodium_Min = +20 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Sodium_Max = +40 # LEAHI-PRS-215, LEAHI-PRS-238 Dialysate_Composition_Sodium_Res = +1 # LEAHI-PRS-215, LEAHI-PRS-238 [Bicarbonate Concentrate Options] # U32 Sodium Bicarbonate # LEAHI-SRS-289 [Dialyzer Type Options] # U32 Diacap Pro 13H # LEAHI-SRS-290 Diacap Pro 16H # LEAHI-SRS-290 Diacap Pro 19H # LEAHI-SRS-290 Optiflux F160NRe # LEAHI-SRS-290 Optiflux F180NRe # LEAHI-SRS-290 Optiflux F200NRe # LEAHI-SRS-290 Optiflux F250NRe # LEAHI-SRS-290 [Dry Bicarb Cartridge Size] # U32 650 g 720 g 760 g 900 g 1100 g 1250 g [Heparin Type Options] # U32 Unfractionated 1,000 # LEAHI-SRS-291 [Vitals Measurement Interval Options] # min (0 = OFF) 0 # LEAHI-PRS-131 5 # LEAHI-PRS-131 10 # LEAHI-PRS-131 15 # LEAHI-PRS-131 20 # LEAHI-PRS-131 30 # LEAHI-PRS-131 60 # LEAHI-PRS-131 [Trancememberance Pressure Ranges] # Adjustment, Trancememberance Window Trancememberance_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? Trancememberance_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? Trancememberance_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? Trancememberance_Adjust_Window_Res = +20 # LEAHI-PRS-255 ? #Monitoring Trancememberance_Monitor_Res = 1 # LEAHI-PRS-269 LEAHI-SRS-222 Trancememberance_Monitor_Min = -650 # LEAHI-PRS-269 LEAHI-SRS-222 Trancememberance_Monitor_Max = +250 # LEAHI-PRS-269 LEAHI-SRS-222 [Arterial Blood Pressure Ranges] # Adjustment, Arterial Window Arterial_Adjust_Window_Def = +120 # LEAHI-PRS-215 ? Arterial_Adjust_Window_Min = +120 # LEAHI-PRS-245 ? Arterial_Adjust_Window_Max = +200 # LEAHI-PRS-245 ? Arterial_Adjust_Window_Res = +40 # LEAHI-PRS-245 ? #Monitoring Arterial_Monitor_Res = 1 # LEAHI-PRS-252 LEAHI-SRS-218 Arterial_Monitor_Min = -400 # LEAHI-PRS-252 LEAHI-SRS-218 Arterial_Monitor_Max = +200 # LEAHI-PRS-252 LEAHI-SRS-218 [Venous Blood Pressure Ranges] # Adjustment, Venous Window Venous_Adjust_Window_Def = +100 # LEAHI-PRS-215 ? Venous_Adjust_Window_Min = +100 # LEAHI-PRS-246 ? Venous_Adjust_Window_Max = +200 # LEAHI-PRS-246 ? Venous_Adjust_Window_Res = +20 # LEAHI-PRS-246 ? # Adjustment, Venous Asymmetric Venous_Adjust_Asymmetric_Def = +20 # LEAHI-PRS-215 ? Venous_Adjust_Asymmetric_Min = +20 # LEAHI-PRS-247 ? Venous_Adjust_Asymmetric_Max = +35 # LEAHI-PRS-247 ? Venous_Adjust_Asymmetric_Res = +5 # LEAHI-PRS-247 ? ##Monitoring Venous_Monitor_Res = 1 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Min = -100 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Max = +500 # LEAHI-PRS-253 LEAHI-SRS-220 [Rinseback Volume Ranges] # mL/min Rinseback_Volume_Def = +300 # LEAHI-PRS-559 Rinseback_Volume_Min = +200 # LEAHI-PRS-238 Rinseback_Volume_Max = +400 # LEAHI-PRS-238 Rinseback_Volume_Res = +10 # LEAHI-PRS-238 [Rinseback Flow Rate Ranges] # mL/min Rinseback_Flow_Rate_Def = +200 # LEAHI-PRS-353 Rinseback_Flow_Rate_Min = +100 # LEAHI-PRS-354 Rinseback_Flow_Rate_Max = +500 # LEAHI-PRS-354 Rinseback_Flow_Rate_Res = +10 # LEAHI-PRS-354 [Ultrafiltration Volume Ranges] # L Ultrafiltration_Volume_Def = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 Ultrafiltration_Volume_Min = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 Ultrafiltration_Volume_Max = +0.8 # LEAHI-PRS-127, LEAHI-PRS-238 Ultrafiltration_Volume_Res = +0.1 # LEAHI-PRS-127 [Substitution Volume Ranges] # L Substitution_Volume_Def = +1.0 # LEAHI-PRS-238 Substitution_Volume_Min = +1.0 # LEAHI-PRS-238 Substitution_Volume_Max = +30.0 # LEAHI-PRS-238 Substitution_Volume_Res = +0.1 # LEAHI-PRS-238 [Minimum RO Rejection Ratio Alarm] Min_RO_Rejection_Ratio_Alarm_Def = +90 # LEAHI-PRS-223 Min_RO_Rejection_Ratio_Alarm_Min = +1 # LEAHI-PRS-223 Min_RO_Rejection_Ratio_Alarm_Max = +99 # LEAHI-PRS-223 Min_RO_Rejection_Ratio_Alarm_Res = +1 # LEAHI-PRS-223 [Disinfection Frequency] Disinfection_Frequency_Def = +3 # LEAHI-PRS-223 Disinfection_Frequency_Min = +1 # LEAHI-PRS-223 Disinfection_Frequency_Max = +5 # LEAHI-PRS-223 Disinfection_Frequency_Res = +1 # LEAHI-PRS-223 [Disinfection Parameters Cycle Time] Disinfection_Parameters_Cycle_Time_Def = +30 # LEAHI-PRS-223 Disinfection_Parameters_Cycle_Time_Min = +1 # LEAHI-PRS-223 Disinfection_Parameters_Cycle_Time_Max = +60 # LEAHI-PRS-223 Disinfection_Parameters_Cycle_Time_Res = +1 # LEAHI-PRS-223 [Maximum Blood Pump Stop Time] Max_Blood_Pump_Stop_Time_Def = +5 # LEAHI-PRS-223 Max_Blood_Pump_Stop_Time_Min = +5 # LEAHI-PRS-223 Max_Blood_Pump_Stop_Time_Max = +30 # LEAHI-PRS-223 Max_Blood_Pump_Stop_Time_Res = +1 # LEAHI-PRS-223 [Inlet Water Cond. Low Threshold] Inlet_Water_Cond_Low_Threshold_Def = +200 # LEAHI-PRS-223 Inlet_Water_Cond_Low_Threshold_Min = +20 # LEAHI-PRS-223 Inlet_Water_Cond_Low_Threshold_Max = +200 # LEAHI-PRS-223 Inlet_Water_Cond_Low_Threshold_Res = +10 # LEAHI-PRS-223 [Inlet Water Cond. High Threshold] Inlet_Water_Cond_High_Threshold_Def = +50 # LEAHI-PRS-223 Inlet_Water_Cond_High_Threshold_Min = +20 # LEAHI-PRS-223 Inlet_Water_Cond_High_Threshold_Max = +200 # LEAHI-PRS-223 Inlet_Water_Cond_High_Threshold_Res = +10 # LEAHI-PRS-223 [Chlorine Water Sample Check] Chlorine_Water_Sample_Check_Def = +240 # LEAHI-PRS-223 Chlorine_Water_Sample_Check_Min = +15 # LEAHI-PRS-223 Chlorine_Water_Sample_Check_Max = +480 # LEAHI-PRS-223 Chlorine_Water_Sample_Check_Res = +5 # LEAHI-PRS-223 [Water Sample Test Result Required] Disabled Enabled